K Number
K112945

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-11-03

(30 days)

Product Code
Regulation Number
876.1075
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lymph nodes and various soft tissue tumors.

Device Description

The EliteCore™ Full Core Biopsy Device is a sterile, disposable device which features a stainless steel cutting cannula with spoon, rotating coring cannula and stylet. The device is comprised of a plastic housing that contains the mechanically actuated components. The key performance attribute of the EliteCore™ Full Core Biopsy Device is the capability of obtaining a full core specimen. A variable throw feature allows the user to choose a throw setting ranging from 13 to 25 mm.

AI/ML Overview

The EliteCore™ Full Core Biopsy Device is a sterile, disposable device for obtaining biopsies from soft tissues. The performance of this device was assessed by comparing it to a predicate device, the Revolution® Full Core Biopsy Device (K090611). The tests conducted aimed to confirm that the quality of samples obtained with the EliteCore™ device is equivalent to that of the predicate device.

1. Table of Acceptance Criteria and Reported Device Performance:

TestAcceptance Criteria (How Equivalent)Reported Device Performance
Package InspectionIdentical testing performed on predicate device with equivalent results.Visually inspected packaging for damage, loose components, secure tip protector, and variable throw set at 22mm. Results were equivalent to predicate device.
Burst TestIdentical testing performed on predicate device with equivalent results.Measured pressure at which sealed tray fails. Results were equivalent to predicate device.
Device InspectionIdentical testing performed on predicate device with equivalent results.Visually inspected device for damage. Results were equivalent to predicate device.
Dimensional (Variable Throw)Identical testing performed on predicate device with equivalent results.Measured variable throw desired vs. actual. Results were equivalent to predicate device.
Dimensional (Variable Throw after Dry Fires)Identical testing performed on predicate device with equivalent results.Measured variable throw after dry fires, desired vs. actual. Results were equivalent to predicate device.
Force to Fire Front Triggers (RH / LH)Identical testing performed on predicate device with equivalent results.Measured force to fire front triggers. Results were equivalent to predicate device.
Force to Fire Rear TriggerIdentical testing performed on predicate device with equivalent results.Measured force to fire rear trigger. Results were equivalent to predicate device.
Tissue Test (Chicken and Liver)Identical testing performed on predicate device with equivalent results.Able to take tissue samples in chicken and liver. Results were equivalent to predicate device.
Tissue Test (Various Throw Distances)Additional testing that was not performed on predicate.Able to take tissue samples at various throw distances. (No direct comparison to predicate stated for this specific aspect, but overall tissue test performance confirmed equivalence). The document states identical testing performed on predicate device with equivalent results.
Tissue Test (1st vs. Last Sample Weight)Identical testing performed on predicate device with equivalent results.Measured 1st sample as compared to last sample for sample weight and verified no loss of effectiveness. Results were equivalent to predicate device.
Mates with Adjustable Coaxial and Able to Take SampleIdentical testing performed on predicate device with equivalent results.The device successfully mated with an adjustable coaxial and was able to take a sample. The results were equivalent to the predicate device.
Verify Weld Integrity (Visual Inspection after Functional Testing)Identical testing performed on predicate device with equivalent results.Visually inspected device for damage along weld joints after functional testing. Results were equivalent to predicate device.
Dimensional (Cannula, Spoon, and Stylet Migration)Identical testing performed on predicate device with equivalent results.Measured cannula, spoon, and stylet for migration. Results were equivalent to predicate device.
Dimensional (Joint Strength for Cannula, Spoon, and Stylet)Identical testing performed on predicate device with equivalent results.Measured the joint strength for the cannula, spoon, and stylet. Results were equivalent to predicate device.

2. Sample Size Used for the Test Set and Data Provenance:

The document does not explicitly state the specific number of devices or biopsy attempts used for each test. Instead, it broadly states that "Identical testing performed on predicate device w/ equivalent results" for most tests. The tissue tests mention "chicken and liver" and "various throw distances," indicating a lab-based, prospective evaluation with animal tissue. No country of origin for the data is specified, but given the submission to the FDA, it is presumed to be conducted by the manufacturer, Inrad, Inc., located in Kentwood, MI, USA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This information is not provided in the document. The performance testing appears to be based on objective physical measurements and visual inspections rather than expert interpretation of biopsy samples.

4. Adjudication Method for the Test Set:

Not applicable. The tests are focused on device performance metrics (e.g., dimensions, forces, ability to take tissue samples) rather than diagnostic accuracy that would require an adjudication process.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is a biopsy instrument, not an AI-powered diagnostic tool. The performance evaluation is based on engineering and basic tissue procurement capabilities.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This is a physical medical device, not an algorithm.

7. The Type of Ground Truth Used:

The ground truth for most tests was based on objective physical measurements (e.g., pressure, force, dimensions, weight) and visual inspections for damage or proper function. For tissue tests, the ground truth was the ability to successfully obtain a tissue sample and verify its properties (e.g., weight, no loss of effectiveness). The comparison was made against the performance of a legally marketed predicate device.

8. The Sample Size for the Training Set:

Not applicable. This device is not an AI/ML device requiring a training set. The performance testing is for a physical medical instrument.

9. How the Ground Truth for the Training Set was Established:

Not applicable. (See #8)

{0}------------------------------------------------

K112945

NOV - 3 2011

Image /page/0/Picture/2 description: The image shows the word "INRAD" in bold, black letters. To the left of the word is a black square with a white diagonal line running from the top left corner to the bottom right corner. A small "R" with a circle around it is in the top right corner of the image.

September 30, 2011

Contact Information

Ryan Goosen President Phone: 616-301-7800, ext. 105 Fax: 616-301-7799 E-mail: rgoosen@inrad-inc.com

Company Information

Inrad, Inc 4375 Donker Court SE Kentwood, MI 49512 Phone: 616-301-7800 616-301-7799 Fax:

Device Name(s)

EliteCore™ Full Core Biopsy Device EliteCore™ Full Core Biopsy Device w/HiLiter®

Device Summary

Trade or Proprietary Name:EliteCore™ Full Core Biopsy Device
EliteCore™ Full Core Biopsy Device w/HiLiter®
Common or Usual Name:Biopsy Instrument
Classification Name:Gastroenterology-urology biopsy instrument(21 CFR 876.1075, Product code KNW)

Name of Predicate(s) or Legally Marketed Device(s)

K090611– Revolution® Full Core Biopsy Device

{1}------------------------------------------------

Device Description

The EliteCore™ Full Core Biopsy Device is a sterile, disposable device which features a stainless steel cutting cannula with spoon, rotating coring cannula and stylet. The device is comprised of a plastic housing that contains the mechanically actuated components. The key performance attribute of the EliteCore™ Full Core Biopsy Device is the capability of obtaining a full core specimen. A variable throw feature allows the user to choose a throw setting ranging from 13 to 25 mm.

Indications for Use

The device is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lymph nodes and various soft tissue tumors.

Substantial Equivalence

The EliteCore™ Full Core Biopsy Device has the same intended use as the Revolution® Full Core Biopsy Device with the exception of the omission of the word breast. The device and the predicate device have the same technological characteristics in terms of design and materials.

Performance Testing Summary

Performance testing confirms that the quality of samples obtained with the EliteCore™ Full Core Biopsy Device is equivalent to that of the predicate device as shown below:

TestInvolvedHow Equivalent
-Package inspectionVisually inspect packagingfor damage, loose componentstip protector secure, and variablethrow set at 22mm.Identical testingperformed on predicatedevice w/ equivalentresults.
-Burst testMeasure pressure at whichsealed tray fails.Identical testingperformed on predicatedevice. w/ equivalentresults.
-Device inspectionVisually inspect devicefor damageIdentical testingperformed on predicatedevice w/ equivalentresults.

{2}------------------------------------------------

K112945

-Dimensional

-Tissue test

-Tissue test cont.

Measure variable throw desired vs. actual.

Measure variable throw after dry fires, desired vs. actual.

Measure force to fire front triggers (RH / LH)

Measure force to fire rear trigger.

Able to take tissue sample in chicken and liver.

Identical testing performed on predicate device w/ equivalent results.

Identical testing performed on predicate device w/ equivalent results.

Identical testing performed on predicate device w/ equivalent results.

Additional testing that was not performed on predicate.

Identical testing performed on predicate device w/ equivalent results.

Able to take tissue samples at various throw distances.

Measure 1st sample as compared

to last sample for sample weight

and verify no loss of effectiveness.

Identical testing performed on predicate device w/ equivalent results.

Identical testing performed on predicate device w/ equivalent results.

Mates with adjustable coaxial Identical testing And able to take sample. performed on predicate

Section 5 - 3

{3}------------------------------------------------

K//2945

device w/ equivalent results.

Identical testing. performed on predicate device w/ equivalent results.

Identical testing performed on predicate device w/ equivalent results.

ldentical testing performed on predicate device w/ equivalent results.

-Verify weld integrity

Visually inspect device for damage along weld joints after functional testing.

-Dimensional

Measure cannula, spoon, and stylet for migration.

Measure the joint strength for the cannula, spoon, and stylet.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three lines representing the staff and a snake winding around it. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Mr. Ryan Goosen President Inrad Inc. 4375 Donker CT SE GRAND RAPIDS MI 49512

NOV - 3 2011

Re: K112945

Trade/Device Name: EliteCore™ Full Core Biopsy Device EliteCore™ Full Core Biopsy Device w/HiLiter® Regulation Number: 21 CFR§ 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: II Product Code: KNW Dated: September 30, 2011 Received: October 4, 2011

Dear Mr. Goosen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{5}------------------------------------------------

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Huker Lemur MD

Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

SECTION 4: INDICATIONS FOR USE STATEMENT

Indications for Use

12945

510(k) Number (if known):

Device Name:

Indications for Use:

EliteCore™ Full Core Biopsy Device EliteCore™ Full Core Biopsy Device w/HiLiter®

The device is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lymph nodes and various soft tissue tumors.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

)Division Sloחס-ח(
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K112945

Section 4 - 1

§ 876.1075 Gastroenterology-urology biopsy instrument.

(a)
Identification. A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification. (1) Class II (performance standards).(2) Class I for the biopsy forceps cover and the non-electric biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.