K Number
K112881

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-05-22

(235 days)

Product Code
Regulation Number
882.1870
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nexstim Navigated Brain Stimulation System (NBS System) 4 is indicated for non-invasive mapping of the primary motor cortex of the brain to its cortical gyrus. The NBS System 4 provides information that may be used in the assessment of the primary motor cortex for pre-procedural planning.

Nexstim NEXSPEECH®, when used together with the Nexstim NBS System 4, is indicated for noninvasive localization of cortical areas that do not contain essential speech function. Nexstim NEXSPEECH® provides information that may be used in pre-surgical planning in patients undergoing brain surgery. Intraoperatively, the localization information provided by NEXSPEECH® is intended to be verified by direct cortical stimulation.

The Nexstim NBS System 4 and NBS System 4 with NEXSPEECH® are not intended to be used during a surgical procedure. The Nexstim NBS System 4 and NBS System 4 with NEXSPEECH® are intended to be used by trained clinical professionals.

Device Description

The NBS System 4 combines magnetic resonance imaging-based, three dimensional localization of cortical motor areas of the brain with noninvasive TMS and simultaneous electromyography measurement to locate areas of the brain that are capable of evoking muscle responses when stimulated.

NEXSPEECH® is intended to be used in conjunction with the NBS System 4 for localization and assessment of cortical areas of speech function for pre-procedural planning purposes.

AI/ML Overview

Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

Acceptance Criteria and Device Performance

The provided document focuses on establishing substantial equivalence to a predicate device for the Nexstim NBS System 4 and Nexstim NBS System 4 with NEXSPEECH®. The "acceptance criteria" are primarily demonstrated through conformance to various standards and the clinical study for NEXSPEECH®.

The primary "acceptance criteria" for NEXSPEECH® as a standalone function is its effectiveness in non-invasive localization of cortical areas that do not contain essential speech function, specifically assessed by its ability to identify negative language areas that are unlikely to show a response to direct cortical stimulation (DCS).

Table of Acceptance Criteria and Reported Device Performance (NEXSPEECH®-specific)

Acceptance Criterion (Clinical Effectiveness)Reported Device Performance (NEXSPEECH®)
Non-invasive localization of cortical areas that do not contain essential speech function for pre-surgical planning.High sensitivity for localizing language areas.
Relatively low specificity for localizing language areas.
Very high overall Negative Predictive Value (NPV) = 0.84.
Particularly high NPV in classical Broca area (NPV = 1.0).
Negative responses can be used to identify brain regions where DCS responses are unlikely to be obtained.
Acceptable safety profile for NEXSPEECH® use.No adverse events related to NEXSPEECH® use reported in the patient population.

Study Details

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: 20 patients (age 25-70, 10 females).
  • Data Provenance: Retrospective and prospective. The pre-operative Nexstim NBS with NEXSPEECH® mapping was conducted, and then compared to intraoperative DCS data from the same patients. The patients were recruited at departments of neurosurgery at 2 university hospitals (country of origin not explicitly stated, but Nexstim Oy is based in Helsinki, Finland).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Ground Truth Establishment: The ground truth was established by "intraoperative language mapping performed utilizing an object naming task and a simultaneous direct electrical stimulation of the exposed cortex to identify essential language areas." The DCS was performed "according to the normal clinical practice used at each study site" by neurosurgeons.
  • Number of Experts: Not explicitly stated, but it would involve at least the neurosurgeon(s) performing the DCS and potentially other clinical staff involved in the awake surgery and assessment.
  • Qualifications of Experts: Neurosurgeons performing Direct Cortical Stimulation (DCS). Specific experience levels are not detailed, but it's implied they are trained clinical professionals at university hospitals.

4. Adjudication method for the test set:

  • The text describes a "quantitative and qualitative analysis" comparing the locations of positive or negative language responses from NEXSPEECH® to those from DCS. There's no explicit mention of an adjudication panel for discrepant cases or a specific adjudication method like 2+1. The comparison appears to be direct between the two mapping methods.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • A classic MRMC comparative effectiveness study, where human readers evaluate cases with and without AI assistance to measure improvement, was not performed for NEXSPEECH®.
  • The study compared the device's output (NEXSPEECH® mapping) directly against an invasive gold standard (DCS), not against human interpretation of data with and without AI assistance. The device itself provides the "pre-procedural planning information," it doesn't assist a human reader in interpreting separate data.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • Yes, a standalone clinical study was performed for NEXSPEECH®. The study directly assessed the performance of the NEXSPEECH® module in localizing cortical areas relative to Direct Cortical Stimulation (DCS), which is considered the gold standard. The device generates the mapping information itself. The output of the NEXSPEECH® system (localization of speech areas) was compared to the output of DCS.

7. The type of ground truth used:

  • Expert Consensus / Gold Standard (Pathology/Clinical Outcomes): The primary ground truth for the NEXSPEECH® effectiveness study was Direct Cortical Stimulation (DCS) performed during awake surgery. DCS is considered the "gold standard for intraoperative language mapping." This is a clinical outcome from an invasive procedure.

8. The sample size for the training set:

  • The provided text does not specify a separate sample size for a training set. The clinical study described is the effectiveness study (test set). It's possible the device was developed and internally validated using other data, but those specifics are not in this document.

9. How the ground truth for the training set was established:

  • As no training set details are provided in this document, the method for establishing its ground truth is not mentioned.

{0}------------------------------------------------

K112881

MAY 2 2 2012

Submission Date:

29 September 2011

Submitter:

Nexstim Oy Elimaenkatu 9b 00510 Helsinki, Finland

Submitter and Official Contact: Mr. Rainer Harjunpää Director, Quality and Regulatory Affairs Nexstim Oy Elimaenkatu 9b 00510 Helsinki, Finland

+011 358 (9) 2727 1710

rainer.harjunpaa(@nexstim.com

Manufacturing Site:

Nexstim Oy Elimaenkatu 9b 00510 Helsinki, Finland

Trade Name:

Nexstim Navigational Brain Stimulation (NBS) System 4, and Nexstim NBS System 4 with NEXSPEECH®

Stimulator, electrical, evoked response Classification Name:

21 CFR §882.1870, 21 CFR §882.4560, 21 CFR §890.1375 Classification Regulation:

GWF, HAW, IKN Product Code:

SubstantiallyEquivalent Devices:New Nexstim ModelPredicate510(k) NumberPredicateManufacturer / Model
Nexstim NavigationalBrain Stimulation(NBS) System 4K091457Nexstim Oy / eXimiaNavigated BrainStimulation System
Nexstim NBS System 4with NEXSPEECH®K924226Radionics, Inc. /Ojemann CorticalStimulator
K091457Nexstim Oy / eXimiaNBS System

{1}------------------------------------------------

Device Description:

The NBS System 4 combines magnetic resonance imaging-based, three dimensional localization of cortical motor areas of the brain with noninvasive TMS and simultaneous electromyography measurement to locate areas of the brain that are capable of evoking muscle responses when stimulated.

NEXSPEECH® is intended to be used in conjunction with the NBS System 4 for localization and assessment of cortical areas of speech function for pre-procedural planning purposes.

Indications for Use:

The Nexstim Navigated Brain Stimulation System (NBS System) 4 is indicated for non-invasive mapping of the primary motor cortex of the brain to its cortical gyrus. The NBS System 4 provides information that may be used in the assessment of the primary motor cortex for preprocedural planning.

Nexstim NEXSPEECH®, when used together with the Nexstim NBS System 4, is indicated for noninvasive localization of cortical areas that do not contain essential speech function. Nexstim NEXSPEECH® provides information that may be used in pre-surgical planning in patients undergoing brain surgery. Intraoperatively, the localization information provided by NEXSPEECH® is intended to be verified by direct cortical stimulation.

The Nexstim NBS System 4 and NBS System 4 with NEXSPEECH® are not intended to be used during a surgical procedure. The Nexstim NBS System 4 and NBS System 4 with NEXSPEECH® are intended to be used by trained clinical professionals.

Technology Comparison: The Nexstim NBS System and NBS System with NEXSPEECH® employ the same technological characteristics as the predicate devices.

{2}------------------------------------------------

Summary of Performance Testing:

Biocompatibility

Patient contact materials which are part of the Nexstim NBS System 4 and NBS System 4 with NEXSPEECH® were designed to comply with the following standard:

  • ISO 10993-1: 2009, Biological evaluation of medical devices -. Part 1: Evaluation and testing within a risk management process,
    and were determined to be safe to use with patients.

Software Verification and Validation

Software for the Nexstim NBS System 4 and NBS System 4 with NEXSPEECH® was designed and developed according to a robust software development process, and were rigorously verified and validated.

Software information is provided in accordance with internal documentation and the following Standards and guidance documents:

  • FDA guidance: The content of premarket submissions for software e contained in medical devices, 11 May 05;
  • FDA guidance: Off-the-shelf software use in medical devices, 09 . Sen 99:
  • FDA guidance: General principles of software validation; Final . guidance for industry and FDA staff, 11 Jan 02;
  • IEC 60601-1-4: 2000, Medical electrical equipment Medical . electrical equipment - Part 1-4: General requirements for safety -Collateral Standard: Programmable electrical medical systems; and
  • IEC 62304: 2006. Medical device software Software life cycle . processes.

Test results indicate that the Nexstim NBS System 4 and NBS System 4 with NEXSPEECH® comply with its predetermined specifications, and the Standards and guidance documents.

{3}------------------------------------------------

Electrical SafetyThe Nexstim NBS System 4 and NBS System 4 with NEXSPEECH®were tested for performance in accordance with the followingStandards:IEC 60601-1: 1988, Aml: 1991, Am2: 1995, Medical electricalequipment - Part 1: General requirements for safety; IEC 60601-1-1: 2000, Medical electrical equipment - Part 1-1:General requirements for safety - Collateral standard: Safetyrequirements for medical electrical systems; IEC 60950-1: 2005, Information technology equipment – Safety -Part 1: General requirements; and UL 2601-1: 1997, Medical electrical equipment - Part 1: Generalrequirements for safety. Test results indicated that the Nexstim NBS System 4 and NBS System4 with NEXSPEECH® comply with the Standards.
ElectromagneticCompatibility (EMC)TestingThe Nexstim NBS System 4 and NBS System 4 with NEXSPEECH®was tested for performance in accordance with the following Standard:IEC 60601-1-2: 2007, Medical electrical equipment – Part 1-2:General requirements for basic safety and essential performance –Collateral standard: Electromagnetic compatibility – Requirementsand tests. Test results indicated that the Nexstim NBS System 4 and NBS System4 with NEXSPEECH® comply with the Standards.
Performance Testing– BenchThe Nexstim NBS System 4 and NBS System 4 with NEXSPEECH®were tested for performance in accordance with internal documentationand the following FDA Guidance Documents and Standards:Guidance for Industry and Food and Drug Administration StaffClass II Special Controls Guidance Document: RepetitiveTranscranial Magnetic Stimulation (rTMS) Systems; and IEC 60601-2-40:1998, Medical electrical equipment - Part 2-40:Particular requirements for the safety of electromyographs andevoked response equipment.

Test results indicated that the Nexstim NBS System 4 and NBS System 4 with NEXSPEECH® comply with its predetermined specification and the applicable Standard.

{4}------------------------------------------------

The Nexstim NBS System 4 and NBS System 4 with NEXSPEECH® Performance Testing were tested for usability in accordance with the following Standards: - Usability IEC 60601-1-6: 2010, Medical electrical equipment – Part 1-6: . General requirements for basic safety and essential performance -Collateral standard: Usability; and IEC 62366: 2007, Medical devices – Application of usability . engineering to medical devices. Test results indicated that the Nexstim NBS System 4 and NBS System 4 with NEXSPEECH® comply with the applicable Standards.

The clinical performance of the Nexstim NBS System 4 and NBS Performance Testing System 4 with NEXSPEECH® was established by literature review and appropriate clinical study and validation testing.

The test results indicated that the Nexstim NBS System 4 is safe and effective for non-invasive mapping of the primary motor cortex of the brain to its cortical gyrus. Furthermore, the test results indicated that the Nexstim NBS System 4 with NEXSPEECH® is safe and effective for noninvasive localization of cortical areas that do not contain essential speech function.

NEXSPEECH® Clinical Study Description

Nexstim has conducted a clinical trial to assess the safety and to establish the effectiveness of Nexstim Navigated Brain Stimulation (NBS) System with NEXSPEECH when used for preoperative mapping of essential language cortex that could be at risk during a surgical procedure. As the gold standard for intraoperative language mapping, direct cortical stimulation (DCS) performed during an awake surgery was selected as the reference method for the comparison (Tharin, Golby 2007). In the study, the mapping results obtained using Nexstim NBS System with NEXSPEECH module were compared to those obtained by invasive DCS performed during awake surgery.

Preoperative speech mapping with NEXSPEECH was performed utilizing an object naming task time-locked to a rapid-rate transcranial magnetic stimulation train to non-invasively create a transient virtual lesion and thereby identify essential language areas (Lioumis 2012). Intraoperative language mapping was performed utilizing an object naming task and a simultaneous direct electrical stimulation of the exposed cortex to identify essential language areas (Picht 2006). The DCS was performed according to the normal clinical practice used at each study site.

  • Clinical

Performance Testing

  • Clinical

{5}------------------------------------------------

Subject Population

Speech mapping with Nexstim NBS with NEXSPEECH and intraoperative cortical stimulation was performed in 20 patients (age 25-70, 10 females) with a lesion requiring surgical intervention in the vicinity of essential language areas recruited at departments of neurosurgery at 2 university hospitals.

Comparison to Reference Method

The locations of areas relevant to speech production were determined prior to surgery non-invasively with Nexstim NBS System and NEXSPEECH. The patients were subsequently operated by a neurosurgeon that performed a language mapping with DCS during an awake surgery procedure.

Locations with positive or negative language response during direct cortical stimulation in the awake surgery were compared to those found in mapping with Nexstim NBS and NEXSPEECH module. A quantitative and qualitative analysis was carried out to evaluate the concordance of results obtained with both methods.

Localization Assessment

In order to evaluate the anatomical location of stimulation effects, each individual brain was divided into 37 individually named anatomical regions (Corina 2010). The NBS data and intraoperative DCS mapping data were both projected on the brain surface divided in this manner. See Figure 1 for schematic representation of the anatomical areas.

Image /page/5/Figure/8 description: The image shows a diagram of the lateral surface of the human brain, with various regions labeled. The diagram includes labels such as PSFG, DPrG, MSFG, ASFG, PMFG, and MMFG in the frontal lobe. Other labeled regions include the parietal lobe (SPL, AnG), temporal lobe (MSTG, MMTG, MITG), and occipital lobe (DLOG, VLOG).

Figure 1. Schematic representation of the anatomical areas used for assessing concordance of speech area localization by NBS NEXSPEECH and DCS.

{6}------------------------------------------------

NexSpeech® Clinicial Study Conclusions

Safety

No adverse events related to NEXSPEECH use were reported in the patient population.

Effectiveness

The results of the present clinical trial demonstrate that NEXSPEECH has high sensitivity but relatively low specificity for localizing language areas. NEXSPEECH mapping has a very high overall negative predictive value everywhere in the brain (NPV=0.84), and particularly so in classical Broca (NPV = 1.0)

Due to the high sensitivity of NEXSPEECH in obtaining speech responses. negative responses obtained with NEXSPEECH can be used to identify brain regions where DCS responses are unlikely to be obtained. This information on negative speech locations can be utilized for planning procedures. With this strategy, any positive NEXSPEECH responses should be verified by intraoperative DCS.

Nexstim concludes that the clinical performance of NEXSPEECH speech mapping is sufficient for noninvasive localization of cortical areas that do not contain essential speech function. Intraoneratively, the localization information provided by NEXSPEECH® should be verified by direct cortical stimulation.

References

Corina DP, Loudermilk BC, Detwiler L, Martin RF, Brinkley JF, Ojemann G. Analysis of naming errors during cortical stimulation mapping: implications for models of language representation. Brain Lang. 2010 Nov;115(2):101-12.

Lioumis P. Zhdanov A. Mäkelä N. Lehtinen H. Wilenius J. Neuvonen T, Hannula H, Deletis V, Picht T, Mäkelä JP. A novel approach for documenting naming errors induced by navigated transcranial magnetic stimulation. J Neurosci Methods. 2012 Mar 15;204(2):349-54.

Picht, T., Kombos, T., Gramm, H.J., Brock, M., Suess, O., 2006. Multimodal protocol for awake craniotomy in language cortex tumour surgery, Acta Neurochir (Wien) 148, 127-137; discussion 137-128.

Tharin S and Golby A. Functional brain mapping and its applications to neurosurgery. Neurosurgery 60:185-201, 2007

{7}------------------------------------------------

Conclusion

Verification and validation activities were conducted to establish the performance and safety characteristics of the Nexstim NBS System 4 and Nexstim NBS System 4 with NEXSPEECH®. The results of these activities demonstrate that the Nexstim NBS System 4 and Nexstim NBS System 4 with NEXSPEECH® is safe and effective when used in accordance with its intended use and labeling.

Therefore, the Nexstim NBS System 4 and NBS System 4 with NEXSPEECH® are considered substantially equivalent to the predicate device.

{8}------------------------------------------------

NDC C て

Comparison of Device Specifications – NBS System 4
CharacteristicNexstim Oy eXimia NBS System (K091457)Nexstim Oy NBS System 4
Intended useThe Nexstim eXimia Navigated Brain Stimulation System (NBSSystem) is indicated for non-invasive mapping of the primarymotor cortex of the brain to its cortical gyrus. The NBS Systemprovides information that may be used in the assessment of theprimary motor cortex for pre-procedural planning.The NBS System is not intended to be used during a surgicalprocedure.The NBS System is intended to be used by trained clinicalprofessionals.The Nexstim Navigated Brain Stimulation System (NBS System) isindicated for non-invasive mapping of the primary motor cortex ofthe brain to its cortical gyrus. The NBS System providesinformation that may be used in the assessment of the primarymotor cortex for pre-procedural planning.The Nexstim NBS System is not intended to be used during asurgical procedure.The Nexstim NBS System is intended to be used by trained clinicalprofessionals.
Tracking system accuracy1.6 mm (mean error in localization of the tool)1.6 mm (mean error in localization of the tool)
Navigation principleBased on anatomy (MRI picture) and calculated electric fieldBased on anatomy (MRI picture) and calculated electric field
TMS stimulus modeSingle pulseSingle pulse
Coil configurationFigure 8-shaped focal coilFigure 8-shape focal and cooled focal coils
Maximum applicatorsurface temperatureFocal coil: 41 °CFocal coil: 41 °CCooled focal coil: 43 °C
Stimulation OutputMaximum electric field ~150 Volts/meter (V/m) measured 25 mmbelow the coil.Maximum electric field ~172 V/m measured 25 mm below the coil.
Maximum output in motorthreshold (MT) units~2.2 times the motor threshold (MT) of a healthy adult hand musclewhere 1 MT corresponds to 45% stimulation intensity.~2.5 times the MT of a healthy adult hand muscle where 1 MTcorresponds to 40% stimulation intensity.
Current waveformBiphasicBiphasic
Pulse length~280 μs~230 μs
Trigger input/output5V TTL input and output5V TTL output
EMG assessmentMuscle response to stimulation measured with eXimia EMGMuscle response to stimulation measured with eXimia EMG
Number of EMG channelsUp to 6 channelsUp to 6 channels
EMG measurement range10 - 500 Hz10 - 500 Hz
Sampling rate3 kHz3 kHz
CharacteristicNexstim Oy eXimia NBS System (K091457)Nexstim Oy NBS System 4
Common mode rejection ratio (CMRR)> 90 dB at 10 - 250 Hz> 90 dB at 10 - 250 Hz
NBS system with TMS electrical ratingVoltage: 100-120 Vac220-240 VacFrequency: 50/60 HzVoltage: 100-120 Vac220-240 VacFrequency: 50/60 Hz
EMG electrical ratingVoltage: 120-240 VacFrequency: 50/60 HzVoltage: 120-240 VacFrequency: 50/60 Hz
WeighteXimia weight: 300 kgTMS weight: 140kgFloor stand: 60kgTotal weight: 223 kgNBS System 4 has TMS II stimulator housed in the same cart with the whole NBS system.
DimensionsTMS cabinet: 72 cm (width) x 60 cm (depth) x 98 cm (height)NBS System cart: 80 cm (width) x 70 cm (depth) x 120 cm (height)All housed in the NBS System 4 cart: 80 cm (width) x 70 cm (depth) x 209 cm (max. height)

Page 11 of 13

{9}------------------------------------------------

Page 12 of 13

{10}------------------------------------------------

0(k) Summ

Stocedure.
The Nexstim NBS System with NEXSPEE
ntended to be used by trained clinical
professionals. exstim NBS System with NEXSPE
ended to be used during a surgical Response to stimulation measured with
CMT for for determination in in in in in in in in in in in in in in in in mana
Response qualificanse by sponse of and the films for
offexstim Oy NBS System 4 v
NEXSPEECH® -10 Hz in 0,1 Hz st inguage are Yes mparison of Device Specifications - Nexstim NBS System 4 with NEXSPEE es
requency selectable from 5、10、2
7 and 100Hz mapping: locate sensorime
and descending montor protocorporations
it the safest online the total to the same
the safest some integral profile to the the sorimotor cortex and descen
or pathways. egra Radionics Inc. Oje.
ortical Stimulator (K924 ch mapping: identify esser
hage sites to maximise thise thinks
umour resection and minim
anent language deficits. d for intraoperative con
tion mapping to aid in
ns in the vicinity of es le response measured w
hal/3rd Party EMG. Resp tial language si ential cortex (es ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------exstim Oy eXimia Navigated
Stimulation System (K09145 ally trigged single pulses (no r Iotor corte ട് IS mode support Stimulation respons
detection Area of brain to be
stimulated rTMS frequenc
range haracteristic ntended use

Page 13 of

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Nexstim Ov c/o Mr. Rainer Harjunpää Director of Ouality and Regulatory Affairs Elimäenkatu 9B. FI-00510 Helsinki. Finland

MAY 2 2 2017

Re: K112881

Trade/Device Name: Nexstim Navigated Brain Stimulation (NBS) System 4, Nexstim NBS System 4 with NEXSPEECH (TM) Regulation Number: 21 CFR 882.1870 Regulation Name: Evoked Response Electrical Stimulator Regulatory Class: Class II Product Code: GWF, HAW, IKN Dated: May 3, 2012 Received: May 10, 2012

Dear Mr. Harjunpää:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{12}------------------------------------------------

Page 2 - Mr. Rainer Harjunpää

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Erh. Rur, mo

Malvina B. Eydelman, A Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{13}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name:

Indications for Use:

K112881

Nexstim Navigated Brain Stimulation (NBS) System 4 and NBS System 4 with NEXSPEECH®

The Nexstim Navigated Brain Stimulation System (NBS System) 4 is indicated for non-invasive mapping of the primary motor cortex of the brain to its cortical gyrus. The NBS System 4 provides information that may be used in the assessment of the primary motor cortex for pre-procedural planning.

Nexstim NEXSPEECH®, when used together with the Nexstim NBS System 4, is indicated for noninvasive localization of cortical areas that do not contain essential speech function. Nexstim NEXSPEECH® provides information that may be used in presurgical planning in patients undergoing brain surgery. Intraoperatively, the localization information provided by NEXSPEECH® is intended to be verified by direct cortical stimulation.

The Nexstim NBS System 4 and NBS System 4 with NEXSPEECH® are not intended to be used during a surgical procedure. The Nexstim NBS System 4 and NBS System 4 with NEXSPEECH® are intended to be used by trained clinical professionals.

AND/OR . Over-The-Counter Use Prescription Use X (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Kristen Bowsher

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

K112881 510(k) Number_

§ 882.1870 Evoked response electrical stimulator.

(a)
Identification. An evoked response electrical stimulator is a device used to apply an electrical stimulus to a patient by means of skin electrodes for the purpose of measuring the evoked response.(b)
Classification. Class II (performance standards).