K Number
K112863
Date Cleared
2012-03-29

(181 days)

Product Code
Regulation Number
876.5980
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

To deliver liquid nutrition formula from storage container to an enteral access device (feeding tube)

Device Description

Acta Enteral Sets are constructed from flexible, medical grade, Non-DEHP tubing and have various different connectors at each end of the tubing. One configuration may include an Enteral solution container constructed from same PVC material as the tubing with a Christmas tree connector to connect to a gastric tube (or an extension set) or another configuration may simply be an extension set which connects to patient gastric tube to the nutrition liquid delivery system. The following three configurations are most common, however, other configurations may be manufactured based on customer demand. (Further details, photos, drawings provided in Device Description, Section XI)

  1. A complete solution delivery system: This consists of, a solution container, integrated delivery set and a Christmas tree connector to connect to the gastric tube or to an extension set
  2. Enteral Administration Set: This consists of a delivery set with a spike to connect to a nutrition solution storage container and a Christmas tree connector to connect to gastric tube or an extension set
  3. Enteral Extension Set: This consists of flexible PVC tubing set with proximal connector which can accept a standard Christmas tree connector and a distal connector designed to fit a custom gastric tube
AI/ML Overview

The provided text describes a medical device submission (K112863) for an "Enteral Set." This document focuses on establishing substantial equivalence to predicate devices, rather than presenting a study with acceptance criteria and device performance results in the way a clinical trial or algorithm performance study would.

Therefore, many of the requested sections (e.g., sample size for test set, number of experts for ground truth, MRMC study, standalone performance) are not applicable or cannot be extracted from this type of regulatory submission. The document primarily details the device's design, intended use, and material testing to demonstrate equivalence.

Here's a breakdown of the information that can be extracted, and where the information is not present:

1. Table of Acceptance Criteria and the Reported Device Performance

Acceptance CriteriaReported Device Performance
BiocompatibilityMeets all the requirements for biocompatibility (ISO 10993-1)
PerformanceMeets applicable performance criteria (ISO 8536-4) for leak test and tensile strength.
DEHP ContentTested for DEHP content. (Result not quantitatively specified, but implied to meet requirements for "Non-DEHP tubing" as stated in Device Description)
ExtractablesTested for extractables. (Result not quantitatively specified, but implied to meet requirements)
LeachablesTested for leachables. (Result not quantitatively specified, but implied to meet requirements)
Physico-chemicalTested for physico-chemical properties for plastics. (Result not quantitatively specified, but implied to meet requirements)

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not specified. The document refers to general testing of construction materials and manufactured sets, not a specific clinical or performance test set size.
  • Data Provenance: Not specified for the performance tests. The manufacturing location is listed as Yangzhou City, Jiangsu Province, China.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
Not applicable. This document does not describe a study involving expert-established ground truth for a diagnostic or AI-based device. It concerns a physical medical device (enteral set).

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI or diagnostic imaging device.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an AI or algorithm-based device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for this device's performance appears to be based on compliance with established international standards for medical devices and materials:

  • ISO 10993-1 for biocompatibility.
  • ISO 8536-4 for specific performance criteria (leak test, tensile strength).
  • Chemical testing for DEHP, extractables, leachables, and physico-chemical properties.

8. The sample size for the training set
Not applicable. This is not a machine learning or AI device.

9. How the ground truth for the training set was established
Not applicable.

{0}------------------------------------------------

K112863 Page 1 of 3

MAR 29 2012

Acta Medical, LLC

Date Submitted: Submitted By:

September 28th, 2011 Acta Medical, LLC Inc 15 Minneakoning Road, Suite 203 Tel: 908-399-0279 Fax:908-788-9788

Submitter Contact:

Aaron Compton Acta Medical, LLC Contact email: acatutd@satx.rr.com Tel:214-533-1416 Fax:972-638-8549

Common Name of Device: Enteral Set

Predicate Device:

Promedic Pediatric Enteral feeding tubes and Extension sets (K092628) Dynarex Enteral Sets For Gravity and Pump Use(K082863) Truecare Biomedix Intravascular Administration Set (K111351)

Tubes, Gastrointestinal (and accessories)

Panel:

Product Code:

,

KNT

Device Classification: Class II

Classification Reference: 21CFR 876.5980

Appropriate Classification Panel: Gastroenterology/Urology

Manufacturing Location: Yangzhou Wei De Li Trade Co. Ltd.

Li Xin Bridge Touqiao Township. Yangzhou City, Jiangsu Province. China Tel: +86-514-87897887 Fax: +86-514-87889967

{1}------------------------------------------------

K11 2063 Page 2/3

Name & Model Numbers of Devices:

    1. Enteral Administration Set with Solution Container, 1200ml capacity (VDENT001)
    1. Enteral Administration Set (VDENTS001)
    1. Enteral Extension Set with Minitube Connector(VDENT8-1222)

Device Description:

Acta Enteral Sets are constructed from flexible, medical grade, Non-DEHP tubing and have various different connectors at each end of the tubing. One configuration may include an Enteral solution container constructed from same PVC material as the tubing with a Christmas tree connector to connect to a gastric tube (or an extension set) or another configuration may simply be an extension set which connects to patient gastric tube to the nutrition liquid delivery system. The following three configurations are most common, however, other configurations may be manufactured based on customer demand. (Further details, photos, drawings provided in Device Description, Section XI)

    1. A complete solution delivery system: This consists of, a solution container, integrated delivery set and a Christmas tree connector to connect to the gastric tube or to an extension set
    1. Enteral Administration Set: This consists of a delivery set with a spike to connect to a nutrition solution storage container and a Christmas tree connector to connect to gastric tube or an extension set
    1. Enteral Extension Set: This consists of flexible PVC tubing set with proximal connector which can accept a standard Christmas tree connector and a distal connector designed to fit a custom gastric tube

Device Intended Use: Enteral Set

To deliver liquid nutrition formula from storage container to an enteral access device (feeding tube)

Substantial Equivalence:

Acta Medical, LLC Enteral Set is substantially equivalent to predicate devices as follows:

{2}------------------------------------------------

K112863 Page 3/3

A complete solution delivery system and Enteral Administration Set is substantially equivalent to predicate device Dynarex Enteral Set for Gravity and Pump Use (K082863) in its design, intended use, function, materials, performance

Enteral Extension Set is substantially equivalent to Promedic, Pediatric Enteral Feeding tubes and Extension sets (K092628) in its design, intended use, function, materials, performance

Enteral Set is substantially equivalent to Truecare Biomedix Intravascular administration set (K111351) in its materials and manufacturer. Enteral Sets are constructed from same materials (non-DEHP PVC, Silicone, ABS) and manufactured by same manufacturer (Yangzhou Wei De Li) at the same manufacturing location as K111351,

Acta Medical, LLC Enteral Sets construction materials have been tested for biocompatibility, DEHP, extractables and leachables, physico chemical tests for plastics. Furthermore, the manufactured sets have been tested per ISO 8536-4 for leak test, tensile strength. Acta Medical, LLC Enteral Sets meets all the requirements for biocompatibility (ISO 10993-1) and applicable performance criteria (ISO 8536-4)

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes on its wing, symbolizing health, services, and human needs. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Aaron Compton Acta Medical L.L.C. 929 Arbor Downs Drive PLANO TX 75023

MAR 2 9 2012

Re: K112863

Trade/Device Name: Enteral Set Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: KNT Dated: March 22, 2012 Received: March 22, 2012

Dear Mr. Compton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{4}------------------------------------------------

Page 2 -

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerély wours.

Benjamin K. Eviko

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K112863

Device Name: Enteral Set

Indications For Use: To deliver liquid nutrition formula from storage container to an enteral access device (feeding tube)

Prescription Use _ x Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
---------------------

Division of Reproductive, Gastro-Renal, and
Urological Devices

510(k) NumberK112863
Page1 of 1

§ 876.5980 Gastrointestinal tube and accessories.

(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.