K Number
K112815

Validate with FDA (Live)

Date Cleared
2012-02-29

(155 days)

Product Code
Regulation Number
890.3860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of the Pride Mobility Products Corporation Quantum Bariatric 451 is to provide mobility to persons limited to a seated position that have the capability of operating a Powered Wheelchair.

Device Description

The Quantum Bariatric 451 is a Powered Wheelchair having a digital controller, electrical system, motors, batteries, seating, and frame. The Quantum Bariatric 451 is equipped with electronic regenerative disc brakes, 8A off-board battery charger, removable 12 volt batteries, front anti-tip wheels, and rear caster wheels.

The Quantum Bariatric 451 is designed with ultimate safety, stability, and performance in mind. The Powered Wheelchair is designed for, but not limited to Pride Mobility Products Corporation, providers/retailers and their consumers.

AI/ML Overview

This document primarily details the substantial equivalence of the Quantum Bariatric 451 powered wheelchair to a predicate device and its compliance with relevant testing standards, rather than presenting a study proving a device meets specific acceptance criteria based on performance metrics such as sensitivity, specificity, or accuracy.

Therefore, many of the requested elements for describing an acceptance criteria study, particularly those related to diagnostic performance or AI model validation, are not applicable or cannot be extracted from this document.

However, I can provide information based on the document's content:

1. A table of acceptance criteria and the reported device performance

Based on the document, the "acceptance criteria" are compliance with established testing standards and demonstrating substantial equivalence to a predicate device. The "device performance" in this context refers to successfully meeting these standards and exhibiting similar characteristics to the predicate device.

Acceptance Criteria CategorySpecific Acceptance Criteria (Standards / Equivalence)Reported Device Performance
Safety & Performance StandardsRESNA WC Vol.1 2009 - Requirements and Test Methods for WheelchairsCompliance
RESNA WC Vol. 2 2009 - Additional Requirements for Wheelchairs with Electrical SystemsCompliance
ANSI/RESNA WC Vol. 2-2008 Section 21 - Requirements and Test Methods for Electromagnetic CompatibilityCompliance
CAL 117 - Flammability TestingCompliance
BS EN1021- Furniture - Assessment of the Ignitability of Upholstered Furniture, Part 2Compliance
ISO 10993 - Biological evaluation of medical devicesCompliance
Substantial EquivalencePerformance, maneuverability, stability, and structure compared to predicate (Pride Mobility Jazzy Frontie - K092961)"Substantially equivalent" with similar performance characteristics and mechanisms for intended use and stability. Differences in weight capacity, drive wheels, and dimensions do not raise new safety/effectiveness questions.

2. Sample size used for the test set and the data provenance

  • Sample Size: The document does not specify a "sample size" in the context of a test set for performance metrics the way one would for an AI or diagnostic device. The testing refers to engineering and safety standards applied to the device itself.
  • Data Provenance: Not applicable in the context of clinical or AI data. The "data" refers to the results of non-clinical engineering and safety tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. The ground truth for this type of device (powered wheelchair) is established by adherence to recognized engineering, safety, and flammability standards, and comparison to existing legally marketed predicate devices, rather than expert interpretation of a test set.

4. Adjudication method for the test set

Not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies for discrepancies in expert opinion on ground truth. This document concerns engineering and safety compliance.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This document is about a powered wheelchair, not an AI or diagnostic device that would involve human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This concept is not applicable to a powered wheelchair. The device itself is a "standalone" product in the sense that it operates, but not in the context of an algorithm's performance without human interaction.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" here is compliance with established international and national standards for medical devices and wheelchairs, along with the mechanical and functional characteristics of the predicate device (Jazzy Frontie K092961). It's essentially engineering standards compliance and predicate device comparison.

8. The sample size for the training set

Not applicable. This device does not involve a "training set" in the context of machine learning or AI.

9. How the ground truth for the training set was established

Not applicable, as no training set (for AI) is involved.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Pride Mobility Products Corp. The word "Pride" is written in a bold, sans-serif font, with the "i" dotted with a circle. The words "Mobility Products Corp." are written in a smaller, sans-serif font below the word "Pride". The logo is black and white.

K112815

FEB 2 9 2012

182 Susquehanna Ave Exeter, PA 18643 570-655-5574 FAX 655-2990

Exhibit 1 510(k) Summary Pride Mobility Products Corporation Quantum Bariatric 451

Submitter's Name & Address:

Pride Mobility Products Corporation 182 Susquehanna Avenue Exeter, Pa. 18643 Phone: (570) 655-5574 Facsimile: (570) 655-1470

Contact Person:

Kimberly Blake

Date Updated:

12/12/2011

Name of Device and Proprietary Name: Quantum Bariatric 451

Common or Usual Name: Powered Wheelchair

Classification Name:

Physical Medicine / Powered Wheelchair

Product Code:

ITI

Comparison to Predicate Devices:

The Quantum Bariatric 451 is substantially equivalent to the Pride Mobility Jazzy Frontie (K092961), when comparing performance, maneuverability, stability, and structure. The performance characteristics and the position of the electronics and drive mechanisms are similar to achieve the same intended use function that enables the user to maintain optimum stability without hindering performance. Please refer to Exhibits 4A, 7, 7A, and additional information provided for comparisons and testing data.

The major differences between the Quantum Bariatric 451 and the Jazzy Frontie (K092961) are in the weight capacity, drive wheels, and dimensions:

  • Jazzy Frontie weight capacity is 300 lbs versus 600 lbs for the Bariatric . 451.
    FDA/1400-5 Rev. 0 Date: 4/23/10 App'd: K. Blake

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Pride Mobility Products Corp. The word "Pride" is written in a bold, stylized font, with a registered trademark symbol next to it. Below "Pride" is the text "Mobility Products Corp."

  • Jazzy Frontie has 9" drive wheels versus 14" on the Bariatric 451. �
  • Jazzy Frontie is smaller dimensionally (34.75" L and 22.75" W) versus . the larger Bariatric 451 (46.6" L, 29" W).

Device Description:

The Quantum Bariatric 451 is a Powered Wheelchair having a digital controller, electrical system, motors, batteries, seating, and frame. The Quantum Bariatric 451 is equipped with electronic regenerative disc brakes, 8A off-board battery charger, removable 12 volt batteries, front anti-tip wheels, and rear caster wheels.

The Quantum Bariatric 451 is designed with ultimate safety, stability, and performance in mind. The Powered Wheelchair is designed for, but not limited to Pride Mobility Products Corporation, providers/retailers and their consumers.

Intended Use:

The intended use of the Pride Mobility Products device is to provide mobility to persons limited to a seated position that have the capability of operating a Powered Wheelchair.

Non-Clinical Testing:

Compliance to applicable Testing Standards is as follows (Refer to 7F for FDA-3654): RESNA WC Vol.1 2009 - Requirements and Test Methods for Wheelchairs (Including Scooters)

RESNA WC Vol. 2 2009 - Additional Requirements for Wheelchairs (Including Scooters) with Electrical Systems

ANSI/RESNA WC Vol. 2-2008 Section 21 - Requirements and Test Methods for Electromagnetic Compatibility.

CAL 117 - Flammability Testing

BS EN1021- Furniture - Assessment of the Ignitability of Upholstered Furniture, Part 2 ISO 10993 -Biological evaluation of medical devices

Discussion of Clinical Testing Performed: N/A

Conclusions:

The Quantum Bariatric 451 Powered Wheelchair has the same intended use and similar technological characteristics as the Jazzy Frontie (K092961), moreover, the nonclinical testing and the predicate comparisons demonstrate that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Thus, the Quantum Bariatric 451 is substantially equivalent to the predicate device, has passed all the necessary testing, and is considered to be safe for user operation.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is facing to the right, and its wings are curved upwards. The logo is simple and clean, and it is easily recognizable.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Pride Mobility Products Corporation % Ms. Kimberly Blake 182 Susquehanna Avenue Exeter, Pennsylvania 18643

FEB 2 9 2012

Re: K112815

Trade/Device Name: Quantum Bariatric 451 Regulation Number: 21 CFR 890. Regulation Name: Powered wheelchair Regulatory Class: Class II Product Code: ITI Dated: February 3, 2012 Received: February 3, 2012

Dear Ms. Blake:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Kimberly Blake

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you uesire specific advice tor your downters/CDRHOffices/ucm115809.htm for go to mtb.//www.ida.gov//rbout/Drical Health's (CDRH's) Office of Compliance. Also, please the Center for Devices and readorogies. In a lease to premarket notification" (21CFR Part note the regulation childed, "Misolanding of corrects under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Tou may obtain other general mionnations one Jone Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson

Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Pride Mobility Products Corp. The word "Pride" is written in a bold, sans-serif font, with the registered trademark symbol next to the "e". Below the word "Pride" is the text "Mobility Products Corp." in a smaller, sans-serif font.

Exhibit 3

Indications for Use

510(k) Number (if known): K

Device Name: Quantum Bariatric 451

Indications for Use: The intended use of the Pride Mobility Products Indications "Tor" OSC." The "MC1 is to provide mobility to persons limited to a Corporation Quantain Banather of operating a Powered Wheelchair.

Prescription Use_ X (Part 21 CFR 801 Subpart D)

AND / OR

Over-The-Counter Use × (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Evaluation (ODE) Concurrence of CDRH, Office of

(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K112815

§ 890.3860 Powered wheelchair.

(a)
Identification. A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.(b)
Classification. Class II (performance standards).