K Number
K112799

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-12-23

(87 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MED-RX Extension Set with T-Connector is intended to be used for administration or withdrawal of fluids to or from a neonatal patient as directed by a physician.

Device Description

The MED-RX Extension Set with T-Connector is offered in two configurations, identical except for the tubing length. The proximal end of the MED-RX Extension Set with T-Connector features a T-Connector with an injection port and an integral rotating male luer lock. The distal end has a female luer lock fitting. The inner and outer diameter of both devices is consistent at 0.040" and 0.095" respectively, while the lengths available are either 4" or 38". Each of the luer fittings (distal female luer lock and proximal rotating male luer lock) are provided with protective caps. The MED-RX Extension Set with T-Connector is provided sterile and is single use.

AI/ML Overview

This document describes the premarket notification for the "MED-RX Extension Set with T-Connector." The provided text focuses on demonstrating substantial equivalence to a predicate device and does not involve AI or algorithms, human readers, or image analysis. Therefore, many of the requested categories related to AI performance are not applicable.

Here's an analysis of the provided information based on the request:

1. Table of Acceptance Criteria and Reported Device Performance

Test DescriptionStandardAcceptance Criteria (Implied by "Pass")Reported Device Performance
Tensile Strength - Tubing/T-ConnectorISO 8536-4:2007PassPass
Tensile Strength - Tubing/Female Luer LockISO 8536-4:2007PassPass
Leakage under PressureISO 8536-4:2007PassPass
Liquid LeakageISO 8536-4:2007PassPass
Particulate ContaminationISO 8536-4:2007Samples meet contamination index limitSamples met contamination index limit
Chemical RequirementsISO 8536-4:2007 (Clauses 5 & 7)PassPass
Natural Rubber Latex ContentModified Lowry MethodDevice does not contain natural rubber latexDevice does not contain natural rubber latex
Sterilization Method of ValidationANSI/AMMI/ISO 11135-1: 2007Validated to a Sterility Assurance Level of 1 x 10^-6Validated to a Sterility Assurance Level of 1 x 10^-6
EO Sterilization ResidualsISO 10993-7: 2008PassPass
Bacterial EndotoxinsANSI/AAMI ST72: 2002PassPass
ISO MEM Elution with L-929 Mouse Fibroblast Cells (Cytotoxicity)ISO 10993-5: 2009Considered non-toxicProduct code 10-1053XLU is considered non-toxic.
ISO Guinea Pig Maximization Sensitization TestISO 10993-10:2002Did not elicit a sensitization responseProduct code 10-1053XLU did not elicit a sensitization response.
ISO Intracutaneous Reactivity TestISO 10993-10:2002Considered a non-irritantProduct code 10-1053XLU would be considered a non-irritant.
ISO Acute Systemic Injection TestISO 10993-11: 2006Requirements of the test metThe findings indicate that the requirements of the ISO Acute Systemic Injection Test have been met.
ASTM Hemolysis Assay - Extract MethodASTM F-756-00Considered non-hemolytic and passesProduct code 10-1053XLU is considered non-hemolytic and passes the test.
Materials Mediated Rabbit Pyrogen TestUSP 32: 2009 <151>Determined to be non-pyrogenicProduct code 10-1053XLU is determined to be non-pyrogenic.

2. Sample size used for the test set and the data provenance
The document does not specify the exact sample sizes (number of units) used for each non-clinical test. The data provenance is internal testing performed by Benlan Inc., as part of their 510(k) submission to the FDA. This is likely retrospective testing for regulatory submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is a medical device in vitro testing, not an AI or diagnostic tool requiring expert ground truth for interpretation.

4. Adjudication method for the test set
Not applicable. This is not a study involving human interpretation or AI output that would require adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a medical device (extension set) that does not involve AI or human readers for diagnostic interpretation.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is a medical device (extension set) and does not involve an algorithm.

7. The type of ground truth used
The "ground truth" for the performance tests is established by the specified ISO, ANSI/AAMI, ASTM, and USP standards. These standards define the acceptable physical, chemical, and biological properties of the device.

8. The sample size for the training set
Not applicable. This is not an AI/machine learning study, so there is no training set.

9. How the ground truth for the training set was established
Not applicable. As there is no training set for an AI/machine learning model, no ground truth was established for it.

{0}------------------------------------------------

Set with T-Connector

DEC 2 3 2011

SECTION 5 - 510(K) SUMMARY

Date of Summary: September 23, 2011

Benlan Inc
2760 Brighton RoadTel (905) 829-5004
Oakville, ON L6H 5T4Fax (905) 829-5006
Official Contact:Cheryl Brown - QA / RA Manager
Proprietary Name:MED-RX Extension Set with T-Connector
Common Name:Intravascular Administration Set
Classification Name:Intravascular Administration Set, 880.5440 Intravascular Administration
Set.
Class:Class II
Product Code:FPA
Predicate Device:Baxter Interlink T-Connector Extension Set (K060074)

Device Description

The MED-RX Extension Set with T-Connector is offered in two configurations, identical except for the tubing length. The proximal end of the MED-RX Extension Set with T-Connector features a T-Connector with an injection port and an integral rotating male luer lock. The distal end has a female luer lock fitting. The inner and outer diameter of both devices is consistent at 0.040" and 0.095" respectively, while the lengths available are either 4" or 38". Each of the luer fittings (distal female luer lock and proximal rotating male luer lock) are provided with protective caps. The MED-RX Extension Set with T-Connector is provided sterile and is single use.

Indications for Use

The MED-RX Extension Set with T-Connector is intended to be used for administration or withdrawal of fluids to or from a neonatal patient as directed by a physician.

Substantial Equivalence

The information provided in the premarket notification demonstrates that the proposed device is substantially equivalent to legally marketed devices. The proposed MED-RX Extension Set with T-Connector is substantially equivalent to the predicate Baxter Interlink T-Connector Extension Set (K060074). Both devices have the same intended use for intravascular administration or withdrawal of fluids to or from a neonatal patient as directed by a physician. Both devices are provided sterile and are single use.

A comparison of features and principles of operation between the proposed device and predicate device is provided in Table 1 below.

{1}------------------------------------------------

K112799
page 2 of 5

n Set with T-Connector

Table 1: Comparison between MED-RX Extension Set with T-Connector and Baxter Interlink T-Connector Extension Set (K060074)

ATTRIBUTEPROPOSED DEVICE –MED-RX Extension Set with T-ConnectorPREDICATE DEVICE –Baxter Interlink T-Connector ExtensionSet (K060074)
General Indications
Indications for UseFor administration or withdrawal offluids to or from a neonatal patient asdirected by a physicianSame
Type of PlacementIntravascularSame
Intended for singlepatient useYesYes
PrescriptionYesYes
Intended PopulationNeonatalSame
Intended Environment ofUseHospitalSame
Non PyrogenicYesYes
Physical Specifications
Tubing outer diameter0.095"0.090"
Tubing inner diameter0.040"0.040"
Approximate length4" or 38"6"
Design Features
Distal ConfigurationFemale luer lock fittingSame
Proximal ConfigurationT-Connector with injection site androtating male luer lockT-connector with injection site and maleluer slip adaptor or rotating male luerlock adaptor
ClampSlide clampSame
CapsProtective luer caps as necessarySame
CompatibilityCompatible with standard luer fittingsSame
Material Composition
TubingPolyvinyl ChlorideSame
ConnectorsMedical grade plasticMedical grade plastic
Packaging and Sterilization
SterileYesYes
Sterilization MethodEthylene Oxide (EO)Gamma Radiation
Packaging ConfigurationMedical grade peelable paper/ polypouchPolypropylene pouch

Summary of Differences

There are no significant differences between the proposed MED-RX Extension Set with T-Connector and the predicate device Baxter Interlink T-Connector Extension Set (K060074). Similarities between the proposed device and the predicate devices include identical indications for use and duration of use. The MED-RX Extension Set with T-Connector and the Baxter Interlink T-Connector Extension Set are sterile, single use devices, packaged in peelable pouches.

{2}------------------------------------------------

sion Set with T-Connector

page 3 of 5

The proposed MED-RX Extension Set with T-Connector is to be offered in two different lengths: 4″ and 38″, whereas the predicate device is only available in a 6″ length. The additional lengths of extension sets to be offered for the proposed device can be considered to have no impact on safety or effectiveness as the additional length is merely intended to increase the amount of accessible tubing for use. The proposed device has a slightly larger tubing outer diameter than the predicate device but the dimensions are effectively equivalent. Both devices feature luer connectors with protective caps as required and slide clamps are placed on each set. The design features of both the proposed device and the predicate device can be also considered equivalent as the configuration of the devices are identical.

Therefore, any minor differences between the proposed device and the predicate have been evaluated to have no impact on safety or effectiveness of the MED-RX Extension Set with T-Connector. Therefore the proposed device can be considered substantially equivalent to legally market devices.

Non-Clinical Test Summary

Verification of functional performance of the MED-RX Extension Set with T-Connector has been performed as per ISO 8536-4: 2007. The MED-RX Extension Set with T-Connector has successfully completed all required performance testing following the applicable guidelines of ISO 8536-4: 2007 including tensile strength, resistance to leakage under pressure, resistance to liquid leakage, particulate contamination, and chemical requirements. The MED-RX Extension Set with T-Connector was also tested for natural rubber latex content. Please refer to Table 2.

TestStandardResults
Tensile Strength -Tubing/T-ConnectorISO 8536-4:2007Pass
Tensile Strength - Tubing/Female Luer LockISO 8536-4:2007Pass
Leakage under PressureISO 8536-4:2007Pass
Liquid LeakageISO 8536-4:2007Pass
Particulate ContaminationISO 8536-4:2007Samples met contamination index limit.

Table 2: Non-Clinical Test Summary

{3}------------------------------------------------

K112799

n Set with T-Connector

TestStandardResults
Chemical RequirementsISO 8536-4:2007 perClause 5 & 7Pass
Natural Rubber Latex ContentModified Lowry MethodDevice does not containnatural rubber latex

Summary of Sterilization

Each MED-RX Extension Set with T-Connector is individually packaged using a medical grade peelable synthetic polymer reinforced paper with a film backing, and sterilized using ethylene oxide. Please see Table 3 for a summary.

Table 3: Sterilization Summary

Test DescriptionStandardResults
Method of ValidationANSI/AMMI/ISO 11135-1: 2007Validated to a Sterility AssuranceLevel of 1 x 10-6
EO Sterilization ResidualsISO 10993-7: 2008Pass
Bacterial EndotoxinsANSI/AAMI ST72: 2002Pass

Summary of Biocompatibility Tests

Biocompatibility testing was successfully completed on sterile finished devices. The MED-RX Extension Set with T-Connector is classified as a prolonged duration, indirect blood path contacting device. A summary of the testing completed and the relevant standards are listed in Table 4.

3 of 5

{4}------------------------------------------------

Benlan™
MED-RX® Extension Set with T-Connector

K1127991

page ్ర 50

Table 4: Biocompatibility Test Summary

Test DescriptionStandardResults
ISO MEM Elution with L-929Mouse Fibroblast Cells(Cytotoxicity)ISO 10993-5: 2009Product code 10-1053XLU is considerednon-toxic.
ISO Guinea Pig MaximizationSensitization Test (Methodof Biomaterial Extracts)ISO 10993-10:2002Product code 10-1053XLU did not elicit asensitization response.
ISO IntracutaneousReactivity TestISO 10993-10:2002Product code 10-1053XLU would beconsidered a non-irritant.
ISO Acute Systemic InjectionTestISO 10993-11: 2006The findings indicate that therequirements of the ISO Acute SystemicInjection Test have been met.
ASTM Hemolysis Assay -Extract MethodASTM F-756-00Product code 10-1053XLU is considerednon-hemolytic and passes the test.
Materials Mediated RabbitPyrogen TestUSP 32: 2009 <151>Product code 10-1053XLU is determinedto be non-pyrogenic.

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

DEC 2 3 2011

Ms. Cheryl Brown Quality Assurance / Regulatory Affairs Manager Benlan Incorporated 2760 Brighton Road Oakville, Ontario CANADA L6H 5T4

Re: K112799

Trade/Device Name: MED-RX Extension Set with T-Connector Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: September 23, 2011 Received: September 27, 2011

Dear Ms. Brown:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2 - Ms. Brown

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety /ReportaProblem /default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov /MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Anthony D.m
Anthony D. Watson, B.S., M.S., M.P.

nthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

KII2799

SECTION 4 - INDICATIONS FOR USE

510(K) Number (If Known): Not Known.

Device Name: MED-RX Extension Set with T-Connector

Indications For Use:

The MED-RX Extension Set with T-Connector is intended to be used for administration or withdrawal of fluids to or from a neonatal patient as directed by a physician.

Prescription Use:
(Part 21 CFR 801 Subpart D)
AND/OR
Over-the-Counter Use
(21 CFR 801 Subpart C)
N/A

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED.

Concurrence of CDRH, Office of Device Evaluation (ODE)

for Richard Chapman Dec 21, 2016

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

510(k) Number: K112799

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.