K Number
K112777

Validate with FDA (Live)

Date Cleared
2012-02-24

(151 days)

Product Code
Regulation Number
880.5570
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sol-M Exchangeable Needles are sterile hypodermic needles intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin.

Device Description

The Sol-M Exchangeable Needles are a sterile, single use, standard hypodermic needle. The device is available in 16 to 31 gauge and in lengths from 5/16" to 1-1/2". In addition, the needle tip is available in a regular or short bevel. Each needle device consists of a stainless steel cannula sealed with epoxy glue into a polypropylene hub. The assembly has a protective polypropylene needle cap/shield. The Sol-M Exchangeable Needles are sterilized by Ethylene Oxide Gas and supplied sterile in blister pack. One hundred blister packs are packaged in a chipboard box. Each Blister pack and chipboard box is labeled with the contents and the appropriate information per the FDA's Quality System Regulation and Labeling requirements.

AI/ML Overview

The Sol-M Exchangeable Needles are intended for injecting fluids into or withdrawing fluids from parts of the body below the skin's surface.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Standard Reference)Reported Device Performance
Sterile Hypodermic Needles for Single Use (ISO 7864)Met appropriate requirements
Conical Fittings with a 6% (Luer) Taper for Syringes, Needles, and Certain Other Medical Equipment (ISO 594)Met appropriate requirements
Stainless Steel Needle Tubing for Manufacture of Medical Devices (ISO 9626)Met appropriate requirements
Hypodermic Needles for Single Use Colour Coding for Identification (ISO 6009)Met appropriate requirements
Physical, mechanical, and biological specificationsSubstantially equivalent to the predicate device (DuoPross Needles)
Accelerated aging to verify performance over device lifePerformance verified
Product specifications and requirementsMet all product specifications and requirements

2. Sample size used for the test set and the data provenance:

The document does not specify a separate test set sample size or its provenance. The testing appears to have been conducted based on meeting international standards and comparing to a predicate device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

Not applicable. The ground truth was established through adherence to international standards and direct comparison of physical, mechanical, and biological specifications. This type of device does not typically involve expert clinical judgment for ground truth establishment in this regulatory context.

4. Adjudication method for the test set:

Not applicable. The evaluation was based on compliance with established engineering and medical device standards.

5. Multi-reader multi-case (MRMC) comparative effectiveness study:

No MRMC comparative effectiveness study was conducted. The submission focused on demonstrating substantial equivalence to a predicate device through technical and performance comparisons rather than clinical efficacy against human readers.

6. Standalone (algorithm only without human-in-the-loop performance) study:

Not applicable. This is a physical medical device (hypodermic needle), not an algorithm or AI.

7. Type of ground truth used:

The ground truth was based on:

  • Compliance with international standards: ISO 7864, ISO 594, ISO 9626, and ISO 6009 define the acceptable performance characteristics for hypodermic needles.
  • Comparison to a predicate device's specifications: The DuoPross Needles (K052445) served as the benchmark for physical, mechanical, and biological equivalence.
  • Verification of product specifications and requirements: Internal testing ensured the device met its own defined design and performance criteria.

8. Sample size for the training set:

Not applicable. This is a physical medical device, not an AI or machine learning model. There is no concept of a "training set" in this context.

9. How the ground truth for the training set was established:

Not applicable, as there is no training set for a physical device.

{0}------------------------------------------------

K112777 page 1 of

510(k) Summary (As required by 21 CFR 807.92(a)) FEB 2 4 2012

September 19, 2011 Date Prepared:

Submitter Information A.

Sol-Millennium Medical, Inc. 5415 Sugarloaf Parkway Suite 2203 Lawrenceville, GA 30043

Phone Number: Contact:

404-973-2200 Karen Dunlap President

Trade Name:

Sol-M Exchangeable Needles

  • Device Information B.
Trade/Proprietary Name:Sol-M Exchangeable Needles
Common name of device:Syringe Cannula
Classification Name:Needle, Hypodermic, SingleLumen
Product Code:80 FMI
Regulatory Class:II
Classification Number:880.5570
Reason for 510(k):New Device

{1}------------------------------------------------

K112777

,. >:్ర
l
. .
C.Predicate Device:DuoPross Needles
Predicate 510(k) #:K052445
Predicate product code:FMI

D. Device Description

The Sol-M Exchangeable Needles are a sterile, single use, standard hypodermic needle. The device is available in 16 to 31 gauge and in lengths from 5/16" to 1-1/2". In addition, the needle tip is available in a regular or short bevel.

Each needle device consists of a stainless steel cannula sealed with epoxy glue into a polypropylene hub. The assembly has a protective polypropylene needle cap/shield.

The Sol-M Exchangeable Needles are sterilized by Ethylene Oxide Gas and supplied sterile in blister pack. One hundred blister packs are packaged in a chipboard box. Each Blister pack and chipboard box is labeled with the contents and the appropriate information per the FDA's Quality System Regulation and Labeling requirements.

E. Statement of Indications for Use

The Sol-M Exchangeable Needles are sterile hypodermic needles intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin.

:

{2}------------------------------------------------

  • F. Comparison of Required Technological Characteristics:
    Information was submitted to demonstrate that there are no significant differences in technological characteristics between the Sol-M Exchangeable Needles and the DuoPross Needles. This included:

    1. Label Review
    • Both devices had similar device descriptions a.
    • b. Both devices were labeled as EO Sterilized, non pyrogenic, non toxic, single use, latex free and with a five year shelf life.
    1. Intended Use Comparison
    • a. The intended use for both devices is the same.
    1. Material Comparison
    • a. Both devices are fabricated from the same materials.
    1. Physical, mechanical and biological specifications
    • a. A comparison of physical, mechanical and biological specifications showed that the Sol-M Exchangeable Needles are substantially equivalent to the predicate device.
  • Summary and Conclusion of Nonclinical and Clinical Tests: G.

The Sol-M Exchangeable Needles met the appropriate requirements contained in the following standards:

    1. ISO 7864, Sterile Hypodermic Needles for Single Use;
    1. ISO 594, Conical Fittings with a 6% (Luer) Taper for Syringes, Needles, and Certain Other Medical Equipment;
    1. ISO 9626, Stainless Steel Needle Tubing for Manufacture of Medical Devices;
    1. ISO 6009, Hypodermic Needles for Single Use Colour Coding for Identification;

{3}------------------------------------------------

K112777

page 4 of 4

H. Discussion of Clinical Tests:

None submitted

  • Conclusions Demonstrating Safety, Effectiveness and I. Performance:
    The device has been tested and found to meet all product specifications and requirements. Accelerated aging was used to verify the performance of the product over the life of the device.

Instructions for Use detail how to use the devices and the conditions of use. Product labeling clearly shows that the device is for single patient use only.

The Sol-M Exchangeable Needles have been found to be safe and effective for their intended use.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus symbol.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Sol-Millennium Medical, Inc. C/O Mr. Jim Barley President Abrimed Consulting 33822 Malaga Drive Dana Point, California 92629

FEB 2 4 2012

Re: K112777

Trade/Device Name: Sol-M Exchangeable Needles Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: December 27, 2011 Received: February 22, 2012

Dear Mr. Barley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Bartley

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

ef

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{6}------------------------------------------------

Indications For Use

510(k) Number (if known):

Device Name: Sol-M Exchangeable Needles

Indications For Use:

The Sol-M Exchangeable Needles are sterile hypodermic needles intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Rijl M. for R-Z-C feb 22, 2012

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

K112777 510(k) Number:

Page 1 of ﭩﭗ

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).