K Number
K112533

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-12-08

(101 days)

Product Code
Regulation Number
888.3060
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Zavation Cervical Plate System is intended for anterior screw fixation of the cervical spine (C2-C7) as an adjunct to fusion. These implants have been designed to provide stabilization for the treatment of the following indications: degenerative disc disease (defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fractures or dislocations), spinal stenosis, deformity (i.e., kyphosis, lordosis or scoliosis), tumor, pseudarthrosis or failed previous fusion.

Device Description

The Zavation Cervical Plate System consists of self-tapping/self-drilling screws and plates. Screws are available in a variety of diameter and length combinations. Plates are available in a variety of lengths.

AI/ML Overview

The provided text describes a 510(k) submission for the Zavation Cervical Plate System, a medical device. This type of submission is for establishing substantial equivalence to a predicate device, not for proving novel clinical effectiveness through extensive studies as one might see for a new drug or a highly innovative device.

Therefore, many of the typical acceptance criteria and study elements listed in the prompt (like sample sizes for test and training sets, expert ground truth, adjudication methods, MRMC studies, standalone algorithm performance, or specific effect sizes) are not applicable to this type of regulatory submission for this device.

The study presented here is a mechanical performance study comparing the new device against a predicate device.

Here's a breakdown based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Acceptance Criteria (Implied)Reported Device Performance
Material CompositionComply with ASTM F136 for titanium alloy (Ti-6Al-4V).Components manufactured from titanium alloy (Ti-6Al-4V) as described by ASTM F136.
Mechanical PerformancePerform equally or superior to the predicate device in static and dynamic loading.Mechanical test results demonstrated that the Zavation Cervical Plate System performs as well as or better than the predicate device in static compression bending/torsion, and dynamic compression bending.
Design CharacteristicsPossess similar basic design, material, sizes, and intended use as the predicate (Technological Characteristics).Same basic design (plate designed fixation system with various screw sizes), material (titanium alloy), sizes (variety of plate and screw sizes), and intended use.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated as human or animal subjects as this was a mechanical bench test. The "worst-case, cervical plate construct" was used for testing. The number of constructs tested for each mechanical test (static and dynamic) is not specified.
  • Data Provenance: Not applicable in the context of human data. The tests were performed according to ASTM F1717, an established standard for spinal implant testing. These are laboratory-generated data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Not applicable. Ground truth in this context refers to the pass/fail criteria of the ASTM F1717 standard and comparison of mechanical output values against the predicate device's performance. No human experts were used to establish "ground truth" for the mechanical characteristics of the device in the way a radiologist reviews images.

4. Adjudication Method for the Test Set

  • Not applicable. This was a mechanical study, not a clinical study requiring human adjudication of outcomes or diagnoses. The "adjudication" was based on objective measurements adhering to ASTM F1717 and comparison to predicate device data.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No. This was a mechanical performance study, not a clinical study involving human readers or comparative effectiveness in a clinical setting.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

  • Not applicable. This device is a physical implant, not an AI algorithm. Therefore, "standalone algorithm performance" is not relevant.

7. The Type of Ground Truth Used

  • The "ground truth" for this submission is based on engineering standards (ASTM F1717) and the performance data of legally marketed predicate devices. The new device's mechanical performance in specific tests (static compression bending, torsion, and dynamic compression bending) was compared to that of the predicate devices. The aim was to demonstrate equivalent or superior performance.

8. The Sample Size for the Training Set

  • Not applicable. There is no "training set" in the context of a mechanical performance study for a physical medical device. This term is relevant for machine learning or AI models.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As there is no training set, there is no ground truth for it to be established.

{0}------------------------------------------------

DEC - 8 2011

510K Summary

Date:August 27, 2011
Submitter:Zavation LLC501 Avalon WayBrandon, MS 39047Phone: 601-919-1119Fax: 800-447-1302
Contact person:John Walker
Type of 510(k) submission:Traditional
Trade name:Zavation Cervical Plate System
Common name:Anterior Cervical Plate
Classification regulation:888.3060
Device classification:Class II
Classification Panel:Orthopedic
Product code:KWQ
Basis for submission:New device

Device Description:

The Zavation Cervical Plate System consists of self-tapping/self-drilling screws and plates. Screws are available in a variety of diameter and length combinations. Plates are available in a variety of lengths.

Intended Use:

The Zavation Cervical Plate System is intended for anterior screw fixation of the cervical spine (C2-C7) as an adjunct to fusion. These implants have been designed to provide stabilization for the treatment of the following indications: degenerative disc disease (defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fractures or dislocations), spinal stenosis, deformity (i.e., kyphosis, lordosis or scoliosis), tumor, pseudarthrosis or failed previous fusion.

{1}------------------------------------------------

· Materials:

The Zavation Cervical Plate System components are manufactured from titanium alloy (Ti-6Al-4V) as described by ASTM F136.

Predicate Device:

Window Cervical Plate System (K993066) Trestle Alphatec Spine (K070681) Slim-Loc Depuy Spine (K013877)

Technological Characteristics:

The Zavation Cervical Plate System possesses the same technological characteristics as the predicate. These include: basic design (plate designed fixation system having various screw diameters and lengths), material (titanium alloy), sizes (variety of plate and screw sizes), and intended use (as described above).

Performance Data:

Static compression bending and torsion, and dynamic compression bending were performed according to ASTM F1717 on a worst-case, cervical plate construct. The mechanical test results demonstrated that the Zavation Cervical Plate System performs as well as or better than the predicate device.

Conclusion:

The Zavation Cervical Plate System is substantially equivalent to the device referenced above and is therefore safe and effective for its intended use.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized symbol resembling an abstract human figure with outstretched arms, accompanied by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around the symbol. The symbol and text are presented in black against a white background.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room ~WO66-G609 Silver Spring, MD 20993-0002

DEC - 8 2011

Zavation, LLC % Mr. John Walker 501 Avalon Way Brandon, Mississippi 39047

Re: K112533

Trade/Device Name: Zavation Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWO Dated: November 02, 2011 Received: November 03, 2011

Dear Mr. Walker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

Page 2 - Mr. John Walker

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

ely yours,
for

Det

m

Det U-

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K112533

Device Name: Zavation Cervical Plate System

Indications For Use:

The Zavation Cervical Plate System is intended for anterior screw fixation of the cervical spine (C2-C7) as an adjunct to fusion. These implants have been designed to provide stabilization for the treatment of the following indications: degenerative disc disease (defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fractures or dislocations), spinal stenosis, deformity (i.e., kyphosis, lordosis or scoliosis), tumor, pseudarthrosis or failed previous fusion.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRA ///H , Office of Device Evaluation (ODE)
-------------------------------------------------------------------------

Page 1 of 1

(Division Sign-Off) Division of Surgical, Orthopedic, /and Restorative Devices

KII2533 510(k) Number.

§ 888.3060 Spinal intervertebral body fixation orthosis.

(a)
Identification. A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.(b)
Classification. Class II.