K Number
K112525

Validate with FDA (Live)

Date Cleared
2011-09-29

(29 days)

Product Code
Regulation Number
870.4260
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The D731 MICRO Ph.I.S.I.O. with 27 micron screen with phosphoryIcholine coating and the D733 MICRO Ph.I.S.I.O. with 40 micron screen with phosphoryIcholine coating are recommended for use in the arterial line of the extracorporeal circuit during any procedure that requires cardiopulmonary bypass. The filters are used to trap and remove gaseous emboli as well as particulate debris that may be introduced through the arterial line. The device should not be used longer than 6 hours. Contact with blood for longer periods is not advised.

Device Description

The D731/D733 MICRO Ph.I.S.I.O. Arterial Filters are sterile, non-pyrogenic disposable filter for use in the arterial line of the cardiopulmonary bypass circuit with flow rate not exceeding 6.0 liters/minute. The D731/D733 MICRO Ph.I.S.I.O. are Arterial Filters with 27 and 40 micron filters screen, respectively. The devices are designed to remove potentially harmful gaseous emboli, aggregated blood constituents, and particulate debris, greater than the pore size, from the arterial line perfusate.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information for the D731/D733 MICRO Ph.I.S.I.O. Arterial Filters, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state quantitative acceptance criteria or detailed reported device performance values in a table format. Instead, it refers to broad compliance with guidance documents and established specifications. The main "acceptance criteria" appear to be meeting the performance of the predicate device and relevant industry standards.

Acceptance Criteria CategoryReported Device Performance
BiocompatibilityMet requirements of ISO 10993-1 and FDA May 1st, 1995 Memorandum (cross-referenced data from unmodified device #K051232).
PackagingMet requirements (cross-referenced data from unmodified device #K051232).
In Vitro Performance (Aged Devices)Performance testing demonstrated compliance with performance specifications defined in "Guidance for Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submission" (November 29, 2000). The results met established specifications when compared to the predicate device (AF 620/640 Ph.I.S.I.O. Arterial Filters) at 6 LPM.
SterilityEffectiveness of production techniques assured device is sterile.
Non-pyrogenicityEffectiveness of production techniques assured device is non-pyrogenic.
FunctionalityFunctionally equivalent to the predicate device.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated. The document mentions "Applicable tests were carried out," "In vitro testing was carried out," and "Testing supplied... includes performance testing that demonstrates compliance with performance specifications." It does not provide the number of devices tested.
  • Data Provenance: The device manufacturer is Sorin Group Italia S.r.l., based in Italy. The tests were likely conducted by the manufacturer or a contract research organization on their behalf. The document does not specify the country of origin of the data beyond the manufacturer's location. The tests are described in the "NON CLINICAL TEST RESULTS" and "IN VITRO TEST RESULTS" sections, suggesting a retrospective approach where existing data or newly generated lab data was compiled for the 510(k) submission, rather than a prospective clinical trial.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. The tests described are non-clinical and in-vitro, focusing on device performance specifications rather than expert-derived ground truth.

4. Adjudication Method for the Test Set

This information is not applicable as the described tests are non-clinical and do not involve human interpretation or adjudication in the way a clinical study with expert readers would.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

This information is not applicable. The document describes a medical device (arterial filter) and its non-clinical performance, not an AI or imaging device that would involve human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This information is not applicable. The device is a physical arterial filter, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the non-clinical tests appears to be established performance specifications and regulatory guidance.

  • For biocompatibility: ISO 10993-1 and FDA May 1st, 1995 Memorandum.
  • For in vitro performance: "Guidance for Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submission" (November 29, 2000), and comparison to the predicate device (AF 620/640 Ph.I.S.I.O. Arterial Filters).
  • For sterility and non-pyrogenicity: Effectiveness of production techniques to assure these properties.

8. The Sample Size for the Training Set

This information is not applicable. This is a physical medical device, not an AI or machine learning algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable. See point 8.

{0}------------------------------------------------

K 112525 Pg 10

SEP 2 9 2011

D731/ D733 MICRO Ph.I.S.I.O. Arterial Filters Sorin Group Italia S.r.I.

Special 510(k) August 31, 2011

510(k) SUMMARY

SUBMITTER:Sorin Group Italia86, Via Statale 12 Nord41037 Mirandola (MO) Italy
CONTACT PERSON:Luigi VecchiPhone: 39 0535 29811Fax: 39 0535 25229
DATE PREPARED:August 31, 2011
DEVICE TRADE NAME:D731 MICRO 27 Ph.I.S.I.O. Arterial Filter, SorinD731 MICRO 27 Arterial Filter with 27 micronscreen with phosphorylcholine coating
D733 MICRO 40 Ph.I.S.I.O. Arterial Filter, SorinD733 MICRO 40 Arterial Filter with 40 micronscreen with phosphorylcholine coating
COMMON NAME:Arterial Filter
CLASSIFICATION NAME:Cardiopulmonary Bypass Arterial Line Blood Filter
UNMODIFIED DEVICE:D731 MICRO 27 Ph.I.S.I.O. Pediatric Arterial Filter,Dideco D731 MICRO 27 Pediatric Arterial Filterwith 27 micron screen with phosphorylcholinecoating
D733 MICRO 40 Ph.I.S.I.O. Pediatric Arterial Filter,Dideco D733 MICRO 40 Pediatric Arterial Filterwith 40 micron screen with phosphorylcholinecoating (#K051232)
PREDICATE DEVICE:Sorin AF 620 Ph.I.S.I.O. Arterial Filter
Sorin AF 640 Ph.I.S.I.O. Arterial Filter (#K093986)

DEVICE DESCRIPTION:

The D731/D733 MICRO Ph.I.S.I.O. Arterial Filters are sterile, non-pyrogenic disposable filter for use in the arterial line of the cardiopulmonary bypass circuit with flow rate not exceeding 6.0 liters/minute. The D731/D733 MICRO Ph.I.S.I.O. are Arterial Filters with 27 and 40 micron filters screen, respectively. The devices are designed to remove potentially harmful gaseous emboli, aggregated blood constituents, and particulate debris, greater than the pore size, from the arterial line perfusate.

{1}------------------------------------------------

Kn 2525 Pg 2 of 4

D731/ D733 MICRO Ph.I.S.I.O. Arterial Filters Sorin Group Italia S.r.I.

Special 510(k) August 31, 2011

The D731/D733 MICRO Ph.I.S.I.O. Arterial Filters are a modified version of the currently marketed D731/D733 MICRO Ph.I.S.I.O. Pediatric Arterial Filters The modifications consist of: an increase of the recommended maximum blood flow rate from 5 to 6 LPM, a change of the device trade name from D731/D733 MICRO Ph.I.S.I.O. Pediatric Arterial Filters to D731/D733 MICRO Ph.I.S.I.O. Arterial Filters, and a change of the filter screen pore size from 20 to 27 micron for the D731 MICRO Ph.I.S.I.O. As a consequence of these modifications, the labeling has been updated.

The modified device has unchanged intended use, materials, operating principles, manufacturing process, control mechanisms, sterilization process and fundamental scientific technology.

The manufacturing process in regards to the coating is also unchanged with respect to the unmodified device.

INDICATION FOR USE:

The D731 MICRO Ph.I.S.I.O. with 27 micron screen with phosphoryIcholine coating and the D733 MICRO Ph.I.S.I.O. with 40 micron screen with phosphoryIcholine coating are recommended for use in the arterial line of the extracorporeal circuit during any procedure that requires cardiopulmonary bypass. The filters are used to trap and remove gaseous emboli as well as particulate debris that may be introduced through the arterial line. The device should not be used longer than 6 hours. Contact with blood for longer periods is not advised.

TECHNOLOGICAL CHARACTERISTICS:

The D731/D733 MICRO Ph.I.S.I.O. Arterial filters have the same design features, manufacturing process, control mechanisms, and operating principles when compared to the unmodified device. The D731/D733 MICRO Ph.I.S.I.O. Arterial filters utilize the same materials, filtering media and the same main blood flow path as the unmodified devices.

Except for the increase of the maximum blood flow rate from 5 LPM to 6 LPM and the change of the filter screen pore size of the D731 MICRO Ph.I.S.I.O. (modified from 20 to 27 micron), the modified D731/D733 MICRO Ph.I.S.I.O. Arterial filters are identical to the current marketed D731/D733 MICRO Ph.I.S.I.O. Pediatric Arterial filters.

No change to the intended use has been made as a result of these modifications. There are no differences in packaging type and material between the D731/D733 MICRO Ph.I.S.I.O. Arterial filters and the unmodified devices, D731/D733 MICRO Ph.I.S.I.O. Pediatric Arterial filters

Compared to the predicate device AF 620/640 Ph.I.S.I.O Arterial Filters, the modified D731/D733 MICRO Ph.I.S.I.O. Arterial filters have the same operating principles. control mechanisms, sterilization process, and fundamental scientific technology of the predicate device.

{2}------------------------------------------------

K112525 pg 3 of 4

D731/ D733 MICRO Ph.I.S.I.O. Arterial Filters Sorin Group Italia S.r.l.

Special 510(k) August 31, 2011

The AF 620/640 Ph.1.S.I.O Arterial Filters predicate device, have also the same recommended maximum blood flow rate of 6 LPM of the modified D731/D733 MICRO Ph.I.S.I.O Arterial filters as well as the same intended use.

The D731/D733 MICRO Ph.I.S.I.O. Arterial filters are ethylene oxide sterilized and have a non-pyrogenic fluid path. They are for single use only.

NON CLINICAL TEST RESULTS:

Applicable tests were carried out in accordance with the requirements of ISO 10993-1 and the FDA May 1st, 1995 Memorandum on the use of the ISO 10993 standard for biocompatibility testing of raw materials.

The D731/D733 MICRO Ph.I.S.I.O. Arterial Filters have exactly the same materials of the unmodified device, D731/D733 MICRO Ph.I.S.I.O. Pediatric Arterial Filters (#K051232). The same formulation of Ph.l.S.l.O. material has been also used to coat all blood contact surfaces.

As no new materials are used with respect to the unmodified device, this 510(k) cross references biocompatibility data for the D731/D733 MICRO Ph.I.S.I.O. Pediatric Arterial Filters (#K051232).

As no packaging changes have been introduced, this 510(k) cross references packaging data previously submitted for the unmodified device (#K051232).

IN VITRO TEST RESULTS:

In vitro testing was carried out in accordance with the relevant requirements of "Guidance for Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submission" Final Guidance for Industry, dated November 29, 2000.

Testing supplied in the 510(k) premarket notification for the D731/D733 MICRO Ph.I.S.I.O. Arterial Filters includes performance testing that demonstrates compliance with performance specifications.

The tests were carried out on sterilized devices were aged (3 +1 years, worst case) comparing the D731/D733 MICRO Ph.I.S.I.O. modified Arterial Filters vs. the predicate device, AF 620/640 Ph.I.S.I.O. Arterial Filters operated at same max blood flow rate (6 LPM). The results of these tests met established specifications.

CONCLUSIONS:

The D731/D733 MICRO Ph.I.S.I.O. modified Arterial Filters are substantially equivalent to the predicate device in terms of functionality.

{3}------------------------------------------------

K112525 pg 4 of 4

D731/ D733 MICRO Ph.I.S.I.O. Arterial Filters Sorin Group Italia S.r.I.

Special 510(k) August 31, 2011

Additional testing has also demonstrated the effectiveness of production techniques to assure that the device is sterile and non-pyrogenic.

In conclusion, test results of this study suggest the D731/D733 MICRO Ph.I.S.I.O. Arterial Filters are equivalent to the AF 620/640 Ph.I.S.I.O. Arterial Filters with respect to device function.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows a logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

SEP 2 9 2011

Sorin Group Italia s.r.1. c/o Mr. Barry Sall Principal Consultant Parexel International Consulting 195 West Street Waltham. MA 02451

Re: K112525

Trade/Device Name: D731 MICRO 27 Ph.I.S.I.O. and D733 MICRO 40 Ph.I.S.I.O. Arterial Filters Regulation Number: 21 CFR 870.4260 Regulation Name: Cardiopulmonary Bypass Arterial Line Blood Filter Regulatory Class: Class II Product Code: DTM Dated: August 31, 2011 Received: August 31, 2011

Dear Mr. Sall:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be

{5}------------------------------------------------

Page 2 - Mr. Barry Sall

found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA is issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1.15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

LL

Bram D. Zuckerman, M.D.
Director
Division of Cardiovascular Devices
Office of Device Evaluation
Center for Devices and
Radiological Health

Enclosure

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a handwritten alphanumeric string. The string appears to be "KM2525". The characters are written in a cursive style, with some connections between the letters and numbers. The writing is dark and clear against a white background.

D731/ D733 MICRO Ph.I.S.I.O. Arterial Filters Sorin Group Italia S.r.I.

Special 510(k) August 31, 2011

K112525 510(k) Number (if known):

Device Name: D731 MICRO 27 Ph.I.S.I.O., Sorin D731 MICRO 27 Arterial Filter with 27 micron screen with phosphorylcholine coating and D733 MICRO 40 Ph.I.S.I.O., Sorin D733 MICRO 40 Arterial Filter with 40 micron screen with phosphoryIcholine coating

Indication for Use:

The D731 MICRO 27 Ph.I.S.I.O. with 27 micron screen with phosphorylcholine coating and the D733 MICRO 40 Ph.I.S.I.O. with 40 micron screen with phosphorylcholine coating are recommended for use in the arterial line of the extracorporeal circuit during any procedure that requires cardiopulmonary bypass. The filters are used to trap and remove gaseous emboli as well as particulate debris that may be introduced through the arterial line. The device should not be used longer than 6 hours. Contact with blood for longer periods is not advised.

Prescription Use × (Part 21CFR 801 Subpart D) Over-the- Counter Use AND/OR (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

LL

(Division Sign-Off) Division of Cardiovascular Devices

510(k) Number_K) | S255_______________________________________________________________________________________________________________________________________________________

§ 870.4260 Cardiopulmonary bypass arterial line blood filter.

(a)
Identification. A cardiopulmonary bypass arterial line blood filter is a device used as part of a gas exchange (oxygenator) system to filter nonbiologic particles and emboli (blood clots or pieces of foreign material flowing in the bloodstream which will obstruct circulation by blocking a vessel) out of the blood. It is used in the arterial return line.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submissions.”