K Number
K112403

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-05-08

(260 days)

Product Code
Regulation Number
880.6880
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AMSCO Chimeron Medium Steam Sterilizers (Models 39SL and 39CSL) are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in the following configuration:

  • Prevacuum equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.
    The AMSCO Chimeron Medium Prevacuum Steam Sterilizers are equipped with the following factory-programmed sterilization cycles and cycle values (Table A-1):
Device Description

The AMSCO Chimeron Medium Steam Sterilizers (Models 39SL and 39CSL) are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in the following configuration:

  • Prevacuum equipped with Prevacuum, Gravity, Liquid, Leak Test and DART . (Bowie-Dick) cycles.
    The chamber sizes of the proposed AMSCO Chimeron Medium Steam Sterilizer Models are as follows:
    · 26" x 26" x 39" (Models 39SL and 39CSL)
AI/ML Overview

The provided text describes the validation of a steam sterilizer and not an AI/ML powered medical device. Therefore, many of the requested criteria related to AI/ML device studies (such as sample size for test set, data provenance, number of experts, adjudication method, MRMC studies, standalone performance, ground truth for training set, and training set sample size) are not applicable.

However, I can extract information related to the acceptance criteria and the study that proves the device meets those criteria for the steam sterilizer.

Acceptance Criteria and Device Performance for AMSCO Chimeron Medium Steam Sterilizer

The STERIS® AMSCO Chimeron Medium Steam Sterilizer (Models 39SL and 39CSL) was validated to meet the requirements of ANSI/AAMI ST8: 2008, December 2008. The primary acceptance criterion is achieving a sterility assurance level (SAL) of at least 10-6 (meaning a probability of survival of 1 in a million) for all validated cycles. Additionally, specific performance metrics related to temperature, moisture retention, and uniform color change (for DART cycle) were evaluated.

Table of Acceptance Criteria and Reported Device Performance

Criterion / Performance MetricAcceptance Criteria (from ANSI/AAMI ST8: 2008)Reported Device Performance (as demonstrated by verification studies)
Sterility Assurance Level (SAL)At least 10-6 probability of survival for all validated cyclesAchieved SAL of at least 10-6 for all validated cycles.
Empty Chamber Testing (Prevac, Gravity, Liquid cycles)Provide steady-state thermal conditions consistent with predicted SAL.Sterilizer is capable of providing steady state thermal conditions within the chamber that are consistent with the predicted sterility assurance level (SAL) in the load.
GRAVITY cycles (fabric test pack)Achieve Fo value of at least 12; moisture retention < 3% increase; no wet spots.Achieved SAL of at least 10-6 through Fo value of at least 12, moisture retention of less than 3% increase, and no evidence of wet spots.
GRAVITY cycles (full load instrument trays)Achieve SAL of at least 10-6 (using half-cycle analysis); moisture retention < 20% increase; no wet spots.Achieved SAL of at least 10-6 using half-cycle cycle analysis, moisture retention of less than 20% increase, and no evidence of wet spots on the outer wrapper.
PREVAC cycles (fabric test pack)Achieve Fo value of at least 12; moisture retention < 3% increase; no wet spots.Achieved SAL of at least 10-6 through Fo value of at least 12, moisture retention of less than 3% increase, and exhibited no wet spots.
PREVAC cycles (full load instrument trays)Achieve SAL of at least 10-6 (using half-cycle analysis); moisture retention < 20% increase; no wet spots.Achieved SAL of at least 10-6 using half-cycle analysis, moisture retention of less than 20% increase, and no evidence of wet spots on the outer wrapper.
LIQUID cycles (three 1,000 ml flasks)Achieve Fo value of at least 12; water loss not exceeding 50 ml; automatic flask closure sealing.Achieved SAL of at least 10-6 through Fo value of at least 12, water loss not exceeding 50 ml, and automatic sealing of the flask closure.
DART cycle (Bowie-Dick Test Pack)Demonstrate uniform color change throughout the test sheet.Demonstrated a uniform color change throughout the test sheet.
Software ValidationMeet FDA's moderate level of concern recommendations for software in medical devices.Performed according to FDA's moderate level of concern recommendations (Guidance for the Content for Premarket Submissions for Software Contained in Medical Devices (05/11/05)).
Safety and Performance Requirements (National Safety Codes)Compliance with relevant national safety codes and standards (e.g., UL, CSA, ASME).Complies with ANSI/UL 61010-1, CAN/CSA C22.2 No. 61010-1, UL 61010A-2-041, CAN/CSA C22.2 No. 1010.2.041, and ASME Boiler Pressure Vessel Code, Section VIII.

Study Details:

  1. Sample Size Used for the Test Set and Data Provenance:

    • The document does not explicitly state a "test set" in the context of an AI/ML algorithm. Instead, it refers to "verification studies" and "validation studies" for physical sterilizer cycles.
    • For the Liquid cycle, it was "validated with 3 1000ml bottles" for the 26" x 26" x 39" sterilizer size, as a guideline for validation, but the full load capacity is 80 bottles (Table B-3).
    • For other cycles, "fabric test pack" and "full load instrument trays" were used, as described in specific sections of the ANSI/AAMI ST8 standard (e.g., Section 5.5.2.1, 5.5.4.1). The exact number of packs/trays per validation run is not specified but is based on the ANSI/AAMI ST8 standard for "full load" validation.
    • The data provenance is not specified as "country of origin" but implies internal STERIS validation in accordance with US/international standards (ANSI/AAMI). The studies are prospective validations performed to demonstrate the sterilizer's performance.
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    • Not applicable as this is not an AI/ML medical device where ground truth is established by expert review. The ground truth for sterilization is objective: achievement of a specific sterility assurance level, physical parameters (temp, time), and absence of specific failure modes (e.g., wet spots). These are measured and compared against established scientific and regulatory standards.
  3. Adjudication Method for the Test Set:

    • Not applicable. Performance is measured against physical and biological indicators according to established standards.
  4. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done:

    • No, a MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic or interpretive AI/ML devices where human readers' performance is augmented or replaced.
  5. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • The device is a physical steam sterilizer. The "algorithm" here refers to the pre-programmed sterilization cycles and control software. The verification studies assess the performance of the sterilizer's system (including its controls and cycles) independently to ensure effective sterilization. So, in a sense, the device's performance is standalone in fulfilling its function. The software validation was performed according to FDA guidance.
  6. The Type of Ground Truth Used:

    • The ground truth used is primarily based on biological indicators (complete kill) and physical measurements (time-at-temperature sufficient to produce an Fo value of at least 12, moisture retention characteristics, water loss, uniform color change, etc.), all defined by the ANSI/AAMI ST8: 2008 standard as objective measures of successful sterilization. This is a form of objective standard/outcome data as defined by the AAMI standard.
  7. The Sample Size for the Training Set:

    • Not applicable. This is not an AI/ML device that requires a training set. The cycles are engineered and validated, not "trained" in the machine learning sense.
  8. How the Ground Truth for the Training Set was Established:

    • Not applicable.

{0}------------------------------------------------

STERIS®

510(k) Summary For AMSCO Chimeron Medium Steam Sterilizer K112403

STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 639-4459

Contact:

Tim Stuthers Director, Regulatory Affairs Telephone: (440) 392-7222 Fax No: (440) 357-9198

Summary Date:

December 29th, 2011

STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600

{1}------------------------------------------------

Device Name 1.

Trade Name:AMSCO Chimeron Medium Steam Sterilizer
Models:39SL and 39CSL
Common/Usual Name:Steam Sterilizer
Classification Name:Steam Sterilizer (21 CFR 880.6880)Product Code FLE

2. Predicate Device

K020747, AMSCO Century Medium Steam Sterilizer, product code [FLE] cleared May 17th, 2002.

Description of Device 3.

The AMSCO Chimeron Medium Steam Sterilizers (Models 39SL and 39CSL) are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in the following configuration:

  • Prevacuum equipped with Prevacuum, Gravity, Liquid, Leak Test and DART . (Bowie-Dick) cycles.
    The chamber sizes of the proposed AMSCO Chimeron Medium Steam Sterilizer Models are as follows:

· 26" x 26" x 39" (Models 39SL and 39CSL)

Intended Use 4.

The AMSCO Chimeron Medium Steam Sterilizers (Models 39SL and 39CSL) are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in the following configuration:

  • Prevacuum equipped with Prevacuum, Gravity, Liquid, Leak Test and DART . (Bowie-Dick) cycles.
    The AMSCO Chimeron Medium Prevacuum Steam Sterilizers are equipped with the following factory-validated sterilization cycles (Table B-1):

{2}------------------------------------------------

Table B-1

:

CyclesSterilizeTemperatureSterilizeTimeDry TimeFactoryDefaultRecommended Load
Prevac270°F(132°C)4 minutes20minutesYesFabric Packs. Refer toTable B-2 forrecommendedquantities.
Prevac270°F(132°C)4 minutes30minutesYesDouble wrappedinstrument trays,maximum weight 25lbs (11.3 kg) each.Refer to Table B-2 forrecommendedquantities.
Prevac270°F(132°C)4 minutes5 minutesNoSingle Fabric Pack
Prevac275°F(135°C)3 minutes30minutesYesDouble wrappedinstrument trays,maximum weight 25lbs (11.3 kg) each.Refer to Table B-2 forrecommendedquantities.
Prevac(ImmediateUse)270°F(132°C)4 minutes1 minutesYesUnwrapped instrumenttray with a singleinstrument or multipleunwrapped instrumenttrays, maximumweight 25 lbs (11.3 kg)each. Refer to TableB-2 for recommendedquantities.
Gravity250°F(121°C)30minutes15minutesNoFabric Packs. Refer toTable B-2 forrecommendedquantities.
Gravity270°F(132°C)15minutes30minutesNoDouble wrappedinstrument trays,maximum weight 25lbs (11.3 kg) each.Refer to Table B-2 forrecommended
Gravity250°F(121°C)30minutes30minutesNoDouble wrappedinstrument trays,maximum weight 25lbs (11.3 kg) each.Refer to Table B-2 forrecommendedquantities.
Gravity270°F(132°C)25minutes15minutesNoFabric Packs. Refer toTable B-2 forrecommendedquantities.
Gravity.(ImmediateUse)270°F(132°C)3 minutes1 minutesYesUnwrapped instrumenttray with a singleinstrument or multipleunwrapped instrumenttrays, non-porous load,maximum weight 25lbs (11.3 kg) each.Refer to Table B-2 forrecommendedquantities.
Gravity(ImmediateUse)270°F(132°C)10minutes1 minutesYesUnwrapped instrumenttray with a singleinstrument or multipleuniwrapped instrumenttrays, non-porous load,maximum weight 25lbs (11.3 kg) each.Refer to Table. B-2 forrecommendedquantities.
Liquid*250°F(121°C)45minutesN/ANoValidated with 31000ml bottles. Referto Table B-3 forrecommended full loadquantities.
DART Warm-UpDART270°F(132°C)3 ½minutes1 minuteYesBowie-Dick Test Pack,DART Test Pack
Leak TestN/AN/AN/AYesN/A

. .

.

.

3

·

{3}------------------------------------------------

. :

.

*The liquid cycle is for non-patient contact use only.

.

.

·

.

{4}------------------------------------------------

Table B-2 lists STERIS' full loads by sterilizer size:

Table B-2 AMSCO Chimeron Medium Steam Sterilizer full load per sterilizer size

:Accessful of Concessfore & Manager & Particle Particle Particle Particle Particle Particle Particle Particle Particle Particle Particle Particle Pro------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ﺎ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤ
-4441------------------LALBURG BLOGS BLOGS & C1And and would and would be the count of the commend to the commend1------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- Automotive Auction Ana------------All condiction in------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------.

Table B-3 is a guideline for liquid cycle processing. The liquid cycle is for nonpatient contact use only.

Table B-3 AMSCO Chimeron Medium Steam Sterilizers Liquid Cycle Guideline

Sterilizer SizeNumber of Containers for full loadNumber of Containers for validationVolume of Liquid In One ContainerMinimum Recommended Sterilize Time at 250°F (121°C)
26" x 26" x 39"8031000 ml45 minutes

Description of Safety and Substantial Equivalence ഗ്

···

A summary of the technological characteristics of the device subject of this r summary of the venire of the predicate devices is included in Table B-4.

5

{5}------------------------------------------------

Table B-4 Summary of the Proposed and Predicate Device Technological Characteristics

and the comments of the comments of the country of the country of the country of the country of the country of the country of the country of the country of the country of the

::

General SterilizerFeaturesAMSCO Chimeron MediumSteam Sterilizer(K112403)AMSCO Century MediumSteam Sterilizer(K020747)
Intended UseA steam sterilizer intended forsterilization of non-porous and porous,heat- and moisture-stable materials inhealthcare facilities.A steam sterilizer intended forsterilization of non-porous and porous,heat- and moisture-stable materials inhealthcare facilities.
Operating PrincipleSteam is the sterilizing agent.Steam is the sterilizing agent.
Sterilization CyclesOfferedPrevacGravityLiquidPrevacGravityLiquidSFPP
Chamber Sizes26" x 26" x 39"26" x 26" x 39"26" x 26" x 49"26" x 26" x 61"
Chamber DoorType 316L stainless steelHinge or Horizontal SlidingType 316L stainless steelHinge or Horizontal Sliding
Shell AssemblyType 316L stainless steelASME certifiedType 316L stainless steelASME certified
Control TechnologyEmbedded ControllerTouch Screen320 x 240 Pixel DisplayInk on Paper PrinterEmbedded ControllerTouch Screen320 x 240 Pixel DisplayInk on Paper Printer
Process MonitorsChamber TransducerDual Element Chamber Drain SensorChamber TransducerDual Element Chamber Drain Sensor
Safety DevicesPressure Relief ValveChamber Float SwitchEmergency Stop Switch on Powered,Sliding Door UnitsPressure Relief ValveChamber Float SwitchEmergency Stop Switch on Powered,Sliding Door Units

б

. .

·

.

{6}------------------------------------------------

Effectiveness

Effectiveness of sterilizer function and exposure time recommendations was demonstrated by complete kill of biological indicators and by verifying an appropriate safety factor or sterility assurance level (SAL) of at least 10° probability of survival. STERIS validates its sterilization cycles using recommended practices, standards and guidelines developed by independent organizations such as the Association for the Advancement of Medical Instrumentation (AAMI). The AMSCO Chimeron Steam Sterilizers have been validated to meet the requirements of ANSI/AAMI ST8: 2008, December 2008.

The results of the AMSCO Chimeron Steam Sterilizer verification studies demonstrate that the sterilizers perform as intended. The results are summarized as follows:

  • Empty chamber testing was performed as described in Section 5.4.2.5 of ● ANSI/AAMI-ST8, for the Prevac, Gravity, and Liquid cycles. These cycles demonstrated that the sterilizer is capable of providing steady state thermal conditions within the chamber that are consistent with the predicted sterility assurance level (SAL) in the load. The sterilizer meets the requirements of Sections 4.4.2 of ANSI/AAMI-ST8.
  • All GRAVITY cycles were validated using the fabric test pack, described in . Section 5.5.2.1 of ANSVAAMI-ST8, and were qualified according to Section 5.5.2 ANSI/AAMI-ST8. These cycles demonstrated a sterility assurance level of at least 10 6 through achievement of a time-at-temperature sufficient to produce an Fo value of at least 12, moisture retention of less than 3% increase in presterilization test pack weight, and no evidence of wet spots.
  • All GRAVITY cycles were validated using full load instrument trays, described . in 5.5.4.1 of ANSVAAMI-ST8, and were qualified according to Section 5.5.4 ANSI/AAMI-ST8. These cycles demonstrated a sterility assurance level of at least 10 d using half-cycle cycle analysis, moisture retention of less than 20% increase in presterilization weight of the towel, and no evidence of wet spots on the outer wrapper.
  • All PREVAC cycles were validated using the fabric test pack, described in . Section 5.5.2.1 of ANSVAAMI-ST8, and were qualified according to Section 5.5.2 ANSI/AAMI-ST8. These cycles demonstrated a sterility assurance level of at least 10 6 through achievement of a time-at-temperature sufficient to produce an Fo value of at least 12, moisture retention of less than 3% increase in presterilization test pack weight, and exhibited no wet spots.
  • All PREVAC cycles were validated using full load instrument trays, described . in 5.5.4.1 of ANSVAAMI-ST8, and were qualified according to Section 5.5.4 ANSVAAMI-ST8. These cycles demonstrated a sterility assurance level of at

{7}------------------------------------------------

least 106 using half-cycle analysis, moisture retention of less than 20% increase in presterilization weight of the towel, and no evidence of wet spots on the outer wrapper.

  • All LIOUID cycles were validated using three 1,000 ml flasks, described in . Section 5.5.3.1 of the ANSVAAMI-ST8, and were qualified according to Section 5.5.3 of ANSI/AAMI-ST8. These cycles demonstrated a sterility assurance level of at least 10th through achievement of a time-at-temperature sufficient to produce an Fo value of at least 12, water loss not exceeding 50 ml, and automatic sealing of the flask closure. The liquid cycle is for non-patient contact use only.
  • The DART cycle was validated using the Bowie-Dick Test Pack, as described in . 5.6.1.1 of the ANSI/AAMI-ST8, was qualified according to Section 5.6.1 of the ANSI/AAMI-ST8. and demonstrated a uniform color change throughout the test sheet. The DART cycle uses the same dynamic air removal routines as the 270°F Prevac cycle.
  • The software validation for the cycle operation was performed according to . FDA's moderate level of concern recommendations provided in the document "Guidance for the Content for Premarket Submissions for Software Contained in Medical Devices (05/11/05). "

Safety

STERIS sterilizers including the AMSCO Chimeron Steam Sterilizers have been designed, constructed and tested to meet the safety and performance requirements of various national safety codes and standards. The AMSCO Chimeron Steam Sterilizers comply with the following requirements:

  • ANSI/UL 61010-1 2nd Edition (7/22/2005) "Safety Requirements for � Electrical Equipment for Measurement, Control and Laboratory Use - Part 1 General Requirements.
  • CAN/CSA C22.2 No. 61010-1 2nd Edition (7/12/2004) "Safety . Requirements for Electrical Equipment for Measurement, Control and Laboratory Use - Part 1 General Requirements".
  • UL 61010A-2-041 1st Edition (3/27/2002) "Particular Requirements for . Autoclaves Using Steam for Treatment of Medical Materials and for Laboratory Processes"
  • CAN/CSA C22.2 No. 1010.2.041 (R2004) "Particular Requirements for . Autoclaves Using Steam for Treatment of Medical Materials and for Laboratory Processes'.
  • ASME Boiler Pressure Vessel Code, Section VIII (Division 1) (7/1/2007)

{8}------------------------------------------------

Hazards - Failure of Performance

Failure of the sterilization process can lead to incidence of cross contamination, the transmission of potentially infectious organisms from one infected person to another who was not otherwise infected prior to the incident.

To avoid failure, the user must ensure that the materials, instruments and devices to be sterilized are thoroughly cleaned, the manufacturer's instructions for use are followed, the cycle to be used for each type of sterilizer load has been validated, the sterilizer has been maintained in accordance with the sterilizer manufacturer's recommended maintenance schedule and is operating properly, and each sterilizer load is monitored with available and validated biological and chemical sterilization process indicators.

The technology designed into STERIS steam sterilizers including the AMSCO Chimeron Steam Sterilizers provide control safeguards that abort the cycle and give appropriate signals, alerts and warnings when required conditions have not been met or when a malfunction occurs.

Conclusion

The AMSCO Chimeron Medium Steam Sterilizer has been validated to meet the requirements of ANSVAAMI ST8: 2008, December 4, 2008.

The results of the AMSCO Chimeron Medium Steam Sterilizer validation studies demonstrate that the sterilizer performs as intended.

Based on the information provided in this premarket notification, it can be concluded that the AMSCO Chimeron Medium Steam Sterilizer is substantially equivalent to the predicate device and is safe and effective when used as intended.

{9}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The logo is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in capital letters.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Robert F. Sullivan Senior Director FDA Regulatory Affairs STERIS Corporation 5960 Heisley Road Mentor, Ohio 44060

MAY - 8 2012

Re: K112403

Trade/Device Name: AMSCO Chimeron Medium Steam Sterilizer Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: II Product Code: FLE Dated: April 26, 2012 Received: April 27, 2012

Dear Mr. Sullivan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{10}------------------------------------------------

Page 2 - Mr. Sullivan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance: Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.goy/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony D. Watson, B.S., M.S., M.B.A.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

Indications for Use

510(k) Number (if known): K112403

Device Name:

AMSCO Chimeron Medium Steam Sterilizer

Indications For Use:

The AMSCO Chimeron Medium Steam Sterilizers (Models 39SL and 39CSL) are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in the following configuration:

  • · Prevacuum equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.
    The AMSCO Chimeron Medium Prevacuum Steam Sterilizers are equipped with the following factory-programmed sterilization cycles and cycle values (Table A-1):
AMSCO Chimeron Medium Prevaccum Steam Sterilizer factory validated sterilization cycles and cycle values
CyclesSterilize TemperatureSterilize TimeDry TimeRecommended Load
Prevac270°F (132°C)4 minutes20 minutesFabric Packs. Refer toTable A-2 for recommendedquantities.
Prevac270°F (132°C)4 minutes30 minutesDouble wrapped instrumenttrays, maximum weight 25lbs (11.3 kg) each andFabric Packs. Refer toTable A-2 for recommendedquantities.
Prevac270°F (132°C)4 minutes5 minutesSingle Fabric Pack
Prevac275°F (135°C)3 minutes30 minutesDouble wrapped instrumenttrays, maximum weight 25lbs (11.3 kg) each. Refer toTable A-2 for recommendedquantities.
Table A-11
--------------

AMSCO Chimeron Medium Prevacuum Steam Sterilizer factory-validated sterilization

{12}------------------------------------------------

CyclesSterilizeTemperatureSterilizeTimeDry TimeRecommended Load
Prevac(ImmediateUse)270°F (132°C)4 minutes1 minutesUnwrapped instrument traywith a single instrument ormultiple unwrappedinstrument trays, maximumweight 25 lbs (11.3 kg)each. Refer to Table A-2for recommendedquantities.
Gravity250°F (121°C)30 minutes15 minutesFabric Packs. Refer toTable A-2 for recommendedquantities.
Gravity270°F (132°C)15 minutes30 minutesDouble wrapped instrumenttrays, maximum weight 25lbs (11.3 kg) each. Refer toTable A-2 for recommendedquantities.
Gravity250°F (121°C)30 minutes30 minutesDouble wrapped instrumenttrays, maximum weight 25lbs (11.3 kg) each. Refer toTable A-2 for recommendedquantities.
Gravity270°F (132°C)25 minutes15 minutesFabric Packs. Refer toTable A-2 for recommendedquantities.
Gravity(ImmediateUse)270°F (132°C)3 minutes1 minutesUnwrapped instrument traywith a single instrument ormultiple unwrappedinstrument trays, non-porous load, maximumweight 25 lbs (11.3 kg)each. Refer to Table A-2for recommendedquantities.
Gravity(ImmediateUse)270°F (132°C)10 minutes1 minutesUnwrapped instrument traywith a single instrument ormultiple unwrappedinstrument trays, non-porous load, maximumweight 25 lbs (11.3 kg)each. Refer to Table A-2for recommendedquantities.
CyclesSterilize TemperatureSterilize TimeDry TimeRecommended Load
Liquid*250°F (121°C)45 minutesN/AValidated with 3 1000ml bottles. Refer to Table A-3 for recommended full load quantities.
DART Warm-Up270°F (132°C)3 minutes1 minuteN/A
DART270°F (132°C)3 ½ minutes1 minuteBowie-Dick Test Pack, DART Test Pack
Leak TestN/AN/AN/AN/A

1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 -

{13}------------------------------------------------

*The liquid cycle is for non-patient contact use only.

.

.

.

:

:

.

. . . . .

:

· .

and the same of the same

.

The following table lists STERIS' full loads by sterilizer size:

Table A-2
AMSCO Chimeron Medium Steam
Sterilizer full load per sterilizer size

. .

.

:

1

గా

:

.

:

. . . . . . . .

:

.

:

.

12008 1112007 19618 18000A P W A W W A A L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L LAnd Children Color
1

{14}------------------------------------------------

The following table is a guideline for liquid cycle processing:

Table A-3
AMSCO Chimeron Medium Steam
Sterilizer Liquid Cycle Guideline
Sterilizer SizeNumber ofContainersfor fullloadNumber ofContainersfor validationVolume ofLiquid In OneContainerMinimumRecommendedSterilize Time at250°F (121°C)
26" x 26" x 39"8031000 ml45 minutes

*The liquid cycle is for non-patient contact use only.

The AMSCO Chimeron Medium Steam Sterilizer is offered in the following chamber sizes:

26" x 26" x 39" (Models 39SL and 39CSL) ●

Prescription Use (Part 21 CFR 801 Subpart D)

:

AND/OR

Over-The-Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

4

Suan Rummerfer Ecn

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K112403

§ 880.6880 Steam sterilizer.

(a)
Identification. A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.(b)
Classification. Class II (performance standards).