K Number
K112389

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-07-20

(337 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SurgiCase Orthopaedics system is intended to be used as a surgical instrument to assist in pre-operative planning and/or in guiding the marking of bone and/or guide surgical instruments in non-acute, non-joint replacing osteotomies for upper extremity orthopedic surgical procedures.

The system is to be used for adult patients.

SurgiCase Guides are intended for single use only.

Device Description

The SurgiCase Orthopaedics system is composed of two components: SurgiCase Connect (software) and SurgiCase Guides (hardware).

The SurgiCase Orthopaedics system is intended to be used as a surgical instrument to transfer a pre-surgical plan to the lower and upper extremity during orthopaedic surgical procedures.

SurgiCase Connect is a medical device for Materialise and a surgeon for pre-surgical simulation and evaluation of surgical treatment options. This includes transferring, visualizing, measuring, annotating and editing medical data.

The SurgiCase Guides are patient specific templates that are based on a pre-surgical software plan and are designed to fit a specific patient. All guides are individually designed and manufactured for each patient using a design and manufacturing process with strict procedures and work instructions to guarantee guides that consistently perform in a safe and effective way. In surgery these guides are used to assist a surgeon in guiding the marking of bone and/or guiding surgical instruments according to the pre-surgical plan.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the SurgiCase Orthopaedics system:

The provided 510(k) summary (K112389) for the SurgiCase Orthopaedics system does not explicitly detail specific acceptance criteria with quantifiable metrics for device performance (e.g., "accuracy of +/- 1mm") nor does it describe a formal clinical or standalone comparative study with human readers or a detailed statistical analysis of performance against such criteria.

Instead, the submission focuses on demonstrating substantial equivalence to predicate devices through a comparison of intended use, materials, and performance characteristics, and relies on non-clinical testing for validation. This is a common approach for certain types of medical devices, especially when establishing equivalence to existing technology.

However, based on the limited information provided, we can infer some aspects and present what is available:

1. Table of Acceptance Criteria and Reported Device Performance

As mentioned, specific quantifiable acceptance criteria are not explicitly stated in this document. The "performance" described is more qualitative and relates to successful completion of non-clinical tests to demonstrate safety and effectiveness, and accuracy adequate to perform as intended.

Acceptance Criteria (Inferred/General)Reported Device Performance
Substantial equivalence to predicate devicesDevice comparison showed substantial equivalence.
Software validation for intended useSurgiCase Connect software validated.
Accuracy of guides for surgical planning/guidanceTesting verified accuracy and performance of guides is adequate.
Biocompatibility of SurgiCase GuidesBiocompatibility tests performed and met requirements.
Sterilization dimensional stability of GuidesSterilization dimensional stability tests performed and met requirements.
Debris test results for GuidesDebris tests performed and met requirements.
Packaging and shipment integrity for GuidesPackaging and shipment tests performed and met requirements.
Cleaning validation for GuidesCleaning validation tests performed and met requirements.

2. Sample Size for the Test Set and Data Provenance

The document states: "SurgiCase Guides were validated through non-clinical studies using bone models and cadaver specimens."

  • Sample Size for Test Set: Not specified. The number of bone models and cadaver specimens used is not provided.
  • Data Provenance: The validation was non-clinical, using bone models and cadaver specimens. The country of origin for these specimens is not mentioned, nor whether the data was retrospective or prospective (though for non-clinical lab testing, "prospective" would be the more fitting description of how the tests were conducted).

3. Number of Experts Used to Establish Ground Truth and Qualifications

This information is not provided in the document. Since the validation was entirely non-clinical using bone models and cadaver specimens, the concept of "experts establishing ground truth" in the diagnostic imaging sense (e.g., radiologists reviewing images) does not directly apply here. Instead, ground truth would be physical measurements and objective assessments against known parameters of the models/specimens, likely performed by engineers, technicians, and potentially surgeons involved in the study design.

4. Adjudication Method for the Test Set

This information is not applicable/provided. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving human readers or expert review of data where there might be disagreements. Since the validation was non-clinical with bone models and cadavers, and no human reader interpretation of images is described as part of the primary validation for the stated performance, an adjudication method is not relevant.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document does not mention any studies involving human readers, either with or without AI assistance, or comparisons between them. The focus is on the device's standalone performance in non-clinical settings.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, implicitly. The non-clinical validation tests using bone models and cadaver specimens assess the performance of the SurgiCase Guides (hardware) and the SurgiCase Connect software (algorithm for planning/design) in isolation from a live surgical scenario involving a human surgeon's real-time interaction. The software is validated for its intended use, and the guides are tested for accuracy. This can be considered a form of standalone performance assessment as it evaluates the device's ability to "perform as intended" without human intervention in the measurement of its accuracy.

7. The Type of Ground Truth Used

The ground truth for the non-clinical validation was likely based on:

  • Physical measurements: Precise measurements taken on the bone models and cadaver specimens to assess the accuracy of the guides and the outcomes of the simulated procedures. This would involve comparing the guided cuts/markings to the pre-surgical plan.
  • Known parameters of the models: For engineered bone models, the "ground truth" of anatomical features and target osteotomy locations would be precisely known.
  • Biocompatibility standards: For biocompatibility, the ground truth would be established regulatory standards and test results.
  • Sterilization efficacy: For sterility, established protocols and detection limits.

It is not pathology, expert consensus in a diagnostic sense, or outcomes data from real patients.

8. The Sample Size for the Training Set

This information is not provided in the document. The filing describes the product as an "Image processing system and software for simulating/evaluating implant placement and surgical treatment options." While image processing software often involves machine learning that requires training data, the document does not specify any ML/AI components in detail or reference a training set. The descriptions point more towards conventional computational geometry and visualization software. If a machine learning component were present, its training data size and provenance would be crucial.

9. How the Ground Truth for the Training Set Was Established

Since information about a training set or specific machine learning components is not provided, how its ground truth was established is also not described.

{0}------------------------------------------------

K112389.

510(k) Summary

JUL 2002012

Submitter information

Company nameMaterialise N.V.
Establishment registration number3003998208
Street AddressTechnologielaan 15
CityLeuven
Postal code3001
CountryBelgium
Phone number+32 16 39 62 80
Fax number+32 16 39 66 06
Contact nameAlexandra Razzhivina
Contact titleRegulatory officer
Contact e-mail addressregulatory.affairs@materialise.be

Submission date

The date of the Traditional 510(k) submission is 16th of August, 2011.

Submission information

Trade NameSurgiCase Orthopaedics, SurgiCase Connect, SurgiCase Guides
Common NameImage processing system and software for simulating/evaluating implant placement and surgical treatment options
Classification NameRadiological image processing system
Product codePBE

Predicate devices

Manufacturer

.

Trade or proprietary or model nameSurgiCase
510(k) numberK073449
Decision date2008/04/16
Product codeLLZ
ManufacturerMaterialise N.V.
Trade or proprietary or model nameSurgiCase Guide
510(k) numberK103136
Decision date2011/04/18
Product codeJEY, MON

Materialise N.V.

{1}------------------------------------------------

Device Information

Description and functioning of the device

The SurgiCase Orthopaedics system is composed of two components: SurgiCase Connect (software) and SurgiCase Guides (hardware).

The SurgiCase Orthopaedics system is intended to be used as a surgical instrument to transfer a pre-surgical plan to the lower and upper extremity during orthopaedic surgical procedures.

SurgiCase Connect is a medical device for Materialise and a surgeon for pre-surgical simulation and evaluation of surgical treatment options. This includes transferring, visualizing, measuring, annotating and editing medical data.

The SurgiCase Guides are patient specific templates that are based on a pre-surgical software plan and are designed to fit a specific patient. All guides are individually designed and manufactured for each patient using a design and manufacturing process with strict procedures and work instructions to guarantee guides that consistently perform in a safe and effective way. In surgery these guides are used to assist a surgeon in guiding the marking of bone and/or guiding surgical instruments according to the pre-surgical plan.

Intended use

The SurgiCase Orthopaedics system is intended to be used as a surgical instrument to assist in pre-operative planning and/or in guiding the marking of bone and/or guide surgical instruments in non-acute, non-joint replacing osteotomies for upper extremity orthopedic surgical procedures.

The system is to be used for adult patients.

SurgiCase Guides are intended for single use only.

Summary of technological characteristics

Device comparison showed that the proposed device is substantially equivalent in intended use, materials and performance characteristics to the predicate devices.

Performance data

Non-clinical tests

The SurgiCase Connect software has been validated for its intended use to determine substantial equivalence to the predicate devices.

SurgiCase Guides were validated through non-clinical studies using bone models and cadaver specimens. Biocompatibility test, sterilization dimensional stability test, debris test and packaging and shipment test and cleaning validation test were performed to assess the safety and effectiveness of the SurgiCase Guides. Testing verified that the accuracy and performance of the device is adequate to perform as intended.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. The logo is simple, using only black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Materialise N.V. % Ms. Alexandra Razzhivina Technologielaan 15 Leuven, Belgium 3001

JUL 30 2012

Re: K112389 Trade/Device Name: Surgicase Connect, Surgicase Guide Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: PBF Dated: July 06, 2012

Received: July 09, 2012

Dear Ms. Razzhivina:

This letter corrects our substantially equivalent letter of July 20, 2012 for the product code and regulation number for your medical device. We have reviewed your Section 510(k) and regulation namber 10. Jour 10 market the device referenced above and have determined premarket hourseantially equivalent (for the indications for use stated in the enclosure) to the device is substantially of a ransmarketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been the onatined in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general market as wisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your (1 wire), it thay of babyer. Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not I vactar agencies: "10 and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical

{3}------------------------------------------------

Page 2 - Ms. Alexandra Razzhivina

device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Molkerson

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K112389

Device Name: The SurgiCase Orthopaedics system

Indications for Use:

The SurgiCase Orthopaedics system is intended to be used as a surgical instrument to assist in rrie Surgicales Online and/or in guiding the marking of bone and/or guide surgical instruments pre operative planning of replacing osteotomies for upper extremity orthopedic surgical procedures.

The system is to be used for adult patients.

SurgiCase Guides are intended for single use only.

Pete. Rums

(Division Sign-(Division of Surgical, Orthopedic. and Restorative Devices

510(k) Number K112381

Prescription Use _ × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

N/A