K Number
K112101

Validate with FDA (Live)

Date Cleared
2012-07-17

(361 days)

Product Code
Regulation Number
862.1155
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Polymed Therapeutics' Fastep™ hCG Pregnancy Serum/Urine Test is a rapid chromatographic immunoassays for the visual, qualitative detection of human chorionic gonadotropin (hCG) in serum or urine specimen to aid in the early detection of pregnancy. For Professional Use Only. The test kits are for health care professionals use including professionals at physician's office labs (POLs) and Point-of-Care site (POC).

Device Description

The Fastep™ hCG Pregnancy Serum/Urine Test are distributed in Cassette format. Each test reagent strip contains mouse monoclonal anti-a-hCG antibody coated membrane and a dried chemical pad containing mouse monoclonal anti-B-hCG anybody colloidal gold conjugate. The control antibodies are goat anti-mouse IgG.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Fastep™ hCG Pregnancy Serum/Urine Test, based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

The 510(k) summary does not explicitly state pre-defined acceptance criteria (e.g., "Sensitivity must be >= 95%"). Instead, it compares the device's performance to a predicate device (Teco Diagnostics One-Step Urine/Serum Combo Pregnancy Card Test) and declares "substantial equivalency." However, we can infer the performance metrics considered acceptable based on the reported results and the statement of substantial equivalency.

CharacteristicAcceptance Criterion (Inferred from Predicate)Reported Device Performance (Fastep™ hCG Test)
Accuracy (Urine Sample)Substantially equivalent to Teco One-Step Combo Card Test (100% Sensitivity, 100% Specificity reported for predicate)Sensitivity: 100.0% Specificity: 100.0%
Accuracy (Serum Sample)Substantially equivalent to Teco One-Step Combo Card Test (98.3% Sensitivity, 100% Specificity reported for predicate)Sensitivity: 98.3% Specificity: 100.0%
Analytical Sensitivity (Detection Limit)Detects serum or urinary hCG at a concentration of 20 mIU/ml or greater.Detects serum or urinary hCG at a concentration of 20 mIU/ml or greater.
Cross-reactivity (Specificity)No cross-reaction with other hormones at specific levels (hLH, hFSH, hTSH, ß-core hCG)No cross-reaction with hLH (300 mIU/ml), hFSH (1,000 mIU/ml), hTSH (1,000 μIU/mL), or ß-core hCG (up to 8.53 pmol/L).
Precision (at POL sites)Consistent results across different lots, sites, and operators for negative, cutoff (20 mIU/ml), and various positive hCG concentrations.At 20 mIU/ml: - Serum: 96.9% Positive, 3.1% Negative - Urine: 96.9% Positive, 3.1% Negative (100% agreement for negative and higher positive concentrations)
InterferencePerformance not affected by specified pH, specific gravity, or presence of common substances (e.g., Acetaminophen, Albumin, Glucose).Performance not affected by urine pH 3.0-8.5, specific gravity 1.00-1.03, or specified concentrations of 15 common substances.

2. Sample Size Used for the Test Set and the Data Provenance

  • Sample Size: 145 clinical specimen sets (referring to the combined urine and serum sets for accuracy correlation studies).
    • Urine: 145 samples (59 positive, 86 negative)
    • Serum: 145 samples (59 positive, 86 negative - derived from the 58 positive, 1 negative, 86 negative results)
  • Data Provenance: Not explicitly stated as retrospective or prospective, nor is the country of origin specified beyond "clinical specimens." It's implied these are human clinical samples (urine and serum). The "POLs site study" suggests some data collected from Physician's Office Labs, which would be clinical settings.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

  • The 510(k) summary does not specify the number of experts or their qualifications for establishing the ground truth.

4. Adjudication Method for the Test Set

  • The 510(k) summary does not explicitly state an adjudication method. For the accuracy studies, the clinical serum specimens were "quantitatively confirmed by Abbott Architect i2000 instrument." This implies the Abbott Architect i2000 results served as the reference standard for the clinical specimens.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. This device is a rapid chromatographic immunoassay, which is a point-of-care test, not an AI-powered diagnostic imaging or interpretation device. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Yes, effectively. This device is the standalone "algorithm" (biochemical reaction and visual interpretation) for detecting hCG. Its performance metrics (sensitivity, specificity, precision, cross-reactivity, interference) are all measures of its standalone capability to detect hCG in samples and provide a visual result. The "human-in-the-loop" aspect is the visual reading by a healthcare professional, but the performance data presented is for the device's ability to produce the correct positive/negative signal.

7. The Type of Ground Truth Used

  • For the accuracy studies (clinical specimens): The ground truth for hCG concentration was established by quantitative confirmation using the Abbott Architect i2000 instrument (predicate device K093318). This is a well-established laboratory instrument for quantitative hCG measurement.
  • For sensitivity, cross-reactivity, precision, and interference studies: The ground truth was established using known concentrations of hCG (e.g., 0 mIU/ml, 10 mIU/ml, 15 mIU/ml, 20 mIU/ml, 40 mIU/ml, 100 mIU/ml) and known concentrations of interfering substances, often in "spiked" samples. This represents a form of analytical gold standard or controlled experimental conditions.

8. The Sample Size for the Training Set

  • This information is not provided in the 510(k) summary. For in vitro diagnostic (IVD) devices like this, the concept of a "training set" in the context of machine learning (where data is used to train an algorithm) is not directly applicable. The device's components (antibodies, reagents) are developed and optimized through R&D, and then validated with test sets.

9. How the Ground Truth for the Training Set Was Established

  • As mentioned above, the concept of a "training set" as commonly understood in AI/machine learning doesn't directly apply here. For the development and optimization of the test, the "ground truth" would have been established through well-characterized reference materials, known concentrations of hCG, and thorough analytical testing during the R&D phase to ensure the antibodies and test format functioned as intended. The 510(k) summary describes the validation testing, not the development process.

{0}------------------------------------------------

510(k) SUMMARY AS REQUIRED BY SECTION 807.92(C)

JUL 1 7 2012

The Assigned 510(k) number is: K112101

Date of Summary: July 13, 2012

Common Name: hCG (Human Chorionic Gonadotropin) Pregnancy Serum/Urine Combo Test

Regulatory Information:

    1. Regulation section: 21 CFR part 862.1155, Human Chorionic Gonadotropin test system
    1. Classification: Class II
    1. Product Code: JHI, radioimmunoassay, human chorionic gonadotropin
    1. Panel: Clinical Chemistry 75

Applicant and Initial Importer:

Polymed Therapeutics, Inc. 3040 Post Oak Blvd., Ste 1110 Huston, TX 77056 Tel.: (713) 777-7088 Fax: (713) 777-7091

Contact Persons:

Primary Contact:

Mrs. Feng-Yu Lee Correspondent for this Application Polymed Therapeutics, Inc. c/o IVDD Regulatory Consultant 27001 La Paz Road Suite 266B Mission Viejo, CA 92691 Tel: (949) 951-3046 Fax: (949) 951-2042 fengyulee@ivddreg.com

Alternate Only:

Terri Wallace . Polymed Therapeutics, Inc 3040 Post Oak Blvd., Ste 1110 Huston, TX 77056 Tel .: (713) 777-7088 Fax: (713) 777-7091 terri@polymedt.com

Identification / Product Name:

Fastep™ hCG Pregnancy Serum/Urine Test

{1}------------------------------------------------

510(k) SUMMARY (Cont.)

Description:

The Fastep™ hCG Pregnancy Serum/Urine Test are distributed in Cassette format. Each test reagent strip contains mouse monoclonal anti-a-hCG antibody coated membrane and a dried chemical pad containing mouse monoclonal anti-B-hCG anybody colloidal gold conjugate. The control antibodies are goat anti-mouse IgG.

Intended Use:

The Polymed Therapeutics' Fastep™ hCG Pregnancy Serum/Urine Test is rapid chromatographic immunoassays for the visual, qualitative detection of human chorionic gonadotropin (hCG) in serum or urine specimen to aid in the early detection of pregnancy. For Professional Use Only.

The test kits are for health care professionals use including professionals at physician's office labs (POLs) or point-of-care site (POC).

Predicate Kit:

Teco Diagnostics One-Step Urine/Serum Combo Pregnancy Card Test is used as predicate device for Polymed Therapeutics' Fastep104 hCG Pregnancy Serum/Urine Test to compare their performance of required studies. The clinical serum specimens used for accuracy study were quantitatively confirmed by Abbott Architect i2000 instrument.

510(k) numbers for predicate devices are:
Abbott Architect i2000 instrumentK093318
Teco Diagnostics One-Step Urine/Serum Combo Pregnancy Testk96446

Performance:

The product performance characteristics of Polymed Therapeutics' Fastep" hCG Pregnancy Serum/Urine Test was evaluated in the blind-labeled spiked control studies and blind-labeled clinical specimen correlation study (including POLs site study). The results of these studies demonstrate Polymed Therapeutics' Fastep104 hCG Pregnancy Serum/Urine Test to be substantially equivalent to the performance characteristics of Teco's One Step Pregnancy Urine/Serum Test. Correlation studies, using 145 clinical specimen sets, produced a minimum of 98.3% accuracy.

Accuracy Results:

A. Urine sample

Teco One-Step Combo Card Test
+-Total
FastepTMCassette Panel+59059
-08686
Total5986145

Sensitivity = (59 / (59+0)) x100 = 100.0%; Specificity = (86 / (86+0)) x 100 = 100.0%

{2}------------------------------------------------

510(k) SUMMARY (Cont.)

B. Serum sample

51 positive and 86 negative quantitative confirmed serum specimens were evaluated. The results of the accuracy finding calculated against predicate device are summarized and presented in the form of a 2x2 table (shown below):

Teco One-Step Combo Card Test
+-Total
FastepTMCassette Panel+58058
-18687
Total5986145

Sensitivity = 58/(58+1)) x100 = 98.3%; Specificity = (86/(86+0)) x 100 = 100.0%

Sensitivity and Cross-reactivity

The Fastep™ hCG Pregnancy Serum/Urine Test detects serum or urinary hCG at a concentration of 20mlU/ml or greater. The test has been standardized to the WHO Fourth International Standard 75/589.

Cross-reactivity (Specificity) evaluated at negative (0 mIU/ml) and positive (20 mIU/ml) hCG specimens showed no cross-reaction:

Compounds (level)% Non-cross- reactivity
hCG (20 mIU/ml)100%
hLH (300 mIU/ml)1,500%
hFSH (1,000mIU/ml)5,000%
hTSH (1,000μIU/mL)5%

Precision (POL sites)

Lab spiked hCG urine and serum controls were used to evaluate the precision performance of
Fastep™ hCG Pregnancy Serum/Urine Cassette and Dipstick Tests. 3 lots devices were 3 Point-of-care sites by 9 operators.

A. Serum controls

Levels(mIU/ml)No. ofNegativeNo. ofPositive%Negative%Positive
02250100%0%
102250100%0%
152250100%0%
2072183.1%96.9%
4002250%100%
10002250%100%

B. Urine controls

Levels(mIU/ml)No. ofNegativeNo. ofPositive%Negative%Positive
02250100%0%
102250100%0%

{3}------------------------------------------------

152250100%0%
2072183.1%96.9%
4002250%100%
10002250%100%

Interference

The performance of Fastep™ hCG Serum/Urine Pregnancy test at negative and cutoff points are not affected when the pH range of urine specimens is at 3.0 to 8.5 and the specific gravity range of urine specimens is at 1.00 to 1.03.

The following substances were added to hCG free and 20mIU/ml hCG spiked serum and urine samples. None of the substances at the concentration tested interfered with the assay.

SubstancesCon. inSerumCon. inUrine
Acetaminophen20 mg/dl20 mg/dl
AcetylsalicylicAcid20 mg/dl20 mg/dl
Albumin2000 mg/dl2000 mg/dl
Ascorbic Acid20 mg/dl20 mg/dl
Atropine20 mg/dl20 mg/dl
Bilirubin40 mg/dl2 mg/dl
Caffeine20 mg/dl20 mg/dl
EDTA80 mg/dl80 mg/dl
Ethanol1%1%
Gentesic Acid20 mg/dl20 mg/dl
Glucose2 g/dl2 g/dl
Hemoglobin125 mg/dl1 mg/dl
Methanol1 %1 %
Salicylic Acid20 mg/dl20 mg/dl
Triglyceride1200 mg/dlN/A

*: ß-core hCG level up to 8.53 pmol/L does not interfere with the assay.

Conclusion:

Of these specimen sets tested in serum and urine device, results of Accuracy, POL site study demonstrate the substantial equivalency between Polymed Therapeutics' Fastep" hCG Pregnancy Serum/Urine Test and the Teco One- Step Pregnancy Urine/Serum Test panel. It is also demonstrated that Polymed Therapeutics' Fastep™ hCG Pregnancy Serum/Urine Test is safe and effective in detecting human chorionic gonadotropin (hCG) in serum or urine sample to aid in the early determination of pregnancy.

Polymed Therapeutics, Inc

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is an abstract image of an eagle.

10903 New Hampshire Avenue Silver Spring, MD 20993

Polymed Therapeutics, Inc c/o Feng-Yu Lee 27001 La Paz Road Suite 266B Mission Viejo, CA 92691

JUL 17 2012

K112101 Re:

Trade Name: Fastep™ hCG Pregnancy Serum/Urine Cassette Tests Regulation Number: 21 CFR §862.1155 Regulation Name: Human chorionic gonadotropin (hcg) test system Regulatory Class: Class II Product Codes: JHI Dated: June 20, 2012 Received: July 9, 2012

Dear Feng-Yu Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{5}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, 1 please note the regulation entitled, "Misbranding by reference to premarket notification" (21 prodo note the rogarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical

Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ...

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm

Sincerely vours.

C. Hsiung, Ph.D.

Counney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use Form

510(k) Number (if known): K11210 |

Device Name: Polymed Therapeutics' Fastep™ hCG Pregnancy Serum/Urine Test

Indications for Use:

The Polymed Therapeutics' Fastep™ hCG Pregnancy Serum/Urine Test is a rapid chromatographic immunoassays for the visual, qualitative detection of human chorionic gonadotropin (hCG) in serum or urine specimen to aid in the early detection of pregnancy. For Professional Use Only.

The test kits are for health care professionals use including professionals at physician's office labs (POLs) and Point-of-Care site (POC).

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Juandria Ono

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K 112101

Page 1 of 1

§ 862.1155 Human chorionic gonadotropin (HCG) test system.

(a)
Human chorionic gonadotropin (HCG) test system intended for the early detection of pregnancy —(1)Identification. A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class II.(b)
Human chorionic gonadotropin (HCG) test system intended for any uses other than early detection of pregnancy —(1)Identification. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class III.(3)
Date PMA or notice of completion of a PDP is required. As of the enactment date of the amendments, May 28, 1976, an approval under section 515 of the act is required before the device described in paragraph (b)(1) may be commercially distributed. See § 862.3.