K Number
K111992

Validate with FDA (Live)

Date Cleared
2012-01-12

(183 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Curos is intended for use on swab-able luer access valves as a disinfecting cleaner prior to line access and to act as a physical barrier to contamination between line accesses. Curos ™ will disinfect the valve three (3) minutes after application and act as a physical barrier to contamination for up to seven (7) days (168 hours) if not removed. The effectiveness of Curos Protectors were tested in vitro against Staphylococcus aureus. Staphylococcus epidermidis, Escherichia coli and Pseudomonas aeruginosa, Candida glibrata, Candida albicans and was found to have >4 log reduction. The Curos Port Protector may be used in the home or healthcare facility.

Device Description

The Curos™ Port Protector device is a single use, sterile device that contains 70% lsopropyl Alcohol and is intended to be used as a disinfectant for needleless luer activated valves

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and the study that proves the device meets those criteria:

1. Table of Acceptance Criteria and Reported Device Performance

OrganismAcceptance Criteria (bacterial count reduction (Δ Log))3 minute exposure (bacterial count reduction (Δ Log))7 day (168 hours) exposure (bacterial count reduction (Δ Log))
Staphylococcus aureus≥ 4.06.06.9
Staphylococcus epidermidis≥ 4.06.87.3
Escherichia coli≥ 4.05.25.2
Pseudomonas aeruginosa≥ 4.05.15.1
Candida Albicans≥ 4.05.6>4.8*
Candida Glabrata≥ 4.05.4>5.3*

Note: The entire population was killed, the reduction in bacteria count was limited to the population quantified by the positive controls used in the testing.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample size (number of times each test was run or number of devices tested) for the efficacy test set. It mentions "a total of 4 bacteria, 2-gram negative and 2 gram positive" and "two selected fungus/yeast micro-organisms." This implies a set of 6 distinct microorganisms were tested.

The data provenance is a non-clinical performance test data and appears to be retrospective in the context of this 510(k) submission, as it was conducted to support the updated indications for use. The country of origin of the data is not specified, but the submission is to the US FDA by a company based in San Diego, California, suggesting the studies were likely conducted in the US or under US regulatory guidelines.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. The ground truth for this type of disinfectant efficacy study is typically established by laboratory testing using standardized microbiological methods to quantify bacterial reduction, rather than expert consensus on individual cases.

4. Adjudication Method for the Test Set

This information is not applicable to this type of study. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies where human readers are making subjective assessments that need to be reconciled. For a quantitative microbiology test, laboratory standards and controls dictate the results.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

A MRMC comparative effectiveness study was not done. This type of study focuses on comparing human reader performance, often with and without AI assistance, which is not relevant to the Curos Port Protector's disinfectant efficacy.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

A standalone performance study was done. The efficacy testing quantifies the disinfectant capability of the "Curos Port Protector" device itself, without human intervention being a variable in the outcome of the log reduction, beyond proper application in the test setup. This is a direct measure of the device's inherent antimicrobial properties.

7. Type of Ground Truth Used

The ground truth used is based on quantitative microbiological laboratory results (bacterial count reduction, Δ Log) against specific microorganisms. This is a direct measurement of the device's ability to disinfect.

8. Sample Size for the Training Set

This information is not provided and is not applicable. The Curos Port Protector is a medical device that uses chemical and physical barriers for disinfection, not an AI/machine learning algorithm. Therefore, there is no "training set" in the context of AI model development. The data presented is performance data for the physical product.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the reasons stated above (no AI/machine learning involved, thus no training set).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services USA logo, which features a stylized eagle. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

Ivera Medical, Inc. Don Canal Regulatory Affairs, Quality Manager 3525 Del Mar Heights Road, Suite #430 San Diego, California 92130

March 11, 2022

Re: K111992

Trade/Device Name: Curos Port Protector Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: QBP

Dear Don Canal:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated January 12, 2012 and the correction letter dated December 14, 2018. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under regulation 880.5440.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Payal Patel, OHT3: Office of GastroRenal, Ob-Gyn, General Hospital and Urology Devices, 240-402-6029, Payal.Patel(@fda.hhs.gov.

Sincerely,

Pror R

Payal Patel Assistant Director for General Hospital Devices DHT3C: Division of Drug Delivery and General Hospital Devices and Human Factors OHT3: Office of GastroRenal, Ob-Gyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol is a circular emblem with a stylized depiction of a caduceus, a symbol associated with medicine and healthcare. The text is arranged in three lines, with "FDA" in a bold, sans-serif font, followed by "U.S. FOOD & DRUG" and "ADMINISTRATION" in a smaller, sans-serif font.

December 14, 2018

Ivera Medical, Inc. Don Canal RA/QA 2731 Loker Avenue West Carlsbad, California 92010

Re: K111992

Trade/Device Name: Curos Port Protector Regulatory Class: Unclassified Product Code: OBP Dated: July 13, 2011 Received: July 13, 2011

Dear Don Canal:

This letter corrects our substantially equivalent letter of January 12, 2012.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov

{2}------------------------------------------------

K111992 - Don Canal

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice

(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

FDA Tina Kiang -

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{3}------------------------------------------------

Indications for Use Statement 4.

The Curos is intended for use on swab-able luer access valves as a disinfecting cleaner prior to line access and to act as a physical barrier to contamination between line accesses. Curos ™ will disinfect the valve three (3) minutes after application and act as a physical barrier to contamination for up to seven (7) days (168 hours) if not removed. The effectiveness of Curos Protectors were tested in vitro against Staphylococcus aureus. Staphylococcus epidermidis, Escherichia coli and Pseudomonas aeruginosa, Candida glibrata, Candida albicans and was found to have >4 log reduction. The Curos Port Protector may be used in the home or healthcare facility.

Prescription Use

AND/OR Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS UNE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Khel C. Chapa 1/9/12

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K111992

Page 14 of 139

Title: <

Ivera Medical Curos 510(k) Notification

Confidential

{4}------------------------------------------------

JAN 1 2 2012

510(k) Summary

General Company Information

Name:Ivera Medical Corporation
Contact:Don CanalConsultant RAQA
Address:Ivera Medical Corporation3525 Del Mar Heights RoadSuite #430San Diego, CA 92130
Telephone:760-612-6090
Fax:858-228-1770

Date Prepared: October 23, 2011

General Device Description

The Curos™ Port Protector device is a single use, sterile device that contains 70% lsopropyl Alcohol and is intended to be used as a disinfectant for needleless luer activated valves

Common Name:Pad, Alcohol
Trade Name:CurosTM Port Protector
Classification:Unclassified Device, product Code LKE

Predicate Devices

K080466 Curos Port Protector, Ivera Medical Corporation K083508 SwabCap, Excelsior Medical Corporation

Intended Use (Indications)

The Curos is intended for use on swab-able luer access valves as a disinfecting cleaner prior to line access and to act as a physical barrier to contamination between line accesses. Curos ™ will disinfect the valve three (3) minutes after application and act as a physical barrier to contamination for up to seven (7) days (168 hours) if not removed. The effectiveness of Curos Protectors were tested in vitro against Staphylococcus aureus, Staphylococcus epidermidis, Escherichia coli and Pseudomonas aeruginosa, Candida glibrata, Candida albicans and was found to have >4 log reduction. The Curos Port Protector may be used in the home or healthcare facility.

{5}------------------------------------------------

Comparison with Predicate Device

There is no change to the device for this 510(k) notification, the only change is to the Intended Use (indications) to reflect the test data for disinfection which includes 3 minutes to 7 days (168 hours), as described below. The materials of construction and technological characteristics are equivalent to the predicate device.

CharacteristicSubject DevicePredicate DevicePredicate Device
Device nameCuros Port ProtectorCuros Port ProtectorSwabCap
Common NameAlcohol, disinfecting padAlcohol, disinfecting padAlcohol, disinfecting pad
ManufacturerIvera MedicalIvera MedicalExcelsior Medical Corporation
510(k) numberK110826K110826K083508
Regulationnumber,product codeUnclassified,Preamendment device,product code: LKBUnclassified,Preamendment device,product code: LKBUnclassified,Preamendment device, product code:LKB
Indications foruseThe Curos is intended for use on swab-able luer access valves as a disinfecting cleaner prior to line access and to act as a physical barrier to contamination between line accesses. Curos ™ will disinfect the valve three (3) minutes after application and act as a physical barrier to contamination for up to seven (7) days (168 hours) if not removed. The effectiveness of Curos Protectors were tested in vitro against Staphylococcus aureus, Staphylococcus epidermidis, Escherichia coli and Pseudomonas aeruginosa, Candida glibrata. Candida albicansThe Curos Port Protector is a device containing 70% Isopropyl Alcohol when left in place for 5 to15 minutes, the Curos Port Protector decontaminates the injection Port; thereafter the Curos Port Protector provides a physical barrier during the intended use.SwabCap is intended for use on swab-able luer access valves as a disinfecting cleaner prior to line access and to act as a physical barrier to contamination between line accesses. SwabCap will disinfect the valve five (5) minutes after application and act as a physical barrier to contamination for up to ninety-six (96) hours under normal conditions if not removed.
CharacteristicSubject DevicePredicate DevicePredicate Device
and was found to have >4log reduction. The CurosPort Protector may be usedin the home or healthcarefacility.
Disinfectant -activeingredient70% Isopropyl Alcohol70% Isopropyl Alcohol70% IsopropylAlcohol
Length0.40 inches0.40 inches0.50 inches
Diameter0.54 inches0.54 inches0.60 inches
User PopulationHome and hospital useHome and hospital useHome and hospitaluse
Colorants Used(type, amount,concentration)Translucent green, moldedplastic, 3% concentrationTranslucent green,molded plastic, 3%concentrationOrange pigment, %concentrationunknown
Provided SterileYesYesYes
Single UseDeviceYesYesYes
Plastic Housingto remain inplaceYesYesYes

Subject Device to Predicate Technological Comparison Table

{6}------------------------------------------------

Substantial Equivalence Performance Testing

Ivera medical has provided non-clinical performance test data that demonstrates the pre-defined acceptance criteria for a disinfecting device has been met. This acceptance criteria is defined as a bacteria count reduction of > 4 log reduction of 2 selected gram positive bacteria, 2 selected gram negative bacteria, and two selected fungus/yeast micro-organisms for a period of time from 3 minutes up to 168 hours (7 days). The efficacy testing was completed using a total of 4 bacteria, 2-gram negative and 2 gram positive as recommended in Draft Guidance for Industry and FDA Staff Premarket Notification [510(k)] Submissions for Medical Devices that Include Antimicrobial Agents DRAFT GUIDANCE. This guidance document is being distributed for comment purposes only. Document issued on: July 19, 2007. The test results are summarized in Table 1.

{7}------------------------------------------------

OrganismAcceptanceCriteria(bacterial countreduction ( $\Delta$ Log))3 minute exposure(bacterial countreduction ( $\Delta$ Log))7 day (168 hours)exposure(bacterial countreduction ( $\Delta$ Log))
Staphylococcus aureus$\geq$ 4.06.06.9
Staphylococcus epidermis$\geq$ 4.06.87.3
Escherichia coli$\geq$ 4.05.25.2
Pseudomonas aeruginosa$\geq$ 4.05.15.1
Candida Albicans$\geq$ 4.05.6>4.8*
Candida Glabrata$\geq$ 4.05.4>5.3*

Table 1 - Efficacy Test Results

  • The entire population was killed, the reduction in bacteria count was limited to the population quantified by the positive controls used in the testing.

The Ivera Curos Port Protector is sterilized using a validated Gamma sterilization process which complies with ISO11137-1:2006/(R) Sterilization of health care products -Radiation - Part 2: Establishing the sterilization dose. Recognition number 14-225.

ISO11137-2:2006 Sterilization of health care products – Radiation – Part 1: requirements for development of validation and routine control of sterilization process for medical devices. Recognition number 14-297.

11137-3:2006/(R) 2010 10/04/2010 AAMI ANSI ISO 14-298 - Radiation - Part 3: Guidance on Dosimetric Aspects. Recognition number 14-298.FDA recognized standard ISO11137 Sterilization Standard.

Ivera Medical has completed testing to demonstrate the Curos Port Protector materials of construction meet FDA recognized standard ISO10993 for biocompatibility.

Conclusion

The analysis arguments and test results demonstrate the Curos™ device is safe for its intended use and is substantially equivalent to the predicate devices.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.