K Number
K111698

Validate with FDA (Live)

Date Cleared
2011-08-19

(63 days)

Product Code
Regulation Number
888.3330
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PROFEMUR® E Cementless Hip Stem is intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients.

Indications For Use:
The PROFEMUR® E Cementless Hip Stem is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:

  1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
  2. inflammatory degenerative joint disease such as rheumatoid arthritis;
  3. correction of functional deformity; and,
  4. revision procedures where other treatments or devices have failed

The PROFEMUR® E Cementless Hip Stems are single use components, intended for use in conjunction with associated ceramic or metal femoral heads as part of an uncemented total hip arthroplasty.

Device Description

The modular PROFEMUR® E Cementless Hip Stems hip stem is an uncemented distal femoral implant designed to couple with Wright Medical Technology's PROFEMUR® CoCr modular necks. The stem design was developed from the previously marketed PROFEMUR® Z Hip Stem (STEM) K021346) and possesses the identical modular neck taper socket and is made of the identical Titanium alloy (ASTM F620) as previous hip stem designs by Wright.

The PROFEMUR® E Cementless Hip Stem is made of the same Titanium alloy and features a corundum blast surface finish similar to the PROFEMUR® Z Hip Stem (K021346). The PROFEMUR® E Cementless Hip Stem shares the rectangular cross-section of the predicate devices, and is designed to feature additional tapered lateral fins on each stem side.

AI/ML Overview

Here's an analysis of the provided text regarding the PROFEMUR® E Cementless Hip Stem, focusing on the acceptance criteria and the study proving it.

Important Note: The provided text is a 510(k) Summary. A 510(k) submission generally aims to demonstrate substantial equivalence to a predicate device, rather than proving that a new device independently meets specific performance acceptance criteria through clinical studies. The focus is on showing the new device is as safe and effective as a legally marketed predicate device. Therefore, the "acceptance criteria" here are implicitly related to demonstrating this substantial equivalence, and the "study" is primarily non-clinical testing.


1. Table of Acceptance Criteria and Reported Device Performance

Based on the 510(k) summary, the device's acceptance criteria are framed in terms of its substantial equivalence to the predicate device, particularly in its mechanical performance.

Acceptance Criterion (Implicit)Reported Device Performance
Mechanical Performance Equivalent to Predicate Device:
Proximal Fatigue PerformanceSimilar to PROFEMUR® Z Hip Stem (K021346)
Distal Fatigue PerformanceSimilar to PROFEMUR® Z Hip Stem (K021346)
Fretting PerformanceSimilar to PROFEMUR® Z Hip Stem (K021346)
Distraction PerformanceSimilar to PROFEMUR® Z Hip Stem (K021346)
Material Composition Equivalence:
Constructed from same Titanium alloyMade of identical Titanium alloy (ASTM F620)
Surface Finish Equivalence:
Features a similar surface finishFeatures a corundum blast surface finish similar to predicate
Design Feature Equivalence (Modular Neck Taper Socket):
Possesses identical modular neck taper socketPossesses identical modular neck taper socket
Intended Use & Indications for Use Equivalence:
Identical to existing Wright PROFEMUR® hip stemsIdentical to those for all existing Wright PROFEMUR® hip stems

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated as a number of devices or clinical subjects. The testing was non-clinical (i.e., laboratory testing on devices).
  • Data Provenance: The document does not specify a country of origin for the data beyond stating "Wright Medical Technology, Inc. 5677 Airline Rd. Arlington, TN 38002" as the submitter. The testing itself involved mechanical evaluations of the device, not clinical data from patients. The study was non-clinical non-human.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This was a non-clinical, mechanical evaluation comparing a device to a predicate. It did not involve expert consensus on medical images or diagnoses.

4. Adjudication Method for the Test Set

Not applicable. This was a non-clinical, mechanical evaluation.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The submission explicitly states: "Clinical data was not provided for the class III modular hip stem."

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This device is a physical, implantable medical device, not a software algorithm.

7. The Type of Ground Truth Used

The "ground truth" in this context is the established mechanical performance and material specifications of the predicate device (PROFEMUR® Z Hip Stem, K021346) and other previously cleared PROFEMUR® hip stems. The new device's performance was compared against these established standards.

8. The Sample Size for the Training Set

Not applicable. There is no concept of a "training set" for this type of non-clinical mechanical testing of a medical implant.

9. How the Ground Truth for the Training Set Was Established

Not applicable. See point 8.

{0}------------------------------------------------

K111698(pg 1/2)

Image /page/0/Picture/1 description: The image shows the logo for Wright Medical Technology. The logo consists of a stylized "W" made up of three black bars, followed by the words "WRIGHT MEDICAL TECHNOLOGY" stacked on top of each other. Below the logo is the text "A Wright Medical Group Company."

AUG 1 9 2511

A Wright Medical Group Company

510(k) Summary of Safety and Effectiveness

In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the use of the PROFEMUR® E Cementless Hip Stem.

(a)(1) Submitted By Submitter's Name:

Date:

Contact Person:

(a)(2) Device Name Proprietary Name:

Common Name:

Classification Name and Reference:

Subject Device Product Code and Panel Code:

(a)(3) Predicate Device Predicate Proprietary Name: Predicate Classification and Number:

Wright Medical Technology, Inc. 5677 Airline Rd. Arlington, TN 38002 800-238-7188 (phone), 901-867-4190 (fax)

August 19, 2011

Gregory Neal Regulatory Affairs Specialist II

PROFEMUR® E Cementless Hip Stem

Hip Stem

21 CFR 888.3320 Hip joint metal/metal semiconstrained. with an uncemented acetabular component, prosthesis Class III

Orthopedics/87/ JDL. KWA, I.PH, LZO

STEM Hip Replacement System 888.3358 LPH Hip joint metal/polymer/ metal semiconstrained porous-coated uncemented prosthesis Class II (510k-K021346)

PRO-FEMUR Hip System 888.3353 LZO Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Class II (510k-K012091)

Metal TRANSCEND Articulation System 888.3320 JDL Hip joint metal/metal semiconstrained, with a cemented acetabular component prosthesis Class III

888.3330 KWA Hip joint metal/metal semiconstrained, an uncemented acetabular with component prosthesis Class III (510k-K004043)

(a)(4) Device Description

The modular PROFEMUR® E Cementles Hip Stems hip stem is an uncemented distal femoral implant designed to couple with Wright Medical Technology's PROFEMUR® CoCr modular necks. The stem design was developed , from the previously marketed PROFEMUR® Z Hip Stem (STEM) K021346) and possesses the identical modular neck taper socket and is made of the identical Titanium alloy (ASTM F620) as previous hip stem designs by Wright,

{1}------------------------------------------------

(a)(5) Intended Use of the Device Intended Use

The PROFEMUR® E Cementless Hip Stem is intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients.

Indications for Use

  • non-inflammatory degeneralive joint disease such as osteoarthritis. avascular necrosis, ankylosis, protrusio 1. acetabuli, and painful hip dysplasia;
  • inflammatory degenerative joint disease such as rheumatoid arthritis; 2.
    1. correction of functional deformity: and,
  • revision procedures where other treatments or devices have failed 4.

The PROFEMUR® E Cementless Hip Stem is a single use component, intended for use in conjunction with associated ceramic or metal femoral heads as part of an uncemented total hip arthroplasty.

(a)(6) Technological Characteristics of the Device

The indications for use for the PROFEMUR® E Cementless Hip Stem are identical to those for all existing Wright PROFEMUR® hip stems. The indications are applied to all PROFEMUR® hip stems and have been cleared in 510/k)s: K041114, K041586, K051995, K052915, K053588, K060663, K081090, K091423, and K 100866.

The PROFEMUR® E Cementless Hip Stem is made of the same Titanium alloy and features a corundum blast surface finish similar to the PROFEMUR® Z Hip Stem (K021346). The PROFEMUR® E Cementless Hip Stem shares the rectangular cross-section of the predicate devices, and is designed to feature additional tapered lateral fins on cach stem side.

(b)(1) Nonclinical Testing

The PROFEMUR® E Cementless Hip Stem represents a minor change to the PROFEMUR® Z Hip Stem (K021346). Proximal and distal fatigue. fretting, and distraction evaluations were conducted on the PROFEMUR® E Cementless Hip Stem. The results were similar to the PROFEMUR® Z Hip Stem (K021346).

(b)(2) Clinical Testing

Clinical data was not provided for the class III modular hip stem.

(b)(3) Conclusions

The indications for use of the PROFEMUR® E Cementless Hip Stem are identical to the previously cleared predicate devices. The design features and materials of the subject devices are substantially equivalent to those of the predicate devices. The fundamental scientific technology of the subject device has not changed relative to the predicate devices. The safety and effectiveness of the PROFEMUR® E Cementless Hip Stem is adequately supported by the substantial equivalence information, and analysis data provided within this Premarket Notification.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure with outstretched arms.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Wright Medical Technology, Inc. % Mr. Gregory Neal Regulatory Affairs Specialist II 5677 Airline Road Arlington, Tennessee 38002

AUG 19 2011

Re: K111698

Trade/Device Name: PROFEMUR® E Cementless Hip Stem Regulation Number: 21 CFR 888.3330 Regulation Name: Hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis Regulatory Class: Class III Product Code: KWA, JDL, LPH, LZO Dated: July 21, 2011 Received: July 22, 2011

Dear Mr. Neal:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

Page - 2 - Mr. Gregory Neal

comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark A. Millman

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

KIII698 (pg 1/1)

Image /page/4/Picture/1 description: The image shows the word "WRIGHT." in a bold, sans-serif font. Below the word is a graphic of two stylized, overlapping shapes that resemble wings or chevrons. The overall design is simple and monochromatic, with a focus on the typography and the geometric shapes.

Indications for Use

Device Name: PROFEMUR® E Cementless Hip Stem

Indications For Use:

The PROFEMUR® E Cementless Hip Stem is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:

  • non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, 1. ankylosis, protrusio acetabuli, and painful hip dysplasia;
  • inflammatory degenerative joint disease such as rheumatoid arthritis; 2.
  • correction of functional deformity; and, 3.
  • revision procedures where other treatments or devices have failed 4.

The PROFEMUR® E Cementless Hip Stems are single use components, intended for use in conjunction with associated ceramic or metal femoral heads as part of an uncemented total hip arthroplasty.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

for. m melker
Page

(Division Sign-Off) (Division Sign-Old)
Division of Surgical, Orthopedic, Division of Surg
and Restorative Devices

K111698
510(k) Number

Page 1 of 1 1

§ 888.3330 Hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis.

(a)
Identification. A hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis is a two-part device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a femoral and an acetabular component, both made of alloys, such as cobalt-chromium-molybdenum. The femoral component is intended to be fixed with bone cement. The acetabular component is intended for use without bone cement (§ 888.3027).(b)
Classification. Class III.(c)
Date PMA or notice of completion of PDP is required. A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 18, 2016, for any hip joint metal/metal semi-constrained prosthesis with an uncemented acetabular component that was in commercial distribution before May 28, 1976, or that has, on or before May 18, 2016, been found to be substantially equivalent to a hip joint metal/metal semi-constrained prosthesis with an uncemented acetabular component that was in commercial distribution before May 28, 1976. Any other hip joint metal/metal semi-constrained prosthesis with an uncemented acetabular component shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.