(158 days)
OncoTrac™ is a software based Picture Archiving and Communication System (PACS) used with general purpose computing hardware for the display and visualization of medical image data. It provides for communication, storage, processing, rendering, and display of DICOM compliant image data derived from various sources including CT and MRI, measurement of lesions identified by trained users, tabulation of measurements, categorization of tumor response in accordance with user selected standards, and generation of a structured imaging report.
OncoTrac™ is intended for use as a diagnostic, review, and analysis tool by trained professionals such as physicians, technologists, and nurses. When interpreted by a trained physician, reviewed images may be used as an element for diagnosis. OncoTrac™ does not provide or claim any automatic detection or automatic diagnosis of abnormal anatomy, structure, or function.
OncoTrac™ is not intended for use for mammography.
OncoTrac™ is a software based Picture Archiving and Communication System (PACS) used with general purpose computing hardware for the display and visualization of medical image data. It provides for communication, storage, processing, rendering, and display of DICOM compliant image data derived from various sources including CT and MRI.
OncoTracTM runs on either a native or virtualized Microsoft Windows platform. Available functions include communication, storage, processing, rendering, and display of DICOM compliant image data derived from various sources including CT and MRI. measurement of lesions identified by trained users, tabulation of measurements, categorization of tumor response in accordance with user selected standards, and generation of a structured imaging report. The user controls these functions with a system of interactive menus and tools.
The provided text is a 510(k) summary for the OncoTrac™ software, dated November 18, 2011. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed study showing the device meets specific acceptance criteria. Based on the provided text, a comprehensive study proving acceptance criteria has not been detailed in a way that allows for the creation of the requested table and detailed analysis.
Here's what can be extracted and what information is missing:
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not explicitly state formal acceptance criteria with numerical targets. Instead, it relies on comparison to a predicate device (GE AW Server K081985) and internal testing for software quality. The "features" table (Document 2) serves as the closest resemblance to acceptance criteria, where the OncoTrac™ is compared functionally to the predicate.
| Feature | Acceptance Criteria (Implied by Predicate) | Reported Device Performance (OncoTrac™) |
|---|---|---|
| DICOM compliance | Yes | Yes |
| 2D Imaging | 2D image viewer | 2D image viewer |
| Interactive user controls | Yes | Yes |
| Lesion measurement | Yes | Yes |
| Body region(s) | Multiple | Multiple |
| Modalities | CT, MRI | CT, MRI |
| Quantitative oncology response assessment | RECIST, WHO | RECIST, WHO |
| Generation of structured report | ||
| Prescription Use | Yes | Yes |
| Intended Users | Trained Professionals | Trained Professionals |
| Software Testing | Passed critical tests for patient safety, acceptable overall performance. | Passed all tests considered critical in terms of patient safety and demonstrated an overall acceptable performance. |
| Hazard Analysis | Conducted | Conducted (level of concern classified as moderate) |
Missing Information/Cannot Determine from Text:
- Specific quantitative performance metrics: The document confirms feature presence but doesn't provide success rates, accuracy figures, or error margins for tasks like lesion measurement accuracy, response assessment concordance, or reporting completeness/accuracy.
- Detailed acceptance criteria for software itself: While "passed all tests considered critical" is stated, the specific criteria for these tests (e.g., bug count, performance benchmarks, usability scores) are not provided.
2. Sample Size Used for the Test Set and Data Provenance:
The document mentions that "The OncoTrac™ software has been extensively tested on Windows 64 bit systems by members of the development and quality control teams." However, it does not specify:
- The sample size of medical images or patient cases used for testing.
- The provenance of this data (e.g., country of origin, retrospective or prospective nature).
- Whether a specific "test set" in the context of machine learning model evaluation was used, as this is a PACS software, not an AI model in the modern sense.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:
Not applicable or not mentioned in the provided text. The document refers to "lesions identified by trained users" and "categorization of tumor response in accordance with user selected standards," implying human interpretation. However, there's no mention of a formal ground truth establishment process involving a specific number of experts or their qualifications for a dedicated test set used to validate the software's performance against a gold standard.
4. Adjudication Method for the Test Set:
Not applicable or not mentioned. Given the nature of a PACS for display and measurement, and the lack of a defined test set for "ground truth" establishment by multiple experts, an adjudication method is not described.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:
No, an MRMC comparative effectiveness study is not mentioned. The 510(k) relies on substantial equivalence to a predicate device based on functional comparison, not on demonstrating improved human reader performance with or without AI assistance. The OncoTrac™ is described as a tool that "does not provide or claim any automatic detection or automatic diagnosis of abnormal anatomy, structure, or function."
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:
The concept of "standalone performance" as it relates to an AI algorithm is not applicable here. OncoTrac™ is described as a software-based PACS for display, visualization, measurement, and reporting, guided by "trained users" and "trained professionals." It's an interactive tool, not an autonomous diagnostic algorithm. The software performs measurements and categorizations based on user input, not entirely independently.
7. The Type of Ground Truth Used:
The document refers to "user selected standards" for tumor response categorization (RECIST, WHO). This implies that the software's ability to categorize is based on widely accepted clinical standards, where the "ground truth" would be established by the consensus of the medical community on how to apply these standards. For lesion measurements, the "ground truth" would be the manually identified and measured lesion dimensions by a trained user, which the software then processes, tabulates, and reports. There's no mention of pathology or outcomes data being directly used as a ground truth for validating the software's core functions.
8. The Sample Size for the Training Set:
Not applicable. The OncoTrac™ described in this 510(k) is an image viewing, processing, and reporting software, not a machine learning model that requires a "training set" in the conventional sense. The software's development likely involved standard software engineering practices, testing, and debugging, but not a dataset for training an AI model.
9. How the Ground Truth for the Training Set Was Established:
Not applicable, as there is no "training set" for an AI model.
{0}------------------------------------------------
OncoTrac™ 510(k)
Image /page/0/Picture/1 description: The image shows the text "K111642" on the first line. The second line says "Page 1 of 4". The text appears to be handwritten in black ink on a white background. The image likely represents a page number within a larger document.
NOV 1 8 2011
510(k) Summary 5
Date Prepared:
May 23, 2011
Submitter/Importer/Distributer Name: Submitter's Address:
Submitter's Phone: Submitter's Fax: Contact: Manufacturer:
Proprietary Name: Common Name: Classification:
Substantially Equivalent to:
Translational Sciences Corporation Translational Sciences Corporation. One Mifflin Place, Suite 400 Cambridge, MA 02138 617.297.2577 617.576.9132 Howard Pinsky, Chief Executive Officer Mint Medical Systems, GmbH Im Neuenheimer Feld 582 D-69120 Heidelberg Germany OncoTracTM Software PACS 892.2050 Picture archiving and communications system, Product Code LLZ, (Class II)
Tradename: AW Server Manufacturer: General Electric Medical Systems 510(k) Number: K081985
Confidential
13
{1}------------------------------------------------
OncoTrac™ 510(k)
Translational Sciences Corporation
Device Description:
OncoTrac™ is a software based Picture Archiving and Communication System (PACS) used with general purpose computing hardware for the display and visualization of medical image data. It provides for communication, storage, processing, rendering, and display of DICOM compliant image data derived from various sources including CT and MRI.
KIII642
Page 2044
OncoTracTM runs on either a native or virtualized Microsoft Windows platform. Available functions include communication, storage, processing, rendering, and display of DICOM compliant image data derived from various sources including CT and MRI. measurement of lesions identified by trained users, tabulation of measurements, categorization of tumor response in accordance with user selected standards, and generation of a structured imaging report. The user controls these functions with a system of interactive menus and tools.
The OncoTrac™ software has been extensively tested on Windows 64 bit systems by members of the development and quality control teams. A hazard analysis has been conducted and the level of concern has been classified as moderate. The release version of the software passed all tests considered critical in terms of patient safety and demonstrated an overall acceptable performance.
Confidential
14
{2}------------------------------------------------
OncoTrac™ 510(k
| K111642 | ||||
|---|---|---|---|---|
| k) | Page 3 of 4 |
| Feature | OncoTracTM | GE AW Server(primary predicate) |
|---|---|---|
| 510k number | K081985 | |
| DICOM compliance | yes | yes |
| 2D Imaging | 2D image viewer | Same plus 3D viewer |
| Interactive user controls | yes | yes |
| lesion measurement | yes | yes |
| body region(s) | multiple | multiple |
| modalities | CT, MRI | CT, MRI |
| Quantitative oncology response assessment | RECISTWHO | RECISTWHO |
| Generation of structured report | same | |
| Prescription Use | Yes | same |
| Intended Users | Trained Professionals | same |
Substantial Equivalence Comparisons to Predicate Devices:
Confidential
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
. .....
.
{3}------------------------------------------------
OncoTrac™ 510(k)
Intended Use:
OncoTrac™ is a software based Picture Archiving and Communication System (PACS) used with general purpose computing hardware for the display and visualization of medical image data. It provides for communication, storage, processing, rendering, and display of DICOM compliant image data derived from various sources including CT and MRI, measurement of lesions identified by trained users, tabulation of measurements. categorization of tumor response in accordance with user selected standards, and generation of a structured imaging report.
OncoTrac™ is intended for use as a diagnostic, review, and analysis tool by trained professionals such as physicians, technologists, and nurses. When interpreted by a trained physician, reviewed images may be used as an element for diagnosis. OncoTrac™ does not provide or claim any automatic detection or automatic diagnosis of abnormal anatomy, structure, or function.
OncoTrac™ is not intended for use for mammography.
Confidential
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Allen Green, MD, PHD, NJ, LLC Counsel Translational Sciences Corporation One Mifflin Place, Suite 400 CAMBRIDGE MA 02138
NOV 1 8 2011
Re: K111642
Trade/Device Name: OncoTrac Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 2, 2011 Received: November 4, 2011
Dear Mr. Green:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{5}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary Pastel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
KIII642
Indications for Use
Applicant: Translational Sciences Corporation. One Mifflin Place, Suite 400 Cambridge, MA 02138
510(k) Number (if known): KN1642
Device Name: OncoTrac™
Indications for Use:
Onco Trac™ is a software based Picture Archiving and Communication System (PACS) used with general purpose computing hardware for the display and visualization of medical image data. It provides for communication, storage, processing, rendering, and display of DICOM compliant image data derived from various sources including CT and MRI, measurement of lesions identified by trained users, tabulation of measurements, categorization of tumor response in accordance with user selected standards, and generation of a structured imaging report.
OncoTrac™ is intended for use as a diagnostic, review, and analysis tool by trained professionals such as physicians, technologists, and nurses. When interpreted by a trained physician, reviewed images may be used as an element for diagnosis. OncoTrac™ does not provide or claim any automatic detection or automatic diagnosis of abnormal anatomy, structure, or function.
OncoTracTM is not intended for use for mammography.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Richard D. O'Shea
(Division Sign Off)
Division of Radialo
10K Klll 642
Confidential
12
N/A