K Number
K111518

Validate with FDA (Live)

Date Cleared
2012-10-03

(489 days)

Product Code
Regulation Number
880.2800
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The STERRAD® Chemical Indicator Strip (PN 14100) is a process indicator (ISO 11140-1:2005) to differentiate processed from unprocessed packages and is intended for use with medical device packages to be sterilized in STERRAD® Sterilization Systems:

MODELCYCLE
STERRAD® 100SStandard
STERRAD® 50Standard
STERRAD® 200Standard
STERRAD® NX®Standard
Advanced
STERRAD® 100NX®Flex
EXPRESS
DUO
Device Description

STERRAD® Chemical Indicator Strip is a through-put process indicator strip to be used with STERRAD® Sterilization Systems. The STERRAD® Sterilization System utilizes hydrogen peroxide gas plasma to achieve rapid, low temperature sterilization of medical devices.

STERRAD® Chemical Indicator Strip functions by means of a chemical reaction. Exposure of the chemical indicator to the STERRAD® Sterilizer cycle results in a recognizable color change from red to yellow. The yellow (or lighter) chemical indicator bar indicates that the load has been exposed to hydrogen peroxide.

AI/ML Overview

The provided text describes the STERRAD® Chemical Indicator Strip, a through-put process indicator for STERRAD® Sterilization Systems. The primary function of the device is to differentiate processed from unprocessed packages by a recognizable color change from red to yellow upon exposure to hydrogen peroxide.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Chemical Indicator Functionality (response to hydrogen peroxide exposure)Passed
End Point / Post Processing Color StabilityPassed
Shelf lifePassed (based on prior submission K103218 as design has not changed)
BiocompatibilityPassed

2. Sample sized used for the test set and the data provenance

The document does not specify the exact sample size used for the performance tests (Chemical Indicator Functionality, End Point / Post Processing Color Stability, and Biocompatibility). It only states that "Performance testing was conducted to demonstrate the functionality of Chemical Indicator Strip in the STERRAD® 100NX® DUO Cycle". The data provenance is not explicitly stated in terms of country of origin, however, it is a nonclinical study, likely conducted by the manufacturer, Advanced Sterilization Products, which is located in Irvine, California, USA. The studies mentioned are nonclinical, and the nature of the device (chemical indicator) suggests lab-based testing rather than clinical trials with human subjects.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. The ground truth for a chemical indicator is determined by its physical and chemical properties and its observable color change. There's no mention of human expert evaluation for establishing the ground truth of the indicator's performance. The "ground truth" here is the confirmed exposure or non-exposure to hydrogen peroxide in a controlled sterilization cycle.

4. Adjudication method

Not applicable. Since there's no mention of human experts establishing ground truth or evaluating device performance in the context of diagnostic interpretation, there is no adjudication method described. The performance is based on objective chemical reactions and color changes.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a chemical indicator strip, not an AI-powered diagnostic tool for human interpretation. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is irrelevant and was not conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the testing described is effectively a standalone performance evaluation of the chemical indicator strip. The device's functionality is assessed based on its inherent chemical reaction and observable color change, independent of any human interpretation or intervention beyond observing the result. The "Chemical Indicator Functionality" study directly evaluates the "algorithm-only" performance of the chemical indicator.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth used for the performance studies is the controlled exposure or non-exposure to hydrogen peroxide in a STERRAD® Sterilization System. For example, to test "Chemical Indicator Functionality," the strip would be subjected to a validated STERRAD® cycle, and its color change would be compared against the known outcome of that cycle (i.e., whether it was successfully exposed to hydrogen peroxide).

8. The sample size for the training set

Not applicable. This device is a chemical indicator, not a machine learning or AI model that requires a training set. Its functionality is based on a chemical reaction, not a learned algorithm.

9. How the ground truth for the training set was established

Not applicable. As explained above, there is no training set for this device.

{0}------------------------------------------------

510(k) Summary

OCT 3 2012

Applicant's Name, Address, Telephone, FAX, Contact Person Advanced Sterilization Products 33 Technology Drive Irvine, CA 92618

Contact Person

Nancy Chu Manager. Regulatory Affairs (949) 453-6435 Telephone (949) 789-3900 Fax

Date: September 10, 2012

1. CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME

Classification Name: Common/Usual Name: Product Classification: Proprietary Name:

Sterilization Process Indicator Chemical Sterilization Process Indicator II STERRAD® Chemical Indicator Strip

2. PREDICATE DEVICE

STERRAD® Chemical Indicator Strip (K103218)

3. INDICATIONS FOR USE

The STERRAD® Chemical Indicator Strip (PN 14100) is a process indicator (ISO 11140-1:2005) to differentiate processed from unprocessed packages and is intended for use with medical device packages to be sterilized in STERRAD® Sterilization Systems:

MODELCYCLE
STERRAD® 100SStandard
STERRAD® 50Standard
STERRAD® 200Standard
STERRAD® NX®Standard
Advanced
STERRAD® 100NX®Flex
EXPRESS
DUO

{1}------------------------------------------------

KIII 218

4. · DESCRIPTION OF DEVICE

STERRAD® Chemical Indicator Strip is a through-put process indicator strip to be used with STERRAD® Sterilization Systems. The STERRAD® Sterilization System utilizes hydrogen peroxide gas plasma to achieve rapid, low temperature sterilization of medical devices.

STERRAD® Chemical Indicator Strip functions by means of a chemical reaction. Exposure of the chemical indicator to the STERRAD® Sterilizer cycle results in a recognizable color change from red to yellow. The yellow (or lighter) chemical indicator bar indicates that the load has been exposed to hydrogen peroxide.

5. SUMMARY OF NONCLINICAL TESTS

Performance testing was conducted to demonstrate the functionality of Chemical Indicator Strip in the STERRAD® 100NX® DUO Cycle and the results show that the Chemical Indicator Strip performs as a through-put process indicator by responding as expected to hydrogen peroxide exposure. Product shelf-life was presented in the STERRAD® Chemical Indicator Strip, EXPRESS cycle 510(k) (K103218) and was not repeated for this submission since the product's design has not changed.

StudyResult
Chemical Indicator FunctionalityPassed
End Point / Post Processing Color StabilityPassed
Shelf lifePassed
BiocompatibilityPassed

6. OVERALL PERFORMANCE CONCLUSIONS

The nonclinical studies demonstrate that the STERRAD® Chemical indicator Strip is as safe and as effective as the predicate, when used with the STERRAD® 100NX DUO Cycle, for differentiating processed from unprocessed packages within the indications for use for the chemical indicator thus establishing that the STERRAD® Chemical Indicator Strip is substantially equivalent to the predicate device, the STERRAD® Chemical Indicator Strip (K103128).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

OCT

3 2012

Advanced Sterilization Products
Ms. Nancy Chu
Manager, Regulatory Affairs
33 Technology Drive
Irvine, California 92618

Re: K111518 Trade/Device Name: Sterrad® Chemical Indicator Strip (PN 14100) Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: JOJ Dated: September 21, 2012 Received: September 24, 2012

Dear Ms. Chu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Ms. Chu

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

For
Elizabeth
Stuttard

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K111518

Device Name:

STERRAD® Chemical Indicator Strip

Indications for Use:

The STERRAD® Chemical Indicator Strip (PN 14100) is a process indicator (ISO 11140-1:2005) to differentiate processed from unprocessed packages and is intended for use with medical device packages to be sterilized in STERRAD® Sterilization Systems:

MODELCYCLE
STERRAD® 100SStandard
STERRAD® 50Standard
STERRAD® 200Standard
STERRAD® NX®Standard
Advanced
STERRAD® 100NX®Flex
EXPRESS
DUO

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use ___ V (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Signature

(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division Control Dental Devices Division of Anesthoolors 277
Infection Control, Dental Devices

111518

510(k) Number;

Page 1 of 1

(Posted November 13, 2003)

§ 880.2800 Sterilization process indicator.

(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).