K Number
K111514

Validate with FDA (Live)

Date Cleared
2011-12-16

(198 days)

Product Code
Regulation Number
882.5050
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SBV2 System is an activity monitor designed and intended for documenting physical movements associated with applications in physiological monitoring. The device's intended use is to analyze limb activity associated with movement during sleep and to extract information about certain sleep parameters from these movements. SBV2 can also be used to assess activity in any instance where quantifiable analysis of physical motion is desirable. The use of SBV2 is indicated for adults 22 years of age and over.

Device Description

The SBV2 System is a device that monitors activity. It relies on the measurement and analysis of wrist movements to detect and characterize sleep/wake periods. The device allows some aspects of sleep derived from the analysis of activity to be reported. The SBV2 System is graphically depicted in Figure 1.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the SBV2 System, based on the provided text:

Acceptance Criteria and Device Performance

Acceptance Criteria (Target Performance)Reported Device Performance (SBV2)
Phase 1: Differentiation of Activity States
Identify "Awake, up and about"95% accuracy
Identify "In bed, awake or asleep"95% accuracy
Identify "Off wrist"87% accuracy
Phase 2: Sleep/Wake Classification vs. Polysomnography (in-bed)
Sensitivity (correctly identify sleep)88%
Specificity (correctly identify wake)55%
Overall SBV2 vs. Polysomnography Agreement93% (weighted average of 95% in-bed state detection and 88% in-bed-asleep detection)

Detailed Study Information

1. A table of acceptance criteria and the reported device performance
(See table above)

2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Phase 1 (Differentiation of Activity States):
    • Sample Size: 180 participants
    • Data Provenance: Not explicitly stated, but the study was conducted by Archinoetics, LLC and Fatigue Science. Given the later mention of the Kettering Sleep Disorders Center in Ohio for Phase 2, it's plausible the data originates from the US. The data was "de-identified." This phase was retrospective, using existing actigraph data.
  • Phase 2 (Sleep/Wake Classification vs. Polysomnography):
    • Sample Size: 50 patients
    • Data Provenance: Conducted at the Wallace Kettering Health Networks Sleep Disorders Center in Kettering, Ohio, USA. This phase involved simultaneous collection of actigraphy data and polysomnography, suggesting a prospective study design.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Phase 1: "Research personnel at Archinoetics, LLC," described as having "substantial expertise in the visual examination and classification of raw actigraphy data." The exact number of experts is not specified.
  • Phase 2:
    • Dr. Donna Arand, Board Certified Sleep Specialist, Kettering Hospital.
    • Dr. John Caldwell, Experimental Psychologist, Fatigue Science.
    • Dr. Chris Russell, Senior Scientist, Archinoetics, LLC.
    • Total: 3 experts.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

The document does not explicitly describe an adjudication method for disagreements among experts. For Phase 1, it implies the "research personnel" collectively established the ground truth. For Phase 2, ground truth was established by polysomnography and compared to expert actigraphy scoring, but the method for reconciling potential disagreements among the three mentioned experts or between their expert actigraphy scoring and the SBV2 algorithm is not detailed. However, the study validated the SBV2 against expert human actigraphy scoring (Phase 1) and against polysomnography (Phase 2), suggesting that polysomnography was the ultimate ground truth for in-bed sleep/wake.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader, multi-case (MRMC) comparative effectiveness study evaluating human readers with and without AI assistance was not conducted or reported in this document. The study focused on the standalone performance of the SBV2 system against human expert actigraphy and polysomnography.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Yes, a standalone performance study was conducted. The performance metrics (accuracy, sensitivity, specificity) listed in the table are for the SBV2 algorithm's classification of sleep/wake states based on actigraphy data, without human-in-the-loop intervention for classification.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Phase 1: Expert human actigraphy scoring ("visual examination and classification of raw actigraphy data").
  • Phase 2: Polysomnography (PSG) is explicitly stated as the "accepted medical standard for conducting sleep evaluations" and was used as the ground truth for sleep/wake classification.

8. The sample size for the training set

The document does not provide information about the sample size used for training the SBV2 algorithm. The reported studies are validation studies demonstrating the performance of an already-developed algorithm.

9. How the ground truth for the training set was established

The document does not provide information on how the ground truth for any potential training set was established.

{0}------------------------------------------------

< 111514

DEC 1 6 2011

1. 510(k) Summary - Basic Information

  • 1.1 Sponsor
    Submitter: Address:

Company:

Voice:

FAX:

Cell:

Email:

Fatigue Science 700 Bishop Street, Suite 2000 Honolulu, Hawaii, 96813

Contact:

Marc Goodman (on behalf of Fatigue Science) Noblitt & Rueland 5405 Alton Parkway, Suite A530 Irvine, CA 92604 (949) 398-5222 (949) 398-5223 (949) 872-1011 marcgood@ca.rr.com

Date Prepared:

November 21, 2011

1.2 Device Name

SBV2TM System Proprietary Name: Common Name: Activity Recording Device Classification Name: GWQ, Electroencephalograph, 882.1400, Class II

1.3 Identification of Legally Marketed Device

SBV2 is substantially equivalent to the ActiGraph device (K040554).

1.4 Device Description

The SBV2 System is a device that monitors activity. It relies on the measurement and analysis of wrist movements to detect and characterize sleep/wake periods. The device allows some aspects of sleep derived from the analysis of activity to be reported. The SBV2 System is graphically depicted in Figure 1.

Figure 1: SBV2 System Graphic Depiction

Image /page/0/Figure/17 description: The image shows a diagram of a data recording system. The system includes an SBV2 data recording unit, a MultiCharger, an ANT Dongle, and a PC Computer. The SBV2 unit is connected to the ANT Dongle, which is connected to the PC Computer.

{1}------------------------------------------------

1.4.1 Components

The following SBV2 System components (depicted in Figure 1) are briefly described as follows:

  • SBV2 (Data Recording Unit); wristwatch-like device used to capture and store . accelerometer data from user.
  • ANT Dongle; A PC peripheral that uses a proprietary wireless protocol to transfer . stored accelerometer data to SleepAnalyzer software application.
  • PC Computer; standard PC used to host the SleepAnalyzer application. .
  • MultiCharger; coin cell battery charger used to fully charge the Data Recording . Unit battery prior to use.

1.5 Intended Use

The SBV2 System is an activity monitor designed and intended for documenting physical movements associated with applications in physiological monitoring. The device's intended use is to analyze limb activity associated with movement during sleep and to extract information about certain sleep parameters from these movements. SBV2 can also be used to assess activity in any instance where quantifiable analysis of physical motion is desirable. The use of the SBV2 is indicated for adults 22 years of age and over.

1.6 Comparison to Cleared Device

SBV2 is substantially equivalent to the ActiGraph (K040554). Table 1 compares the features of SBV2 with those of ActiGraph.

CharacteristicSBV2ActiGraph
Usage Characteristics
Intended UseThe SBV2 System is an activity monitordesigned and intended for documenting physicalmovements associated with applications inphysiological monitoring. The device's intendeduse is to analyze limb activity associated withmovement during sleep and to extractinformation about certain sleep parameters fromthese movements. SBV2 can also be used toassess activity in any instance wherequantifiable analysis of physical motion isdesirable. The use of the SBV2 is restricted toadults 22 years of age and over.The ActiGraph is a small limb worn activitymonitor designed for documenting physicalmovements associated with applications inphysiological monitoring. The device's intendeduse is to analyze limb activity associated withmovement during sleep. The unit can also beused to assess activity in any instance wherequantifiable analysis of physical motion isdesirable.
Principle ofOperationAccelerometer used to collect actigraphy data todetermine a person's activity level during thecourse of a day.same
ComponentsData recording unit (SBV2 activity monitor)worn by patientPC software application prepares datarecording unit, retrieves collected data, andgenerates report.same
Data CollectionMethodData recording unit is attached to patient's limbor torso and worn continuously during datacollection period.same
Sampling Interval16 Hz1 Hz and up (web site indicates 30 Hz)
Recording Time1 to 30 days. Typically 7 days.2 hours to 180 days (end of battery life)

Table 1: Comparison of SBV2 to ActiGraph (K040554)

510(k) Summary (per 21 CFR 807.92)

{2}------------------------------------------------

CharacteristicSBV2ActiGraph
Data Recording Unitto PC InterfaceWireless communications using ANT protocol in2.4 GHz ISM Band.RS232 or USB
CommercialDistributionIndicated for Over the Counter Use. Patientreceives data recording unit from dispensingagent such as pharmacist.Prescription Use
ReportingDispensing agent retrieves collected data fromdata recording unit and generates report basedon collected data.Same
Physical Characteristics
Size1.5"x1.35"x.4"2" x 2" x 0.5" (does not include wrist band)
Weight1 oz.1.5 oz, with battery 2.5 oz
Batteryrechargeable 3.7V Lithium Coin Cellsame
Recharging MethodMultiChargerconnection to USB port
AccelerometerSensitivity2.4 milliGs16 milliGs
EnclosurePolycarbonatesame
Wrist BandSantopreneNylon w Velcro
MoisturesusceptibilityWater resistantsame
StorageTemperature-40°C to +60°C-10°C to +50°C
OperatingTemperature0°C to +40°C0°C to +45°C

Differences between SBV2 and ActiGraph 1.6.1

The meaningful differences between SBV2 and the ActiGraph predicate device are their means of commercial distribution and their methods for communicating between data recording unit and PC software application.

  • SBV2 is intended for Over-The-Counter commercial distribution. ActiGraph is ● indicated for Prescription Use.
  • . SBV2 uses wireless communications between its data recording unit and its PC software. ActiGraph uses hardwired serial communications.

2. Performance Information

ਾਂ

Performance data produced by a two-phase investigation demonstrates substantial equivalence between polysomnographic sleep/wake classification and SBV2 actigraphic sleep/wake classification.

2.1 Substantial Equivalence Evaluations

The Fatigue Science SBV2 was validated against expert human actigraphy scoring and against polysomnography as a device for accurately measuring sleep/wake periods. Polysomnography is considered to be the accepted medical standard for conducting sleep evaluations. There were two phases of the evaluation. A detailed account may be found in the Fatigue Science White Paper entitled Validation of the Fatigue Science SBV2 authored by Russell, Caldwell, Arand, Myers, Wubbels, and Downs (2010).

2.1.1 Overview

The first phase of the scoring validation effort was conducted by research personnel at Archinoetics, LLC. These personnel possessed substantial expertise in the visual

{3}------------------------------------------------

examination and classification of raw actigraphy data into various states such as "awake and moving about" versus "lying quietly in bed or elsewhere". The second phase of the classification yalidation effort was conducted by Dr. Donna Arand, Board Certified Sleep Specialist, Kettering Hospital; Dr. John Caldwell, Experimental Psychologist, Fatigue Science; and Dr. Chris Russell, Senior Scientist, Archinoetics, LLC; in conjunction with the Wallace Kettering Health Networks Sleep Disorders Center in Kettering Ohio.

2.1.2 Phase 1 Summary

Since the SBV2 is designed to be worn 24-hours per day for several days at a time, it was first necessary to ensure that the "up-and-about periods" could be correctly differentiated from the "lying-in-bed periods." Using de-identified actigraph data from 180 participants, we determined the agreement between algorithmic versus human classification of actigraphy into the following states: 1) Awake and moving about, 2) In bed-Awake or Asleep, or 3) Actigraph off wrist. Results indicated that the SBV2 algorithm accurately differentiated among the 3 states of interest between 87% and 95% of the time (see Table 1).

Table 1. Identification accuracy of "up & about" vs "in bed" vs "off wrist"
-------------------------------------------------------------------------------------------
AWAKE, up and aboutIN BED, awake or asleepOFF WRIST
95%95%87%

2.1.3 Phase 2 Summary

Since once the "in-bed" state was determined to have occurred it was of utmost importance to show correct actigraphy-based classification of sleep and wake episodes, a second study focused on the comparison between in-bed polysomnography and in-bed actigraphy. Via the collection of actigraphy data simultaneously from 50 patients undergoing polysomnographic assessment at the Kettering Sleep Disorders Center, the specificity and sensitivity of actigraphic sleep/wake assessments were determined. Results indicated that SBV2 sleep/wake classification corresponded to night-time polysomnographic findings with a sensitivity of 88% and a specificity of 55% (see Table 2).

THE SE STAULTIONAL AND POST LA LE POLL CARTER STERES MICA
Total accuracyensitivity
0400005 201
Table 2: Classification accuracy of SBV2 vs polysomnography sleep/wakeand the contract of the contract and the comments of the comments of the contribution of the comments of the contribution of the comments of the contribution of the contribut
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

2.1.4 Final Result

Considering that for most people the majority of in-bed time would be spent sleeping, the in-bed sensitivity (correct identification of sleep) was weighted slightly more heavily than the in-bed specificity (correct identification of wake) by averaging the "in-bed state detection" of 95% with the "in-bed-asleep detection" of 88%. This yielded an overall SBV2 vs polysomnography agreement level of 93%.

{4}------------------------------------------------

2.2 Sleep Report Summary

The device allows some aspects of sleep derived from the analysis of activity to be described in a sleep report. The sleep report uses a set of parameters that describe the normal ranges for an average adult user. Any further interpretations of sleep from the report should only be performed by a qualified health care professional such as a physician with training in the normal and abnormal physiology of sleep. These normal ranges are supported by the following data.

2.2.1 Sleep Duration

The normal range of sleep duration for adults is set to 7-9 hours according to supporting data from Table 1 in Berger et al. (2005), Sleep/wake disturbances in people with cancer and their caregivers: State of the Science. Oncology Nursing Forum, 32(6):E98-E126. In addition, both the National Sleep Foundation and the Mayo Clinic state that 7-9 is the normal range for adults. The following table appears at http://www.sleepfoundation.org/article/howsleep-works/how-much-sleep-do-we-really-need:

How Much Sleep Do You Really Need?
AgeSleep Needs
Newborns (0-2 months)12-18 hours
Infants (3 to 11 months)14 to 15 hours
Toddlers (1-3 years)12 to 14 hours
Preschoolers (3-5 years)11 to 13 hours
School-age children (5-10 years)10 to 11 hours
Teens (10-17)8.5-9.25 hours
Adults7-9 hours

2.2.2 Sleep Latency

The normal range of sleep latency is set to 10-20 minutes according to supporting data found in a published chapter by Kryger et al. in "Principles and practice of Sleep Medicine, 2nd edition, page 965."

Source: National Steep Foundation

Sleep Efficiency Range 2.2.3

The normal sleep efficiency range is stated as 95-80% according to supporting data found in Table 1 in Berger et al. (2005), Sleep/wake disturbances in people with cancer and their caregivers: State of the Science, Oncology Nursing Forum. 32(6):E98-E126; as well as in Table 2 in Bonnet and Arand (2007). EEG arousal norms by age, Journal of Clinical Sleep Medicine, 3(3):271-274.

Bedtime Deviation 2.2.4

There are no established norms for this measure, however it is widely considered a general good-sleep-hygeine practice to minimize the deviation in bedtimes from night to night in order to promote quality restorative sleep.

{5}------------------------------------------------

Therefore, this metric (without displaying the norm) is shown in the sleep report to present users with an assessment of how well they are following this recommended practice (of adhering to a consistent bedtime).

2.3 Obtaining References

All referenced materials will be provided to dispensers in the form of hard copies in a binder. Below is a list of included references:

  • => Russel, C.A, Caldwell, J.A., et al. (2011), "Validation of the Fatigue Science SBV2." Fatigue Science White Paper.
  • Carskadon, M.A. (1994). Guidelines for the Multiple Sleep Latency Test (MSLT): A Standard Measure of Sleepiness, in Kryger et al. (eds.), "Principles and practice of Sleep Medicine," pp 962-966. Philadelphia: WB Saunders Co.

  • Berger et al., (2005). Sleep/wake disturbances in people with cancer and their caregivers: State of the Science, "Oncology Nursing Forum," 32(6):E98-E126

  • Bonnet and Arand (2007), EEG arousal norms by age, "Journal of Clinical Sleep Medicine," 3(3):271-274

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a sans-serif font.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Sleep Performance D.B.A. Fatigue Science c/o Mr. Mark Goodmand, Senior Associate Noblit & Rueland 5405 Alton Parkway, Suite A530 Irvine, CA 92604

Re: K111514

Trade/Device Name: SVB2 System Regulation Number: Unclassified Regulation Name: Sleep Assessment Device Regulatory Class: Unclassified Product Code: LEL Dated: November 21, 2011 Received: November 23, 2011

DEC 1 6 2011

Dear Dr. Downs:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{7}------------------------------------------------

Page 2 - J. Hunter Downs III, Ph.D.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

for

Sincerely vours.

Malvina B. Eydelman, M.D.
Director
Division of Ophthalmic, Neurological,
and Ear, Nose and Throat Devices
Office of Device Evaluation

Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Indications for Use Statement

510 (k) Number (if known) K111514

Device Name SBV2 System

Indications The SBV2 System is an activity monitor designed and intended for for Use documenting physical movements associated with applications in physiological monitoring. The device's intended use is to analyze limb activity associated with movement during sleep and to extract information about certain sleep parameters from these movements. SBV2 can also be used to assess activity in any instance where quantifiable analysis of physical motion is desirable. The use of SBV2 is indicated for adults 22 years of age and over.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

Page 1 of 1

Kii514
510(k) Number

§ 882.5050 Biofeedback device.

(a)
Identification. A biofeedback device is an instrument that provides a visual or auditory signal corresponding to the status of one or more of a patient's physiological parameters (e.g., brain alpha wave activity, muscle activity, skin temperature, etc.) so that the patient can control voluntarily these physiological parameters.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter when it is a prescription battery powered device that is indicated for relaxation training and muscle reeducation and prescription use, subject to § 882.9.