K Number
K111407

Validate with FDA (Live)

Date Cleared
2012-01-18

(244 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ASC TriPort 15 Laparoscopic Access Device is intended for use as a multiple instrument and/or camera port during minimally invasive abdominal laparoscopic surgery.

Device Description

The ASC TriPort 15 Laparoscopic Access Device is a faparoscopic multi-instrument port which performs the following two functions:

  • Retracting a small abdominal incision to allow multiple laparoscopic instruments to pass . through to the abdomen at the same time during laparoscopic surgery.
  • Ensuring that pneumoperitoneum is maintained in the abdomen during the surgical . procedure, whether or not one or more laparoscopic instruments are passing through the device.

The ASC TriPort 15 Laparoscopic Access Device is comprised of the following three components:

  • An introducer component, which delivers the Distal Ring through a pre-made incision, . into the abdominal cavity.
  • A refractor component, which retracts an abdominal incision to allow the passage of . laparoscopic instruments.
  • A valve component which maintains the pneumoperitoneum established for the surgical . procedure.

The minor design changes to the cleared TriPort+ (K110004) device to create the TriPort 15 are as follows:

    1. Changing the valve configuration from a four valve (3 x 5mm & 1 x 10mm) to a three valve (2 x 5mm & 1 x 15mm) configuration.
    1. The introduction of a new colorant for the 15mm Valve Cap. Note that this portion of the device does not contact the patient.

The changes made to the parent TriPort+ Laparoscopic Access Devices to produce the TriPort 15 are minor and do not represent changes to its intended use, or the operating principles or mechanism of action of the device.

The ASC TriPort 15 Laparoscopic Access Device is sterile and disposable.

AI/ML Overview

The provided text is a 510(k) summary for the ASC TriPort 15 Laparoscopic Access Device, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with detailed acceptance criteria and expert reviews in terms of clinical performance metrics like sensitivity, specificity, or reader improvement.

Therefore, the requested information, which is typical for a study validating the diagnostic accuracy or clinical effectiveness of a device, is not present in the provided text.

Specifically, the document states:
"Sterilization and shelf life testing, biocompatibility testing, design validation testing, and animal testing of the TriPort 15 was conducted to verify that the changes to the technological characteristics do not affect safety or effectiveness. In all instances, the TriPort 15 functioned as intended."
And:
"Performance data demonstrate that the TriPort 15 is as safe and effective as the predicate device."

This indicates that the "acceptance criteria" were related to verifying safety and effectiveness through non-clinical performance testing (sterilization, shelf life, biocompatibility, design validation, animal testing) to ensure the minor changes from the predicate device did not introduce new safety concerns or alter its intended function.

Since this is a submission for a laparoscopic access device (a surgical tool, not a diagnostic or AI-driven device), the nature of the "acceptance criteria" is fundamentally different from a diagnostic device that would involve human reader performance, ground truth establishment by experts, and statistical metrics like sensitivity and specificity.

Therefore, I cannot provide a table of acceptance criteria and reported device performance in the format requested (which implies clinical performance metrics), nor details about sample sizes for test sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, training set details, or ground truth establishment relevant to an AI/diagnostic device.

The study described is an engineering validation and non-clinical performance study to demonstrate that the modified device (TriPort 15) is as safe and effective as its predicate (TriPort+), primarily through bench testing and animal studies, not through clinical trials involving human subjects or expert interpretations of images.

{0}------------------------------------------------

K11140.7

Page 1 of (2)

JAN 1 8 2012

510(k) SUMMARY Advanced Surgical Concepts' (ASC) TriPort 15 Laparoscopic Access Device

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Hogan Lovells US LLP Columbia Square 555 Thirteenth Street, NW Washington, DC 20004 USA Contact Person: Jonathan Kahan +1 202 637 5794 Telephone: January 17, 2012 Date Prepared:

ﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮ 2 - 1

Name of Device and Name/Address of Sponsor

TriPort 15 Laparoscopic Access Device

Advanced Surgical Concepts Unit 4, Sunnybank Centre Upper Dargle Road Bray, Co. Wicklow Ireland Contact Person: Ed Hyland +353 (0)1 2864777 Telephone:

Common or Usual Name

Laparoscopic single port access device (Product Code OTJ)

Classification Name

21 C.F.R. §876.1500, Endoscope and accessories Class II

Predicate Device

ASC TriPort+ Laparoscopic Access Device (K110004)

Intended Use / Indications for Use

The ASC TriPort 15 Laparoscopic Access Device is intended for use as a multiple instrument and/or camera port during minimally invasive abdominal laparoscopic surgery.

Technological Characteristics

The ASC TriPort 15 Laparoscopic Access Device is a faparoscopic multi-instrument port which performs the following two functions:

{1}------------------------------------------------

KIII 407

Page 2 of 2

  • Retracting a small abdominal incision to allow multiple laparoscopic instruments to pass . through to the abdomen at the same time during laparoscopic surgery.
  • Ensuring that pneumoperitoneum is maintained in the abdomen during the surgical . procedure, whether or not one or more laparoscopic instruments are passing through the device.

The ASC TriPort 15 Laparoscopic Access Device is comprised of the following three components:

  • An introducer component, which delivers the Distal Ring through a pre-made incision, . into the abdominal cavity.
  • A refractor component, which retracts an abdominal incision to allow the passage of . laparoscopic instruments.
  • A valve component which maintains the pneumoperitoneum established for the surgical . procedure.

The minor design changes to the cleared TriPort+ (K110004) device to create the TriPort 15 are as follows:

    1. Changing the valve configuration from a four valve (3 x 5mm & 1 x 10mm) to a three valve (2 x 5mm & 1 x 15mm) configuration.
    1. The introduction of a new colorant for the 15mm Valve Cap. Note that this portion of the device does not contact the patient.

The changes made to the parent TriPort+ Laparoscopic Access Devices to produce the TriPort 15 are minor and do not represent changes to its intended use, or the operating principles or mechanism of action of the device.

The ASC TriPort 15 Laparoscopic Access Device is sterile and disposable.

Performance Data

Sterilization and shelf life testing, biocompatibility testing, design validation testing, and animal testing of the TriPort 15 was conducted to verify that the changes to the technological characteristics do not affect safety or effectiveness. In all instances, the TriPort 15 functioned as intended.

Substantial Equivalence

The ASC TriPort 15 Laparoscopic Access Device is as safe and effective as the predicate device. The TriPort 15 device has the same intended use and similar indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the TriPort 15 model and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the TriPort 15 is as safe and effective as the predicate device. Thus, the ASC TriPort 15 Laparoscopic Access Device is substantially equivalent.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird-like figure, with its wings spread and its body forming a flowing, abstract shape.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JAN 1 8 2012

Advanced Surgical Concepts % Hogan Lovells US LLP Mr. Jonathan S. Kahan 355 Thirteenth Street, NW Washington, District of Columbia 20004

Re: K111407

Trade/Device Name: TriPort 15 Laparoscopic Access Device Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: OTJ Dated: January 17, 2012 Received: January 17, 2012

Dear Mr. Kahan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the each and to review is substantially equivalent (for the indical commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclasion with the provisions of the Medical Device Amendments, or to
and Cosmetic Act (Act) that do not require approvise of the Federal Food, Drug, and Cosmetic Act (Act (Act (Act that do not require approval of the Federal Food, Drug,
You may, therefore, market the device, subject to the a premarket approval application You may, therefore, market the device, subject to the general controls provisions of PMA).
general controls provisions of the Act include result general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 – Mr. Jonathan S. Kahan

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. .

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson

Mark N. Melke Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Sṭatement

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: TriPort 15 Laparoscopic Access Device

Indications for Use:

The ASC TriPort 15 Laparoscopic Access Device is intended for use as a multiple instrument and/or camera port during minimally invasive abdominal laparoscopic surgery.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page | of |

Neil R. Poole for mxm
(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K111407

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.