(109 days)
The Instinct Java spinal fixation system is designed for spinal fixation procedures in skeletally mature patients performed through a posterior approach. The Instinct Java spinal fixation system is indicated for the temporary realignment and stabilization of one or more intervertebral segments from the thoracic spine to the sacrum until bony fusion is obtained.
The Instinct Java spinal fixation system is indicated to achieve bony fusion via osteosynthesis at thoracic, lumbar and/or lumbosacral levels of the spine in documented cases of degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, fracture, spinal stenosis, kyphotic or lordotic spinal deformities, scoliosis, tumor and pseudoarthrosis, or for revision of a failed previous fusion.
The Instinct Java System is a temporary implant system used to correct spinal deformity in skeletally mature patients and facilitate the biological process of spinal fusion. This system is intended for non-cervical posterior use in the thoracic, lumbar and sacral areas of the spine. The Instinct Java spinal fixation system is indicated to achieve bony fusion via osteosynthesis at thoracic, lumbar and/or lumbosacral levels of the spine in documented cases of degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, fracture, spinal stenosis, kyphotic or lordotic spinal deformities, scoliosis, tumor and pseudoarthrosis or for revision of a failed previous fusion.
The system consists of implants and instruments. The implants consist of monoaxial and polyaxial screws of varying diameters and lengths, blockers, pre-contoured and straight rods, and transverse connectors of varying lengths. All implants are made of titanium alloy, with the exception of one commercially pure titanium component within the transverse connectors.
Re-usable surgical instruments are provided to facilitate placement of the implants.
In addition, the Instinct Java System is compatible with the transverse connectors currently in the market as part of the Sequoia Spinal System, identified in K082032. Furthermore, the Instinct Java may be connected to the NexLink Band & In-Line Rod Connector, identified in K062505, K060634, K052566, K052247, K031985.
The implants and instruments are provided non-sterile. Instructions for Use are provided that contain validated cleaning and sterilization instructions for the user.
This system is intended to provide stabilization until a solid spinal fusion develops. The system may then be removed, per the surgeon's discretion.
Here's a breakdown of the acceptance criteria and study information based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Bench Test Requirements) | Reported Device Performance |
|---|---|
| Requirements of ASTM F-1717:2004 for static performance | Confirmed through static performance testing |
| Requirements of ASTM F-1717:2004 for dynamic compression bending | Confirmed through dynamic compression bending |
| Requirements of ASTM F-1717:2004 for static torsion | Confirmed through static torsion testing |
| Validation of cleaning and sterilization instructions | Cleaning and sterilization instructions were validated for the non-sterile components of the system. |
2. Sample Size Used for the Test Set and Data Provenance
The document describes bench testing for the Instinct Java System, not a clinical study involving a "test set" in the context of diagnostic AI or clinical trials. Therefore, information about a test set sample size, country of origin, or retrospective/prospective data is not applicable. The testing was performed on the device components themselves.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts
This information is not applicable as the study involves bench testing of a physical medical device (spinal fixation system), not an AI algorithm requiring expert-established ground truth from clinical data.
4. Adjudication Method for the Test Set
This is not applicable for a bench testing study of a physical medical device.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document describes bench testing of a physical medical device, not an AI system.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
No, a standalone algorithm performance study was not done. The device is a physical spinal fixation system, not a software algorithm.
7. The Type of Ground Truth Used
The "ground truth" for the device's performance was established through adherence to specified ASTM standards (ASTM F-1717:2004) for mechanical performance (static, dynamic compression bending, static torsion) and validated cleaning and sterilization protocols for the non-sterile components.
8. The Sample Size for the Training Set
This information is not applicable. The document describes bench testing of a physical medical device, not an AI system requiring a training set.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable for a physical medical device. The "ground truth" for the device's design and manufacturing is adherence to engineering specifications and regulatory standards.
{0}------------------------------------------------
Page 1 of 2
AUG 26 2011
zimmer
spine
510(k) SUMMARY Zimmer Spine İnstinct™ Java® System
KIII301 510(k) Number
| Date of Summary Preparation: | August 17, 2011 |
|---|---|
| Submitter: | Zimmer Spine, Inc.7375 Bush Lake RoadMinneapolis, MN 55439 |
| Company Contact: | Elsa A. LinkeRegulatory Affairs |
| Manufacturer: | Zimmer SpineCité Mondiale23, parvis des Chartrons33080 BordeauxFrance |
| Device Name: | Instinct Java System |
| Common Name: | Spinal Fixation System |
| Classification Name: | Pedicle Screw Spinal System |
| Product Code: | MNI, MNH, NKB |
| Regulation Number: | 888.3070 |
| Device Classification: | Class III |
| Predicate Devices: | Zimmer Spine Sequoia® Spinal System, K082032Zimmer Spine Silhouette® Spinal System, K042702Diamondback Spinal System, K100377 |
Description of Device:
The Instinct Java System is a temporary implant system used to correct spinal deformity in skeletally mature patients and facilitate the biological process of spinal fusion. This system is intended for non-cervical posterior use in the thoracic, lumbar and sacral areas of the spine. The Instinct Java spinal fixation system is indicated to achieve bony fusion via osteosynthesis at thoracic, lumbar and/or lumbosacral levels of the spine in documented cases of degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, fracture, spinal stenosis, kyphotic or lordotic spinal deformities, scoliosis, tumor and pseudoarthrosis or for revision of a failed previous fusion.
{1}------------------------------------------------
The system consists of implants and instruments. The implants consist of monoaxial and rne oyoten consiste on of varying diameters and lengths, blockers, pre-contoured and straight rods, and transverse connectors of varying lengths. All implants are made of titanium alloy, with the exception of one commercially pure titanium component within the transverse connectors.
Re-usable surgical instruments are provided to facilitate placement of the implants.
In addition, the Instinct Java System is compatible with the transverse connectors currently in addition, the market as part of the Sequoia Spinal System, identified in K082032. Furthermore, the Instinct Java may be connected to the NexLink Band & In-Line Rod Connector, identified in K062505, K060634, K052566, K052247, K031985.
The implants and instruments are provided non-sterile. Instructions for Use are provided that contain validated cleaning and sterilization instructions for the user.
This system is intended to provide stabilization until a solid spinal fusion develops. The system may then be removed, per the surgeon's discretion.
Intended Use:
The Instinct Java spinal fixation system is designed for spinal fixation procedures in skeletally mature patients performed through a posterior approach. The Instinct Java spinal Skelonaly maters pationed for the temporary realignment and stabilization of one or more intervertebral segments from the thoracic spine to the sacrum until bony fusion is obtained.
The Instinct Java spinal fixation system is indicated to achieve bony fusion via r to motifies at thoracic, lumbar and/or lumbosacral levels of the spine in documented osteooynthould at there was (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, fracture, spinal the diso commined by history and deformities, scoliosis, tumor and pseudoarthrosis, or for revision of a failed previous fusion.
Comparison of Technological Characteristics:
The Instinct Java Spinal System shares the same technological characteristics as the rne instinut dava Opinal Cycleracteristics include similar design, materials, range of sizes, predioute devices. The and intended use. Determination of substantially equivalent technical requirements, and intensed to the predicate devices was confirmed through static performanoo ending, dynamic compression bending and static torsion testing in compression beneming, equirements of ASTM F-1717:2004. In addition, cleaning and sterilization instructions were validated for the non-sterile components of the system.
Substantial Equivalence:
The Instinct Java Spinal System is substantially equivalent to the predicate devices in design, materials, function and intended use.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with three intertwined strands instead of snakes. The symbol is black and the background is white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
AUG 26 2011
Zimmer Spine, Inc. % Ms. Elsa A. Linke 7375 Bush Lake Road Minneapolis, Minnesota 55439
Re: K111301
Trade/Device Name: Instinct™ Java® System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNI, MNH Dated: August 18, 2011 Received: August 19, 2011
Dear Ms. Linke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 - Ms. Elsa A. Linke
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Erine S. Keith
- Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use Statement
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Instinct™ Java® System
Indications for Use:
The Instinct Java spinal fixation system is designed for spinal fixation procedures in skeletally mature patients performed through a posterior approach. The Instinct Java spinal fixation matem is indicated for the temporary realignment and stabilization of one or more intervertebral segments from the thoracic spine to the sacrum until bony fusion is obtained.
The Instinct Java spinal fixation system is indicated to achieve bony fusion via osteosynthesis at thoracic, lumbar and/or lumbosacral levels of the spine in documented cases of degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, fracture, spinal stenosis, kyphotic or lordotic spinal deformities, scoliosis, tumor and pseudoarthrosis, or for revision of a failed previous fusion.
Prescription Use _ × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
KIII 30 ! 510(k) Number_
N/A