K Number
K111243

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-05-16

(13 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Fujinon ultrasonic processor SU-8000 is intended to be used in combination with Fujinon/Fujifilm ultrasonic endoscope, video processor, light source, monitor, recorder, and various peripheral devices. The product is intended to provide ultrasonic images of submucosal and peripheral organs of the upper gastrointestinal tract for observation, recording and to aid in diagnosis during endoscopic evaluation.

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical Application: General Application
Specific: Tran-esoph. (non-Card.), Other (Specify)
Mode of Operation: B, M, PWD, Color Doppler, Combined1
Other includes gastro-intestinal tract and surrounding organs

Device Description

Fujinon Ultrasonic Processor SU-8000 is a new ultrasonic processor, which can be used with the previously-cleared ultrasonic gastrointestinal endoscopes, EG-530UR and EG-590UT via K063847.

Fujinon Ultrasonic Processor SU-8000 consists of a scan engine and a function box and SU-8000 is controlled by keyboard (model: CP-8000) SU-8000 connects to ultrasonic endoscopes, which emits ultrasound in a body cavity by driving ultrasonic transducers of the endoscope. SU-8000 also processes the reflection of ultrasonic signals the transducer and converts it to an ultrasound image.

SU-8000 is used in combination with ultrasonic endoscopes, video processor, light source, cart, recorder, foot switch and other peripheral devices (e.g. external monitor and printer), which is the same as the legally marketed device, SU-7000.

AI/ML Overview

I am unable to provide a structured response about acceptance criteria and study details based on the provided text. The document is a 510(k) summary for the Fujinon Ultrasonic Processor SU-8000, which states:

  • "No clinical test was conducted."
  • The device was evaluated in accordance with various IEC safety and performance standards (e.g., IEC 60601-1, IEC 60601-1-2, etc.).
  • Substantial equivalence to a legally marketed device (SU-7000) was established based on having the same Indications for Use and very similar Functional and Technical requirements.

This type of submission focuses on demonstrating safety and effectiveness through adherence to recognized standards and substantial equivalence to existing devices, rather than through a new clinical study with specific acceptance criteria and performance metrics typically associated with AI/algorithm validation. Therefore, the requested information (such as sample size, data provenance, number of experts, adjudication, MRMC study, standalone performance, ground truth details for test/training sets) is not present in this document.

{0}------------------------------------------------

FUJIFILI

MAY 16 2011

FUJIFILM GROUP
FUJINON INC.
10 High Point Drive
Wayne, NJ 07470
Tel: (973) 633-5600

K/// 2.43
page 1 of 2

510(k) Summary

Date: April 14, 2011

Submitter's Information:

Fujinon Inc. 10 High Point Drive Wayne, NJ 07470 USA

Contact Person:

Name:Gina Walljasper
Title:Director, Quality and Regulatory Compliance
Telephone:(973) 633-5600
Facsimile:(973) 633-8818
E-Mail:gina.walljasper@fujinon.com

Identification of the Proposed Device:

Proprietary/Trade Name:
Common Name:

Fujinon Ultrasonic Processor SU-8000 Ultrasonic Processor for Ultrasonic Gastrointestinal Endoscopy Class 2

Device Class: Classification Information:

Classification NameCFR SectionProduct Codes
Gastroscope21 CFR 876.1500FDS
Ultrasonic Pulsed Doppler Imaging System21 CFR 892.1550IYN
Ultrasonic Pulsed Eco Imaging System21 CFR 892.1560IYO
Diagnostic Ultrasonic Transducer21 CFR 892.1570ITX

l. INDICATIONS FOR USE

The Fujinon ultrasonic processor SU-8000 is intended to be used in combination with Fujinon/Fujifilm ultrasonic endoscope, video processor, light source, monitor, recorder, and various peripheral devices. The product is intended to provide ultrasonic images of submucosal and peripheral organs of the upper gastrointestinal tract for observation, recording and to aid in diagnosis during endoscopic evaluation.

ll. DEVICE DESCRIPTION

Fujinon Ultrasonic Processor SU-8000 is a new ultrasonic processor, which can be used with the previously-cleared ultrasonic gastrointestinal endoscopes, EG-530UR and EG-590UT via K063847.

Fujinon Ultrasonic Processor SU-8000 consists of a scan engine and a function box and SU-8000 is controlled by keyboard (model: CP-8000) SU-8000 connects to ultrasonic endoscopes, which emits ultrasound in a body cavity by driving ultrasonic transducers of

Fujimon Ultrasonic Processor SU-8000

Image /page/0/Picture/21 description: The image shows the text "39 of 1027" at the top. Below that is the word "FUJINON" in large, bold, black letters. The text appears to be a page number or some other kind of identifier.

{1}------------------------------------------------

FUJKFILM

FUJIFILM GROUP
FUJINON INC.
10 High Point Drive
Wayne, NJ 07470
Tel: (973) 633-5600
K111243 page 2 of =

the endoscope. SU-8000 also processes the reflection of ultrasonic signals the transducer and converts it to an ultrasound image.

SU-8000 is used in combination with ultrasonic endoscopes, video processor, light source, cart, recorder, foot switch and other peripheral devices (e.g. external monitor and printer), which is the same as the legally marketed device, SU-7000.

= SUMMARY OF STUDIES

Fujinon Ultrasonic Processor (SU-8000) was evaluated in accordance with following safety and performance requirements in addition to the applicable quality system regulations:

IEC 60601-1Medical electrical equipment - Part 1: General requirements forsafety
IEC 60601-1-1Medical electrical equipment - Part 1-1: General requirements forsafety - Collateral standard: Safety requirements for medicalelectrical systems
IEC60601-1-2Medical electrical equipment - Part 1-2: General requirements forthe basic safety and essential performance - Collateral standard:Electromagnetic compatibility - Requirements and tests
IEC60601-2-18Medical electrical equipment - Part 2-18: Particular requirementsfor the safety of endoscopic equipment
IEC 60601-2-37Medical electrical equipment - Part 2-37: Particular requirementsfor the basic safety and essential performance of ultrasonic medicaldiagnostic and monitoring equipment

No clinical test was conducted.

IV. SUBSTANTIAL EQUIVALENCE

Fujinon Ultrasonic Processor SU-8000 is substantially equivalent to the following device:

Legally Marketed Device$§$ 510(k) #
Fujinon Ultrasonic Endoscope & Processor(EG-530UR, EG-530UT with SU-7000)K063847

The proposed device, Fujinon Ultrasonic Processor SU-8000 has the same Indications for Use and very similar Functional and Technical requirements as our legally marketed device, Ultrasonic Processor (SU-7000) via K063847.

V. CONCLUSION

Fujinon Ultrasonic Processor SU-8000 is substantially equivalent to the legally marketed device and conforms to applicable medical device safety and performance standards.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" are arranged in a circular pattern around the caduceus symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Fujinon, Inc. c/o Mr. Mark Job Regulatory Affairs Specialist Regulatory Technology Services, Inc. 1394 25th Street, N.W. BUFFALO MN 55313

MAY 1 6 2011

Re: K11243

Trade/Device Name: Fujinon Ultrasonic Processor SU-8000 Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FDS, IYN, IYO and ITX Dated: May 2, 2011 Received: May 3, 2011

Dear Mr. Job:


We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device ----Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the Fujinon Ultrasonic Processor SU-8000, as described in your premarket notification:

Transducer Model Number

EG-530UR and EG-530UT

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

If you have any questions regarding the content of this letter, please contact Mary Beth O'Brien, M.S.R.N., at (301) 796-6657.

Sincerely yours,

Roque M. Diaz

lerbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosures

{4}------------------------------------------------

K///243

Indications For Use Statement

510(k) Number (If Known):

): Killys

Device Name:

Fujinon Ultrasonic Processor SU-8000

Indications for Use:

The ultrasonic processor (SU-8000) is intended to be used in combination with Fujinon/Fujifilm ultrasonic endoscope, video processor, light source, monitor, recorder, and various peripheral devices. The product is intended to provide ultrasonic images of submucosal and peripheral organs of the upper gastrointestinal tract for observation, recording and to aid in diagnosis during endoscopic evaluation.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Vorza M. Why

ductive, Gastro-Renal, and ાં 24

{5}------------------------------------------------

K///243

Diagnostic Ultrasound Indications For Use

510(k) Number (If Known):

System Name: Fujinon Ultrasonic Processor SU-8000

With All Ultrasonic Endoscopes (EG-530UR and EG-530UT) Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
GeneralSpecificBMPWDCWDColorDopplerCombined1Other
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ(Thyroid, Breast, Testes, etc.)
Neonatal Cephalic
GeneralAdult Cephalic
ApplicationTrans-rectal
Trans-vaginal
Trans-urethral
Tran-esoph. (non-Card.)NNNNN1
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Specify)2NNNNN1
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Tran-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Specify)

= new indication; P = previously cleared by FDA; E = added under this appendix

Combined modes includes B+M, B+PWD, B+CD+PWD modes

2 Other includes gastro-intestinal tract and surrounding organs

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K111243

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Prescription Use
(Per 21 CFR 801.109)

{6}------------------------------------------------

Diagnostic Ultrasound Indications For Use

11242

510(k) Number (If Known):

System Name: Fujinon Ultrasonic Processor SU-8000

Ultrasonic Endoscope (EG-530UT) Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
GeneralSpecificBMPWDCWDColorDopplerCombined1Other
OphthalmicOphthalmic
GeneralApplicationFetal
Abdominal
Intra-operative
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ(Thyroid, Breast, Testes, etc.)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Tran-esoph. (non-Card.)NNNNN1
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Specify)2NNNNN1
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Tran-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Specify)

w indication; P = previously cleared by FDA; E = added under this appendix

1 Combined modes includes B+M, B+PWD, B+CD+PWD modes

2 Other includes gastro-intestinal tract and surrounding organs

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

(División Sign-Off) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number: K111243
Prescription Use
(Per 21 CFR 801.109)

{7}------------------------------------------------

Diagnostic Ultrasound Indications For Use

510(k) Number (If Known): System Name: Fujinon Ultrasonic Processor SU-8000 Transducer: Ultrasonic Endoscope (EG-530UR)

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
GeneralSpecificBMPWDCWDColorDopplerCombined¹Other
OphthalmicOphthalmic
GeneralApplicationFetal
Abdominal
Intra-operative
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ
(Thyroid, Breast, Testes, etc.)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Tran-esoph. (non-Card.)NNN
Musculo-skeletal
(Conventional)
Musculo-skeletal
(Superficial)
Intravascular
Other (Specify)²NNN
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Tran-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
PeripheralPeripheral vessel
VesselOther (Specify)

N= new indication; P = previously cleared by FDA; E = added under this appendix

1 Combined modes includes B+M, B+PWD, B+CD+PWD modes

² Other includes gastro-intestinal tract and surrounding organs

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Joseph R. Whitely

(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K111243

Prescription Use (Per 21 CFR 801.109)

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.