K Number
K111232

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-01-26

(269 days)

Product Code
Regulation Number
888.3020
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GAP Endo-Exo Medullary System is indicated as a temporary implant to assure alignment, stabilization and fixation of: long bones that have been surqically prepared (osteotomy) for correction of deformities or fractures caused by trauma or disease. The GAP Endo-Exo Medullary System is used for pediatric patients (child and adolescent) age 2 to 21. It can be used to correct the following conditions:

  • Diaphyseal facture of the femur, tibia and humerus .
  • Fractures of the femoral neck .
  • . Subtrochanteric, intertrochanteric and combination fractures
  • . Correction of deformities (OI, Coxa vara, Coxa valga)
  • . Nonunions and malunions
Device Description

The GAP Endo-Exo Medullary System consists of an intramedullary cannulated nail linked to various types of plates via and/or mechanical screws creating a a combined laq Endoamedullary/ Exomedullary osteosythesis device. 3 and 4mm cortical screws are used to secure the nail to the bone.

AI/ML Overview

The provided document describes the GAP Endo-Exo Medullary System, an intramedullary nail system for pediatric patients. It focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed study on specific acceptance criteria and device performance as would be expected for an AI/ML medical device.

Therefore, many of the requested sections regarding AI/ML study specifics (e.g., sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, training set details) are not applicable to this document as it pertains to a mechanical medical device with no AI/ML component described.

Here's the information that can be extracted from the provided text, adapted for the context of this device:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Test Standard)Reported Device Performance
ASTM Standard 1264: Static 4-point bending testResults indicated the device is functionally safe for its intended use.
ASTM Standard 1264: Torsion testResults indicated the device is functionally safe for its intended use.
ASTM Standard 1264: 4-point bending fatigue testResults indicated the device is functionally safe for its intended use.

2. Sample size used for the test set and the data provenance

  • Sample Size: Not explicitly stated. The tests were biomechanical, likely using multiple samples of the device and possibly bone simulants or cadaveric bone, but specific numbers were not provided.
  • Data Provenance: The tests were "Biomechanical Testing" performed to demonstrate safety. The document does not specify the country of origin of the test data (e.g., where the lab was located), nor does it indicate if the data was retrospective or prospective in a clinical sense, as these were laboratory biomechanical tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • This question is not applicable as the document describes a mechanical device. "Ground truth" in the context of biomechanical testing typically refers to established engineering principles, material properties, and adherence to test standards, rather than expert consensus on medical images or diagnoses. The expertise would lie in the conducting and interpretation of the biomechanical tests according to ASTM standards.

4. Adjudication method for the test set

  • This question is not applicable as the document describes a mechanical device. Adjudication methods like 2+1 or 3+1 are used for reconciling expert opinions on diagnoses or image interpretations, which is not relevant to biomechanical testing of an intramedullary nail.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This question is not applicable. The device is a mechanical intramedullary nail, not an AI/ML algorithm or software. No MRMC study was conducted or described, nor would it be relevant for this type of device. The document explicitly states: "No clinical testing was provided as a basis for substantial equivalence. A clinical data report based on equivalent products can be found in Appendix H." This further confirms that no clinical studies (including those involving human readers or AI assistance) were part of the submission for this particular device's performance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This question is not applicable. The device is a mechanical intramedullary nail, not an AI/ML algorithm.

7. The type of ground truth used

  • For the biomechanical tests, the "ground truth" was based on adherence to the specified ASTM standard 1264 "Standard Specifications and Test Method L for Intramedullary Fixation Devices" and the resulting performance (strength, durability) being within acceptable ranges for safety and intended use. This is a scientific/engineering ground truth defined by established test methods and safety criteria.

8. The sample size for the training set

  • This question is not applicable. The device is a mechanical intramedullary nail, not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established

  • This question is not applicable. The device is a mechanical intramedullary nail, not an AI/ML algorithm that requires a training set.

{0}------------------------------------------------

JAN 2 6 2012

K111232

Montreal, January 16th, 2012

510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS

Applicant: Pega Medical Inc. 1111 Autoroute Chomedey Laval, Quebec H7W 5J8 Canada

Contact Person: Ariel R. Duiovne

Proprietary Name: GAP Endo-Exo Medullary System

Common Name: Intramedullary Nail

Device Classification: Class II

Classification Name: Rod, Fixation, Intramedullary And Accessories 21 CFR 888-3020

Device Product Code: HSB

Establishment Reqistration Number: 9048931

Intended Use: The GAP Endo-Exo Medullary System is indicated as a temporary implant to assure alignment, stabilization and fixation of: long bones that have been surqically prepared (osteotomy) for correction of deformities or fractures caused by trauma or disease. The GAP Endo-Exo Medullary System is used for pediatric patients (child and adolescent) age 2 to 21. It can be used to correct the following conditions:

  • Diaphyseal facture of the femur, tibia and humerus .
  • Fractures of the femoral neck .
  • . Subtrochanteric, intertrochanteric and combination fractures
  • . Correction of deformities (OI, Coxa vara, Coxa valga)
  • . Nonunions and malunions

Page 1 of 2

{1}------------------------------------------------

K111232

Description: The GAP Endo-Exo Medullary System consists of an intramedullary cannulated nail linked to various types of plates via and/or mechanical screws creating a a combined laq Endoamedullary/ Exomedullary osteosythesis device. 3 and 4mm cortical screws are used to secure the nail to the bone.

Basis for substantial equivalent:

The GAP Endo-Exo Medullary System is claimed to be substantially equivalent in design and function to the following predicate devices:

  • Fassier-Duval Telescopic IM System (Stainless steel or Ti 1. pediatric nail), Pega Medical Inc. (K041393/K020885)
    1. Synthes Adolescent Lateral Entry Femoral Nail System (Ti alloy nail and screws), Synthes USA. (K070843)
    1. Titanium Pediatric Femoral Nail (Ti alloy nail and screws) Biomet Inc. (K993956)

The intended uses of these devices are the same as the GAP Endo-Exo Medullary System

Summary of Technologies: The technological characteristics of the GAP Endo-Exo Medullary System are the same or similar to the ones of the predicate devices.

Biomechanical Testing: A static 4 point bending test, torsion test and 4 point bending fatique test based on the ASTM standard 1264 "Standard Specifications and Test Method L for Intramedullary Fixation Devices" were performed to demonstrate safety. The results indicated that the GAP Endo-Exo Medullary System is functionally safe for its intended use.

Clinical Testing: No clinical testing was provided as a basis for substantial equivalence. A clinical data report based on equivalent products can be found in Appendix H.

Page 2 of 2

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Pega Medical Inc % Mr. Ariel R. Dujovne 1111 Autoroute Chomedey Laval, Quebec H7W 5J8 Canada

JAN 2 6 2012

Re: K111232

Trade/Device Name: GAP Endo-Exo Medullary System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary Fixation Rod Regulatory Class: Class II Product Code: HSB Dated: January 16. 2012 Received: January 18, 2012

Dear Mr. Dujovne:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 - Mr. Ariel R. Dujovne

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N Melkerson Director Division of Surgical, Orthopedic & Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement

510(k) Number (if known):

Device Name: GAP ENDO-EXO MEDULLARY SYSTEM

Indications for Use:

The GAP Endo-Exo Medullary System is indicated as a temporary implant to assure alignment, stabilization and fixation of: long bones that have been surgically prepared (osteotomy) for correction of deformities or fractures caused by trauma or disease. The GAP Endo-Exo Medullary System is used for pediatric patients (child and adolescent) age 2 to 21. It can be used to correct the following conditions:

  • Diaphyseal facture of the femur, tibia and humerus .
  • Fractures of the femoral neck .
  • Subtrochanteric, intertrochanteric and combination fractures .
  • Correction of deformities (OI, Coxa vara, Coxa valga) ●
  • Nonunions and malunions .

Over-The-Counter Use _ Prescription Use no AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Tan N. Phyp

Page 1 of 1

for (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K\1232

N/A