K Number
K111172

Validate with FDA (Live)

Date Cleared
2011-08-29

(125 days)

Product Code
Regulation Number
878.4630
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 4- Series Unit is a therapeutic product designed for individuals who require ultraviolet radiation for diagnosed skin disorders.

Device Description

The 4 Series Unit is a partial-body phototherapy panel ultraviolet light source controlled by an integrated digital timer or a microprocessor controller. The operator interface consists of two main components: a LCD, and a Membrane with 4 buttons. When the operator sets the treatment time or the desired dose using the operator interface located on the control panel, the treatment begins by illuminating the lamps which emit ultraviolet light. The spectral output at peak wavelengths of 305nm,311nm,350nm,or 365nm depending on lamp. The 4 Series Unit is a therapeutic product designed for individuals who require ultraviolet radiation for diagnosed skin disorders.

AI/ML Overview

This document describes a 510(k) premarket notification for the Daavlin Distributing Company's 4 Series Phototherapy Unit. The primary argument for substantial equivalence relies on the device being nearly identical to a previously cleared predicate device, the Daavlin 1 Series Phototherapy Unit (K100378).

Here's an analysis of the provided information regarding acceptance criteria and performance studies:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance
Design ConfigurationSame as predicate device (1 Series Phototherapy Unit)"constructed in the same design configuration as the predicate device"
Energy Sources (UV lamps)Identical to predicate device"utilizing identical energy sources (UV lamps)"
Material CompositionIdentical to predicate device"materials of identical composition"
Intended UseSame as predicate device"The intended use... are the same of the predicate device."
Indications for UseSame as predicate device"general and specific indications for use... are the same of the predicate device."
Spectral OutputSame as predicate device"spectral output... are the same of the predicate device." (Explicitly mentioned: "The spectral output at peak wavelengths of 305nm, 311nm, 350nm, or 365nm depending on lamp.")
Mode of OperationSame as predicate device"mode of operation... are the same of the predicate device."
LabelingSame as predicate device"labeling... are the same of the predicate device."
AccessoriesSame as predicate device"accessories... are the same of the predicate device."
General Operating PrinciplesSame as predicate device"general operating principals... are the same of the predicate device."
Treatment AreaSimilar to predicate device, but larger"The treatment area, while still partial body, is similar, but larger on the 4 Series unit than on the predicate device."
Overall Performance EquivalenceEquivalent spectral/power outputs compared to predicate"The 4 Series Phototherapy Unit performance data is the same as that of the claimed predicate devices. The Controller, UV lamps and cabinet construction used in the production of the predicate device and the 4 Series Phototherapy Unit are identical, giving equivalent spectral/power outputs."

2. Sample Size Used for the Test Set and the Data Provenance

This document does not describe a test set or clinical study in the traditional sense (e.g., patient data, prospective/retrospective studies). The claim of equivalence is based on the technical and design similarities to a predicate device. Therefore, information about sample size and data provenance for a test set is not applicable here.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Not applicable. No test set requiring expert ground truth establishment is described.

4. Adjudication Method for the Test Set

Not applicable. No test set requiring adjudication is described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The submission focuses on device equivalence through design and technical specifications, not human reader performance with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No, a standalone performance study was not done. This device is a phototherapy unit (medical equipment for light therapy), not an AI algorithm.

7. The Type of Ground Truth Used

The "ground truth" in this submission is the established performance and safety of the predicate device, the Daavlin 1 Series Phototherapy Unit (K100378). The entire argument revolves around demonstrating that the new device is fundamentally the same and performs identically or equivalently to this predicate.

8. The Sample Size for the Training Set

Not applicable. This device is a physical phototherapy unit; it does not involve machine learning algorithms or a "training set" in the computational sense.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As noted above, there is no training set for this type of device. The ground truth for the predicate device would have been established through its own regulatory clearance process, likely involving engineering testing and demonstrating safety and effectiveness for its intended use.

{0}------------------------------------------------

KIIII172

pg 1 of 2

510(k) Summary

ਾਂ ਕ

AUG 2 9 2011

The 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92

The Daavlin Distributing Company Submitter's Name: Registration Number: 1526255 205 West Bement Street Address: Bryan, Ohio 43506 419.636.6304 Telephone: Michele Thiel Contact: Date Prepared: 31 March 2011 4 Series Phototherapy Unit Device Trade Name: 4 Series Partial- Body Phototherapy Unit Device Common Name: Class II Device Classification: FTC Product Code: CFR 878.4630 Regulation Number: Ultraviolet lamp for dermatologic/skin disorders Regulation Name: Daavlin Distributing Company Predicate Device: 1 Series Phototherapy Unit K100378 Device Description:

{1}------------------------------------------------

K111172 pg 2 of 2

The 4 Series Unit is a partial-body phototherapy panel ultraviolet light source controlled by an integrated digital timer or a microprocessor controller. The operator interface consists of two main components: a LCD, and a Membrane with 4 buttons. When the operator sets the treatment time or the desired dose using the operator interface located on the control panel, the treatment begins by illuminating the lamps which emit ultraviolet light. The spectral output at peak wavelengths of 305nm,311nm,350nm,or 365nm depending on lamp. The 4 Series Unit is a therapeutic product designed for individuals who require ultraviolet radiation for diagnosed skin disorders.

Predicate Device Comparison:

The 4 Series Unit is constructed in the same design configuration as the predicate device, utilizing identical energy sources (UV lamps) and materials of identical composition. The 4 Series varies from the predicate device, in that the body is slightly larger and therefore contains more lamps to maintain the same performance. The intended use, general and specific indications for use, spectral output, mode of operation, labeling, accessories, and general operating principals of the 4 Series ultraviolet light-emitting medical device are the same of the predicate device. The treatment area, while still partial body, is similar , but larger on the 4 Series unit than on the predicate device.

Intended Use:

The 4 Series Unit is a therapeutic product designed for individuals who require ultraviolet radiation for diagnosed skin disorders.

Performance Data:

The 4 Series Phototherapy Unit performance data is the same as that of the claimed predicate devices. The Controller, UV lamps and cabinet construction used in the production of the predicate device and the 4 Series Phototherapy Unit are identical, giving equivalent spectral/power outputs.

Conclusion:

On the basis of the information provided in this Summary, the Daavlin Distributing Company believes the 4 Series Phototherapy Unit is substantially equivalent to the legally commercialized predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the circumference. Inside the circle is a stylized emblem featuring three angled lines that resemble a bird in flight or a symbolic representation of human services. The emblem is black, and the text is also in a dark color, providing a clear contrast against the likely white or light background.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Daavlin Distributing Company % Ms. Michele Thiel Management Representative 205 West Bement Street Bryan, Ohio 43506

AUG 2 9 2011

Re: K111172

Trade/Device Name: 4 Series Phototherapy Unit Regulation Number: 21 CFR 878.4630 Regulation Name: Ultraviolet lamp for dermatologic disorders Regulatory Class: II Product Code: FTC Dated: August 22, 2011 Received: August 22, 2011

Dear Ms. Thiel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

Page 2 - Ms. Michele Thiel

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

F. N. Malkerson

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

KIIII72

pg 1 of 1

Indication for Use

510(k) Number

4 Series Phototherapy Unit Device Name

Indications for Use

The 4- Series Unit is a therapeutic product designed for individuals who require ultraviolet radiation for diagnosed skin disorders.

Prescription Use X

OR (per 21 CFR 801.109)

Over-the-Counter Use_

(Please do not write below this line - Continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nel Rl Sl for mxm

(Division Sign-C Division of Surgical, Orthopedic, and Restorative Devices

. :0(k) Number K111172

§ 878.4630 Ultraviolet lamp for dermatologic disorders.

(a)
Identification. An ultraviolet lamp for dermatologic disorders is a device (including a fixture) intended to provide ultraviolet radiation of the body to photoactivate a drug in the treatment of a dermatologic disorder if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug.(b)
Classification. Class II.