K Number
K111162

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-01-13

(263 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Slim Onebody System is intended for immediate loading in partially or fully edentulous mandibles and maxillae to serve as temporary support for provisional prosthetic device during the healing phase (maximum six months) of permanent endosseous dental implants.

Device Description

The Slim Onebody System is a mini implant made of Ti-6A1-4V ELI alloy(ASTM F136) and Pure Titanium Gr4(ASTM F67). It serves as a temporary support for provisional prosthetic device during the healing phase of permanent endosseous dental implantation. Its surface is S.L.A(Sand-Blasted, Large Grit, Acid-etched). It is comprised of diameter ranging from 2.0 - 3.0mm and lengths ranging from 10mm -14mm. The product consists of a FIX type and a Ball Type implant.

AI/ML Overview

The provided document describes a 510(k) summary for the "Slim Onebody System," an endosseous dental implant. This submission focuses on non-clinical testing to demonstrate substantial equivalence to a predicate device, rather than a clinical study establishing specific acceptance criteria for performance metrics typically found in AI/ML device evaluations.

Therefore, many of the requested categories are not applicable to the information provided in this 510(k) summary.

Here's an analysis based on the available information:

Acceptance Criteria and Device Performance (Not applicable in the AI/ML sense)

The document does not define specific performance metrics like sensitivity, specificity, or AUC with acceptance thresholds. Instead, it relies on demonstrating that the device meets established engineering and biocompatibility standards, and compares its physical and material properties to a predicate device.

Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria Category (Not AI/ML specific)Acceptance Criteria (Implicit from testing)Reported Device Performance
BiocompatibilityMeet ISO7405 (2008) standards for dental devicesNon-toxic, non-sensitizing to biological tissues
- ISO Cytotoxicity NegativePassed
- ISO Systemic toxicity NegativePassed
- ISO Pyrogenicity NegativePassed
- ISO Sensitization NegativePassed
- ISO Intracutaneous reactivity NegativePassed
- ISO Irritation NegativePassed
Mechanical PerformanceWithstand anticipated forces (FDA guidance)
- ISO FatigueStrong enough
- ISO Static compressiveStrong enough
- Torque ForceStrong enough
Substantial EquivalenceSimilar raw material, mechanical/physical properties, shape, and intended use as predicate device"Slim Onebody System is substantial equal with the predicate devices."

Study that Proves the Device Meets Acceptance Criteria

The studies conducted were non-clinical bench tests and biocompatibility evaluations.

  1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not specified. Standard practice for these tests involves a certain number of samples, but the exact count isn't in this summary.
    • Data Provenance: The tests were conducted on the "Slim Onebody System" itself, presumably by Dentium Co., Ltd., which is based in Korea.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. For non-clinical bench and biocompatibility testing, "ground truth" is established by adherence to recognized international standards (ISO, ASTM) and validated testing methodologies, rather than expert consensus on interpreted data like in AI/ML. Labs performing these tests are accredited and their staff qualified, but this isn't documented as "expert" ground truth in the traditional sense for these evaluations.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. Adjudication methods like 2+1 or 3+1 (where multiple experts independently assess and an arbitrator resolves disagreements) are relevant for subjective interpretations, often in clinical imaging or diagnostic studies. Non-clinical tests typically have objective pass/fail criteria based on quantitative measurements.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an AI/ML device. It is a physical medical device (dental implant).
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This is not an AI/ML device.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For biocompatibility: Ground truth is established by the test results themselves demonstrating compliance with ISO standards (e.g., cell viability in cytotoxicity, lack of systemic reaction in systemic toxicity).
    • For mechanical testing: Ground truth is established by quantitative measurements meeting specified force/fatigue thresholds derived from FDA guidance and engineering principles.
  7. The sample size for the training set:

    • Not applicable. This device does not have a "training set" as it is not an AI/ML product.
  8. How the ground truth for the training set was established:

    • Not applicable. As above, no training set.

{0}------------------------------------------------

JAN 1 3 2012

510(k) Summary

12/14/2011

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92.

1. Company

SubmitterOfficial Correspondent
NameDentium Co., Ltd.Dentium USA
Address27-5 Iui-dong, Yeongtong-gu,Suwon-si, Gyeonggi-do, 443-766,Korea6761 Katella Avenue Cypress, CA90630, USA
Phone+82-31-207-2200714-226-0229
Fax+82-31-207-3883714-226-0019
Contact personLee, Sungwon / R&Dswlee(@dentium.com)Eunkyung Son / Doctorjson@dentiumusa.com

2. Device Name

Proprietary name:Slim Onebody System
Common name:Endosseous Dental Implant
Classification name:Implant, Endosseous, Root-Form
Product Code:DZE
Regulation Number:21CFR872.3640
Device Class:Class II

3. Predicated Device

IMTEC, Sendax MDI and MDI Plus K031106

4. Description

The Slim Onebody System is a mini implant made of Ti-6A1-4V ELI alloy(ASTM F136) and Pure Titanium Gr4(ASTM F67). It serves as a temporary support for provisional prosthetic device during the healing phase of permanent endosseous

510(k) Summary.

Page 1 of 4

{1}------------------------------------------------

dental implantation. Its surface is S.L.A(Sand-Blasted, Large Grit, Acid-etched). It is comprised of diameter ranging from 2.0 - 3.0mm and lengths ranging from 10mm -14mm. The product consists of a FIX type and a Ball Type implant.

5. Substantial Equivalence

Trade nameSlim Onebody SystemSendax MDI
ManufacturerDentium Co., Ltd.IMTEC(SENDAX)
510(k)New deviceK031106
MaterialsTitanium AlloyTitanium, Titanium Alloy
FormRoot-Form, ThreadedRoot-Form, Threaded
SterilizationSterileSterile
Length10mm~14mm10mm~18mm
Diameter2.0~ 3.01.8 ~ 2.4
Indicationsfor useThe Slim Onebody System isintended for immediate loading inpartially or fully edentulousmandibles and maxillae to serveas temporary support forprovisional prosthetic deviceduring the healing phase(maximum six months) ofpermanent endosseous dentalimplants.The MDI and MDI PLUS areself-tapping titanium threadscrews indicated for long-termintra-bony applications.Additionally, the MDI may alsobe used for inter-radiculartransitional applications.These devices will permitimmediate splinting stability andlong-term fixation of new orexisting crown and bridgeinstallations, for full partialendentulism, and employingminimally invasive surgical
UsePrescriptionintervention.Prescription

Raw material, mechanical and physical properties, shape, and intended use are similar to the predicated devices. The differences between Slim Onebody and predicate devices are the slight mechanical and physical characteristics. However, the slight differences do not affect the application of the device. Therefore, we state that Slim Onebody is substantial equal with the predicate devices.

{2}------------------------------------------------

6. Indication for Use

The Slim Onebody System is intended for immediate loading in partially or fully edentulous mandibles and maxillae to serve as temporary support for provisional prosthetic device during the healing phase (maximum six months) of permanent endosseous dental implants.

7. Non-Clinical Testing

Non-clinical test data was used to support the decision of safety and effectiveness. Non-clinical testing consisted of performance of fatigue testing in accordance with the FDA guidance Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments.

Biocompatibility testing on the proposed Slim Onebody System has been completed. Requirements for biological evaluation of the proposed device were based on the ISO7405 (2008), "Dentistry-Evaluation of biocompatibility of medical devices used in dentistry." The biocompatibility test results show that the materials used in the design and manufacture of the components of the proposed device are non-toxic and non-sensitizing to biological tissues consistent with its intended use. The following biocompatibility tests were completed:

-ISO Cytotoxicity-ISO Systemic toxicity
-ISO Pyrogenicity-ISO Sensitization
-ISO Intracutaneous reactivity-ISO Irritation

stemic toxicity nsitization plantation

The proposed Slim Onebody System was evaluated using the following performance bench testing to confirm the performance characteristics:

-ISO Fatigue -ISO Static compressive - --Torque Force

The testing indicates that the device is strong enough to withstand the anticipated forces

8. Clinical Testing

{3}------------------------------------------------

Dentir

Non-clinical test data was used to support the decision of safety and effectiveness.

9. Review

All test results demonstrate that the materials chosen, the manufacturing process, and the design utilized for the Slim Onebody System met the established specifications necessary for consistent performance according to its intended use.

Slim Onebody System has been subjected to safety, performance, and product validations prior to release. Safety tests including biocompatibility have been performed to ensure the devices comply with the applicable International and US regulations.

10. Conclusions

All of the data consistent with the recommendations in the FDA guidance document Root-form Endosseous Dental Implants testing of the implants demonstrated that the Slim Onebody System is substantially equivalent to the predicate device implant.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Dentium Company, Limited C/O Dr. Eunkyung Son Dentium USA 6761 Katella Avenue Cypress Cypress, California 90630

JAN 1 3 2012

Re: K11162

Trade/Device Name: Slim Onebody System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: DZE Dated: December 20, 2011 Received: December 21, 2011

Dear Dr. Son:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.

{5}------------------------------------------------

Page 2 - Dr. Son

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing ( very and suitements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem /default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Wh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{6}------------------------------------------------

Indications for Use

k11162 510(k) Number:

Slim Onebody System Device Name:

Indications for Use:

The Slim Onebody System is intended for immediate loading in partially or fully edentulous mandibles and maxillae to serve as temporary support for provisional prosthetic device during the healing phase (maximum six months) of permanent endosseous dental implants.

V Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Penner

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K111162

Page 1 of 1

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.