K Number
K110989

Validate with FDA (Live)

Date Cleared
2011-11-03

(209 days)

Product Code
Regulation Number
874.1820
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Neurovision Ink Printed Endotracheal Tube Electrode is intended for use during surgery and parasurgical care only, with any compatible monitoring system, for continuous EMG monitoring and assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation.

Device Description

The Neurovision Ink Printed Endotracheal Tube Electrode is a sterile, single use disposable accessory that connects to a compatible EMG monitor to provide an open airway for patient ventilation and continuous EMG monitoring of the laryngeal musculature during surgery. The ET tube is made of a flexible PVC material with an inflatable low pressure cuff. The Neurovision Ink Printed Endotracheal Tube Electrode is intended for use with any compatible EMG monitoring system, for continuous EMG monitoring of the laryngeal musculature as well as providing an open airway for patient ventilation.

AI/ML Overview

The provided 510(k) summary for the Neurovision Ink Printed Endotracheal Tube Electrode primarily focuses on establishing substantial equivalence to a predicate device through technological characteristics and non-clinical testing. It does not describe a study involving acceptance criteria for device performance in a clinical setting, nor does it involve AI, human readers, or ground truth as typically understood in the context of diagnostic or interpretive devices.

Therefore, many of the requested sections regarding acceptance criteria, study details, and expert involvement cannot be extracted from this document, as they are not applicable to the type of submission made.

Here's an analysis based on the available information:

1. Table of Acceptance Criteria and Reported Device Performance

This 510(k) does not present a table of acceptance criteria for clinical performance (e.g., sensitivity, specificity, accuracy) and corresponding reported device performance. Instead, it demonstrates substantial equivalence through "technological characteristics" and "nonclinical testing."

The document states that the subject device "is substantially equivalent" to the predicate device (NuVasive EMG Endotracheal Tube - K094054). The underlying acceptance criterion for this type of submission is that the new device is as safe and effective as the predicate, based on its similar design, intended use, materials, function, packaging, and sterilization.

The non-clinical tests performed, and their implied "acceptance criteria" through demonstrating substantial equivalence, are:

Acceptance Criterion (Implied)Reported Device Performance (as stated in document)
Wire pull resistanceDemonstrated substantial equivalence
Abrasion resistanceDemonstrated substantial equivalence
Submersion integrityDemonstrated substantial equivalence
Bend test resultsDemonstrated substantial equivalence
Impedance within acceptable rangeDemonstrated substantial equivalence
Biocompatibility (Systemic Toxicity, Intracutaneous Sensitivity, Irritation and Skin Sensitization)Met ISO 10993-11 and ISO 10993-10 standards (implied acceptance)
Sterilization ValidationPassed VDMax method (implied acceptance)
Porous Packaging ValidationConducted according to ISO 11607 (implied acceptance)

2. Sample Size Used for the Test Set and Data Provenance

Not applicable. This submission relies on non-clinical testing and comparison to a predicate device, not a test set derived from patient data.

3. Number of Experts Used to Establish Ground Truth and Qualifications

Not applicable. Since there's no clinical performance study involving diagnostic interpretation or AI, there was no need for experts to establish ground truth in that context. The "experts" involved would have been the engineers and scientists conducting the non-clinical tests and the regulatory professionals assembling the 510(k).

4. Adjudication Method for the Test Set

Not applicable. No test set involving human judgment or interpretation was used.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, an MRMC comparative effectiveness study was not done. This device is an Endotracheal Tube with EMG monitoring electrodes, not an AI or imaging interpretation device.

6. Standalone (Algorithm Only) Performance Study

No, a standalone performance study (algorithm only) was not explicitly performed or reported. The device itself is hardware. Its functionality (EMG monitoring) is contingent on a compatible monitoring system (which is not part of this 510(k)). This 510(k) focuses on the safety and technological equivalence of the tube and its electrodes.

7. Type of Ground Truth Used

The "ground truth" for the non-clinical tests would be the established safety and performance parameters for each test (e.g., a minimum force for wire pull, acceptable limits for impedance, passing criteria for biocompatibility assays). For the overall submission, the "ground truth" for substantial equivalence is the predicate device's established safety and effectiveness.

8. Sample Size for the Training Set

Not applicable. This device is not an AI algorithm and therefore does not have a training set in that context.

9. How Ground Truth for the Training Set Was Established

Not applicable. See point 8.

{0}------------------------------------------------

neurovision
MEDICAL

K110989
NOV - 3 2011 p.1/2

Traditional 510k Neurovision Ink Printed Endotracheal Tube Electrode

510K Summary

. -

In accordance with Title 21 of the Code of Federal Regulations, Part 807 and in particular 21 CFR §807.92, the following summary of information is provided:

Applicant Information

Christine Vergely Neurovision Medical Products, Inc. 2225 Sperry Ave., Suite 1000 Ventura, CA 93003 PH 805-866-6999 Fax: 413-410-4579 christie@neurovisionmedical.com

Device Identification:

Trade or Proprietary Name:Neurovision Ink Printed Endotracheal Tube Electrode.LETEIP-1, LETEIP-2
Common or Usual Name:Endotracheal Tube with Electromyography (EMG)monitoring Electrodes.
Device Class:Class II
Product Code:ETN

Predicate Device

The subject Neurovision Ink Printed Endotracheal Tube Electrode is substantially equivalent to the following predicate devices concurrently distributed commercially in the U.S.:

  • K094054 NuVasive EMG Endotracheal Tube .

Device Description

Device Description The Nedrovision This Thise Chucklashic (EMG) monitoring during surgery. The ET tube is made of a flexible PVC material with an inflatable low pressure cuff. The Neurovision Ink made of a liexble I VO matonal with browided as a sterile, single use disposable accessory Printed Endonatible EMG monitor to provide an open airway for patient ventilation that connects to a a a current Laryngel Nerve (RLN).

Intended Use

The Neurovision Ink Printed Endotracheal Tube Electrode is intended for use atinyoung EMC The Neurovision This Frinted Endottachear Tube Elocatoring system, for continuous EMG
and parasurgerical care only, with any compatible monitoring system, for continuous en w and parasungencal care only, with any only compansion musculature as well as providing an open airway for patient ventilation.

Technological Characteristics of Device in Relation to Predicate Devices

Technological Characterisatis of Deville in heat the Electrode is substantially equivalent to
The subject Neurovision Ink Printed Endotracheal Tube Electrode in the United St The subject Neurovision ink Philed Endoration in the United States. The the predicated device cleared by the rechnically equivalent and have the same technological

{1}------------------------------------------------

neurovision
MEDICAL

K110989

p. 2/2

Traditional 510k Neurovision Ink Printed Endotracheal Tube Electrode

characteristics to its predicate device through comparison in areas including design, intended use, material composition, function, packaging, and sterilization.

CharacteristicsPredicate DeviceSubject Device
Laryngeal SurfaceElectrodeNuVasive EMG EndotrachealTube(K094054)Neurovision Ink PrintedEndotracheal Tube Electrode
YESYES
EndotrachealLocationYESYES
Number ofElectrodes42 or 4
Electrode SurfaceMaterialConductive Silver Ink
Tube & CuffMaterialsPVC
Reinforcing MaterialNone
Tube DimensionsVarious DimensionsVarious Dimensions
Sterilization andPackagingSterile, Single Use OnlySterile, Single Use Only

Nonclinical testing, including wire pull, abrasion, submersion and bend tests as well as impedance measurements demonstrated that the subject Neurovision Ink Printed Endotracheal Tube Electrode is substantially equivalent to the predicate.

Biocompatibility testing included ISO 10993-11 Systemic Toxicity, ISO 10993-10 Intracutaneous Sensitivity and ISO 10993-10 Irritation and Skin Sensitization.

Sterilization Validation was by VDMax method and Porous Packaging Validation was conducted according to ISO 11607.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is in all capital letters and is evenly spaced around the circular border.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Neurovision Medical Products, Inc. c/o Ms. Christine Vergely Regulatory Manager 2225 Sperry Ave., Suite 1000 Ventura, CA 93003

NOV - 3 2011

~ Re: K110989

Trade/Device Name: Neurovision Ink Printed Endotracheal Tube Electrode Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical Nerve Stimulator/Locator Regulatory Class: Class II Product Code: ETN, BTR, GWF Dated: October 17, 2011 Received: October 19, 2011

Dear Ms. Vergely:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Malvina R. Eydelman, M.D.

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K110989

Device Name: Neurovision Ink Printed Endotracheal Tube Electrode

Indications for Use:

The Neurovision Ink Printed Endotracheal Tube Electrode is intended for use during The Nouronolon mit a care only, with any compatible monitoring system, for surgery and paraourgiour sale of the ensessment of the nerves supplying the lanyngeal musculature as well as providing an open airway for patient ventilation.

Prescription Use _ X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) JOE HUTTER

510(k) Number.

(Division Sign-Off) : Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

§ 874.1820 Surgical nerve stimulator/locator.

(a)
Identification. A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.(b)
Classification. Class II.