K Number
K110923

Validate with FDA (Live)

Date Cleared
2012-01-24

(298 days)

Product Code
Regulation Number
866.3120
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The cobas® CT/NG Test is an in vitro nucleic acid amplification test that utilizes Polymerase Chain Reaction (PCR) and nucleic acid hybridization for the qualitative detection of Chlamydia trachomatis (CT) and/or Neisseria gonormoeae (NG) DNA to aid in the diagnosis of chlamydial and gonococcal disease. The test may be used with vaginal swab specimens self-collected in a clinical setting and male urine from both symptomatic and asymptomatic individuals. Specimens to be tested should be collected in cobas® PCR Media.

Ancillary Collection Kits

The cobas® PCR Female Swab Sample Kit is used to collect and transport self-collected vaginal swab specimens in a clinical setting. The cobas® PCR Media serves as a nucleic acid stabilizing transport and storage medium for gynecological specimens. Use this collection kit only with the cobas CT/NG Test. NOTE: This collection kit should not be used for collection of alternative gynecological specimens.

The cobas® PCR Urine Sample Kit is used to collect and transport male urine specimens.

The cobas® PCR Media serves as a nucleic acid stabilizing transport and storage medium for urine specimens. Use this collection kit only with the cobas® CT/NG Test. NOTE: This collection kit should not be used for collection of female urine specimens.

Device Description

The Roche Molecular Systems (RMS) cobas® CT/NG Test consists of five reagent kits:

  • cobas® 4800 System Sample Preparation Kit .
  • cobas® 4800 CT/NG Amplification/Detection Kit .
  • cobas® 4800 CT/NG Controls Kit .
  • . cobas 4800 System Wash Buffer Kit
  • . cobas® 4800 System Control Diluent Kit

Sample Collection Kits to be used for the cobas® CT/NG Test are:

  • . cobas® PCR Female Swab Sample Kit
  • cobas® PCR Urine Sample Kit .

The cobas® CT/NG Test for Chlamvdia trachomatis (CT) and Neisseria gonorrhoeae (NG) is based on two major processes: (1) automated sample preparation to obtain nucleic acids. including CT and NG DNA; (2) simultaneous PCR amplification of target DNA sequences using both CT and NG specific complementary primer pairs and real-time detection of cleaved fluorescent-labeled CT and NG specific oligonucleotide detection probes. Internal control, containing CT and NG DNA, is added to all samples during automated sample preparation and is amplified and detected simultaneously with each sample to monitor the entire process.

The cobas 4800 System utilizes the cobas x 480 Instrument for automated sample preparation, and the automated cobas z 480 Analyzer for automated amplification and detection. The cobas 4800 system software integrates the sample preparation with nucleic acid amplification and detection to generate test results.

AI/ML Overview

Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:

cobas® CT/NG Test: Acceptance Criteria and Study Details

The cobas® CT/NG Test is an in vitro nucleic acid amplification test for the qualitative detection of Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (NG) DNA.

1. Table of Acceptance Criteria and Reported Device Performance

The document describes both non-clinical performance evaluations (analytical sensitivity, inclusivity, specificity, interference, precision) and clinical performance evaluations (reproducibility and a clinical specimen study). The acceptance criteria for analytical and reproducibility studies are generally indicated by the achieved performance (e.g., ≥ 95% detection for LOD, 100% agreement for negative controls, high positive percent agreement for reproducibility). For clinical performance, the reported sensitivity and specificity serve as the primary metrics.

Acceptance CriteriaReported Device Performance
Analytical Sensitivity (LOD): Target concentration detectable as positive in ≥ 95% of replicates.C. trachomatis (CT):- Urine: 0.75 IFU/mL- Vaginal Swabs: 10.00 IFU/mLN. gonorrhoeae (NG):- Urine: 2.25 CFU/mL- Vaginal Swabs: 100.00 CFU/mL
Inclusivity: Detection of additional CT serovars/variants and NG strains around LOD levels with high positive hit rates.CT: Analytical sensitivity for 14 serovars + nvCT ranged from 0.125 IFU/mL to 5.0 IFU/mL (PCR Media + urine) and all showed 100% positive hit rates at 10 IFU/mL in stabilized negative vaginal specimens.NG: Analytical sensitivity for 44 strains ranged from 3.0 CFU/mL to 20 CFU/mL (PCR Media) and was 3.75 CFU/mL (PCR Media + urine). All showed 100% hit rates at 100 CFU/mL in stabilized negative vaginal specimens.
Analytical Specificity: No interference with CT/NG detection or false positives from various microorganisms.No interference with CT/NG detection or false positive results were observed from 184 bacteria, fungi, and viruses, including those commonly found in urogenital tracts.
Interference: No or minimal interference from common OTC products and endogenous substances.- Replens® vaginal moisturizer produced invalid/false negative results in urine panels, but no interference in vaginal swab specimens.- Maximum allowable concentrations of whole blood (> 0.35% v/v in urine, none in vaginal), PBMC cells (> 1 x 10^5 cells/mL), and routine levels of cervical mucus showed no interference.
Precision: Anticipated hit rates (e.g., 100% for positive and 0% for negative at LOD levels) and low CVs for Ct values.CT & NG (PCR Media & PCR Media + Urine):- Negative levels: 0% hit rate- 1x LOD / Neg: 100% hit rate (CT & NG)- 1x LOD / 2.5x LOD: 100% hit rate (CT & NG)- 2.5x LOD / 1x LOD: 100% hit rate (CT & NG)- Overall CV (%) for Ct values: 1.1% to 1.5% (urine panel, CT); 1.6% to 1.8% (swab panel, CT); 1.2% to 1.5% (urine panel, NG); 1.4% to 1.9% (swab panel, NG).
Reproducibility: High percent agreement across lot, site, operator, run, and day.CT (Urine & Swab Panels):- Negative Percent Agreement (NPA): 100%- Positive Percent Agreement (PPA): 100%NG (Urine & Swab Panels):- NPA: 100%- PPA: Lowest overall was 99.52%.
Clinical Sensitivity (Overall):CT: 95.7% (246/257) (95% CI: 92.5%, 97.6%)NG: 99.0% (103/104) (95% CI: 94.8%, 99.8%)
Clinical Specificity (Overall):CT: 99.7% (2585/2594) (95% CI: 99.3%, 99.8%)NG: 99.9% (2744/2747) (95% CI: 99.7%, 100.0%)
Clinical Positive Predictive Value (PPV) at study prevalence:CT: 96.5%NG: 97.2%
Clinical Negative Predictive Value (NPV) at study prevalence:CT: 99.6%NG: 100.0%

2. Sample size used for the test set and the data provenance

  • Clinical Specimen Study:
    • Sample Size: A total of 2,851 evaluable male and female subjects were enrolled for the clinical specimen study.
    • Data Provenance: The study was a multi-center clinical investigation conducted in the United States. Specimens were collected at 12 geographically diverse sites, including OB-GYN practices, public and private STD clinics, and family planning centers. The study was prospective in nature, as subjects were enrolled, and specimens were collected and tested.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The ground truth ("Patient Infected Status") was established using a combination of two commercially available nucleic acid amplification tests (NAATs) for CT and NG, as reference assays. The number and qualifications of experts involved in the interpretation of these reference assays or in establishing a consensus based on them are not explicitly stated in the provided document. The ground truth was primarily based on the concordance of these reference NAAT results.

4. Adjudication method for the test set

The adjudication method for determining "Patient Infected Status" (PIS) was based on the combined results from two reference NAATs.

  • A subject was categorized as infected for CT or NG if a minimum of two positive results (at least one from each reference NAAT) was reported across relevant specimen types (e.g., endocervical swab/urethral swab and/or urine specimen).
  • Specific rules were applied for certain situations, such as females being categorized as non-infected for swab specimens if reference swab and PreservCyt were negative, but urine specimens were positive.
  • A subject was classified as non-infected if at least one of the reference NAATs reported negative results for all sample types.
  • If PIS could not be determined due to missing and/or indeterminate results from the reference tests, the subject was excluded from the analysis. This method implies a form of "consensus" or "composite reference standard" based on the agreement of the two predicate devices, rather than a direct expert panel adjudication in the typical sense of 2+1 or 3+1 for imaging.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This device is a fully automated in vitro diagnostic test (nucleic acid amplification test) and does not involve human readers interpreting results. Therefore, there is no AI assistance or effect size on human reader performance to report. The comparison is between the new device and established reference assays.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the primary clinical performance evaluation is a standalone (algorithm only) performance study. The cobas® CT/NG Test is an automated system for sample preparation, amplification, and real-time detection, generating results without human interpretive input. The reported sensitivity, specificity, and predictive values are for the device operating as a standalone diagnostic system.

7. The type of ground truth used

The ground truth used was a composite reference standard (CRS) or "patient infected status" (PIS). PIS was determined based on the combined results from two different, commercially available nucleic acid amplification tests (NAATs) which served as reference assays. A subject was considered infected if at least two positive results were obtained, with at least one from each reference NAAT, across various relevant specimen types.

8. The sample size for the training set

The document does not explicitly describe a separate training set for the clinical performance evaluation in the context of machine learning or AI. Diagnostic devices like the cobas® CT/NG Test typically undergo analytical validation (e.g., LOD, inclusivity, specificity) and then clinical validation with a distinct set of clinical samples. The "training" for such devices is inherent in their design and optimization prior to the formal validation studies, and not usually characterized by a distinct "training set" of patient data in the way an AI model would have. The data described (analytical and clinical) are primarily for performance evaluation and validation.

9. How the ground truth for the training set was established

As noted above, a distinct "training set" in the AI sense is not described. For the general development of the assay, the "ground truth" would be established through highly characterized reference materials (quantified cultures of CT and NG), which were used extensively in the analytical performance studies (LOD, inclusivity, analytical specificity).

{0}------------------------------------------------

Roche Molecular Systems, Inc. Pleasanton, CA 94588-2722

JAN 2 4 2012

K110923

cobas® CT/NG Test Section 5: 510(k) Summary 510(k) Summary Report

cobas® CT/NG Test 510(k) Summary

Submitted by:Roche Molecular Systems, Inc.4300 Hacienda DrivePleasanton, CA 94588-2722Phone Number: (925) 730-8729Fax Number: (925) 730-8128
Contact:James Bonds
Date of Preparation:January 24, 2012
Device Trade Name:cobas® CT/NG Test
Common Name:Chlamydia trachomatis (CT) and Neisseria gonorrhoea (NG) Test
Type of Test:Nucleic Acid Amplification Test, DNA, Chlamydia trachomatis (CT) andNeisseria gonorrhoea (NG), qualitative
Classification Names:Chlamydia serological reagentsNeisseria spp. Direct serological test reagentsReal Time Nucleic Acid Amplification System
Regulations:866.3120866.3390862.2570
Product codes:MKZ (DNA Probe, Nucleic Acid Amplification, Chlamydia)LSL (DNA Reagents, Neisseria)OOI (Real Time Nucleic Acid Amplification System)
Panel:Microbiology
Predicate Devices - Assay:Gen-Probe APTIMA Combo 2 Assay (K060652)BD ProbeTec Q* Chlamydia trachomatis Amplified DNA Assay (K091724)and BD ProbeTec Q* Neisseria gonorrhoeae Amplified DNA Assay'(K091730)
Predicate Device - CollectionKitsAbbott multi-Collect Specimen Collection Kit (K092704)

510(k) Premarket Notification

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------↓ 510(k) Summary Report Page 1

r

{1}------------------------------------------------

TABLE OF CONTENTS

1.Device Description
2.Intended Use
3.Technological Characteristics
4.Non-Clinical Performance Evaluation
4.1.Analytical Sensitivity
4.2.Inclusivity
4.3.Analytical Specificity
4.4.Interference
4.5.Precision
5.Clinical Performance
5.1.Reproducibility
5.1.1.C. trachomatis (Table 10 and Table 11)
5.1.2.N. gonorrhoeae (Table 12 and Table 13)
5.2.Clinical Specimen Study
5.2.1.Study Design
5.2.2.Determination of Patient Infected Status
5.2.3.Study Results
5.2.4.Conclusion Drawn from Clinical Specimen Study
6.Conclusion

List of Tables

Table 1: Comparison of the cobas® CT/NG Test with the Predicate Devices6
Table 2: Comparison of the cobas® PCR Sample Collection Devices with the Predicate Device7
Table 3: cobas® CT/NG Test Limit of Detection8
Table 4: Summary of CT Serovars/Variant Inclusivity Verification Results9
Table 5: Summary of NG Strains Inclusivity Verification Results10
Table 6: Microorganisms Tested for Analytical Specificity
Table 7: List of Microorganisms Tested Below 1 x 106 copies/mL for Analytical Specificity 13
Table 8: Results from Endogenous Interference Testing
Table 9: In-House Precision Study Hit Rate Analysis
Table 10: C. trachomatis: Percent Agreement by Panel Member for Lot, Site/Instrument, andDay - PCR Media/Urine
Table 11: C. trachomatis: Percent Agreement by Panel Member for Lot, Site/Instrument, andDay - PCR Media/Swab
Table 12: N. gonorrhoeae: Percent Agreement by Panel Member for Lot, Site/Instrument, andDay - PCR Media/Urine
Table 13: N. gonorrhoeae: Percent Agreement by Panel Member for Lot, Site/Instrument, andDay - PCR Media/Swab
Table 14: Determination of Patient Infected Status
Table 15: CT Clinical Performance Compared with Patient Infected Status by Gender, SampleType, and Symptom Status
Table 16: NG Clinical Performance Compared With Patient Infected Status by Gender, SampleType, and Symptom Status
Table 17: CT Positive/Negative Analysis for Female Patient Infected Status
Table 18: CT Positive/Negative Analysis for Male Patient Infected Status
Table 19: NG Positive/Negative Analysis for Female Patient Infected Status
Table 20: NG Positive/Negative Analysis for Male Patient Infected Status
Table 21: Positive Predictive Value and Negative Value for Hypothetical CTPrevalence
Table 22: Positive Predictive Value and Negative Value for Hypothetical NGPrevalence

{2}------------------------------------------------

{3}------------------------------------------------

DEVICE DESCRIPTION 1.

The Roche Molecular Systems (RMS) cobas® CT/NG Test consists of five reagent kits:

  • cobas® 4800 System Sample Preparation Kit .
  • cobas® 4800 CT/NG Amplification/Detection Kit .
  • cobas® 4800 CT/NG Controls Kit .
  • . cobas 4800 System Wash Buffer Kit
  • . cobas® 4800 System Control Diluent Kit

Sample Collection Kits to be used for the cobas® CT/NG Test are:

  • . cobas® PCR Female Swab Sample Kit
  • cobas® PCR Urine Sample Kit .

The cobas® CT/NG Test for Chlamvdia trachomatis (CT) and Neisseria gonorrhoeae (NG) is based on two major processes: (1) automated sample preparation to obtain nucleic acids. including CT and NG DNA; (2) simultaneous PCR amplification of target DNA sequences using both CT and NG specific complementary primer pairs and real-time detection of cleaved fluorescent-labeled CT and NG specific oligonucleotide detection probes. Internal control, containing CT and NG DNA, is added to all samples during automated sample preparation and is amplified and detected simultaneously with each sample to monitor the entire process.

The cobas 4800 System utilizes the cobas x 480 Instrument for automated sample preparation, and the automated cobas z 480 Analyzer for automated amplification and detection. The cobas 4800 system software integrates the sample preparation with nucleic acid amplification and detection to generate test results.

{4}------------------------------------------------

and the same of the same of the same of the same of the states of the states of the states of the states of the states of the states of the states of the states of the states

2. INTENDED USE

Assav

The cobas® CT/NG Test is an in vitro nucleic acid amplification test that utilizes Polymerase Chain Reaction (PCR) and nucleic acid hybridization for the qualitative detection of Chlamvdia trachomatis (CT) and/or Neisseria gonorrhoeae (NG) DNA to aid in the diagnosis of chlamydial and gonococcal disease. The test may be used with vaginal swab specimens self-collected in a clinical setting and male urine from both symptomatic individuals. Specimens to be tested should be collected in cobas® PCR Media.

Ancillary Collection Kits

The cobas® PCR Female Swab Sample Kit is used to collect and transport self-collected vaginal swab specimens in a clinical setting. The cobas® PCR Media serves as a nucleic acid stabilizing transport and storage medium for gynecological specimens. Use this collection kit only with the cobas® CT/NG Test. NOTE: This collection kit should not be used for collection of alternative gynecological specimens.

The cobas PCR Urine Sample Kit is used to collect and transport male urine specimens. The cobas® PCR Media serves as a nucleic acid stabilizing transport and storage medium for . urine specimens. Use this collection kit only with the cobas CT/NG Test. NOTE: This collection kit should not be used for collection of female urine specimens.

3. TECHNOLOGICAL CHARACTERISTICS

The primary technological characteristics and intended use of the RMS cobas® CT/NG Test are substantially equivalent to other legally marketed nucleic acid amplification tests intended for the qualitative detection of Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (NG).

As indicated in Table 1, the RMS cobas® CT/NG Test is substantially equivalent to significant characteristics of the identified predicate devices, the Gen-Probe APTIMA Combo 2 Assay (K060652), the BD ProbeTec Q* Chlamydia trachomatis Amplified DNA Assay (K091724) and the BD ProbeTec Q* Neisseria gonorrhoeae Amplified DNA Assay (K091730).

{5}------------------------------------------------

Submitted Device:RMS cobas® CT/NG TestPredicate Device:Gen-Probe APTIMA Combo2 Assay (K060652)Predicate Devices:BD ProbeTec Q CTAmplified DNA Assay(K091724) and BDProbeTec Q GC AmplifiedDNA Assay (K091730)
GeneralIntended UseQualitative in vitro diagnostictest for the direct qualitativedetection of Chlamydiatrachomatis and/or Neisseriagonorrhoeae in patientspecimenssamesame
SampleTypesMale urinePatient-collected vaginalswabsMale urineMale urethral swabsFemale urineEndocervical swabsClinician-collected vaginalswabsPatient-collected vaginalswabsCervical specimens inPreservCyt® mediaMale urineMale urethral swabsFemale urineEndocervical swabsPatient-collected vaginalswabsCervical specimens inPreservCyt® and SurePath®media
SubjectStatusAsymptomatic andsymptomaticAsymptomatic andsymptomaticAsymptomatic andsymptomatic
SampleCollectionDevicesUrine collection kitSwab collection kitUrine collection kitSwab collection kitUrine collection kitVaginal collection kit
CT AnalyteTargetsCT cryptic plasmid DNACT ompA geneCT ribosomal RNACT cryptic plasmid DNA
NG AnalyteTargetsNG genomic DNANG ribosomal RNANG genomic DNA
SamplePreparationProcedureSemi-automatedSemi-automated/automatedManual/semi-automated
AmplificationTechnologyReal-time PCRRibosomal RNA transcriptionmedicated amplification(TMA)Strand displacement DNAamplification (SDA)
DetectionChemistryPaired reporter andquencher fluorescencelabeled probes (TaqManTechnology) usingfluorescence resonanceenergy transfer (FRET)Photon measurment fromselectively hybridizedchemiluninescent probesreported as Relative LightUnits (RLU)Fluorescent dye labeledprobes using fluorescenceresonance energy transfer(FRET)
Submitted Device:RMS cobas® CT/NG TestPredicate Device:Gen-Probe APTIMA Combo2 Assay (K060652)Predicate Devices:BD ProbeTec Qx CTAmplified DNA Assay(K091724) and BDProbeTec Qx GC AmplifiedDNA Assay (K091730)
ResultAnalysisBased on PCR cyclethreshold (Ct) analysisDetermined by a cut-offbased on the total RLU andkinetic curve typeDetermined by relatingMOTA scores (signalstrength) to pre-determinedcutoff values

Table 1: Comparison of the cobas® CT/NG Test with the Predicate Devices

.

{6}------------------------------------------------

As indicated in Table 2, the cobas® PCR Female Swab Sample Kit and the cobas® PCR Urine Sample Kit are substantially equivalent to those characteristics for the predicate device. The collection kits have the same basic components and intended use, and all are intended as accessories to their respective assays. Data presented in this pre-market notification further support the substantial equivalence of the subject devices to the predicate.

Table 2: Comparison of the cobas® PCR Sample Collection Devices with the Predicate Device

Submitted Device:RMS cobas® PCR FemaleSwab Sample KitSubmitted Device:RMS cobas® PCR UrineSample KitPredicate Device:Abbott multi-CollectSpecimen Collection Kit(K092704)
GeneralIntended UseCollection and transport ofvaginal swab specimensself-collected in a clinicalenvironment for the detectionof Chlamydia trachomatisand Neisseria gonorrhoeaeper the instructions provided.Not intended for home use.Collection and transport ofurine specimens for thedetection of Chlamydiatrachomatis and Neisseriagonorrhoeae per theinstructions provided. Notintended for home use.Collection and transport ofmale and female, swab andurine specimens for thedetection of Chlamydiatrachomatis and Neisseriagonorrhoeae per instructionsprovided. Not intended forhome use.
SampleTypesVaginal swab self-collectedin a clinical settingMale urineMale urineMale urethral swabsFemale urineEndocervical swabsVaginal swabs, cliniciancollected and self-collected

{7}------------------------------------------------

In summary, the intended use, technology, and functionality of the cobas® CT/NG Test as compared to the predicate devices do not raise any new types of safety or effectiveness questions and are substantially equivalent.

4. NON-CLINICAL PERFORMANCE EVALUATION

4.1. Analytical Sensitivity

The analytical sensitivity (Limit of Detection or LOD) for the cobas® CT/NG Test was determined by analyzing dilutions of quantified CT and NG cultures of CT and NG were diluted into negative vaginal swab specimen in cobas® PCR Media and negative urine specimen plus cobas® PCR Media to determine the LOD for vaginal swab and urine specimens, respectively. All levels were analyzed using the full cobas® CT/NG Test workflow across 3 unique lots of cobas® CT/NG Test reagents. This test defines LOD as the target concentration which can be detected as positive in ≥ 95% of the replicates tested with each reagent lot.

The LOD for the CT serovar D culture and NG strain 19424 in cobas® PCR Media, vaginal swab specimens stabilized in cobas® PCR Media and urine specimens diluted into cobas® PCR Média are shown in Table 3.

C. trachomatisN. gonorrhoeae
Specimen TypesLevelsTestedReplicates/LevelLOD(IFU/mL)LevelsTestedReplicates/LevelLOD(CFU/mL)
Urine7192*0.757192*2.25
Vaginal Swabs5192**10.005192**100.00

Table 3: cobas® CT/NG Test Limit of Detection

*Testing included one negative level with 167-168 replicates

** Testing included one negative level with 82-84 replicates

4.2. Inclusivity

The sensitivity of the cobas CT/NG Test was determined for 14 additional CT serovars, the Swedish new variant (nvCT) strain and an additional 44 independently isolated strains of NG. Testing was done to demonstrate that these targets can be detected around the LOD levels determined during analytical sensitivity testing for the CT serovar D culture and NG strain

{8}------------------------------------------------

  1. Panels were prepared as described for the LOD study with the number of panel levels varying from 1 to 5, as required. At least 49 replicates were tested for each panel level using one lot of cobas CT/NG Test reagents. Results are shown in Table 4 and Table 5. In Table 5, all NG strains with identical LOD results are presented as a group, shown in the columns labeled "Numbers of NG Strains." Since LOD evaluation was done with samples stabilized in cobas PCR Media, the LOD for untreated urine is twice the level reported in Table 4 and Table 5.

The analytical sensitivity for all 14 CT serovars plus the nvCT variant (Table 4) ranged from 0.2 IFU/mL (IFU= Inclusion Forming Units) to 5.0 IFU/mL in cobas® PCR Media and from 0.13 IFU/mL to 0.75 IFU/mL in cobas® PCR Media plus negative urine specimen. All CT serovars and the nvCT variant were tested at 10 IFU/mL only in stabilized negative vaginal specimen and all showed 100% positive hit rates at 10 IFU/mL.

The analytical sensitivity for all 44 NG strains ranged from 3.0 CFU/mL to 20 CFU/mL in cobas® PCR Media and was 3.75 CFU/mL in cobas® PCR Media plus urine specimens. All NG strains were tested at 100 CFU/mL only in stabilized negative vaginal specimen. All showed 100% hit rates at 100 CFU/mL (CFU = Colony Forming Units).

LOD Results for C. trachomatis
Serovar Type orVariantVaginal Swabs*Urine
IFU/mL% PosIFU/mL% Pos
A10.0100%0.12598%
B10.0100%0.75100%
Ba10.0100%0.75100%
C10.0100%0.75100%
E10.0100%0.75100%
F10.0100%0.75100%
G10.0100%0.75100%
H10.0100%0.75100%
I10.0100%0.7598%
J10.0100%0.12596%
K10.0100%0.75100%
LV Type 110.0100%0.125100%

Table 4: Summary of CT Serovars/Variant Inclusivity Verification Results

{9}------------------------------------------------

LV Type 210.0100%0.125100%
LV Type 310.0100%0.125100%
nvCT10.0100%0.75100%

Table 5: Summary of NG Strains Inclusivity Verification Results

Numbers of NGStrainsLOD Urine
CFU/mL% Hit Rate
33.7596%
43.7598%
373.75100%
Total = 44
Numbers of NGStrainsLOD Vaginal Swabs*
CFU/mL% Hit Rate
Total = 44100100%

4.3. Competitive Inhibition

Panels were prepared by spiking CT and NG cultures into urine and vaginal specimens stabilized in cobas® PCR Media to various concentration levels to examine the potential for competitive inhibition. Panels were prepared with two strains each of CT and NG. Panels were tested in one run per day over the course of 5 days. Two replicates of each panel member were tested in every run, generating a maximum of 10 test results for each level and CT and NG strain respectively. Average Ct values for each of the panel levels are summarized in Tables 6 and 7. All CT and NG hit rates were 100% for all panel levels in both matrices. Competitive inhibition was not seen in any combination of CT and NG levels in either matrix.

Table 6: Competitive inhibition Study for CT and NG Cultures in Urine Stabilized in
cobas® PČR Media (Ct Values)
Panel LevelStrain 1Strain 2
CT Level/IFU/mLNG Level /(CFU/mL)CTNGCTNG
Low/2Low/635.134.832.534.8
Low/2Medium/2434.933.132.633.0
Medium/8Low/633.534.130.135.0

{10}------------------------------------------------

High/1.00E+05Low/619.434.118.635.0
Low/2High/1.00E+0635.217.632.617.4
High/1.00E+05High/1.00E+0619.518.118.917.7
Table 7: Competitive inhibition Study for CT and NG Cultures in Vaginal Swabs
Collected in cobas® PCR Media (Ct Values)
Panel LevelStrain 1Strain 2
CT Level/IFU/mLNG Level /(CFU/mL)CTNGCTNG
Low/25Low/25036.435.134.134.7
Low/25Medium/100036.233.233.731.8
Medium/100Low/25034.535.132.234.4
High/1.00E+05Low/25023.933.822.933.8
Low/25High/1.00E+0635.722.833.822.0
High/1.00E+05High/1.00E+0623.723.221.922.1

4.4. Analytical Specificity

A panel of 184 bacteria, fungi and viruses, including those commonly found in the male/female urogenital tract, as well as representatives of N. cineria, N. flava, N. lactamica, N. perflava and N. subflava and other phylogenetically related organisms, were tested with the cobas® CT/NG Test to assess analytical specificity. The organisms listed in Table 6 were spiked at high concentrations (microorganisms tested below 1 x 106 copies/mL are listed in Table 9) into CT/NG negative urine specimens plus cobas PCR Media and pooled negative vaginal specimen spiked with CT and NG cultures at 3 times the limit of detection.

Results indicated that none of these organisms interfered with detection of CT and NG or produced a false positive result in the CT/NG negative matrices.

Achromobacter xerosisHelicobacter pyloriNeisseria sicca
Acinetobacter calcoaceticusHepatitis B virus (HBV)Neisseria subflava
Acinetobacter IwoffiHepatitis C virus (HCV)Neisseria subflava 6458
Acinetobacter sp. genospecies 3Human immunodeficiency virusNeisseria subflava 6617
Actinomyces israeliiHuman papillomavirus type 16(CaSki cells)Neisseria subflava 6618
Actinomyces pyogenesHuman papillomavirus type 18(HeLa cells)Neisseria subflava 7441
AdenovirusHerpes Simplex Virus (HSV-1)Neisseria subflava 7452
Aerococcus viridansHerpes Simplex Virus (HSV-2)Neisseria weaverii
Aeromonas hydrophilaKingella dentrificansPantoea agglomerans
Alcaligenes faecalisKingella kingaeParacoccus denitrificans
Bacillus subtilisKlebsiella oxytocaPasteurella maltocida
Bacillus thuringiensisKlebsiella pneumoniae ss ozaenaePediococcus acidilactica
Bacteroides caccaeLactobacillus acidophillusPeptostreptococcus anaerobius
Bacteroides fragilisLactobacillus brevisPeptostreptococcus asacharolyticus
Bacteroides ureolyticusLactobacillus crispatusPeptostreptococcus magnus
Bifidobacterium adolescentisLactobacillus delbrueckii subsp.lactisPlesiomonas shigelloides
Bifidobacterium breveLactobacillus jenseniiPrevotella bivia
Bifidobacterium longumLactobacillus lactis lactisPrevotella corporis
Branhamella catarrhalisLactobacillus orisPrevotella intermedia
Brevibacterium linensLactobacillus parabuchneriPropionibacterium acnes
Campylobacter gracilisLactobacillus vaginalisProteus mirabilis
Campylobacter jejuniLactococcus lactis cremorisProteus vulgaris
Candida albicansLegionella bozemniiProvidencia stuartii
Candida glabrataLegionella pneumophilaPseudomonas aeruginosa
Candida guilliermondiListeria monocytogenesPseudomonas fluorescens
Candida kruseiMicrococcus luteusPseudomonas putida
Candida parapsilosisMobiluncus curtisii subsp. curtisiiRahnella aquatilis
Candida tropicalisMobiluncus curtisii subsp. holmesiiRhizobium radiobacter
Chlamydophila pneumoniaeMobiluncus mulierisRhodospirillum rubrum
Chromobacter violaceumMoraxella catarrhalisRuminococcus productus
ChryseobacteriummeningosepticumMoraxella lacunataSaccharomyces cerevisiae
Citrobacter braakiiMoraxella osloensisSalmonella Choleraesuis
Citrobacter freundiiMorganella morganiiSalmonella Minnesota
Clostridium innocuumMycobacterium aviumSalmonella Typhimurium
Clostridium perfringensMycobacterium gordonaeSerratia denitrificans
Clostridium sporogenesMycobacterium smegmatisSerratia marcescens
Corynebacterium genitaliumMycoplasma genitaliumStaphylococcus aureus
Corynebacterium renaleMycoplasma hominisStaphylococcus epidermidis
Corynebacterium xerosisMycoplasma pneumoniaeStaphylococcus saprophyticus
Cryptococcus neoformansNeisseria cinerea 832Streptococcus agalactiae
CytomegalovirusNeisseria cinerea 3306Streptococcus anginosus
Deinococcus radioduransNeisseria cinerea 3307Streptococcus bovis
Deinococcus radionugnansNeisseria cinerea 3308Streptococcus dysgalactiae
Derxia gummosaNeisseria cinerea 6317Streptococcus equinis
Edwardsiella tardaNeisseria denitrificansStreptococcus mitis
Eikenella corrodensNeisseria elongata ssp.nitroreducansStreptococcus mutans
Enterobacter aerogenesNeisseria flavaStreptococcus pneumoniae
Enterobacter cloacaeNeisseria flavescensStreptococcus pyogenes
Enterococcus aviumNeisseria kochiStreptococcus salivarius
Enterococcus faecalisNeisseria lactamicaStreptococcus sanguis
Enterococcus faeciumNeisseria meningitidis 135Streptomyces griseinus
Epstein Barr VirusNeisseria meningitidis Serogroup ATreponema pallidum
Erwinia herbicolaNeisseria meningitidis Serogroup BTrichomonas vaginalis
Erysipelothrix rhusiopathiaeNeisseria meningitidis Serogroup CUreaplasma urealyticum
Escherichia coliNeisseria meningitidis Serogroup DVeillonela parvula
Ewingella americanaNeisseria meningitidis Serogroup YVibrio parahaemolyticus
Flavobacterium meningosepticumNeisseria mucosaWeissella paramesenteroides
Fusobacterium nucleatumNeisseria perflava 837Yersinia enterocolitica
Gardnerella vaginalisNeisseria perflava 911
Gemella haemolysansNeisseria perflava 6339
Gemella morbillorumNeisseria perflava 6340
Haemophilus ducreyiNeisseria perflava 6341
Haemophilus influenzaeNeisseria polysaccharea

Table 8: Microorganisms Tested for Analytical Specificity

{11}------------------------------------------------

{12}------------------------------------------------

Table 9: List of Microorganisms Tested Below 1 x 106 copies/mL for Analytical Specificity

Microorganism TestedConcentration Tested in Listed Matrix*Negative UrineSpecimen pluscobas®PCR MediaNegative VaginalSpecimen pluscobas®PCR Media
Adenovirus8x105 copies/mL8x105 copies/mL
Chlamydophila pneumoniae1.1x104 copies/mL1.1x104 copies/mL
Gemella morbillorum4.5 x 104 copies/mL
Hepatitis C virus (HCV)5.6 x 104 copies/mL5.6 x 104 copies/mL
Human papillomavirus (HPV) type 16(SiHa cells)5x104 copies/mL5x104 copies/mL
Human papillomavirus (HPV) type 18(HeLa cells)1x104 copies/mL1x104 copies/mL
Neisseria cinerea 33084x105 copies/mL4x105 copies/mL
Prevotella bivia9x104 copies/mL
Prevotella corporis1.4x105 copies/mL
Treponema pallidumNot Tested1x105 copies/mL
Trichomonas vaginalis6.5x105 copies/mL

{13}------------------------------------------------

*Gray cells indicate concentration tested was ≥ 1 x 106 copies/mL in that matrix

4.5. Interference

Interference testing was performed using cobas® PCR Media plus negative urine and negative vaginal swab specimen spiked with CT and NG cultures at ~ 3 x LOD for each target. Eighteen over-the-counter (OTC) products, including contraceptive jelly, lubricants, feminine sprays, antifungal cream and anti-itch cream, as well as whole blood, cervical mucus and PBMC cells were tested for interference. Of the 18 OTC products tested, Replens® vaginal moisturizer produced invalid and/or false negative results in the cobas® PCR Media plus negative urine panel samples. No interference from Replens® vaginal moisturizer was observed with vaginal swab specimens tested.

The levels of whole blood, mucus and PBMC cells shown in Table 10 represent maximum allowable concentrations which will not interfere with cobas® CT/NG Test performance. Concentrations in urine samples were determined using total sample volume, including stabilizing media.

Blood (v/v)PBMC (cells/mL)Mucus
Conc.TestedInterferenceObservedConc.TestedInterferenceObservedConc.TestedInterferenceObserved
Urine stabilized incobas® PCR Media0, 0.25%,0.35%, 0.5%,1%, 3%> 0.35%0, 1.0E+05,1.0E+06,1.0E+07> 1 x 105NTNT
Vaginal Specimenstabilized in cobas® PCRMedia0, 1%,3%, 5%, 10%None0, 1.0E+05,1.0E+06,1.0E+07> 1 x 105Routinelevel*None

Table 10: Results from Endogenous Interference Testing

NT = Not Tested

*Routine level = Quantity of cervical mucus equivalent to amount normally removed prior to sampling

4.6. Precision

Precision of the cobas CT/NG Test was examined in-house using a test panel composed of CT and NG cultures diluted into cobas® PCR Media and cobas® PCR Media mixed with

{14}------------------------------------------------

negative urine. The precision panel was designed to include members with either CT or NG at approximately the LOD for the panel matrix, members with both CT and NG at approximately the LOD and 2.5 x LOD for the panel matrix and a negative level. Testing was done with three unique lots of cobas CT/NG Test reagents and three instruments for a total of 24 runs. A description of the precision panels and the study performance in % hit rate are shown in Table 11. All positive panel levels yielded the anticipated hit rates. All negative panel levels tested negative throughout the study.

PanelNumberPanel MatrixTarget Conc.NTestedN PosCTN PosNGHit Rate95% CI
CTNGLowerUpper
1cobas® PCR MediaNegNeg144000%0.02.5
2cobas® PCR Media1 X LODNeg1441440100%97.5100.0
3cobas® PCR MediaNeg1 X LOD1440144100%97.5100.0
4cobas® PCR Media1 X LOD2.5 X LOD144144144100%97.5100.0
5cobas® PCR Media2.5 X LOD1 X LOD144144144100%97.5100.0
1cobas® PCR Media +UrineNegNeg144000%0.02.5
2cobas® PCR Media +Urine1 X LODNeg1441440100%97.5100.0
3cobas® PCR Media +UrineNeg1 X LOD1440144100%97.5100.0
4cobas® PCR Media +Urine1 X LOD2.5 X LOD144144144100%97.5100.0
5cobas® PCR Media +Urine2.5 X LOD1 X LOD144144144100%97.5100.0

Table 11: In-House Precision Study Hit Rate Analysis

*99.3 Hit Rate for NG. CT Hit Rate is 100%

5. . CLINICAL PERFORMANCE

The clinical performance characteristics of the cobas® CT/NG Test were established in two multi-center clinical investigations conducted in the United States. One study evaluated the reproducibility at one internal and two external testing sites. The other study evaluated the sensitivity, specificity, and predictive values of the cobas® CT/NG Test on clinical specimens.

{15}------------------------------------------------


5.1. Reproducibility

A Reproducibility Study was performed across lot, testing site, operator, run, and day for the cobas 8 CTNG Test using 2 panels prepared from swabs and urine collected in cobas PCR Media. A run for cobas® PCR Media (urine or swab) included 3 replicates of each of 5 panel members and 1 positive and 1 negative control (17 total tests). If cobas PCR Media panels were combined in a run, only 1 positive and 1 negative control were included (32 total tests). The 2 operators at each site performed 2 runs per day, for a total of 3 days of testing per operator per panel type (6 days of testing total for each panel type and reagent lot). Testing was performed with 2 reagent lots (6 days of testing per lot).

Overall. 74 runs were performed. and 72 valid runs were obtained for urine and swab panel types. The 2 invalid runs were due to instrument errors. A total of 1,080 tests were performed on the 5 panel members for each panel type in the valid runs. There was 1 invalid test result in the urine panel type, and 2 invalid test results in the swab panel type. These invalid tests were due to instrument errors.

All valid test results were included in the analyses that reported the percent agreement for CT and NG for each panel type separately. There were no false positive results for either analyte (CT and NG) for both panel types for negative panel members, thus giving negative percent agreement (NPA) of 100% for each analyte.

C. trachomatis (Table 12 and Table 13) 5.1.1.

Percent agreement for the positive panel members was excellent across both panel types and panel members showing a positive percent agreement (PPA) of 100%.

Analysis of variance components of the Ct values from valid tests performed on positive panel members yielded overall CV (%) ranges from 1.1% to 1.5% for the urine panel type and 1.6% to 1.8% for the swab panel type.

Table 6: C. trachomatis: Percent Agreement by Panel Member for Lot, Site/Instrument, and Day - PCR Media/Urine

THE COLLECTION CONTRACT CONTRACTOR COLLECTION CONTROLLERS CONTRACTOR COLLECTION CONTRACTOR COLLECTION CONTRACTOR COLLECTION CONTRACTOR COLLECTION CONTRACTOR COLLECTION CONTRAanel11A 44. 44. 146-46. 44. 446Cice Castel------Actively All American1 40 000 0000
and procession of the confinited and contract and consideration of the differentComments of the American Company Company Company Company Company of Children

{16}------------------------------------------------

.

MemberSDCV %LotSite/InstrumentDay
Negative CT, Negative NGn/an/a2100.0108/1081100.071/711100.072/72
3100.0107/1072100.072/722100.071/71
3100.072/723100.072/72
1x LOD CT, Negative NG0.541.52100.0108/1081100.072/721100.072/72
3100.0108/1082100.072/722100.072/72
3100.072/723100.072/72
Negative CT, 1x LOD NGn/an/a2100.0108/1081100.072/721100.072/72
3100.0108/1082100.072/722100.072/72
3100.072/723100.072/72
1x LOD CT, 2.5x LOD NG0.481.32100.0108/1081100.072/721100.072/72
3100.0108/1082100.072/722100.072/72
3100.072/723100.072/72
2.5x LOD CT, 1x LOD NG0.401.12100.0108/1081100.072/721100.072/72
3100.0108/1082100.072/722100.072/72
3100.072/723100.072/72
  • For Negative samples, Percent Agreement = (number of negative results/total valid results);

For Positive samples, Percent Agreement = (number of positive results/total valid results)

{17}------------------------------------------------

PanelMemberCtSDCtCV%LotSite/InstrumentDay
Negative CT,Negative NGn/an/a2100.0108/1081100.072/721100.072/72
3100.0108/1082100.072/722100.072/72
3100.072/723100.072/72
1x LOD CT,Negative NG0.611.62100.0107/1071100.072/721100.072/72
3100.0108/1082100.071/712100.072/72
3100.072/723100.071/71
Negative CT,1x LOD NGn/an/a2100.0108/1081100.072/721100.072/72
3100.0107/1072100.071/712100.071/71
3100.072/723100.072/72
1x LOD CT,2.5x LOD NG0.661.82100.0108/1081100.072/721100.072/72
3100.0108/1082100.072/722100.072/72
3100.072/723100.072/72
2.5x LOD CT,1x LOD NG0.591.62100.0108/1081100.072/721100.072/72
3100.0108/1082100.072/722100.072/72
3100.072/723100.072/72

Table 7: C. trachomatis: Percent Agreement by Panel Member for Lot, Site/Instrument, and Day - PCR Media/Swab

  • For Negative samples, Percent Agreement = (number of negative results/total valid results); For Positive samples, Percent Agreement = (number of positive results/total valid results)

5.1.2. N. gonorrhoeae (Table 12 and Table 13)

Percent agreement for the positive panel members was excellent across all panel types and panel members. The lowest overall PPA was 99.52% for the "Negative CT, 1 x LOD NG" panel member for PCR Media/Urine panel type.

Analysis of variance components of the Ct values from valid tests performed on positive panel members yielded overall CV (%) ranges from 1.2% to 1.5% for the urine panel type and 1.4% to 1.9% for the swab panel type.

{18}------------------------------------------------

PanelMemberCtSDCtCV %LotSite/InstrumentDay
Negative CT,Negative NGn/an/a2100.0108/1081100.071/711100.072/72
3100.0107/1072100.072/722100.071/71
3100.072/723100.072/72
1x LOD CT,Negative NGn/an/a2100.0108/1081100.072/721100.072/72
3100.0108/1082100.072/722100.072/72
3100.072/723100.072/72
Negative CT,1x LOD NG0.531.5299.1107/1081100.072/721100.072/72
3100.0108/1082100.072/72298.671/72
398.671/723100.072/72
1x LOD CT,2.5x LOD NG0.411.22100.0108/1081100.072/721100.072/72
3100.0108/1082100.072/722100.072/72
3100.072/723100.072/72
2.5x LOD CT,1x LOD NG0.541.52100.0108/1081100.072/721100.072/72
3100.0108/1082100.072/722100.072/72
3100.072/723100.072/72

Table 8: N. gonorrhoeae: Percent Agreement by Panel Member for Lot, Site/Instrument, and Day - PCR Media/Urine

1 For Negative samples, Percent Agreement = (number of negative results/total valid results);

For Positive samples, Percent Agreement = (number of positive results/total valid results)

PanelMemberCtSDCtCV%Percent Agreement ¹LotSite/InstrumentDay
Negative CT,Negative NGn/an/a2100.0108/1081100.072/721100.072/72
3100.0108/1082100.072/722100.072/72
3100.072/723100.072/72
1x LOD CT,Negative NGn/an/a2100.0107/1071100.072/721100.072/72
3100.0108/1082100.071/712100.072/72
3100.072/723100.071/71
Negative CT,1x LOD NG0.681.82100.0108/1081100.072/721100.072/72
3100.0107/1072100.071/712100.071/71
3100.072/723100.072/72

Table 9: N. gonorrhoeae: Percent Agreement by Panel Member for Lot, Site/Instrument, and Day - PCR Media/Swab

{19}------------------------------------------------

PanelMemberCtSDCtCV%LotSite/InstrumentDay
1x LOD CT,2.5x LOD NG0.491.42100.0 108/1081100.0 72/721100.0 72/72
3100.0 108/1082100.0 72/722100.0 72/72
3100.0 72/723100.0 72/72
2.5x LOD CT,1x LOD NG0.711.92100.0 108/1081100.0 72/721100.0 72/72
3100.0 108/1082100.0 72/722100.0 72/72
3100.0 72/723100.0 72/72

1 For Negative samples, Percent Agreement = (number of negative results/total valid results); For Positive samples, Percent Agreement = (number of positive results/total valid results)

5.2. Clinical Specimen Study

5.2.1. Study Design

Performance characteristics of the cobas CT/NG Test were evaluated in a multi-center clinical study conducted in the United States. Specimens were collected at 12 geographically diverse sites including OB-GYN practices, public and private STD clinics, and family planning centers. A total of 2.851 evaluable male and female, symptomatic and asymptomatic subjects were enrolled. Subjects were classified as symptomatic if the subject reported symptoms indicative of CT or NG infection. Specimens collected from each female subject included urine, a clinician- or self-collected vaginal swab, endocervical swabs, and a cervical sample in PreservCyt" Solution (Hologic Corporation, Bedford MA) obtained with a spatula/cytobrush or a broom. Male subjects provided urethral swabs and urine specimens. Specimens were tested using the cobas CT/NG Test and two commercially available nucleic acid amplification tests (NAAT) for CT and NG. The NAATs were used as reference assays in the clinical study.

For females, urine specimens were collected first. If a cervical cytology test was a scheduled part of the visit, that sample was taken next, followed by a vaginal swab and then endocervical swabs. If a cervical cytology test was not a scheduled part of the visit, a vaginal swab was taken next, followed by endocervical swabs and then the cervical specimen. The order of endocervical swab collection as well as clinician- or self-collection of the vaginal swab was alternated to minimize collection bias. For males, the urethral swabs were collected first in alternating order

{20}------------------------------------------------

and then the urine specimen was collected. All specimens were taken and stored according to the manufacturer's instructions.

Determination of Patient Infected Status 5.2.2.

For each subject, a patient infected status (PIS) was determined based on the combined results from the reference assays. A subject was categorized as infected for CT or NG if a minimum of two positive results (at least one from each reference NAAT) was reported, as described in Table 14. Female subjects with positive results on both reference urine specimens and negative results on both reference endocervical swab specimens and the reference cervical sample were categorized as infected for urine and not infected for swab specimens. A subject was classified as non-infected if at least one of the reference NAATs reported negative results for all sample types.

NAAT1Urine/Endocervical orUrethral SwabNAAT2Urine/Endocervical orUrethral SwabNAAT2Cervical Specimen inPreservCytPatient Infected Status
+/++/++ or -Infected
+/++/- or -/++ or -Infected
+/- or -/++/++ or -Infected
+/--/++ or -Infected
-/++/-+ or -Infected
-/+-/++ or -Infected
+/-+/-+Infected
+/-+/--Infected (Urine)Non-Infected (Swabs)
+/- or -/+-/-+ or -Non-Infected
+/+-/-+ or -Non-Infected
-/-+/++ or -Non-Infected
-/-+/- or -/++ or -Non-Infected
-/--/-+ or -Non-Infected

Table 10: Determination of Patient Infected Status

{21}------------------------------------------------

If patient infected status could not be determined due to missing and/or indeterminate results from the reference tests, the subject was excluded from the analysis. PIS could not be determined for one subject.

5.2.3. Study Results

Table 15 through Table 22 summarize the data from the clinical specimen study.

Results from the cobas CT/NG Test were compared with the PIS for calculation of test sensitivity and specificity. A total of 21,988 CT and 21,987 NG results were analyzed by sex. sample type, and the presence of symptoms. The overall sensitivity and specificity for CT was 95.7% and 99.7%. The overall prevalence was 9.0% for CT (6.3% in women and 16.4% in men). The overall sensitivity and specificity for NG was 99.0% and 99.9%. The overall prevalence was 3.6% for NG (1.6% in women. 9.2% in men). Sensitivity, specificity, and prevalence for CT are presented in Table 15. Sensitivity, specificity, and prevalence for NG are presented in Table 16.

A comparison of patient infected status, test results from the reference tests and test results from the cobas CT/NG Test was performed. CT results for female subjects are presented in Table 17, and for male subjects in Table 18. NG results for female subjects are presented in Table 19. and for male subjects in Table 20.

The positive and negative predictive values (PPV and NPV) were calculated using hypothetical prevalence rates and the cobas® CT/NG Test sensitivity and specificity determined in the study. The positive and negative predictive value estimates for CT are presented in Table 21, and for NG in Table 22.

Table 11: CT Clinical Performance Compared with Patient Infected Status by Gender, Sample Type, and Symptom Status

SampleTypeaSymptomStatusbTotal(n)SENS95% CISPEC95% CIPREV(%)PPV(%)NPV(%)
Female
VGSymp107393.8%(75/80)(86.2%,97.3%)99.7%(990/993)(99.1%,99.9%)7.596.299.5
Asymp101094.1%(48/51)(84.1%,98.0%)99.7%(956/959)(99.1%,99.9%)5.094.199.7

{22}------------------------------------------------

Overall208393.9%(123/131)(88.4%,96.9%)99.7%(1946/1952)(99.3%,99.9%)6.395.399.6
Male
Symp29697.3%(72/74)(90.7%,99.3%)99.5%(221/222)(97.5%,99.9%)25.098.699.1
URAsymp47298.1%(51/52)(89.9%,99.7%)99.5%(418/420)(98.3%,99.9%)11.096.299.8
Overall76897.6%(123/126)(93.2%,99.2%)99.5%(639/642)(98.6%,99.8%)16.497.699.5
All Combined285195.7%(246/257)(92.5%,97.6%)99.7%(2585/2594)(99.3%,99.8%)9.096.599.6

³ VG-S = self-collected vaginal swab; UR = urine.

b Symp = symptomatic; Asymp = asymptomatic.

Note: Subjects were designated as being infected with CT if at least 2 NAATs with different target regions gave positive results for the endocervical swab (urethral swab for males) and or the urine specimen. However, females were categorized as non-infected for any swab specimen if the swab specimens and the PreserVCyt specimen (NAAT2) were negative and the urine specimens were positive. Note: Subjects with designated infection status and valid cobas CTING Test results were considered evaluable and included in this summary table.

Note: CI = (score) confidence interval; PREV = prevalence; SENS = sensitivity; SPEC = specificity;

PPV = positive predictive value; NPV = negative predictive value.

Table 12: NG Clinical Performance Compared With Patient Infected Status by
Gender, Sample Type, and Symptom Status
SampleTypeaSymptomStatusbTotal(n)SENS95% CISPEC95% CIPREV(%)PPV(%)NPV(%)
Female
VGSymp107395.7%(22/23)(79.0%,99.2%)99.9%(1049/1050)(99.5%,100.0%)2.195.799.9
VGAsymp1010100.0%(10/10)(72.2%,100.0%)100.0%(1000/1000)(99.6%,100.0%)1.0100.0100.0
Overall208397.0%(32/33)(84.7%,99.5%)100.0%(2049/2050)(99.7%,100.0%)1.697.0100.0
Male
URSymp296100.0%(64/64)(94.3%,100.0%)99.1%(230/232)(96.9%,99.8%)21.697.0100.0
URAsymp472100.0%(7/7)(64.6%,100.0%)100.0%(465/465)(99.2%,100.0%)1.5100.0100.0
Overall768100.0%(71/71)(94.9%,100.0%)99.7%(695/697)(99.0%,99.9%)9.297.3100.0
All Combined285199.0%(103/104)(94.8%,99.8%)99.9%(2744/2747)(99.7%,100.0%)3.697.2100.0

{23}------------------------------------------------

1 .


A L MOND SHIPE & A------------------A & A MINNERIAL AND II the promote and aSamplevoevmotomStatusAn and profitsA . Commend Collection Collection Collection Collection Collection Collection Concession Compressional Concession Compressional Concession Comercial Concession CompressionalSENSA B A - A -95%(------------------------------------------------------------------------------------------------------------------------------------------------------------------------------SPECSರಿಕೆಳE0/010/0NPV10/0
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

ªVG-S = self-collected vaginal swab; UR = urine.

b Symp = symptomatic; Asymp = asymptomatic.

Note: Subjects were designated as being intelled with NG if at least 2 NAATs with different target regions gave positive results for the endocervical swab (urethral swab for males) and/or the urine specimen.

Note: Subjects with designated infection status and valid cobas CT/NG Test results were considered in this summary table.

Note: CI = (score) confidence interval; PREV = prevalence; SENS = sensitivity; SPEC = specificity;

PPV = positive predictive value; NPV = negative predictive value.

{24}------------------------------------------------

NAAT1NAAT2cobas CT/NG TestSymptom Status
Patient InfectedStatusSWURSWURPC PreVGSympAsympTotal
Infected++++++6139100
Infected+-+-++246
Infected+-++++505
Infected++++NA+213
Infected++++++213
Infected+++-++112
Infected++-+++112
Infected+++++-101
Infected++++-+011
Infected+++---011
Infected++-+-+101
Infected+-+++-101
Infected+-+-+-101
Infected+-+---011
Infected-++++-101
Infected-+++--101
Infected-+-+NA-011
Total Infected8051131
Non-Infected------9579111868
Non-Infected----NA-162440
Non-Infected--+---8311
Non-InfectedNANA----099
Non-Infected--++--303
Non-Infected----+-033
Non-Infected-----+213
Non-Infected-+-+--112
Non-Infected-+----022
Non-Infected--NA---022
Non-Infected-NA----202
Non-InfectedNA-----202
Non-Infected+-----011
Non-Infected--++++011
Non-Infected--+-++011
Non-Infected----++101
Non-Infected---NA--101
Total Non-Infected9229591881

Table 13: CT Positive/Negative Analysis for Female Patient Infected Status

  • Symp = symptomatic; Asymp = asymptomatic.

Note: Subjects were designated as being infected with CT if at least 2 NAATs with different target regions gave positive results for the endocervical swab and/or urine specimen. However, females were categorized as non-infected for any swab specimen if the swab specimens and the PreservCyt specimen (NAAT2) were negative and the unine specimens were positive.

Note: Subjects with designated infection status and valid cobas CT/NG Test results were considered evaluable and are included in this summary table.

Note: + denotes Positive; - denotes Negative; NA indicates specimen was not obtained for test result was missing/invalid. Note: SW = endocervical swab; UR = urine; VG = vaginal swab; PC Pre = PreservCyt (pre-aliquot).

{25}------------------------------------------------

PatientInfectedStatusNAAT1NAAT2cobasCT/NG TestSymptom Statusa
SWURSWURURSympAsympTotal
Infected+++++6743110
Infected-+-++336
Infected-++++033
Infected+++-+112
Infected+-+--011
Infected+-+++011
Infected-++-+101
Infected-+-+-101
Infected++++-101
Total Infected7452126
Non-Infected-----218412630
Non-Infected--+--112
Non-Infected---+-112
Non-Infected--++-022
Non-Infected-+---022
Non-Infected----+011
Non-Infected--+++011
Non-Infected+----101
Non-Infected++--+101
Total Non-Infected222420642

Table 14: CT Positive/Negative Analysis for Male Patient Infected Status

4 Symp = symptomatic; Asymp = asymptomatic.

Note: Subjects were designated as being infected with CT if at least 2 NAATs with different target regions gave positive results for the urether swab and/or the urine specimen.

Note: Subjects with designated infection status and valid cobas " CT/NG Test results were considered in this summary table.

Note: + denotes Positive; - denotes Negative.

Note: SW = urethral swab; UR= urine.

Table 15: NG Positive/Negative Analysis for Female Patient Infected Status

NAAT1NAAT2cobas CT/NG TestSymptom Status
Patient InfectedStatusSWURSWURPC PreVGSympAsympTotal
Infected++++++16925
Infected+-+-++303
Infected-+++++112
Infected+++++-101
Infected++++-+101
Infected+++-++101
Total Infected231033
Non-Infected------10179581975
Non-Infected----NA-182644
Non-InfectedNANA----099
Non-Infected+-----415
Non-Infected---+--224
Non-Infected--NA---224
Non-Infected---+--112
Non-Infected-NA----202
Non-InfectedNA-----202

{26}------------------------------------------------

Patient InfectedStatusNAAT1NAAT2cobas CT/NGTestSymptom StatusaTotal
SWURSWURPC PreSympAsymp
Non-Infected-+----011
Non-Infected-----+101
Non-Infected---NA--101
Total Non-Infected105010002050

ª Symp = symptomatic; Asymp = asymptomatic.

Note: Subjects were designated as being infected with NG if at least 2 NAATs with different target regions gave positive results for the endocervical swab and/or the urine specimen.

Note: Subjects with designated infection status and valid cobas CT/NG Test results were considered evaluable and are included in this summary table.

Note: + denotes Positive; - denotes Negative; NA indicates specimen was not obtained for test result was missing/invalid. Note: SW = endocervical swab; UR = urine; VG = vaginal swab; PC Pre = PreservCyt (pre-aliquot).

Table 16: NG Positive/Negative Analysis for Male Patient Infected Status

PatientInfectedStatusNAAT1NAAT2cobasCT/NG TestSymptom StatusaTotal
SWURSWURURSympAsymp
Infected+++++63770
Infected++-++101
Total Infected64771
Non-Infected-----227464691
Non-Infected----+202
Non-Infected-+-+-112
Non-Infected--+--101
Non-Infected+----101
Total Non-Infected232465697

4 Symp = symptomatic; Asymp = asymptomatic.

Note: Subjects were designated as bein NG if at least 2 NAATs with different target regions gave positive results for the urethral swab and/or the urine specimen.

Note: Subjects with designated infection status and valid cobas CT/NG Test results and are included in this summary table.

Note: + denotes Positive; - denotes Negative.

Note: SW = urethral swab; UR= urine.

Table 17: Positive Predictive Value and Negative Predictive Value for Hypothetical CT Prevalence

Prevalence (%)Sensitivity (%)Specificity (%)PPV (%)NPV (%)
195.799.773.6100.0
395.799.789.599.9
595.799.793.699.8
1095.799.796.899.5
1595.799.798.099.2
2095.799.798.698.9
3095.799.799.298.2
5095.799.799.695.9
  • Cverall sensitivity and specificity were estimated by compains the cobas to patient infected status in both female and male subjects.

Note: PPV = positive predictive value; NPV = negative predictive value.

{27}------------------------------------------------

Prevalence (%)Sensitivity (%)Specificity (%)PPV (%)NPV (%)
89.0ශිරි විශ්‍රී ලිපි90.2100.0
3aalinਰੰਤੇ ਰੋ96.6100.0
599.0ರಿದಿ. ರಿ87.9વેત્ત્વે છે. છે. છે. છે છે છે. છ
1099.0વેવી જેવી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામમાં પ્રાથમિક શાળા, પંચાયતઘર, આંಕ್ಕಿ.ರવેલું છે. કે
1599.089.9ਰੇਰੇ 499.8
2099.0છે. વિવેત જેવી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામમાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમ99.699.8
3099.0ਰੇਰੇ ਰੋ99.799.6
പ്പn boag aag a00 0

Table 18: Positive Predictive Value and Negative Predictive Value for 1 More +1. ~+1.

Overall sensitivity and specificity were estimated by comparing the cobas to patient infected status in both female and male subjects.

Note: PPV = positive predictive value; NPV = negative predictive value.

5.2.4. Conclusion Drawn from Clinical Specimen Study

In both symptomatic and asymptomatic patients, the cobas® CT/NG Test offers high sensitivity, specificity, and positive and negative predictive values for the direct, qualitative detection of CT and NG in male urine specimens and self-collected vaginal swabs. These results are of key importance from both diagnostic and public health perspectives.

CONCLUSION 6.

A comparison of the intended use, technological characteristics, and the results of non-clinical analytical and clinical performance studies demonstrate that the cobas® CT/NG Test is safe and effective when used in accordance with the product labeling.

{28}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/28/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol on the right side. To the left of the bird symbol is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion.

10903 New Hampshire Avenue Silver Spring, MD 20993

Mr. James R. Bonds Director, Regulatory Affairs Roche Molecular Systems, Inc. 4300 Hacienda Drive Pleasanton, CA 94588-0900

JAN 2 4 2012

Re: K110923 Trade Name: cobas® CT/NG Test Regulation Number: 21 CFR §866.3120 Chlamydia serological reagents Regulation Name: Regulatory Class: Class I, II Product Code: MKZ, LSL, OOI Dated: January 20, 2012 Received: January 23, 2012

Dear Mr. Bonds:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a

{29}------------------------------------------------

legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.goy/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Vag a
Sallie A. Kintner, M.S., R.D.

A. Hoivat, M Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{30}------------------------------------------------

cobas® CT/NG Test Indications for Use Statement

510(k) Number (if known): Device Name: Indications for Use:

K110923 Roche cobas® CT/NG Test Assav

The cobas® CT/NG Test is an in vitro nucleic acid amplification test that utilizes Polymerase Chain Reaction (PCR) and nucleic acid hybridization for the qualitative detection of Chlamydia trachomatis (CT) and/or Neisseria gonormoeae (NG) DNA to aid in the diagnosis of chlamydial and gonococcal disease. The test may be used with vaginal swab specimens self-collected in a clinical setting and male urine from both symptomatic and asymptomatic individuals. Specimens to be tested should be collected in cobas® PCR Media.

Ancillary Collection Kits

The cobas® PCR Female Swab Sample Kit is used to collect and transport self-collected vaginal swab specimens in a clinical setting. The cobas® PCR Media serves as a nucleic acid stabilizing transport and storage medium for gynecological specimens. Use this collection kit only with the cobas CT/NG Test. NOTE: This collection kit should not be used for collection of alternative gynecological specimens.

The cobas® PCR Urine Sample Kit is used to collect and transport male urine specimens.

The cobas® PCR Media serves as a nucleic acid stabilizing transport and storage medium for urine specimens. Use this collection kit only with the cobas® CT/NG Test. NOTE: This collection kit should not be used for collection of female urine specimens.

Prescription Use X (Per 21 CFR Subpart D) OR

Over-The-Counter Use (Per 21 CFR Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE — CONTINUE ON ANOTHER PAGE IF NEEDED

Concu
Jayass

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off

Office of In Vitro Diagnestie Device Exaluation and Safety ·

Indications for Use Statement Page 1

510(k) K110923

§ 866.3120 Chlamydia serological reagents.

(a)
Identification. Chlamydia serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to chlamydia in serum. Additionally, some of these reagents consist of chlamydia antisera conjugated with a fluorescent dye used to identify chlamydia directly from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of disease caused by bacteria belonging to the genusChlamydia and provides epidemiological information on these diseases. Chlamydia are the causative agents of psittacosis (a form of pneumonia), lymphogranuloma venereum (a venereal disease), and trachoma (a chronic disease of the eye and eyelid).(b)
Classification. Class I (general controls).