K Number
K110899

Validate with FDA (Live)

Date Cleared
2012-02-01

(308 days)

Product Code
Regulation Number
866.3110
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Helicobacter pylori (H. pylori) ELISA IgA test kit is intended for the qualitative detection of IgA antibodies to H. pylori in human serum in the adult population. This test is intended as a second test to aid in the diagnosis of H. pylori in patients with gastrointestinal symptoms, in conjunction with clinical findings. It should be performed and interpreted with another assay for detection of IgG antibodies to H. pylori.

Device Description

The assay requires a total of 90 minutes incubation time. The test uses purified antigen coated on microtiter wells. Serum is added to each well and incubated for 30 minutes at 37°C. If H. pylori IgA antibodies are present they will bind to the antigen in the well. Unbound serum is removed by washing the wells three times. An HRP-conjugated anti-human IgA is then added to each well and incubated for 30 minutes at 37°C. If H. pylori antibody is present, it will bind to the antibody attached to the antigen on the wells are again washed three times to remove any unbound conjugate. A TMB substrate is added to each well and incubated for 30 minutes at 37°C. If enzyme is present, it will react with the substrate to generate a colored product. After the incubation period, Stop Solution and the color intensity is measured the reaction is stopped with a spectrophotometrically.

AI/ML Overview

The document describes the Gold Standard Diagnostics Helicobacter pylori IgA ELISA Test Kit and its performance characteristics.

1. Table of Acceptance Criteria and Reported Device Performance

The provided document doesn't explicitly state "acceptance criteria" through a formal table. However, the study focuses on demonstrating substantial equivalence to a predicate device, and the key performance metrics reported are Percent Positive Agreement, Percent Negative Agreement, and Overall Agreement with the predicate device (Micro Detect Inc. Pylori Detect IgA).

Performance MetricAcceptance Criteria (Implied)Reported Device Performance (Gold Standard Diagnostics H. pylori ELISA IgA)
Percent Positive AgreementDemonstrate substantial agreement with the predicate device. Specific threshold not explicitly stated but generally expected to be high for equivalence.94.5% (C.I. 76.0% - 100%)
Percent Negative AgreementDemonstrate substantial agreement with the predicate device. Specific threshold not explicitly stated.93.8% (C.I. 88.8% - 99.5%)
Overall AgreementDemonstrate substantial agreement with the predicate device. Specific threshold not explicitly stated.94.0% (C.I. 85.9% - 100%)
Intra-Assay CVAcceptable precision for diagnostic assays (e.g., typically <10-15%). Ranges observed: 2.5% - 12.2%.
Inter-Assay CVAcceptable precision for diagnostic assays (e.g., typically <10-15%). Ranges observed: 6.4% - 13.7%.
Cross-ReactivityMinimal to no cross-reactivity with common interfering organisms (e.g., <5-10% inhibition).Mean percent inhibition 0.8% to 10.9% for non-H. pylori organisms. No effects on analytical specificity reported.
Interfering SubstancesNo significant interference at specified concentrations of common interfering substances.No interference noted with hemoglobin, bilirubin, cholesterol, and triglycerides.
Limit of Detection (LoD)Detection of positive results 95% of the time, near cutoff with acceptable precision and accuracy.LoD determined at OD value of 0.558 and unit value of 11.3 (positive 95.8% of the time).

Note: The acceptance criteria for agreement with the predicate are implied by the demonstration of substantial equivalence, rather than explicitly stated numerical targets.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 625 samples.
  • Data Provenance: The samples were "routinely tested for H. pylori," suggesting they were clinical samples. The document does not specify the country of origin, nor explicitly state if they were retrospective or prospective, although "routinely tested" implies existing samples (retrospective).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This type of information is not applicable to this study. The "ground truth" for the comparative effectiveness study was established by comparing the performance of the new device to a legally marketed predicate device (Micro Detect Inc. Pylori Detect IgA). For discrepant samples, a third commercially available assay (Inova QUANTA Lite™ H. pylori IgA ELISA) was used for adjudication. No human expert interpretation of raw data (e.g., radiologists reading images) was involved in establishing the ground truth for the device's diagnostic performance in the primary comparison.

4. Adjudication Method for the Test Set

For discrepant samples between the Gold Standard Diagnostics H. pylori ELISA IgA assay and the Micro Detect Inc. Pylori Detect IgA assay, a third, commercially available ELISA assay (Inova QUANTA Lite™ H. pylori IgA ELISA) was used. The document does not specify a numerical adjudication rule (e.g., 2+1, 3+1), but rather that the "third assay called" the outcome for these discrepant samples.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This study evaluates an automated ELISA test kit, not a device requiring human interpretation of results, and therefore, the concept of "human readers" or "AI assistance" in the context of improving human performance is not applicable.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

Yes, a standalone performance efficacy study was done. The device itself is an automated ELISA test kit, which performs its function without continuous human intervention during the assay process to determine a positive or negative result. The clinical comparison and all non-clinical tests (precision, reproducibility, cross-reactivity, interfering substances, limit of detection) evaluate the device's standalone performance.

7. The Type of Ground Truth Used

The primary "ground truth" for the comparative effectiveness study was the result from a legally marketed predicate device (Micro Detect Inc. Pylori Detect IgA). For discrepant results, a third commercially available ELISA assay (Inova QUANTA Lite™ H. pylori IgA ELISA) was used as an adjudicator. This is a form of reference standard test agreement, rather than expert consensus, pathology, or direct outcomes data (like biopsy results).

8. The Sample Size for the Training Set

The document does not specify a sample size for a "training set." This type of information is typically relevant for machine learning-based devices. For an ELISA kit, development and optimization would involve internal studies, but these are not usually referred to as a "training set" in the same way.

9. How the Ground Truth for the Training Set Was Established

As no "training set" is explicitly mentioned for a machine learning context, the method for establishing its ground truth is not applicable/not provided. The development of an ELISA kit relies on established biochemical principles and extensive internal validation and optimization against known positive and negative samples, calibration standards, and various interfering substances to define its operational parameters and cutoff values.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Gold Standard Diagnostics. The logo is in black and white and features the company name in a bold, sans-serif font. To the right of the logo is the number K110899. Below the logo is the date FEB - 1 2012.

510(k) Summary of Safety and Effectiveness

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirement of SMDA 1990 and 21 CFR 807.92.

    1. Submitter's Name: Gold Standard Diagnostics 2851 Spafford St. Davis, CA. 95618 Address: Phone Number: 530-759-8000 Contact Person: Napoleon Monce September 20, 2011 Date:
    1. Product and Trade Name: Helicobacter pylori IgA ELISA

Common Name or Classification Name: Campylobacter pylori

Product Code: LYR

3) Legally marketed device to which the submitter claims equivalence:

Micro Detect, Inc. Pylori Detect IgA ELISA for the qualitative detection of IgA antibodies against H. pylori in human serum. The test is intended as an aid in the diagnosis of infection by H. pylori in patients with gastrointestinal symptoms. K003794.

4) Description of the device:

The assay requires a total of 90 minutes incubation time. The test uses purified antigen coated on microtiter wells. Serum is added to each well and incubated for 30 minutes at 37°C. If H. pylori IgA antibodies are present they will bind to the antigen in the well. Unbound serum is removed by washing the wells three times. An HRP-conjugated anti-human IgA is then added to each well and incubated for 30 minutes at 37°C. If H. pylori antibody is present, it will bind to the antibody attached to the antigen on the wells are again washed three times to remove any unbound conjugate. A TMB substrate is added to each well and incubated for 30 minutes at 37°C. If enzyme is present, it will react with the substrate to generate a colored product. After the incubation period, Stop Solution and the color intensity is measured the reaction is stopped with a spectrophotometrically.

5) Intended use of the device:

The Helicobacter pylori (H. pylori) ELISA IgA test kit is intended for the qualitative detection of IgA antibodies to H, pylori in human serum in the adult population. This test is intended as a second test to aid in the diagnosis of H. pylori in patients with gastrointestinal symptoms, in conjunction with

{1}------------------------------------------------

clinical findings. It should be performed and interpreted with another assay for detection of IgG antibodies to H. pylori.

6) Comparison with the predicate device:

The Gold Standard Diagnostics H. pylori ELISA IgA Test Kit was compared to a commercially marketed kit by Micro Detect, Inc. the Pylori Detect IgA (K003794) catalog number HpKi-A. Both kits have the same intended use and use the same methodology. Below are tables comparing the reagents provided, the procedures, and their performances.

Table 1: Reagent Comparison

Gold Standard Diagnostics H. pyloriELISA IgA Test KitMicro Detect Inc. Pylori Detect IgA
Antigen coated Microtiter Plate – 96 wellsAntigen coated Microtiter Plate – 96 wells
Wash Solution - 20xDiluent/Wash Concentrate -- 25x
Diluent - Ready to UseDiluent/Wash Concentrate - 25x
IgA Conjugate – Anti Human HRPIgA Conjugate - Anti Human Peroxidase
Substrate - Tetramethylbenzidine (TMB)Substrate - Tetramethylbenzidine (TMB)
Stop Solution - Acid mixtureStop Solution - Sulfuric Acid
H. pylori IgA Positive ControlH. pylori IgA Positive Control
H. pylori IgA Cutoff ControlH. pylori IgA Calibrator
H. pylori IgA Negative ControlH. pylori IgA Negative Control

Table 2: Procedure Comparison

Gold Standard Diagnostics H. pyloriELISA IgA Test KitMicro Detect Inc. Pylori Detect IgA
Dilute Samples 1:101 in DiluentDilute Samples 1:101 in reconstitutedDiluent/Wash
Add 100ul of Samples and ControlsAdd 100ul of Samples and Controls
Incubate for 30 minutes at 37°CIncubate for 20 minutes at RoomTemperature

{2}------------------------------------------------

Wash four times with reconstituted WashSolutionWash three times with reconstituted WashSolution
Add 100ul of ConjugateAdd 100ul of Conjugate
Incubate for 30 minutes at 37°CIncubate for 20 minutes at RoomTemperature
Wash four times with reconstituted WashSolutionWash three times with reconstituted WashSolution
Add 100ul of SubstrateAdd 100ul of Substrate
Incubate for 30 minutes at 37°CIncubate for 15 minutes at RoomTemperature
Add 50ul of Stop SolutionAdd 100ul of Stop Solution
Read with Spectrophotometer at 450nmRead with Spectrophotometer at 450nm

6(b1) Nonclinical tests:

The intra and inter assay precision was calculated by running six patient sera (four positives and two negatives) at three different sites. The results are summarized in the table below:

Sample 1Sample 2Sample 3Sample 4Sample 5Sample 6
Site 1Ave:0.9450.8030.8600.4980.0910.115
SD:0.0240.0340.0390.0250.0060.008
Intra-AssayCV:2.5%4.2%4.5%5.0%6.3%7.1%
Site 2Ave:1.0300.7370.7160.5790.0910.137
SD:0.0600.0570.0410.0460.0040.009
Intra-AssayCV:5.9%7.7%5.7%7.9%4.8%6.9%
Site 3Ave:0.9960.6670.7680.5790.0910.151
SD:0.1080.0350.0570.0420.0100.009
Intra-AssayCV:10.8%5.3%7.4%7.3%12.2%6.1%
Ave:0.9640.7700.8120.5270.0900.126
Inter-AssaySD:0.0620.0630.0740.0490.0070.017
CV:6.4%8.1%9.1%9.4%7.4%13.7%

{3}------------------------------------------------

Reproducibility:

The reproducibility of the assay was done by testing three samples in triplicate (a high negative, low positive and a moderate positive) for five days, twice a day, at three sites with two technicians per site. The results are summarized in the table below:

5 Day Average:Sample 1Sample 2Sample 3
Site 1Tech 1OD:0.6140.7241.545
SD:0.0450.0660.091
CV:7.3%9.1%5.9%
Site 1Tech 2OD:0.5880.7171.494
SD:0.0590.0730.091
CV:10.1%10.2%9.8%
Site 2Tech 1OD:0.5940.7041.546
SD:0.0340.0450.082
CV:5.7%6.4%5.3%
Site 2Tech 2OD:0.6180.8451.741
SD:0.0360.0440.071
CV:5.8%5.2%4.1%
Site 3Tech 1OD:0.3560.4801.046
SD:0.0480.0870.128
CV:13.4%18.0%12.2%
Site 3Tech 2OD:0.4780.6361.323
SD:0.0860.1100.148
CV:18.0%17.3%11.2%

Cross Reactivity:

An adsorption study was performed to evaluate any cross reactivity. Briefly, sera with different levels of antibodies to H. pylori were adsorbed with either H. pylori, Candida albicans, E. coli, Borrelia burgdorferi, Clostridium spp., Campylobacter, Bacillus, Enterobacter, Pseudomonas, Haemophilus Influenza, and Proteus. The identity of the bacteria used were identified by the ATCC and confirmed with mass spectrometry. The bacteria were evaluated at a concentration of 10' cfulml or higher.

Sera with different levels of antibodies to H. pylori, were adsorbed with the recommended organisms. The adsorbed samples were compared to the untreated samples and the mean percent inhibition was calculated. The results are summarized in the following table:

OrganismConcentration(cfu/ml)Mean percentinhibitionCrossReactivity
Helicobacter pylori.96%
Candida albicans2.40x1070.8%None

{4}------------------------------------------------

Escherichia coli6.90x1072.4%None
Borrelia burgdorferi1.00x1084.3%None
Clostridium spp.1.20x1071.7%None
Campylobacter1.50x1093.3%None
Bacillus4.40x10710.9%None
Enterobacter1.80x1084.1%None
Pseudomonas1.45x1083.8%None
HaemophilusInfluenza7.90x1075.8%None
Proteus1.40x1083.1%None

The mean percent inhibition for H. pylori was 96%, and 0.8% to 10.9% with the other organisms. Overall no effects on the analytical specificity were seen on the Gold Standard Diagnostics H. pylori ELISA IgA assay.

Interfering Substance

The effect of potential interfering substances on samples using the Gold Standard Diagnostics H. pylori ELISA IgA assay was evaluated. High levels of hemoglobin, bilirubin, cholesterol and triglycerides in serum samples were tested at the assay cutoff (9-11 units) in triplicate. The recommended concentrations from the guideline "Interference Testing In Clinical Chemistry" from the Clinical and Laboratory Standards Institute were used (see table below). No interference was noted with any of substances tested and the substances tested did not affect the performance of the Gold Standard Diagnostics H. pylori ELISA IgA assay.

SubstanceConcentrationH. pyloriconcentrationMean PercentInhibition
Hemoglobin2 g/L9-11 units9%
Bilirubin342 [Mu]mol/L9-11 units7%
Cholesterol13 mmol/L9-11 units0%
Triglyceride37 mmol/L9-11 units-32%

{5}------------------------------------------------

Leukocytes, intestinal secretions or mucus, fat, and medications used to relieve diarrhea or other gastric symptoms were not tested, therefore, it is not known if these substances will interfere with the assay as they were not evaluated.

Limit of Detection

To determine the Limit of Detection (LoD) for the assay, a pooled positive and a pooled negative sample were tested along with the negative, cutoff and positive controls provided in the kit. Further, the pooled positive sample was diluted up to 4.3 fold. The samples, and dilutions, were then tested 24 times (24 replicates) over several days as described in the Clinical Laboratory Standards Institute CLSI document EP-17. The LoD chosen was one that gave a positive result 95% of the time, was near the cutoff, and gave acceptable precision and accuracy results.

In Summary, the controls supplied in the kit were within their acceptable ranges. The pooled negative sample gave negative results on all 24 replicates giving an average OD value of 0.050 and a corresponding Unit value of 1.1. The pooled positive sample gave positive results on all 24 replicates giving an average OD value of 1.259 and a corresponding Unit value 25.3.

The pooled positive sample was then diluted up to 4.3 fold and each dilution was tested with the assay. The 3.3 fold dilution was positive 95.8% of the time (after 24 replicates), gave acceptable precision and accuracy results, gave an average OD value of 0.558 was close to the average cutoff OD value (0.497) and gave a corresponding unit value of 11.3. Therefore, we can conclude that the LoD of the Gold Standard Diagnostics H. pylori IgA Test can be determined at an OD value of 0.558 and unit value of 11.3.

6(b2): Clinical Comparison:

The performance of the Gold Standard Diagnostics H. pylori ELISA IgA assay was determined by conducting a correlation study using 625 samples being routinely tested for H. pylori. The samples were tested on both the Gold Standard Diagnostics H. pylori ELISA IgA assay and a commercially available ELISA assay (Micro Detect Inc. K003794) as the predicate device. The results are summarized in the following table:

Micro Detect Inc.

PositiveEquivocalNegative
Positive1213126
GSDEquivocal13823
Negative75394

The discrepant samples were further tested on a third assay, the Inova QUANTA Lite™ H. pylori IgA ELISA (which is also commercially available). Of the seven Gold Standard Diagnostics negative samples, Micro Detect Inc. positive samples, the third assay called five samples positive. Of the 26 Gold Standard Diagnostics positive samples, Micro Detect Inc. negative samples, the third assay called two samples borderline, one sample negative and 23 samples positive.

{6}------------------------------------------------

The comparison data produced a percent positive agreement of 94.5% (C.1. 76.0% - 100%), a percent negative agreement of 93.8% (C.I. 88.8% - 99.5%), and an overall agreement of 94.0% (C.I. 85.9% -100%).

6(b3) Conclusion:

From the data and comparison above, it is our contention that the Gold Standard Diagnostic H. pylori IgA ELISA test is substantially equivalent to the commercially marketed Micro Detect, Inc. Pylori Detect IgA kit.

{7}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is a stylized symbol that resembles a bird or a human figure in motion, composed of flowing lines. The symbol is black, and the text is also in a dark color, set against a white background.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

FEB - 1 2012

Gold Standard Diagnostics C/o Napoleon Monce Director, Product Development 2851 Spafford Street Davis, CA 95618

Re: K110899

Trade/Device Name: Gold Standard Diagnostics Helicobacter pylori IgA Test Kit Regulation Number: 21 CFR 866.3110 Regulation Name: Campylobacter fetus serological reagents Regulatory Class: Class I Product Code: LYR Dated: January 18, 2012 Received: January 19, 2012

Dear Mr. Monce:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice

{8}------------------------------------------------

Page 2 – Mr. Monce

requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Sally A. Horvat, M.Sc., Ph.D.

Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Indications for Use

510(k) Number (if known): K110899

Device Name: Gold Standard Diagnostics Helicobacter pylori ELISA IgA Test Kit

Indications For Use:

The Helicobacter pylori (H. pylori) ELISA IgA test kit is intended for the qualitative detection of IgA antibodies to H. pylori in human serum in the adult population. This test is intended as a second test to aid in the diagnosis of H. pylori in patients with gastrointestinal symptoms, in conjunction with clinical findings. It should be performed and interpreted with another assay for detection of IgG antibodies to H. pylori.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Freddie K. Poole

sion Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K110899

Page 1 of

l . I

§ 866.3110

Campylobacter fetus serological reagents.(a)
Identification. Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identifyCampylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases.Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.(b)
Classification. Class I (general controls).