K Number
K110860

Validate with FDA (Live)

Device Name
COMP A-I-R
Date Cleared
2011-09-02

(157 days)

Product Code
Regulation Number
868.5630
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NE-C801 Nebulizer Compressor System is intended to provide air to the pneumatic nebulizer in order to aerosolize medications for inhalation by the patient for respiratory disorders. The system is designed for use with pediatric (defined by the prescribed medication) and adult patients in the home, hospital, and sub-acute settings.

Device Description

The Omron NE-C801 nebulizer compressor system is a standard nebulizer compressor system with an integral compressor and handheld, pneumatic nebulizer intended for general purpose use. It is powered by standard AC. This is a modification of the NE-C28 cleared under K060811.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Omron NE-C801 Nebulizer Compressor system:

It's crucial to understand that this document describes a 510(k) Pre-Market Notification for a modified device, the NE-C801, which is being compared to a predicate device, the NE-C28 (cleared under K060811). The primary goal of a 510(k) is to demonstrate substantial equivalence to a predicate device, not necessarily to independently prove de novo safety and effectiveness through extensive clinical trials.

The provided text therefore focuses on demonstrating that the modifications (minor design changes to the nebulizer, and making the compressor smaller and lighter) do not alter the fundamental safety or effectiveness parameters compared to the predicate. No clinical investigation was deemed necessary.


1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" as a set of specific numerical thresholds the new device must meet to be considered successful. Instead, the performance testing aims to demonstrate equivalence to the predicate device within a 95% confidence interval. The "acceptance criteria" implicitly are that the performance of the NE-C801 is comparable to the NE-C28.

Performance Metric (Feature)Predicate (NE-C28) Performance (K060811)NE-C801 Reported Performance (Mean)Implicit Acceptance Criteria (Comparison to Predicate)
Materials ISO 10993-Identical to K060811Identical materials.
Total Delivered Dose (ug)Comparable to predicate within 95% CI.
Pulmicort401.6391.11(See Confidence Interval below)
Intal12233.312368.89(See Confidence Interval below)
Salbutamol7972.27883.33(See Confidence Interval below)
Total Delivered Dose fraction (%)Comparable to predicate within 95% CI.
Pulmicort80.3%78.2%(See Confidence Interval below)
Intal61.2%61.8%(See Confidence Interval below)
Salbutamol79.7%77.0%(See Confidence Interval below)
Particle size (MMAD) (Microns)Comparable to predicate within 95% CI.
Pulmicort3.733.88(See Confidence Interval below)
Intal2.712.91(See Confidence Interval below)
Salbutamol2.352.54(See Confidence Interval below)
Geometric Std. Dev. (GSD)Comparable to predicate within 95% CI.
Pulmicort2.051.93(See Confidence Interval below)
Intal2.292.27(See Confidence Interval below)
Salbutamol2.592.62(See Confidence Interval below)
Respirable Fraction (% Mass 0.5-5 microns)Comparable to predicate within 95% CI.
Pulmicort62.7%61.9%(See Confidence Interval below)
Intal72.8%70.5%(See Confidence Interval below)
Salbutamol69.2%66.4%(See Confidence Interval below)
Total Respirable Dose (ug 0.5 -5.0 microns)Comparable to predicate within 95% CI.
Pulmicort251.9242.17(See Confidence Interval below)
Intal8894.88729.49(See Confidence Interval below)
Salbutamol5514.45235.02(See Confidence Interval below)
Medication captured on USP Throat (ug)Comparable to predicate within 95% CI.
Pulmicort14.913.34(See Confidence Interval below)
Intal213.4253.66(See Confidence Interval below)
Salbutamol141.4144.94(See Confidence Interval below)
Medication captured on USP Throat Fraction (%)Comparable to predicate within 95% CI.
Pulmicort3.8%3.4%(See Confidence Interval below)
Intal1.7%2.0%(See Confidence Interval below)
Salbutamol1.8%1.8%(See Confidence Interval below)
Medication retained in Device (ug)Comparable to predicate within 95% CI.
Pulmicort104.496.11(See Confidence Interval below)
Intal7533.38000.00(See Confidence Interval below)
Salbutamol1977.82072.22(See Confidence Interval below)
Medication Retained in Device Fraction (%)Comparable to predicate within 95% CI.
Pulmicort20.9%19.2%(See Confidence Interval below)
Intal37.7%40.0%(See Confidence Interval below)
Salbutamol39.6%41.4%(See Confidence Interval below)
Coarse Particle (>4.7 um) Fraction (%)Comparable to predicate within 95% CI.
Pulmicort35.3%37.1%(See Confidence Interval below)
Intal25.0%27.7%(See Confidence Interval below)
Salbutamol24.6%26.6%(See Confidence Interval below)
Fine Particle (<4.7 um) Fraction (%)Comparable to predicate within 95% CI.
Pulmicort60.9%59.6%(See Confidence Interval below)
Intal73.2%70.2%(See Confidence Interval below)
Salbutamol73.7%71.6%(See Confidence Interval below)
Ultra Fine Particle (um) Fraction (%)Comparable to predicate within 95% CI.
Pulmicort5.4%3.5%(See Confidence Interval below)
Intal9.4%8.9%(See Confidence Interval below)
Salbutamol0.19%16.9%(See Confidence Interval below)
Nebulization rate (g/ml)Comparable to predicate.
Pulmicort0.440.37(See Confidence Interval below)
Intal0.350.30(See Confidence Interval below)
Salbutamol0.410.37(See Confidence Interval below)

Note on Acceptance Criteria: The stated "Confidence level of testing" is "The test and number of samples (3) with 3 runs provided a 95% confidence level." This implies that the acceptance criteria for substantial equivalence are met if the NE-C801's performance metrics, with their calculated 95% confidence intervals, are deemed comparable to the predicate (NE-C28) and fall within acceptable ranges for nebulizer performance. The document doesn't provide explicit numerical ranges for acceptance, but rather demonstrates the new device's performance alongside that of the predicate. Visually, the numbers for NE-C801 are generally similar to NE-C28, and the confidence intervals provided in "Table 2 - Particle Characterization results at 95% confidence" support the claim of equivalence.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Three (3) device samples were tested, with 3 runs each, for a total of 9 sample points per each drug type.
  • Data Provenance: The document does not specify the country of origin. The study is prospective in the sense that laboratory tests were conducted on newly manufactured devices for the purpose of this submission. It is not a retrospective analysis of prior performance data or clinical outcomes.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

N/A. This study involves physical performance testing (aerosol characterization) using an 8-stage cascade impactor and USP <601> induction port throat, not human expert assessment, to establish "ground truth." The ground truth here is the measured physical properties of the aerosol.

4. Adjudication Method for the Test Set

N/A. As the ground truth is based on physical scientific measurements, no human adjudication method was employed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. This is a physical performance comparability study of a medical device, not a study involving human readers or AI in a diagnostic context. Therefore, an MRMC study and effect size of human reader improvement with AI assistance are not applicable.

6. Standalone Performance Study (Algorithm Only)

Yes, in the context of device performance. The "performance testing" described in the document is a standalone evaluation of the NE-C801's ability to aerosolize medications and deliver particles of specific characteristics, without human intervention as part of the primary function being tested. It's an "algorithm only" equivalent in that it measures the device's intrinsic output.

7. Type of Ground Truth Used

The ground truth used is based on physical measurements and analytical chemistry/physics of the aerosolized medications. Specifically, it involves:

  • Particle Characterization per Cascade Impactor
  • Total Delivered Dose (ug)
  • Total Delivered Dose fraction (%)
  • Particle size (Mass Median Aerodynamic Diameter - MMAD) (Microns)
  • Geometric Standard Deviation (GSD)
  • Respirable Fraction (% Mass 0.5-5 microns)
  • Total Respirable Dose (ug 0.5 -5.0 microns)
  • Medication captured on USP Throat (ug and fraction %)
  • Medication Retained in Device (ug and fraction %)
  • Coarse Particle Fraction (%), Fine Particle Fraction (%), Ultra Fine Particle Fraction (%)
  • Nebulization rate (g/ml)

8. Sample Size for the Training Set

N/A. This is not an AI/Machine Learning study. There is no concept of a "training set" in this context. The testing involved direct physical measurement of device performance.

9. How the Ground Truth for the Training Set Was Established

N/A. As above, there is no training set for this type of device performance study.

{0}------------------------------------------------

SEP - 2 2011 510(k) Summary Page 4 of 6 31-Aug-11

KI10860

Omron Healthcare, Inc.
1200 Lakeside DriveTel - 847-247-5626
Bannockburn, IL 60015 USAFax - 847-247-5626
Official Contact:Mirna DiPano - Director, Quality & Regulatory
Proprietary or Trade Name:NE-C801 Nebulizer Compressor system
Common/Usual Name:Nebulizer (direct patient interface)
Classification Name/Code:CAFNebulizer (direct patient interface)CFR 868.5630
Device:NE-C801 nebulizer compressor system
Modified Device:Omron- NE-C28 - K060811

Device Description:

The Omron NE-C801 nebulizer compressor system is a standard nebulizer compressor system with an integral compressor and handheld, pneumatic nebulizer intended for general purpose use. It is powered by standard AC. This is a modification of the NE-C28 cleared under K060811.

The modifications to the device do not change the intended use of the predicate device. The modifications to the device do not alter the fundamental scientific technology. As there are no changes in hardware that will impact performance, there is no need to validate the changes through a clinical investigation.

Indications for Use:

The NE-C801 Nebulizer Compressor System is intended to provide air to the pneumatic nebulizer in order to aerosolize medications for inhalation by the patient for respiratory disorders. The system is designed for use with pediatric (defined by the prescribed medication) and adult patients in the home, hospital, and sub-acute settings.

Patient Population:

Pediatric (defined by the prescribed medication) and adult patients

Environment of Use:

Home, hospital, and sub-acute settings

Contraindications: None

{1}------------------------------------------------

510(k) Summary Page 2 of 6 31-Aug-11

Summary of Modifications:

The modifications to the device do not change the intended use of thepredicate device. The modifications to the device do not alter the fundamental scientific technology. As there are no changes in hardware that will impact performance, there is no need to validate the changes through a clinical investigation.

Modifications:

  • . Minor changes to nebulizer design
  • Changes to physical dimensions .
  • . Change to compressor design to make it smaller and lighter weight

There is no change in intended use, including patient population and environment of use. There is no change in contraindications.

Performance Testing:

Verification activities, as required by the risk analysis, for the modification were performedby the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met. Table 1 summarizes the comparison and Table 2 the testing.

Table 1 - Comparison and Equivalence of Model NE-C801 and the Predicate

FeaturesNewPredicate (K060811)
NE-C801NE-C28
Indications for useThe NE-C801 Nebulizer CompressorSystem is intended to provide air tothe pneumatic nebulizer in order toaerosolize medications for inhalationby the patient for respiratorydisorders.The Omron Compressor Nebulizer Systemsinclude a compressor and nebulizer, ModelNE-C28 uses an AC powered compressor.The electrically powered compressor providescompressed air to the supplied pneumaticnebulizer to aerosolize drugs for inhalation bythe patient. The nebulizer which whendriven by the integral air compressor,nebulizes specific inhalable drugs forinhalation by a patient for treatment ofrespiratory disorders such as allergies, asthma,cystic fibrous, COPD, etc.
Environment of UseHome, Hospital, Sub-acute settingsSame
Patient PopulationPediatric (defined by the prescribedmedication) and adult patientsSame - Pediatric and adult
ContraindicationsNoneNone
Principle of OperationPneumatic (gas powered) jetnebulizerPneumatic (gas powered) jet nebulizer
AerosolizationContinuous during inhalation andexhalationContinuous during inhalation and exhalation
Compressed gas sourceNebulizer compressorWall air / oxygen with flow ratecontrolNebulizer compressorWall air / oxygen with flow ratecontrol
Typical flow rate8 Ipm8 Ipm

{2}------------------------------------------------

510(k) Summary Page 3 of 6

31-Aug-11

FeaturesNewNE-C801Predicate (K060811)NE-C28
Components available in kitwith nebulizerMouthpiece / HoseFace MaskOxygen / Delivery tubingAerosol teeMouthpiece / HoseFace MaskOxygen / Delivery tubingAerosol tee
Component / Accessoriesintended useAll are single patient, multi-useAll are single patient, multi-use
Standards metIEC 60601-1IEC 60601-1-2IEC 60601-1IEC 60601-1-2
Operating conditions10°C to 40°C 30% to 85% RH10°C to 40°C 30% to 85% RH
Storage conditions-20°C to 60°C 10% to 95% RH-20°C to 60°C 10% to 90% RH
Dimensions (mm)142(W) x 72(D) x 98(H)165(W) x 260(D) x 121(H)
Weight (kg)0.270 kg2.5 kg
Performance
Materials ISO 10993Identical to K060811
Particle Characterization per Cascade Impactor
Total Delivered Dose (ug)
MeanPulmicort - 391.1Pulmicort - 401.6
Intal - 12368.9Intal - 12233.3
Salbutamol - 7883.3Salbutamol - 7972.2
Total delivered Dose fraction
(%)Pulmicort - 78.2%Pulmicort - 80.3%
Intal - 61.8%Intal - 61.2%
Salbutamol - 77.0%Salbutamol - 79.7%
Particle size (MMAD)
(Microns)Pulmicort - 3.88Pulmicort - 3.73
Intal - 2.91Intal - 2.71
Salbutamol - 2.54Salbutamol - 2.35
Geometric Std. Dev.
(GSD)Pulmicort - 1.93Pulmicort - 2.05
Intal - 2.27Intal - 2.29
Salbutamol - 2.62Salbutamol - 2.59
Respirable Fraction
(% Mass 0.5-5 microns)Pulmicort-61.9%Pulmicort - 62.7%
Intal - 70.5%Intal - 72.8%
Salbutamol - 66.4%Salbutamol - 69.2%
Total Respirable Dose (ug
0.5 -5.0 microns)Pulmicort - 242.2Pulmicort - 251.9
Intal - 8729.5Intal - 8894.8
Salbutamol - 5335.0Salbutamol - 5514.4
Medication captured on USP
Throat (ug)Pulmicort - 13.3Pulmicort - 14.9
Intal - 253.6Intal - 213.4
Salbutamol - 144.9Salbutamol - 141.4
Medication captured on USP
Throat Fraction (%)Pulmicort - 3.4%Pulmicort - 3.8%
Intal - 2.0%Intal - 1.7%
Salbutamol - 1.8%Salbutamol - 1.8%
Medication retained in
Device (ug)Pulmicort - 96.1Pulmicort - 104.4
Intal - 8000.0Intal - 7533.3
Salbutamol - 2072.2Salbutamol - 1977.8
Medication Retained in
Device Fraction (%)Pulmicort - 19.2%Pulmicort - 20.9%
Intal - 40.0%Intal - 37.7%
Salbutamol - 41.4%Salbutamol - 39.6%

{3}------------------------------------------------

510(k) Summary Page 4 of 6 31-Aug-11

FeaturesNewNE-C801Predicate (K060811)NE-C28
Particle Characterization per Cascade Impactor
Coarse Particle (>4.7 um)Fraction (%)Pulmicort - 37.1%Intal - 27.7%Salbutamol - 26.6%Pulmicort - 35.3%Intal - 25.0%Salbutamol - 24.6%
Fine Particle (<4.7 um)Fraction (%)Pulmicort - 59.6%Intal - 70.2%Salbutamol - 71.6%Pulmicort - 60.9%Intal - 73.2%Salbutamol - 73.7%
Ultra Fine Particle (um)Fraction (%)Pulmicort - 3.5%Intal - 8.9%Salbutamol - 16.9%Pulmicort - 5.4%Intal - 9.4%Salbutamol - 0.19%
Nebulization rate (g/ml)Pulmicort - 0.37Intal - 0.30Salbutamol - 0.37Pulmicort - 0.44Intal - 0.35Salbutamol - 0.41
Confidence level of testingThe test and number of samples (3) with 3 runs provided a 95% confidence levelThe test and number of samples (3) with 3 runs provided a 95% confidence level

Table 2 - Particle Characterization results at 95% confidence

A series of aerosol performance tests were performed using an 8 stage cascade impactor at a sampling flow rate of 15 I/min equipped with a USP <601> induction port throat. Aerosol was sampled directly from the outlet. Three (3) device samples were tested with 3 runs each, for a total of 9 sample points per each drug. Aerosol was sampled directly from the outlet. The following table is a summary of results with intervals given for a 95% confidence level.

Mean / Std. DevPulmicort(250 ug/ml)Intal(10 mg/ml)Salbutamol(5 mg/ml)
Total delivered Dose (ug)391.11±16.5112368.89±269.617883.33±116.96
Total Delivered Dose Fraction (%)78.2%±3.3%61.8%±3.1%77.0%±2.5%
Particle Size (MMAD) um3.88±0.282.91±0.142.54±0.28
Geometric Standard Deviation1.93±0.202.27±0.022.62±0.03
Respirable Fraction(0.5-5 um)61.9%±4.0%70.5%±1.3%66.4%±1.6%
Total Respirable Dose(ug between 0.5-5 um)242.17±18.908729.49±497.575235.02±233.61
Medication Captured on USPThroat13.34±3.40253.66±27.42144.94±16.32
Medication Captured on USPThroat Fraction (%)3.4%±0.4%2.0%±0.1%1.8%±0.1%
Medication Retained in Device96.11±9.618000.00±438.212072.22±257.90

{4}------------------------------------------------

510(k) Summary Page 5 of 6 31-Aug-11

Mean / Std. DevPulmicort(250 ug/ml)Intal(10 mg/ml)Salbutamol(5 mg/ml)
Medication Retained in DeviceFraction (%)19.2%$\u00b1$ 1.9%40%$\u00b1$ 2.2%41.4%$\u00b1$ 5.2%
Mean / Std. DevPulmicort(250 ug/ml)Intal(10 mg/ml)Salbutamol(5 mg/ml)
Coarse Particle Fraction (%) (>4.7 um)37.1%$\u00b1$ 2.9%27.7%$\u00b1$ 1.5%26.6%$\u00b1$ 2.3%
Fine Particle Fraction (%) (<4.7 um)59.6%$\u00b1$ 3.5%70.2%$\u00b1$ 1.5%71.6%$\u00b1$ 2.3%
Ultra-Fine Particle Fraction &(<1.0 um)3.5%$\u00b1$ 1.7%8.9%$\u00b1$ 2.1%16.9%$\u00b1$ 5.3%

The differences between the proposed device and the predicates are:

  • . Size
  • . Slight modification to nebulizer

It is our view that these are not significant differences that affect the safety or effectiveness of the intended device as compared to the predicate device.

Substantial Equivalence:

The NE-C801 is viewed as substantially equivalent to the predicate device because:

Indications for Use:

The Indications for Use are as a general purpose nebulizer and compressor system which is identical to predicate - K060811

Patient Population:

The patient population is pediatric and adult which is identical to predicate - K060811. The pediatric population was defined based upon the limits of the prescribed medication.

Environment of Use:

The environment of use is - home, hospital and sub-acute care settings which is identical to predicate -K060811

Technology:

The technology of the compressor is identical compressor technology to predicate- K060811

The technology of the pneumatic (jet-style) continuous flow nebulization is identical nebulizer technology to predicate - K060811

Materials:

The materials in the gas and fluid pathway are identical to predicate- K060811.

{5}------------------------------------------------

510(k) Summary Page 6 of 6 31-Aug-11

Performance Testing:

We have performed a number of performance tests which are identical to those done with the predicate, K060811. These included:

  • IEC 60601-1 ●
  • . IEC 60601-1-2
  • Cascade Impactor particle testing with 3 devices for 3 runs with 3 drugs resulting in a 95% . confidence interval
  • . Compressor testing
    • VOC testing per EPA TO-15 o
    • o PM2.5 testing per NIOSH NMAM 0600
    • Ozone testing per OSHA method ID 214 O

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The department's name is written in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Omron Healthcare. Incorporated C/O Mr. Paul Dryden President ProMedic, Incorporated 24301 Woodsage Drive Bonita Springs. Florida 34134

EP - 2 2001

Re: K110860

Trade/Device Name: NE-C801 Nebulizer Compressor System Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: July 12, 2011 Received: July 13, 2011

Dear Mr. Dryden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class IH (PMA), it may be subject to additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{7}------------------------------------------------

Page 2 – Mr. Dryden

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Anthony D. Watson

Anthony D. Watson. B.S.. M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Indications for Use Statement

Page 1 of 1

K110860 510(k) Number:

Device Name:

NE-C801 nebulizer compressor system

Indications for Use:

The NE-C801 Nebulizer Ciompressor System is intended to provide air to the pneumatic nebulizer in order to aerosolize medications for inhalation by the patient for respiratory The system is designed for use with pediatric (defined by the prescribed disorders. medication) and adult patients in the home, hospital, and sub-acute settings.

Prescription Use XX (Part 21 CFR 801 Subpart D)

or

Over-the-counter use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

L. Shult

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K110860

§ 868.5630 Nebulizer.

(a)
Identification. A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.(b)
Classification. Class II (performance standards).