K Number
K110857

Validate with FDA (Live)

Date Cleared
2011-07-15

(109 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.
A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner (21CFR 880.6250)

Device Description

Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powder-Free Vinyl Patient Examination Glove, 80LYZ, and meets all requirement of ASTM Standard D5250-06.

AI/ML Overview

The provided document describes the acceptance criteria and the study conducted for Shandong Zhanpeng Medical Products Co., Ltd. Synthetic Vinyl Patient Examination Gloves - Powder Free, Yellow Color.

1. Table of acceptance criteria and reported device performance:

Acceptance Criteria CategorySpecific Test/StandardAcceptance CriteriaReported Device Performance
Physical and DimensionsASTM-D-5250-06Meets all requirements (Specifics not detailed)Meets all requirements for Physical and Dimensions Testing
Pinhole/Water LeakFDA 1000 ml. Water Fill TestAOL 2.5, Inspection Level IMeets these requirements
BiocompatibilityPrimary Skin Irritation TestNo primary skin irritant reactionsResults showed no primary skin irritant reactions
BiocompatibilitySkin Sensitization (allergic contact dermatitis) TestNo sensitization reactionsResults showed no sensitization reactions
Powder ContentPowder-Free ClaimNo more than 2 mg powder per gloveConducted to insure gloves meet "powder-free" claims

2. Sample size used for the test set and the data provenance:

  • Physical and Dimensions Testing: Inspection Level S-2, AOL 2.5 (Specific sample size not provided, but these are statistical sampling plans). Data provenance is from testing conducted by the manufacturer, presumably in China.
  • FDA 1000 ml. Water Fill Test: Samplings of AOL 2.5, Inspection Level I (Specific sample size not provided, but these are statistical sampling plans). Data provenance is from testing conducted by the manufacturer, presumably in China.
  • Primary Skin Irritation and Skin Sensitization Testing: Sample size is not specified. Data provenance is from testing conducted by the manufacturer.

The data is retrospective as it was collected as part of the 510(k) submission process to demonstrate substantial equivalence to a predicate device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This is not applicable as the device is a medical glove, and the "ground truth" is established through standardized laboratory tests (e.g., ASTM standards, FDA water fill test, biocompatibility tests) rather than expert interpretation of images or clinical data.

4. Adjudication method for the test set:

Not applicable for this type of device and testing. The evaluation of test results is based on meeting predefined specifications of the ASTM standard and FDA requirements.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is a medical glove; it does not involve AI or human readers for diagnostic interpretation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This device is a medical glove; it does not involve an algorithm.

7. The type of ground truth used:

The ground truth is based on the results of standardized laboratory testing as per ASTM-D-5250-06 and the FDA 1000 ml Water Fill Test, as well as biocompatibility testing. This is objective data derived from physical and chemical properties and biological responses, not expert consensus or pathology in the context of diagnostic imaging.

8. The sample size for the training set:

Not applicable. This device is a medical glove, and its performance is established through direct physical, dimensional, and biocompatibility testing, not through machine learning models requiring training data.

9. How the ground truth for the training set was established:

Not applicable for the reasons stated above.

{0}------------------------------------------------

JUL 1 5 2011

510 (K) SUMMARY

This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21CFR 807.92.

1. Submitter's Identification:

Shandong Zhanpeng Medical Products Co., Ltd. 6 Chuangxin Road Yunhe Industry Park Qingzhou, Shandong, China Date summary prepared: March 24, 2011

Name of the Device: 2.

Shandong Zhanpeng Medical Products Co., Ltd. Synthetic Vinyl Patient Examination Gloves - Powder Free, Yellow Color

3. Predicate Device Informaton:

Shijiazhuang Hongxiang Plastic Products Ltd. Synthetic Vinyl Patient Examination Gloves - Powder Free (K992821)

4. Device Description:

Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powder-Free Vinyl Patient Examination Glove, 80LYZ, and meets all requirement of ASTM Standard D5250-06.

5. Intended Use:

A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

6. Comparison to Predicate Devices:

Shandong Zhanpeng Medical Products Co., Ltd. Synthetic Vinyl Patient Examination Gloves, Powder-Free, yellow color are substantially equivalent in safety and effectiveness to the Shijiazhuang Hongxiang Plastic Products Co., Ltd. and Sunmax Enterprise Shanghai Co., Ltd. Powder-Free Vinyl Patient Examination Gloves, Powder free.

{1}------------------------------------------------

7. Substantial Equivalence are as follows:

The standards used for Shandong Zhanpeng Medical Products Co., Ltd. glove production are based on ASTM-D-5250-06. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AOL 2.5.

The FDA 1000 ml. Water Fill Test was also conducted with samplings of AOL 2.5, Inspection Level I, meeting these requirements, Primary Skin irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.

There are no special labeling claims and we do not claim our gloves as hypoallergenic is conducted to insure that our gloves meet our "powder-free" claims (contain no more than 2 mg powder per glove).

8. Discussion of Clinical Tests Performed:

Not Applicable - There is no hypoallergenic claim.

9. Conclusions:

Shandong Zhanpeng Medical Products Co., Ltd. Synthetic Vinyl Patient Examination Gloves, Powder-Free, yellow color conform fully to ASTM-D-5250-06 standard as well as applicable 21 CFR references, and, meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol consisting of three curved lines that resemble a person with outstretched arms. The symbol is black, and the background is white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Shandong Zhanpeng Medical Products Company, Limited C/O Mr. Frank Liu President Basic Medical Industries, Incorporated 12390 East End Avenue Chino, California 91710

JUL 1 5 2011

Re: K110857

Trade/Device Name: Patient Vinyl Examination Gloves, Powder Free, Non-Sterile Yellow Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove

Regulatory Class: I

Product Code: LYZ

Dated: June 8, 2011 Received: June 9, 2011

Dear Mr. Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. W222

{3}------------------------------------------------

Page 2 - Mr. Liu

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersO ffices/CDRH/CDRHOff ices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Anthony D.m

Cinthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Shandong Zhanpeng Medical Products Co., Ltd. 6 Chuangxin Road, Yunhe Industry Park Qingzhou, Shandong, China

INDICATIONS FOR USE

Shandong Zhanpeng Medical Products Co., Ltd. Applicant:

510(k) Number: K ||0857

Device Name:

Patient Vinyl Examination Gloves, Powder free, Non-Sterile, Yellow Color

Indications of Use:

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner (21CFR 880.6250)

Prescription Use _____________________________________________________________________________________________________________________________________________________________

Over the Counter Use _ X

Factory Initials

Elaine S. Mayhell
(Division Sign-Off)

Division of Anesthesiology, General Hospital ിക്കുന്നത്.

510(k) Number: _______________________________________________________________________________________________________________________________________________________________ KII0857

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.