K Number
K110761

Validate with FDA (Live)

Device Name
MICT TSH
Date Cleared
2012-03-09

(357 days)

Product Code
Regulation Number
862.1690
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MICT® TSH is a magnetic lateral flow immunoassay for the in-vitro quantitative determination of TSH in human serum. The measurements of thyroid stimulating hormone (TSH) produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders. The MICT System is intended for use in clinical laboratories.

Device Description

The MICT® TSH Immunoassay for the quantitative determination of TSH in human serum is designed for use on MICT® Instrument.
MICT® TSH Device Reagents:

  1. Solid phase nitrocellulose membrane device with two detection zones:
    a. Control line region containing a polyclonal goat anti-rabbit IgG antibody, and
    b. Test line region containing a monoclonal anti-TSH antibody
  2. Lyophilized conjugate tube containing the monoclonal anti-TSH antibody covalently coupled to 300 nm super paramagnetic particles.
    The MICT® Immunoassay is a "sandwich" magnetic chromatographic assay.
AI/ML Overview

The provided 510(k) summary describes the MICT® TSH device and its performance characteristics. This is a medical device for in-vitro quantitative determination of TSH in human serum, not an AI/ML device, and therefore does not have acceptance criteria and a study design in the format requested for AI/ML devices. However, I can extract the relevant performance data as presented.

Here's the information based on the provided text, adapted to the closest relevant categories for a non-AI/ML device:

1. Table of Acceptance Criteria and Reported Device Performance

For this type of device, "acceptance criteria" are generally inferred from a comparison to a predicate device and established analytical performance benchmarks for immunoassays. The study demonstrates substantial equivalence to the predicate device (FastPack® TSH).

FeatureAcceptance Criteria (Implied by Predicate)Reported Device Performance (MICT® TSH)
Precision (Within Lab)Demonstrate comparable or better precision to the predicate device.Site 1: 1. 0.191 µIU/mL (14.1%CV)2. 0.29 µIU/mL (9.4%CV)3. 0.83 µIU/mL (9.4%CV)4. 3.7 µIU/mL (7.8%CV)5. 46.7 µIU/mL (8.8%CV)Site 2: 1. 0.191 µIU/mL (13.3%CV)2. 0.29 µIU/mL (10.1%CV)3. 0.87 µIU/mL (7.8%CV)4. 4.1 µIU/mL (9.1%CV)5. 46.6 µIU/mL (9.5%CV)Site 3: 1. 0.187 µIU/mL (14.4%CV)2. 0.29 µIU/mL (9.5%CV)3. 0.87 µIU/mL (7.6%CV)4. 3.95 µIU/mL (9.1%CV)5. 48.4 µIU/mL (7.5%CV)(Predicate: Site 1 Total Within Lab: 0.54 µIU/mL (20.0%CV), 1.54 µIU/mL (9.7%CV), 4.69 µIU/mL (7.0%CV), 46.30 µIU/mL (9.5%CV))
SensitivityComparable or better than predicate device (Predicate Analytical = 0.043 µIU/mL, Functional = 0.13 µIU/mL)LOB = 0.04 µIU/mLLOD = 0.077 µIU/mLLOQ = 0.077 µIU/mL
Method Comparison (Correlation with Predicate)High correlation (e.g., r > 0.95 and slope/intercept close to 1/0)Site 1: MICT = 0.97 FastPack - 0.27 (n=66, r = 0.99)Site 2: MICT = 0.93 FastPack - 0.07 (n=74, r = 0.99)Site 3: MICT = 0.98 FastPack 0.00 (n=64, r = 0.99)Pooled: MICT = 0.96 FastPack - 0.12 (n=204, r = 0.99)
Interfering SubstancesNo significant interference at specified concentrations.Bilirubin: No interference up to 40 mg/dLHemoglobin: No interference up to 1000 mg/dLTriglycerides: No interference up to 3200 mg/dLPhospholipid: No interference up to 800 mg/dL
SpecificityNo significant cross-reactivity with related hormones.No interference from 500 mIU/mL LH and 500 mIU/mL FSH.

2. Sample Size Used for the Test Set and Data Provenance
The method comparison study used a total of n=204 samples.

  • Site 1: n=66
  • Site 2: n=74
  • Site 3: n=64

The data provenance is not explicitly stated in terms of country of origin or whether it's retrospective/prospective. As a clinical laboratory device study, these are typically prospective evaluations of clinical samples.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. This is an immunoassay device, and "ground truth" for TSH levels is established by the comparative measurement of known reference methods or another analytically validated device (the predicate device in this case), not by expert interpretation.

4. Adjudication Method for the Test Set
Not applicable. See point 3.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is an immunoassay device, not an AI/ML device used by human readers for interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This is an automated immunoassay system; its performance is standalone in the sense that the instrument provides the quantitative reading.

7. The Type of Ground Truth Used
The "ground truth" (or reference method) for the method comparison study was the Qualigen FastPack® TSH device. This is a legally marketed predicate device, implying its results are accepted as a valid reference for TSH determination.

8. The Sample Size for the Training Set
Not applicable. This device is an immunoassay, not an AI/ML system that requires a "training set." The device's calibration curve is factory-generated.

9. How the Ground Truth for the Training Set was Established
Not applicable. See point 8. The device uses a factory-generated master curve with a single calibration per lot, implying that the "calibration" (analogous to training for an AI model) is established by the manufacturer using a set of known standards.

{0}------------------------------------------------

K110761

MAR - 9 2012


510(k) Summary

MICT® TSH Using the MICT® Instrument

This summary of 510(k) safety and effectiveness is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

1.Submittername, address,contactMagnaBioSciences LLC6325 Lusk Blvd.San Diego, CA 92121
Telephone:Fax:(858)481-4400(858-481-7410
Contact Person:James L. Wyatt
Date Prepared:February 2012
2.Device nameProprietary name:MICT® TSHUsing the MICT® Instrument
Common name:Magnetic ImmunoChomatographyic Test for thedetermination of TSH
Classification Name:Quantitative Determination of TSH in HumanSerum
3.PredicatedeviceQualigen FastPack® TSH (K052301)
4.DevicedescriptionThe MICT® TSH Immunoassay for the quantitative determination of TSHin human serum is designed for use on MICT® Instrument.
MICT® TSH Device Reagents
1. Solid phase nitrocellulose membrane device with two detectionzones:a. Control line region containing a polyclonal goat anti-rabbitIgG antibody, andb. Test line region containing a monoclonal anti-TSHantibody

{1}------------------------------------------------

MagnaBioSciences, LLC

510(k) Summary (continued)

    1. Lyophilized conjugate tube containing the monoclonal anti-TSH antibody covalently coupled to 300 nm super paramagnetic particles.
      The MICT® Immunoassay is a "sandwich" magnetic chromatographic assay:
  • Step 1: Sample or control (100 uL) is added to the lyophilized . conjugate tube and mixed using an up and down motion in the pipet.

  • Set 2: Transfer the entire contents of the sample/conjugate to the ● sample port on the nitrocellulose test cassette.

  • Step 3: The sample/conjugate will chromatograph along the . nitrocellulose membrane for 30 minutes, but not longer than 45 minutes after starting the test..

  • Step 4: Insert the test cassette into the MICT magnetic Instrument . and press start after barcode is read.

  • Results will automatically print. ●

  • The MICT® TSH is a magnetic lateral flow immunoassay for the in-vitro ડ. Intended use quantitative determination of TSH in human serum. The measurements of thyroid stimulating hormone (TSH) produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders. The MICT TSH immunoassay is designed for use with the MCIT instrument. The MICT System is intended for use in clinical laboratories.

  • The following tables compare the MICT® TSH with the Qualigen Comparison to 6. Predicate Device FastPack® TSH.

{2}------------------------------------------------

510(k) Summary (continued)

MagnaBioSciences, LLC

FeatureMICT® TSHFastPack® TSH
Intended Use.The MICT® TSH is a magnetic lateral flow immunoassay for the in-vitro quantitative determination of TSH in human serum. The measurements of thyroid stimulating hormone (TSH) produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders. The MICT System is intended for use in clinical laboratories.FastPack® TSH Immunoassay is a paramagnetic particle immunoassay for the in-vitro quantitative determination of TSH in human plasma. The measurements of thyroid stimulating hormone (TSH) produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders. The FastPack TSH Immunoassay is designed for use with the FastPack System
Assay Methodology:Sandwich immunoassaySandwich immunoassay
Storage Condition:1 - 30°C2-8 °C
Data AnalysisInternal data reduction via microcomputerInternal data reduction via microcomputer
Test ProcessingSemi-AutomatedSemi-Automated
Sample Type:SerumPlasma
Detector:Magnetic InstrumentPhotomultiplier Tube (PMT)
LabelSuper Magnetic ParticlesAlkaline Phosphatase -Lucegine
Sample Volume100 µL100 µL
Assay Range0.13 to 100 µIU/mL0.13 to 100 µIU/mL
Instrument RequiredMICT® InstrumentQualigen FastPack® System
AntibodyMonoclonal/MonoclonalMonoclonal/Monoclonal
Solid PhaseNitrocellulose MembraneMagnetic Particles
DetectionMagnetic MomentChemiluminescence
CalibrationFactory generated master curve with single calibration per lot, calibration is stable until expiration of the assay lot.Factory generated calibration curve using 6 standards.Recalibration required 14 day at the customer site.
ThroughputSingle SampleSingle Sample
Time to Result30 minutes15 minutes
Reagents Supplied asBox of 20 reagent cassettes with lyophilized conjugate tubesBox of 50 individual Tests

{3}------------------------------------------------

MagnaBioSciences, LLC

1

Performance Characteristics:MICT® TSHFastPack® TSH
FeatureMeanµIU/mLWithin Lab%CVMeanµIU/mLTotal Within Lab%CV
PrecisionSite 1
10.19114.110.5420.0
20.299.421.549.7
30.839.434.697.0
43.7546.77.88.8446.309.5
Site 2
10.19113.3
20.2910.1
30.877.8
44.1546.69.19.5
Site 3
10.18714.4
20.299.5
30.877.6
43.95548.49.17.5
SensitivityLOB = 0.04 µIU/mLLOD = 0.077 µIU/mLLOQ = 0.077 µIU/mLAnalytical = 0.043 µIU/mLFunctional = 0.13 µIU/mL
MethodComparisonThree site performance evaluation of MICT vs. QualigeFastPack TSH Method.Site 1: MICT = 0.97 FastPack -0.27 (n=66, r = 0.99)Site 2: MICT = 0.93 FastPack -0.07 (n=74, r = 0.99)Site 3: MICT = 0.98 FastPack 0.00 (n=64, r = 0.99)Pooled: MICT = 0.96 FastPack -0.12 (n=204, r = 0.99)
InterferingSubstancesNo interference up to:No interference up to:
Bilirubin40 mg/dL40 mg/dL
Hemoglobin1000 mg/dL1000 mg/dL
Triglycerides3200 mg/dL1000 mg/dL
Phospholipid800 mg/dL--
Specificity500 mIU/mL LH500 mIU/mL FSHn.d500 mIU/mL LH500 mIU/mL FSHn.d

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". Inside the circle is a stylized image of an eagle or other bird with outstretched wings. The bird is facing to the left.

Food and Drug Administration

10903 New Hampshire Avenue Silver Spring, MD 20993

MagnaBioSciences, LLC. c/o James Wyatt, PhD Chief Technical Officer 6325 Lusk Blvd. San Diego, CA 92121 .

MAR - 9 2012

K110761 Re:

Trade Name: MICT® TSH and MICT® Instrument Regulation Number: 21 CFR §862.1690 Regulation Name: Thyroid Stimulating Hormone Test System Regulatory Class: II Product Codes: JLW, JJQ Dated: February 28, 2012 Received: March 6, 2012

Dear Dr. Wyatt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). .

{5}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical

Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ...

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm

Sincerely yours,

V

Counney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices

Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use Form

510(k) Number: K110761

Device Name: MICT®TSH

Indications for Use:

The MICT® TSH is a magnetic lateral flow immunoassay for the in-vitro quantitative determination of TSH in human serum. The measurements of thyroid stimulating hormone (TSH) produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders. The MICT System is intended for use in clinical laboratories.

Prescription Use) × (Part 21 CFR 801. Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Autie Charles

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) 110761

2 Page 1 of

{7}------------------------------------------------

Indications for Use Form

510(k) Number (if known): K110761 ·

Device Name: MICT® Instrument

Indications for Use:

The MICT® TSH is a magnetic lateral flow immunoassay for the in-vitro quantitative determination of TSH in human serum. The measurements of thyroid stimulating hormone (TSH) produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders. The MICT System is intended for use in clinical laboratories.

Prescription Use) (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Qute Chuler

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) 110761

Page 1 of 2

§ 862.1690 Thyroid stimulating hormone test system.

(a)
Identification. A thyroid stimulating hormone test system is a device intended to measure thyroid stimulating hormone, also known as thyrotrophin and thyrotrophic hormone, in serum and plasma. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.(b)
Classification. Class II.