Intended Use

This instrument tray is intended to be used to enclose and protect Olympus flexible endoscopes during sterilization. The instrument tray is designed to hold one endoscope at a time and is to be used in conjunction with an FDA cleared sterilization wrap for ethylene oxide sterilization. The instrument tray is an optional accessory to the Olympus endoscopes for which it is designed. Maintenance of sterility depends on the sterilization wrap, not on the instrument tray. This instrument tray is indicated for ethylene oxide sterilization of only compatible Olympus endoscopes. For compatible endoscopes, refer to the table below, "Compatible Olympus Endoscopes".

Device Description

The Olympus Instrument Tray is comprised of plastic lids and bottoms that contain numerous large holes (approximately 7mm in diameter) that permit ready ingress and egress of sterilization gases. The trays are designed to provide protection from physical damage to the flexible endoscope during sterilization and storage.

AI/ML Overview

This document describes the Olympus Instrument Tray, intended to protect flexible endoscopes during ethylene oxide (EtO) sterilization. The study focuses on demonstrating the tray's effectiveness in allowing proper sterilization and aeration of endoscopes.

Here's the breakdown of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance
Sterilization EffectivenessDemonstrate 6-log reduction capability (SAL of 10^-6^) in a 30-minute half cycle for both the instrument tray (worst-case component load) and the worst-case endoscope (URF-P5)- Instrument Tray: Sterilized successfully in a 30-minute half cycle, demonstrating 6-log reduction capability (SAL of 10^-6^).- Worst-Case Endoscope (URF-P5): Sterilized successfully in a 30-minute half cycle, demonstrating 6-log reduction capability (SAL of 10^-6^).
Ethylene Oxide ResidualsResidual concentrations of Ethylene Oxide (EO) and Ethylene Chlorohydrin (ECH) must be within acceptable limits after a 12-hour aeration time, pursuant to ISO10993-7.- Instrument Tray: Residual concentrations of EO and ECH were all within acceptable limits after 12 hours of aeration.- Worst-Case Endoscope (URF-P5): Residual concentrations of EO and ECH were all within acceptable limits after full cycles and appropriate aeration.
BiocompatibilityMaterial (Radel-R) meets the requirements for biocompatibility pursuant to ISO-10993.The trays are constructed of biocompatible Radel-R, which meets the requirements for biocompatibility according to ISO-10993.
Risk AnalysisRisk analysis carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007.Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007. (No specific performance data is provided here, just that the process was followed.)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not explicitly state the numerical sample size for the test set (number of trays or endoscopes tested). However, the testing involved:
    • "worst case component load" for the instrument tray validation.
    • "worst case endoscope (URF-P5)" for the endoscope validation. This implies at least one URF-P5 endoscope was tested, along with an unspecified number of biological indicator organisms and chemical indicators.
  • Data Provenance: The data appears to be prospective as it describes validation testing performed specifically for this submission. The country of origin of the data is not specified, but the applicant and manufacturer are based in Germany, and the correspondent in the USA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • The document does not mention the use of external experts to establish ground truth for the test set. The ground truth (successful sterilization defined by 6-log reduction and acceptable residual levels) was established through objective laboratory testing using FDA cleared biological indicator organisms and chemical indicators, in conjunction with relevant ISO standards.

4. Adjudication Method for the Test Set

  • No adjudication method is described, as the evaluation relies on objective laboratory measurements (biological indicator kill, residual chemical levels) rather than subjective human interpretation.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This device is an instrument tray, not an AI or imaging device that involves human readers or interpretation of results. Therefore, an MRMC comparative effectiveness study was not applicable and not done.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This device is an instrument tray, not an algorithm or AI system. Therefore, a standalone performance study as typically understood for AI was not applicable and not done. The "standalone" performance here relates to the physical and chemical performance of the tray itself in facilitating sterilization.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • The ground truth used for both sterilization effectiveness and residual testing was based on objective biological and chemical indicators and adherence to established regulatory standards (SAL of 10^-6^ and ISO 10993-7).
    • For sterilization: 6-log reduction in biological indicators.
    • For residuals: Measurement of EO and ECH concentrations within acceptable limits.

8. The Sample Size for the Training Set

  • This device does not involve a "training set" as it is not an AI/machine learning device. The validation is based on physical and chemical testing described above.

9. How the Ground Truth for the Training Set was Established

  • Not applicable, as there is no training set for this device.

{0}------------------------------------------------

K110748

510(k) Summary of Safety and EffectivenessFebruary 1, 2012FEB 15 2012
ApplicantOlympus Winter & Ibe GmbHKuehnstrasse 61 * 22045 Hamburg * GermanyEstablishment Registration No: 9610773
Official CorrespondentStacy Abbatiello Kluesner, M.S., RACRegulatory Affairs & Quality AssuranceOlympus America Inc.3500 Corporate ParkwayPO Box 610Center Valley, PA 18034-0610Phone: 484-896-5405FAX: 484-896-7128Email: Stacy.Kluesner@Olympus.comEstablishment Registration No.: 2429304
ManufacturerOlympus Winter and Ibe GmbHKuehnstrasse 61 * 22045 Hamburg * GermanyEstablishment Registration No.: 9610773
Device Trade Name:Olympus Winter & Ibe Instrument Tray WA05991A;abbreviated hereinafter as Olympus Instrument Tray
Common Name:Sterilization Wrap Container, Tray, Cassette & otherAccessories
Regulation Number:21 CFR 880.6850
Regulation Name:Sterilization Wrap
Regulatory Class:II
Product Code:KCT
Classification Panel:General Hospital
Performance Standards:None established
Predicate Devices:K092682 Gyrus ACMI® Flexible Endoscope Storage-Sterilization Trays K033222 Olympus Sterilization Trays

Intended Use:

This instrument tray is intended to be used to enclose and protect Olympus flexible endoscopes during sterilization. The instrument tray is designed to hold one endoscope at a time and is to be used in conjunction with an FDA cleared sterilization wrap for ethylene oxide sterilization. The instrument tray is an optional accessory to the Olympus endoscopes for which it is designed. Maintenance of sterility depends on the sterilization wrap, not on the instrument tray.

This instrument tray is indicated for ethylene oxide sterilization of only compatible Olympus endoscopes. For compatible endoscopes, refer to the table below "Compatible Olympus Endoscopes".

{1}------------------------------------------------

Compatible Olympus Endoscopes are:

CHF-CB30LCYF-V2RENF-V3LF-DP
CHF-CB30SCYF-VA2ENF-VQLF-GP
CHF-P20CYF-VHENF-VT2LF-P
CHF-P60CYF-VHAENF-XPLF-TP
CHF-XP20CYF-VHRENF-VHLF-V
CHF-VENF-GPHYF-1TURF-P5
CYF-5ENF-P4HYF-VURF-V
CYF-5AENF-T3HYF-XP
CYF-V2ENF-V2LF-2

Worst case dimension:

The endoscope URF-P5 offers the most challenging dimensions in regards to maximum length (835 mm), minimum inner diameter (1.2 mm) and number of lumen (1).

Ethylene oxide sterilization parameters:

In chamber conditioning
Temperature................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................ the list of the table is not complete. I will add the missing values and dots. ```json {

{2}------------------------------------------------

Product Description:

TrayItem no.Compatible endoscopes
Olympus InstrumentTrayWA05991AOlympus Flexible Endoscopes of the CHF,CYF, ENF, HYF, LF, and URF series
FeatureThe Olympus Instrument Tray is comprised of plastic lids and bottomsthat contain numerous large holes (approximately 7mm in diameter) thatpermit ready ingress and egress of sterilization gases.The trays are designed to provide protection from physical damage to theflexible endoscope during sterilization and storage.
Intended UseProvided above.
MaterialThe trays are constructed of biocompatible RADEL-R. Radel-R is apolyphenylsulfone plastic that is widely used in medical devices. Radel-Rmeets the requirements for biocompatibility pursuant to ISO-10993 and iscompatible with EtO sterilization.
Patient ContactThe trays do not contact the patient
SterilizationmodalityEtO
Dimensions531mm (L) x 255mm (W) x 86mm (D)

Performance Data:

Validation of "Worst-Case" Challenges of the Instrument Tray:

Validation was performed to assess the worst case component load and the ability of ethylene oxide to penetrate and aerate from the surfaces and silicon holders of the instrument Trav. The PCDs (process challenge devices) were placed in the trays and inoculated with FDA cleared biological indicator organisms, and chemical indicators were placed. The trays were wrapped with two layers of FDA approved sterilization wrap (Kimguard KC600; K082554) and placed into ethylene oxide sterilizer for processing. The system was sterilized successfully in a 30 minute half cycle demonstrating 6 log reduction capability ( SAL of 10-6). Test systems were exposed to 60 minute full cycles and ethylene oxide residual testing was performed, with a 12 hour aeration time. Pursuant to ISO10993-7, residual concentrations of EO and ECH were all within acceptable limits. Therefore the instrument tray is effectively sterilized when the worst case components are loaded and the level of remaining residuals was found to be acceptable.

Validation of 'Worst-Case" Endoscope in the Instrument Trav:

Validation was also performed to assess the worst case endoscope (URF-P5) loaded into the instrument tray. This testing assessed the ability of ethylene oxide to penetrate the worst case endoscope features (lumen length and diameter) and the level of residuals remaining. Validation was performed with FDA cleared accessories, including biological indication wrap (Kimguard KC600; K082554) and ethylene oxide sterilizer. The endoscopes were sterilized successfully in a 30 minute half cycle demonstrating 6 log reduction capability ( SAL of 10-6). The endoscopes were exposed for full cycles and all ethylene oxide residual concentrations of EO and ECG were all within acceptable limits. Therefore, the Instrument Tray was found to not hinder sterilization and aeration of the worst-case endoscope.

Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007. Other standards applied to the design of this device include ANSI/AMMI ST33-1996, AAMI / ANSI / ISO 11135-1, AAMI / ANSI / ISO 10993-1, AAMI / ANSI / ISO 10993-7, and AAMI/ ANSI ST77:2006.

Technological Characteristics and Substantial Equivalence:

{3}------------------------------------------------

The Olympus Winter & Ibe GmbH Instrument Tray is composed of the same materials, constructed the same and uses the same design philosophy as the general hospital trays sold by Gyrus ACMI FGT Endoscopes Storage/Sterilization trays (K092682) and Olympus (K033222). In contrast to the Olympus Sterilization Trays (K033222), the Olympus Instrument Tray is only indicated for EtO sterilization as is the Gyrus ACMI FGT Storage/Sterilization Tray (K092686).

■ Conclusion:

In summary, the Olympus Instrument Tray is substantially equivalent to the predicate devices and raises no new concerns of safety or efficacy when compared to the predicate devices.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines representing its wings and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002

Olympus Winter & Ibe GmbH C/O Stacy A. Kluesner, M.S., RAC Regulatory Affairs & Quality Assurance Olympus America, Inc. 3500 Corporate Parkway Center Valley, Pennsylvania 18034

FEB 1 5 2012

Re: K110748

Trade/Device Name: Olympus Instrument tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: February 3, 2012 Received: February 8, 2012

Dear Ms. Kluesner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the gencral controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 -- Ms. Kluesner

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

for Anthony J. Watson

Anthony D. Watson, B.S., M.S., M.B.A. · Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number:K110748
Device Name:Olympus Instrument tray
Model Numbers:WA05991A

Indications For Use:

This instrument tray is intended to be used to enclose and protect Olympus flexible endoscopes during sterilization. The instrument tray is designed to hold one endoscope at a time and is to be used in conjunction with an FDA cleared sterilization wrap for ethylene oxide sterilization. The instrument tray is an optional accessory to the Olympus endoscopes for which it is designed. Maintenance of sterility depends on the sterilization wrap, not on the instrument tray. This instrument tray is indicated for ethylene oxide sterilization of only compatible Olympus endoscopes. For compatible endoscopes, refer to the table below, "Compatible Olympus Endoscopes".

Compatible Olympus Endoscopes are:

CHF-CB30LCYF-V2RENF-V3LF-DP
CHF-CB30SCYF-VA2ENF-VQLF-GP
CHF-P20CYF-VHENF-VT2LF-P
CHF-P60CYF-VHAENF-XPLF-TP
CHF-XP20CYF-VHRENF-VHLF-V
CHF-VENF-GPHYF-1TURF-P5
CYF-5ENF-P4HYF-VURF-V
CYF-5AENF-T3HYF-XP
CYF-V2ENF-V2LF-2
Prescription Use
(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use
(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

es fa Antary Dudrim

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Page 1 of 2

510(k) Number: K110748

{7}------------------------------------------------

Worst case dimension:

The endoscope URF-P5 offers the most challenging dimensions in regards to maximum length (835 mm), minimum inner diameter (1.2 mm) and number of lumen (1).

Ethylene oxide sterilization parameters: In chamber conditioning Temperature................................................................................................................................................................... Humidity ... ... ... ......................................................................................................................................................... Vacuum set point.............................................................................................................................................................. Time ......................................................................................................................................................................... Exposure Temperature................................................................................................................................................................... Humidity ... ... ... ......................................................................................................................................................... Sterilization gas ............................................................................................................................................................ Ethylene oxide gas concentration............................................................................................................................ 735 mg/l Exposure time................................................................................................................................................................. Aeration Temperature................................................................................................................................................................... Time .........................................................................................................................................................................

Prescription Use_ (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use_ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth F. Clamin-Will

Division of Anesthesiology, General Hospital Infection Control, Dental Device

Page 2 of 2

510(k) Number: K110748

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).