K Number
K110693

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-06-08

(89 days)

Product Code
Regulation Number
870.3640
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Paceart System is intended for use as a 12-lead electrocardiograph, pacemaker artifact analyzer, and transtelephonic ECG receiving station. It also acts as a database for cardiac patients with or without pacemakers or Implantable Cardioverter Defibrillators.

Device Description

The Medtronic Paceart Optima System Software, Model POS12D, is intended for use with the Medtronic Paceart System, previously cleared per K024278. The Paceart System is a 12-lead electrocardiograph, transtelephonic receiving station, and a clinic management tool that organizes patient, device and programmer information. The Paceart System is designed to integrate data from implanted cardiac device programmers, pacemaker transmitters, cardiac event recorders, and remote data networks, such as the Medtronic CareLink® Network, into a single patient record, enabling clinicians to conduct their daily work through commonly formatted data and clinic workflow tools. Note: The Paceart System will also be marketed and referred to as the Paceart Optima System when using the Paceart Optima System Software. The Paceart System consists of hardware and software. There are no design or process changes to the hardware components of the Paceart System as part of this submission. Therefore, the subject of this 510(k) submission is for the Paceart Optima System Software only and contains only information on the software changes.

AI/ML Overview

The provided 510(k) summary for the Medtronic Paceart Optima™ System Software (K110693) does not describe acceptance criteria or a study that proves the device meets specific performance criteria in the way typically seen for AI/ML-driven medical devices.

Instead, this submission is for a software update to an existing system (Paceart System, K024278). The "Summary of Testing" section explicitly states: "Software verification testing and validation testing is performed to demonstrate the Paceart Optima System Software meets established performance criteria and to support equivalency to the referenced predicate device."

This indicates a focus on software functionality and equivalence to a predicate, rather than an independent clinical performance study with defined acceptance metrics for a specific algorithm's output.

Therefore, the following information cannot be extracted from the provided text:

  • A table of acceptance criteria and the reported device performance
  • Sample size used for the test set and the data provenance
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts
  • Adjudication method for the test set
  • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size
  • If a standalone performance study was done
  • The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
  • The sample size for the training set
  • How the ground truth for the training set was established

The submission focuses on:

  1. Software Verification and Validation: Ensuring the software changes function as intended and meet established requirements.
  2. Substantial Equivalence: Demonstrating that the updated software is equivalent in intended use, design, materials, and performance to the predicate device.

Conclusion:

The provided document describes a software update for a clinic management system, not a device with AI/ML components requiring specific performance metrics against a clinical ground truth. The "testing" mentioned refers to standard software verification and validation, aimed at ensuring the new software features work correctly and the overall system remains equivalent to its predicate. No details regarding clinical studies, ground truth establishment, or specific performance metrics are present because they are not typically required for this type of software update submission.

{0}------------------------------------------------

510(k) SUMMARY

K110693

Date Prepared:March 10, 2010
Submitter:Medtronic, Inc.Cardiac Rhythm Disease Management8200 Coral Sea Street NEMounds View, MN 55112
Contact:Rachel U. LibiSr. Principal Regulatory Affairs Specialist
Telephone:(763) 526-1668
Fax:(651) 367-0603
E-mail:rachel.libi@medtronic.com
Proprietary Name:Paceart Optima™ System Software
Common Name:Pacemaker Waveform Analyzer and DigitalElectrocardiograph
Device ClassificationClass II, 21 CFR 870.2340, 870.2920, 870.3640,Pacemaker Waveform Analyzer and DigitalElectrocardiographClass I, 21 CFR 880.6310, Medical Device DataSystem (MDDS)
Product Code:DPS, DXH, KRE and OUG

Summary of Substantial Equivalence

The intended use, design, materials and performance of the Paceart OptimaTM System Software are substantially equivalent to the Paceart® System Software, which was cleared as part of the following predicate device/system.

  • Medtronic Paceart System cleared via 510(k), K024278, on May 29, 2003 .

Device Description

The Medtronic Paceart Optima System Software, Model POS12D, is intended for use with the Medtronic Paceart System, previously cleared per K024278. The Paceart System is a 12-lead electrocardiograph, transtelephonic receiving station, and a clinic management tool that organizes patient, device and programmer information. The Paceart System is designed to integrate data from implanted cardiac device programmers, pacemaker transmitters, cardiac event recorders, and remote data networks, such as the Medtronic CareLink® Network, into a single patient record, enabling clinicians to conduct their daily work through commonly formatted data and clinic workflow tools. Note: The Paceart System will also be marketed and referred to as the Paceart Optima System when using the Paceart Optima System Software.

{1}------------------------------------------------

The Paceart System consists of hardware and software. There are no design or process changes to the hardware components of the Paceart System as part of this submission. Therefore, the subject of this 510(k) submission is for the Paceart Optima System Software only and contains only information on the software changes.

Indications for Use

The Paceart System is intended for use as a 12-lead electrocardiograph, pacemaker artifact analyzer, and transtelephonic ECG receiving station. It also acts as a database for cardiac patients with or without pacemakers or Implantable Cardioverter Defibrillators,

Technological Characteristics

Intended use, design, materials, performance and technological characteristics are substantially equivalent to the predicate devices referenced.

Summary of Testing

Software verification testing and validation testing is performed to demonstrate the Paceart Optima System Software meets established performance criteria and to support equivalency to the referenced predicate device.

Conclusion

Medtronic considers the Paceart Optima System Software to be substantially equivalent to the previously cleared Paceart System Software (reference K024278) through the data and information presented. No new safety or effectiveness issues were identified.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird-like figure. The text is in all caps and appears to be in a sans-serif font.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Medtronic, Inc. c/o Ms. Rachel U. Libi Sr. Principal Regulatory Affairs Specialist 8200 Coral Sea Street Mounds View, MN 55304

JUN - 8 2011

Re: K110693

Trade/Device Name: Paceart Optima™ System Software Regulation Number: 21 CFR 870.3640 Regulation Name: Indirect pacemaker generator function analyzer Regulatory Class: Class II (two) Product Code: KRE, DPS, DXH, and OUG Dated: May 18, 2011 Received: May 19, 2011

Dear Ms. Libi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Rachel U. Libi

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours

B.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices

Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATION FOR USE

510(k) Number (if known): Kl 10693

Device Name: Paceart Optima™ System Software

Indications for Use: The Paceart System is intended for use as a 12-lead electrocardiograph, pacemaker artifact analyzer, and transtelephonic ECG receiving station. It also acts as a database for cardiac patients with or without pacemakers or Implantable Cardioverter Defibrillators.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

ision Sign-Off) vision of Cardiovascular Devices

510(k) Number K110643

§ 870.3640 Indirect pacemaker generator function analyzer.

(a)
Identification. An indirect pacemaker generator function analyzer is an electrically powered device that is used to determine pacemaker function or pacemaker battery function by periodically monitoring an implanted pacemaker's pulse rate and pulse width. The device is noninvasive, and it detects pacemaker pulse rate and width via external electrodes in contact with the patient's skin.(b)
Classification. Class II (performance standards).