K Number
K110684

Validate with FDA (Live)

Date Cleared
2011-06-24

(105 days)

Product Code
Regulation Number
882.5330
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MedCAD AccuShape™ PEEK Patient Specific Cranial Implant is designed individually for each patient and intended to correct defects / replace bony voids in the cranial skeleton.

Device Description

The MedCAD AccuShape™ PEEK Patient Specific Cranial Implant devices are individually sized and shaped implantable prosthetic cranioplasty plates intended to correct defects / replace voids in the cranial skeleton of a specific patient. The implants are designed using the patient's CT imaging data and precision manufactured from implantable grade polyether-ether-ketone (PEEK) material. The devices have a nominal thickness of 3mm, ranging from 2-5mm depending on the anatomical location. The device can be supplied as one or as multiple parts due to material constraints and/or the complexity of the device, with each part ranging in size from 10 x 10 (mm) to 200 x 200 (mm). The implants are provided with 2mm drainage holes spaced 10 mm apart from center to center with an edge margin of 10 mm. The devices are non-pyrogenic and are provided non-sterile for sterilization by the physician prior to implantation. The implants are attached to the native bone with commercially available cranioplasty fasteners.

AI/ML Overview

This document describes the MedCAD AccuShape™ PEEK Patient Specific Cranial Implant, a device intended to correct defects or replace bony voids in the cranial skeleton. It is a Class II device (21 CFR 882.5330).

1. Table of Acceptance Criteria and Reported Device Performance

The provided text describes a 510(k) summary, which focuses on establishing substantial equivalence to predicate devices rather than setting specific performance acceptance criteria like diagnostic accuracy or sensitivity/specificity for an AI algorithm. In this context, "acceptance criteria" are related to demonstrating that the new device shares the same fundamental characteristics and safety/effectiveness profile as the legally marketed predicate devices.

Acceptance Criteria (Related to Substantial Equivalence)Reported Device Performance (MedCAD AccuShape™ PEEK Patient Specific Cranial Implant)
Technological Characteristics: Use of electronic CT images and CAD for implant design.Uses electronic CT imaging data and computer aided design (CAD) to determine patient-specific implant dimensions.
Material: Use of implantable grade polymer.Precision manufactured from implantable grade polyether-ether-ketone (PEEK) material. Emphasized as a key similarity to predicate devices.
Resulting Technological Characteristics: Biocompatibility, sterilization method, strength, stiffness, elasticity, density, radiolucency.Substantially equivalent to predicate devices regarding these characteristics. Biological laboratory tests (including thorough sterilization validation) confirm pyrogen-free and sterile status. Precision measurements validate dimensional accuracy and stability. Material testing demonstrates structural integrity.
Safety and Effectiveness: Demonstration of safety and effectiveness comparable to predicate devices.Non-clinical tests (biological, sterilization, precision measurements, material testing) assure that the device's safety and effectiveness are substantially equivalent to those of the predicate devices. The FDA's 510(k) clearance itself signifies that the device is deemed substantially equivalent for the stated indications for use.
Primary Intended Use: Same as predicate devices.Intended to correct defects / replace bony voids in the cranial skeleton, same as predicate devices.
Principles of Operation: Same as predicate devices.The 510(k) summary states that the conclusion of substantial equivalence is based on similarities in "principles of operation."
Functional Design: Same as predicate devices.The 510(k) summary states that the conclusion of substantial equivalence is based on similarities in "functional design."
Established Medical Use: Same as predicate devices.The 510(k) summary states that the conclusion of substantial equivalence is based on similarities in "established medical use."

2. Sample Size Used for the Test Set and Data Provenance

This is not an AI/CADe type device that uses a "test set" in the conventional sense of evaluating an algorithm's performance on a dataset of patient images. Instead, this device is a patient-specific cranial implant. The "testing" involves demonstrating the physical and biological characteristics of the material and the manufacturing process, and confirming that the final product adheres to design specifications comparable to predicate devices.

  • Sample Size for Non-Clinical Tests: Not explicitly stated as a number of patient cases. The description mentions "Biological laboratory tests including thorough sterilization validation have been conducted" and "Precision measurements have validated the dimensional accuracy and stability of the devices. In addition, material testing has been performed to demonstrate structural integrity." These are likely tests on material samples and manufactured devices, rather than a patient image test set.
  • Data Provenance: Not applicable in the context of patient image data. The "data" here would be laboratory test results and manufacturing quality control data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This is not applicable as there is no "test set" of patient cases for which an AI algorithm is generating a result that needs comparison to an expert-established ground truth. The device itself is the physical implant, designed and manufactured based on a patient's CT data. The "ground truth" for this type of device relates to engineering specifications, material properties, and biological compatibility.

4. Adjudication Method for the Test Set

Not applicable for a physical implant device where performance is established through non-clinical testing of materials and manufacturing processes, rather than an AI algorithm's output on a diagnostic test set.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an implantable medical device, not an AI-assisted diagnostic or assistive tool for human readers interpreting medical images.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI algorithm. Its design uses CAD (Computer-Aided Design) based on patient CT data, which is a common engineering process for custom implants, not an autonomous AI system delivering a diagnostic output.

7. The Type of Ground Truth Used

The "ground truth" for this device is based on engineering specifications, material science standards, and biological testing standards. Specifically:

  • Confirmation of manufacturing precision and dimensional accuracy against design specifications.
  • Validation of material properties (strength, stiffness, elasticity, density) against established PEEK material standards and predicate device characteristics.
  • Demonstration of biocompatibility and sterility against recognized biological testing standards.

8. The Sample Size for the Training Set

Not applicable. This device does not involve machine learning or an AI algorithm that requires a "training set." The design process uses a patient's individual CT data, not a pre-trained model.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Med CAD

5. 510(k) SUMMARY

Submitter:MedCAD824 Exposition Avenue, Suite 3Dallas, Texas 75226
Contact Person:Nancy HairstonPresidentTEL: 214-324-0030 extension 300FAX: 888-264-0726nancy@medicalCAD.com
Date Prepared:March 09, 2011
Trade Name:MedCAD AccuShape™ PEEK Patient Specific Cranial Implant (PSCI)
Common Name:Patient Specific Cranial Implant (PSCI)
Classification Name:Preformed nonalterable cranioplasty plate
Product Code:GXN
Classification:Class II, 21 CFR 882.5330
Predicate Device:K033868 - Synthes Patient Specific Cranial/Craniofacial ImplantsK072601 - OsteoSymbionics Patient-Specific Cranial Implant
Device Description:The MedCAD AccuShape™ PEEK Patient Specific Cranial Implant devices areindividually sized and shaped implantable prosthetic cranioplasty plates intended tocorrect defects / replace voids in the cranial skeleton of a specific patient. Theimplants are designed using the patient's CT imaging data and precisionmanufactured from implantable grade polyether-ether-ketone (PEEK) material.The devices have a nominal thickness of 3mm, ranging from 2-5mm depending onthe anatomical location. The device can be supplied as one or as multiple parts dueto material constraints and/or the complexity of the device, with each part rangingin size from 10 x 10 (mm) to 200 x 200 (mm). The implants are provided with 2mmdrainage holes spaced 10 mm apart from center to center with an edge margin of 10mm.The devices are non-pyrogenic and are provided non-sterile for sterilization by thephysician prior to implantation. The implants are attached to the native bone withcommercially available cranioplasty fasteners.
Statement ofIntended Use:The MedCAD AccuShape™ PEEK Patient Specific Cranial Implant is designedindividually for each patient and intended to correct defects / replace bony voids inthe cranial skeleton.

824 Exposition Ave. Suite 3, Dallas, TX 75226 Phone: 214.453.8864 Fax: 888.264.0726

{1}------------------------------------------------

Med CAD

The MedCAD AccuShape™ PEEK Patient Specific Cranial Implant device is Summary of Technological substantially equivalent to the predicate devices regarding use of electronic CT Characteristics: images and computer aided design in determining patient specific implant dimensions, use of implantable grade polymer as the device material, and resulting technological characteristics including biocompatibility, sterilization method, strength, stiffness, elasticity, density, and radiolucency. Summary of Biological laboratory tests including thorough sterilization validation have been Non-Clinical Test conducted for the MedCAD Patient Specific Cranial Implant devices and material. Data: These tests have proven the devices to be pyrogen-free and sterile, when used in accordance with the product instructions. Precision measurements have validated the dimensional accuracy and stability of the devices. In addition, material testing has been performed to demonstrate structural integrity. Together, these non-clinical tests assure that the MedCAD AccuShape™ PEEK Patient Specific Cranial Implant safety and effectiveness are substantially equivalent to those of the predicate devices. Conclusion: MedCAD considers the MedCAD AccuShape™ PEEK Patient Specific Cranial Implant to be substantially equivalent to the predicate devices listed above. This conclusion is based on the similarities in primary intended use, principles of operation, functional design, materials, test results, and established medical use.

824 Exposition Ave. Suite 3, Dallas, TX 75226 Phone: 214.453.8864 Fax: 888.264.0726

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three stylized lines or shapes, possibly representing a stylized bird or a similar design element.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-00002

VanDuzen dba MedCAD c/o Ms. Diane Rutherford Regulatory Engineer Ken Block Consulting 1201 Richardson, Suite 280 Richardson, TX 75080

JUN 2 4 2011

Re: K110684

Trade/Device Name: MedCAD AccuShape™ PEEK Patient Specific Cranial Implant Regulation Number: 21 CFR 882.5330 Regulation Name: Preformed nonalterable cranioplasty plate Regulatory Class: Class II Product Code: GXN Dated: May 31, 2011 Received: June 01, 2011

Dear Ms. Rutherford:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

Page 2 - Ms. Diane Rutherford

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.html 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket nouification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/deliault.html for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

http://www.fda.gov/MedicalDevices/ResourcesforYour/Industry/default.htm.

Sincerely yours,
Kesia Alexander

Image /page/3/Picture/7 description: The image shows the name and title of Malvina B. Eydelman, M.D. She is the Director of the Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices. The image also mentions the Office of Device Evaluation and the Center for Devices and Radiological Health.

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

510(k) Number: K110684

Device Name: MedCAD AccuShape™ PEEK Patient Specific Cranial Implant

Indications for Use:

The MedCAD AccuShape™ PEEK Patient Specific Cranial Implant is designed individually for each patient and intended to correct defects / replace bony voids in the cranial skeleton.

Prescription Use _ X (21 CFR 801 Subpart D)

AND/OR

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDHR, Office of Device Evaluation (ODE)

Division Sign-Off

Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K110684

§ 882.5330 Preformed nonalterable cranioplasty plate.

(a)
Identification. A preformed nonalterable cranioplasty plate is a device that is implanted in a patient to repair a skull defect and is constructed of a material, e.g., stainless steel or vitallium, that cannot be altered or reshaped at the time of surgery without changing the chemical behavior of the material.(b)
Classification. Class II (performance standards).