K Number
K110656

Validate with FDA (Live)

Date Cleared
2011-08-18

(164 days)

Product Code
Regulation Number
880.5570
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is intended for use in venous blood collection.

Device Description

This device is intended for use in venous blood collection.

IMC blood collection needles are manufactured from stainless steel sharpened at both ends that is attached to the hub. The hub is threaded on one side to connect with the needle holder which is used to guide the needle into an evacuated blood collection tube. This end of the needle is shorter end and is fitted with a protective rubber sleeve. And a needle cap. The opposite end of the needle is 1" or 1 ½" for withdrawing blood. The two needle caps protect the needle and maintain the sterility. The seal between the two needle caps is covered with a perforated paper label that simplifies identification and acts as a seal of integrity.

IMC blood collection needles are a sterile single use disposable product. The needles are nontoxic and non-pyrogenic and are available in a variety of combinations of needle sizes (20 to 22 gauge) and needle lengths (1" and 1 ½').

AI/ML Overview

The document describes the acceptance criteria and performance of the IMC Blood Collection Needle, comparing it to the predicate device, VACUETTE VISIO PLUS blood collection needle (K061483).

1. Table of Acceptance Criteria and Reported Device Performance:

Performance CharacteristicsAcceptance CriteriaIMC Blood Collection Needle Performance
Hub/needle bond strength20G: >54NMeets Standard Criteria
21G: >44NMeets Standard Criteria
22G: >40NMeets Standard Criteria

2. Sample Size Used for the Test Set and Data Provenance:

The document does not explicitly state the specific sample size used for the hub/needle bond strength test. It states "Meets Standard Criteria," implying that an adequate number of samples were tested to demonstrate compliance with the ISO 7894:1993 standard for each gauge. The data provenance is not specified, but given the submission for FDA approval, it is presumably from testing conducted by or for the manufacturer, International Medsurg Connection. The study is a prospective performance test designed to meet pre-defined criteria.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

This information is not applicable to the type of device and test described. The "ground truth" for hub/needle bond strength is established by objective physical measurement against a specified international standard (ISO 7894:1993), not by expert opinion or interpretation.

4. Adjudication Method for the Test Set:

Not applicable. The test involves direct physical measurement and comparison to quantitative criteria, not subjective assessment requiring adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done:

No, an MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic devices that involve human interpretation of images or other data. The IMC Blood Collection Needle is a physical medical device, and its performance is evaluated through engineering and mechanical testing.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

Not applicable. This device does not involve algorithms or human-in-the-loop performance in the context of diagnostic interpretation. The performance reported is that of the device itself.

7. The Type of Ground Truth Used:

The ground truth used is based on objective measurement against an international standard (ISO 7894:1993). The hub/needle bond strength is a quantifiable physical property, and the acceptance criteria are numerical thresholds defined by this standard.

8. The Sample Size for the Training Set:

Not applicable. There is no "training set" in the context of this device. The testing described is a performance validation test, not a machine learning model.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

AUG 1 8 2011

K 110656

8. 510(K) SUMMARY

Trade Name:IMC Blood Collection Needle (non-sterile and sterile)
Common Name:Blood Collection Needle
Classification Name:21 CFR 880.5570
Submitter Information:International Medsurg Connection935 N Plum Grove Rd, STE FSchaumburg, Illinois 60173
Summary Prepared By:Peter KimDirector of Quality AssuranceInternational Medsurg Connection935 N Plum Grove Rd, STE FSchaumburg, Illinois 60173Telephone: 847-619-9926Fax: 847-619-9927e-mail: peterkim@intlmedsurg.com
Date Prepared:December 22, 2010
Predicate Devices:• VACUETTE VISIO PLUS blood collection needle (K061483)

Device Name(s):

IMC Blood Collection Needle (non-sterile and sterile)

Classification Panel:

General Hospital

Legally Marketed Device Under Which Substantial Equivalence is Being Claimed:

International Medsurg Connections, Inc is claiming substantial equivalence of the IMC Blood Collection Needle with the currently marketed:

Description510(k) Number
VACUETTE VISIO PLUS Blood Collection NeedleK061483

Device Description

This device is intended for use in venous blood collection.

IMC blood collection needles are manufactured from stainless steel sharpened at both ends that is attached to the hub. The hub is threaded on one side to connect with the needle holder which is used to guide the needle into an evacuated blood collection tube. This end of the needle is shorter end and is fitted with a protective rubber sleeve. And a needle cap. The opposite end of the needle is 1" or 1 ½" for withdrawing blood. The two needle caps protect the needle and maintain the sterility. The seal between the two needle caps is covered with a perforated paper label that simplifies identification and acts as a seal of integrity.

{1}------------------------------------------------

IMC blood collection needles are a sterile single use disposable product. The needles are nontoxic and non-pyrogenic and are available in a variety of combinations of needle sizes (20 to 22 gauge) and needle lengths (1" and 1 ½').

Statement of Intended Use

Indications For Use: This device is intended for use in venous blood collection

Name/Description

. .. .

DescriptionSizeSterility
20 Gauge Blood collection needle1"and 1 ½"Sterile & Non-sterile
21 Gauge Blood collection needle1"and 1 ½"Sterile & Non-sterile
22 Gauge Blood collection needle1"and 1 ½"Sterile & Non-sterile

New Devices as Compared to Marketed Device(s)

The IMC Blood Collection Needle and the predicate device are intended to be used to inject fluids into or withdraw fluids from parts of the body below the surface of the skin.

Feature/CharacteristicIMC Blood Collection NeedleVACUETTE VISIO PLUS Blood Collection Needle: K061483 (Predicate)
Intended UseThis device is intended for use in venous blood collection.This device is intended for use in venous blood collection.
MaterialNeedle capHDPESimilar
Resistance coverRubber (Synthetic)Similar
HubABSSimilar
Label30g/m2 paperSimilar
Needle (Cannula)SUS 304Similar
Needle capPolypropyleneSimilar
colors
20GYellow colorYellow color
21GDeep green colorGreen color
22GBlack colorBlack color
Length1"and 1 ½"1"and 1 ½"
Gauge20G, 21G & 22G21G & 22G
Cover dimension1" size: 47mm1 ½" size: 47mm1" size: 47mm1 ½" size: 47mm
Cover colorWhite (for all gauges)Clear (for all gauges)

{2}------------------------------------------------

( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) (IMC Blood Callection Naeds.The state of the states of the states and the state of the state of the collection of the14 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1a m close and estate antenant consequent of the program and any aA CIUCALLE VICallery Contractor Complete Come1. 2 . 3 . 3 . 3 . 3 . 3 . 3.

Performance Data:

Performance CharacteristicsTest MethodAcceptanceCriteriaIMC BloodCollection NeedleVACUETTE VISIONPLUS Blood CollectionNeedleK061483
Hub/needle bond strengthISO 7894 : 199320G : >54N21G: >44N22G: >40NMeets StandardCriteriaMeets Standard Criteria

Conclusions:

The indications for use, technology, specification, safety of the IMC Blood Collection Needle and the predicate devices K061483 are essentially the same. The differences between the Blood Collection Needle are minor and do not raise new issues of safety or effectiveness. Hence, the IMC Blood Collection Needles are substantially equivalent to the predicate device.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a black and white logo. The logo consists of a symbol on the right and text on the left. The symbol appears to be three curved lines stacked on top of each other. The text reads "DEPARTMENT OF HEALTH & H.", arranged vertically along the left side of the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Peter Kim Director of Quality Assurance International Medsurg Connection 935 North Plum Grove Road, Suite F Schaumburg, Illinois 60173

AUG 1 8 2011

Re: K110656

Trade/Device Name: IMC Blood Collection Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic single lumen needle Regulatory Class: II Product Code: FMI Dated: August 3, 2011 Received: August 8, 2011

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2- Mr. Kim

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), 1 please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony La Rue

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number :

Device Name: IMC Blood collection needle

Indications For Use: This device is intended for use in venous blood collection.

Name/Description

GaugeSizeSterility
20 Gauge Blood Collection needle1" and 1 ½"Sterile & Non-sterile
21 Gauge Blood Collection needle1" and 1 ½"Sterile & Non-sterile
22 Gauge Blood Collection needle1" and 1 ½"Sterile & Non-sterile

X Prescription Use __ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Ri C. 8/18/4

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

110656 510(k) Number: _______________________________________________________________________________________________________________________________________________________________

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).