K Number
K110620

Validate with FDA (Live)

Date Cleared
2011-05-03

(61 days)

Product Code
Regulation Number
866.2660
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Premier C. difficile GDH is a qualitative enzyme immunoassay screening test to detect Clostridium difficile antigen, glutamate dehydrogenase, in fecal specimens from symptomatic persons suspected of having C. difficile infection (CDI). This test does not distinguish between toxigenic strains of C. difficile. Samples from symptomatic patients that produce positive results with this test must be further tested with an assay designed to detect toxigenic C. difficile strains and assist with the diagnosis of CDI.

Device Description

Premier C. difficile GDH is a qualitative enzyme immunoassay screening test to detect Clostridium difficile antigen, glutamate dehydrogenase, in fecal specimens from symptomatic persons suspected of having C. difficile infection (CDI). The assay consists of Premier C. difficile GDH Microwells coated with polyclonal antibodies specific to C. difficile GDH, Premier C. difficile GDH Enzyme Conjugate, Premier 20X Wash Buffer II, Premier Substrate I, Premier Stop Solution I, Premier C. difficile GDH Sample Diluent/Negative Control, and Premier C. difficile GDH Positive Control.

AI/ML Overview

Here's a summary of the acceptance criteria and study findings for the Premier C. difficile GDH assay, based on the provided text:

Acceptance Criteria and Device Performance

The document implicitly defines acceptance criteria through the reported performance characteristics. The primary measure of clinical performance is the comparison to bacterial C. difficile culture.

Table 1: Acceptance Criteria and Reported Device Performance (Clinical Study)

Performance MetricAcceptance Criteria (Implicit)Reported Device PerformanceComments
Clinical SensitivityHigh % (e.g., above 85-90%)92.3% (95% Cl: 86.0 - 95.9%)Met the expectation for a high sensitivity screening test.
Clinical SpecificityHigh % (e.g., above 90-95%)95.8% (95% Cl: 93.9 – 97.1%)Met the expectation for a high specificity screening test.
Overall CorrelationHigh % (e.g., above 90%)95.2% (93.4 - 96.5%)Good overall agreement with the reference method.
Analytical Sensitivity (LoD)Defined as detectable at a specific concentration with 95% probability8 ng/mL (based on 45 replicates per measurand with 95% probability of positive response)Clear analytical limit of detection established.
Interference TestingNo interference at specified concentrations of common substancesNo interference observed for listed substances (e.g., Barium sulfate, Metronidazole, Vancomycin HCl)Demonstrated robustness against common interfering substances.
Cross-ReactivityNo cross-reactivity with common microorganisms, or identified and notedNo cross-reactivity observed with a wide range of bacteria and viruses, except for Staphylococcus aureus (Cowan strain I) and Clostridium sporogenes.Most common pathogens did not cross-react, but two specific Clostridium species and Staphylococcus aureus were noted as cross-reactive.
Strain ReactivityPositive reactions with a representative panel of C. difficile strainsPositive reactions at 5.7 x 10^7 cells/mL with 30+ strainsDemonstrated ability to detect various C. difficile strains.
Reproducibility100% agreement for moderate positive, high negative, and negative samples100% agreement over 5 non-consecutive days, across 3 sites, 2 operators per site.Excellent reproducibility across different sites and operators.

Study Information

  1. Sample sizes used for the test set and the data provenance:

    • Test Set Sample Size: 733 qualified patient samples.
    • Data Provenance: The data was collected prospectively (clinical trials conducted from November 2011) from independent clinical test sites located in the Midwestern and Southwestern regions of the United States. Gender and age ranges were reported (22 days to 99 years).
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document states that the performance characteristics were determined by comparison to bacterial C. difficile culture. It does not mention the use of human experts to establish ground truth for the clinical test set; rather, the gold standard for diagnosis was a laboratory method (bacterial culture).
  3. Adjudication method for the test set:

    • Since the ground truth was established by bacterial C. difficile culture, an expert adjudication method (like 2+1, 3+1) was not described or necessary. The comparison was directly against the culture results.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was done. This device is an Enzyme Immunoassay (ELISA) kit, which is a laboratory test where results are read spectrophotometrically or visually from a microplate, not an imaging device requiring human interpretation of complex visual data. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Yes, the clinical performance described (sensitivity, specificity, correlation) represents the standalone performance of the Premier C. difficile GDH assay. It is an "algorithm only" in the sense that it is a biochemical assay designed to yield a direct result (positive/negative) based on antigen detection, without human interpretation influencing the diagnostic outcome beyond standard laboratory procedures (e.g., pipetting, reading the spectrophotometer).
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth used for the clinical performance evaluation was bacterial C. difficile culture.
  7. The sample size for the training set:

    • The document does not explicitly state a "training set" in the context of machine learning or AI models. For an ELISA kit, development typically involves analytical studies (sensitivity, specificity, interference, cross-reactivity, strain reactivity) and then clinical validation. The "analytical sensitivity" study used 45 replicates for each measurand. The "reproducibility" panels involved blind-coded samples tested multiple times. These studies contribute to the device's development and validation but are not a "training set" in the common AI sense.
  8. How the ground truth for the training set was established:

    • Again, the concept of a "training set" in the AI sense is not directly applicable. For the analytical studies, the "ground truth" (e.g., known concentration of C. difficile GDH antigen, presence/absence of interfering substances, known microorganisms) was established through controlled laboratory spiking and preparation of contrived samples. For the broader validation, bacterial C. difficile culture served as the reference standard.

{0}------------------------------------------------

K110620

MAY - 3 2011

510(k) Application Premier C. difficile GDH
---------------------------------------------
Description:510(k) Summary Premier C. difficile GDH
Identification:Attachment 002
Date:March 01, 2011
510(k) number:
Date of Preparation:March 01, 2011
Submitter:Meridian Bioscience, Inc
Submitter's address:3471 River Hills Drive
Cincinnati, Ohio 45244
Contact:Susan Bogar
Contact number:(513) 271-3700
Device name:Premier C. difficile GDH
Common name:Enzyme Immunoassay for C. difficile Common Antigen
Classification:Antigen, C. difficile
MCB, CFR Section 866.2660
Predicate device:K030992: TECHLAB C. DIFF CHEK™ - 60 ELISA
Reference comparator:Bacterial culture

Description of the device:

Premier C. difficile GDH is a qualitative enzyme immunoassay screening test to detect Clostridium difficile antigen, glutamate dehydrogenase, in fecal specimens from symptomatic persons suspected of having C. difficile infection (CDI). The assay consists of Premier C. difficile GDH Microwells coated with polyclonal antibodies specific to C. difficile GDH, Premier C. difficile GDH Enzyme Conjugate, Premier 20X Wash Buffer II, Premier Substrate I, Premier Stop Solution I, Premier C. difficile GDH Sample Diluent/Negative Control, and Premier C. difficile GDH Positive Control. ·

Intended Use:

Premier C. difficile GDH is a qualitative enzyme immunoassay screening test to detect Clostridium difficile antigen, glutamate dehydrogenase, in fecal specimens from symptomatic persons suspected of having C. difficile infection (CDI). This test does not distinguish between toxigenic strains of C. difficile. Samples from symptomatic patients that produce positive results with this test must be further tested with an assay designed to detect toxigenic C. difficile strains and assist with the diagnosis of CDI.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo consists of a globe graphic on the left, followed by the word "Meridian" in a bold, sans-serif font. Below "Meridian" is the text "Bioscience, Inc." in a smaller, also bolded, sans-serif font. A horizontal line separates the two lines of text.

510(k) Application Premier C. difficile GDH 510(k) Summary Premier C. difficile GDH Description: Identification: Attachment 002 Date: March 01, 2011

Table 1: Comparison to predicate device.

·

CharacteristicPremier C. difficile GDHTECHLAB C. DIFF CHEK-60
Test FormatELISA MicroplateELISA Microplate
Intended Use
Qualitative/QuantitativeQualitativeQualitative
Target AntigenClostridium difficile glutamatedehydrogenaseClostridium difficile glutamatedehydrogenase
Screening, Diagnostic, orIdentification TestScreeningScreening
Specimen Types
Human Stool UnpreservedYesYes
Reagents/ComponentsPremier C. difficile GDH MicrowellsPremier C. difficile GDH EnzymeConjugatePremier 20X Wash Buffer IIPremier Substrate IPremier Stop Solution IPremier C. difficile GDH SampleDiluent/Negative ControlPremier C. difficile GDH Positive ControlMicroassay plateEnzyme Conjugate20X Wash BufferSubstrateStop SolutionDiluentPositive Control
Diagnostic Marker
AntibodyYesYes
Antibody Sources
Solid phase (microplate)Rabbit polyclonalPolyclonal
Enzyme ConjugateRabbit polyclonalMouse monoclonal
Sample Preparation
Unpreserved liquid/semi-solidstool50 µL of thoroughly mixed stool into200 µL Sample Diluent.Vortex for 15 seconds.50 µL specimen into 200 µL Diluent.Vortex for 10 seconds.
Testing TimeApproximately 60 minutesApproximately 60 minutes
Equipment
General Laboratory EquipmentSpectrophotometerStatFax™- 2200 Incubator/Shaker(optional)Semiautomated Microplate Washer(optional)Spectrophotometer
Reading MethodSpectrophotometricVisualSpectrophotometric
Results Interpretation
Single Wavelength ( A450 nm)Negative: < 0.200Positive: ≥ 0.200Negative : < 0.120Positive: ≥ 0.120
Dual WavelengthNegative: < 0.150Positive: ≥ 0.150A450/630 nmNegative: < 0.080Positive: ≥ 0.080A450/620 nm
Visual ReadN/APositive: Any yellow colorNegative: Colorless

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo consists of a globe graphic on the left, followed by the word "Meridian" in a large, bold font. Below "Meridian" is the text "Bioscience, Inc." in a smaller, bold font. A horizontal line separates "Meridian" from "Bioscience, Inc."

510(k) Application Premier C. difficile GDH 510(k) Summary Premier C. difficile GDH Description: Identification: Attachment 002 Date: March 01, 2011

Performance Comparison, Non-clinical Tests

Analytical Sensitivity

Sensitivity studies were designed to determine within 95% confidence intervals the analytical limit of detection (LoD) of C. difficile GDH antigen diluted in a human stool matrix. The analytical sensitivity of this assay was based on 45 replicates for each measurand and with a stated probability (95%) of obtaining positive responses at the following levels of the measurand when spiked in stool: 8 ng/mL.

Interference Testing

Selected drugs and other non-microbial substances that might be present in stool samples from healthy persons or patients suspected of having C. difficile associated disease were added to three negative stool samples and three positive stool samples. The contrived positive specimens were prepared from a pool of donor stools that were confirmed as negative. The samples were inoculated with C. difficile strain 11186 at 8 ng/mL, the limit of detection for this assay. Potentially interfering substances were added at final concentrations of 5% V/V or greater. Dilution Controls for each sample were prepared by adding a phosphate-buffered saline solution in place of the potentially interfering substance. Each sample was tested in triplicate.

The following substances, at the specified saturated solvent/diluent concentrations, do not interfere with Premier C. difficile GDH test results in the final concentrations listed: Barium sulfate (5 mg/mL), Fecal fat (2.65 mg stearic acid and 1.3 mg palmitic acid/mL), Hemoglobin (3.2 mg/mL), Imodium AD (Loperamide HCl) (6.67 x 10ª mg/mL), Kaopectate® (Bismuth subsalicylate) (0.87 mg/ml), Metronidazole (12.5 mg/mL), Mucin (3.33 mg/mL), Mylanta® (Aluminum hydroxide w/ magnesium hydroxide) (4.2 mg/mL), Pepto-Bismoth subsalicylate) (0.87 mg/mL), Polyethylene glycol (79.05 mg/mL), Prilosec® (Omeprazol) (0.5 mg/mL), Simethicone (0.625 mg/mL), Tagamet® (Cimetidine) (0.5 mg/mL). Tums® (Calcium carbonate) (0.5 mg/mL), Vancomycin HCl (2.5 mg/mL), Whole blood (25%). White blood cells (5%).

Cross-reactivity Study

Potentially crossreactive microorganisms that might be present in stool samples from healthy persons or patients suspected of having C. difficile associated disease were added to a natural negative and contrived positive sample. The contrived positive specimens were prepared from a pool of donor stools that was confirmed negative. The contrived positive sample was prepared by spiking a confirmed negative sample with C. difficile strain 11186 at 8 ng/mL the limit of detection for this assay. Potentially cross-reactive microorganisms were added at final concentrations of 1.2 x 10 °CFU/mL (bacteria or fungi) or final concentrations greater than 1 x 10 °TC/Dg/mL (viruses). Dilution controls for each sample were prepared by adding a 0.85% saline solution in place of the potentially cross-reactive organisms.

The following microorganisms, at the indicated concentrations, do not interfere with Premier C. difficile GDH test results: Aeromonas hydrophila, Bacillus cereus, Bacillus subtilis, Bacteroides fragilis, Campylobacter coli, Campylobacter fetus, Campylobacter jejuni, Candido albicans, Citrobacter freundii, Clostridium butyricum, Clostridium bifermentans, Clostridium histolyticum novyi, Clostridium perfringens, Clostridium septicum, Clostridium sordellii, Clostridium tetani, Enterobacter cloocae, Enterobacter cloocae, Enterococus faecalis, Escherichia coli, Escherichia coli 0157:H7, Escherichia fergusonii, Helicobacter pylori, Klebsiella pneumoniae, Lactococus lactis, Listeria monocytogenes, Peptostreptococcus anaerobius, Plesiomonas shiqelloides, Porphyromonas asaccharolytica, Proteus vulgaris, Pseudomonas aeruginosa, Pseudomonas fluorescens, Salmonella Group B, Salmonella Group C, Salmonella Group E, Serratia liquifaciens, Serratia marcescens, Shigella boydii, Shigella sonnei, Staphylococcus aureus, Staphylococcus epidermidis,

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo features a stylized globe on the left, followed by the word "Meridian" in a bold, italicized font. Below "Meridian" is the text "Bioscience, Inc." in a smaller, non-italicized font. A horizontal line separates "Meridian" from "Bioscience, Inc."

510(k) Application Premier C. difficile GDH
Description:510(k) Summary Premier C. difficile GDH
Identification:Attachment 002
Date:March 01, 2011

Vibrio parahaemolyticus, Yersinio enterocolitica, Adenovirus Type 41, Coxsackievirus Strain B4, Echovirus Strain 11, Rotavirus Strain WA.

Stools spiked with Staphylococus aureus (Cowan strain I) and Clostridium sporogenes were found to be crossreactive with Premier C. difficile GDH.

Strain Reactivity

C. difficile stock cultures from different sources were tested and produced positive reactions at a concentration of 5.7 x 10 ' cells/mL with the Premier C. difficile GDH assay. Strains tested were as follows: 8864, 10463, 43598, 2004052, 2004111, 2004118, 2004205, 2004206, 2005070, 200525, 2006240, 2007431, 2007435, 2007858, 2008016, 2008029, 2008188, 2008341, 2008351, 2009018, 2009065, 2009099, 2009132, 2009277, 11186, B1, 818, B17, BK6, CF1, G1, J7, K12, Y1, 234, 586, 611, 620, 2C62, 2C165, C122, UNC 19904, X15076.

Performance Comparison, Clinical Tests

Clinical trials for the Premier C. difficile GDH assay were conducted from November 2011. Performance characteristics of the Premier C. difficile assay were determined by comparison to bacterial C. difficile culture. Independent clinical test sites located in the Midwestern and Southwestern regions of the United States evaluated a total of 733 qualified patient samples were collected from 337 (46.0%) males and 390 (53.2%) females. Gender was not defined for 6 (1%) patients. The age groups of patients range from 22 days to 99 years. No differences in test performance were observed based on patient age, gender, or geographic location. Overall sensitivity was determined to be 92.3% (95% Cl: 86.0 - 95.9%). Overall specificity was determined to be 95.8% (95% Cl: 93.9 – 97.1%). Subsequent tables show overall assay performance as well as performance by clinical site and patient age.

Premier C. difficile GDH
CulturePositiveNegativeTotal
Positive1089117
Negative26590616
Total134599733
95% CI
Sensitivity108/11792.3%86.0 - 95.9%
Specificity590/61695.8%93.9 - 97.1%
Correlation698/73395.2%93.4 - 96.5%
Table 2: Performance Characteristics for Premier C. difficile GDH

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo consists of a globe on the left, followed by the word "Meridian" in a bold, sans-serif font. Below "Meridian" is the text "Bioscience, Inc." in a smaller, sans-serif font. There is a line separating the two lines of text.

510(k) Application Premier C. difficile GDH
Description:510(k) Summary Premier C. difficile GDH
Identification:Attachment 002
Date:March 01, 2011

Table 3: Performance characteristics by site.

PremierC. difficileGDHPositive SamplesNegative Samples
Clinical SitePremierGDH/CultureSensitivity %95% CIPremierGDH/CultureSpecificity %95% CI
Site 113/1681.3%57.0 - 93.4%84/8796.6%90.3 – 98.8%
Site 228/3093.3%78.7 - 98.2%132/14094.3%89.1 - 97.1%
Site 344/4695.7%85.5 - 98.8%147/15396.1%91.7 – 98.2%
Site 415/15100.0%79.6 - 100%169/17596.6%92.7 - 98.4%
Site 58/1080.0%49.0 - 94.3%58/6195.1%86.5 - 98.3%

Table 4: Performance Data (patients less than 2 years of age)

CulturePremier C. difficile GDH
PositiveNegativeTotal
Positive21223
Negative35457
Total245680
95% CI
Sensitivity21/2391.3%73.2-97.6%
Specificity54/5794.7%85.6-98.2%
Correlation75/8093.8%86.2-97.3%

Table 5: Performance Data (patients 2 years of age and greater)

Premier C. difficile GDH
CulturePositiveNegativeTotal
Positive87794
Negative23536559
Total110543653
95% CI
Sensitivity87/9492.6%85.4 - 96.3%
Specificity536/55995.9%93.9 - 97.2%
Correlation623/65395.4%93.5 - 96.8%

·

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo features a globe graphic on the left, followed by the word "Meridian" in a bold, italicized font. Below "Meridian" is the text "Bioscience, Inc." in a smaller, also bold and italicized font. A horizontal line separates the two lines of text.

510(k) Application Premier C. difficile GDH
Description:510(k) Summary Premier C. difficile GDH
Identification:Attachment 002
Date:March 01, 2011

Table 6: Overall results by patient age.

PremierC. difficileGDHPositive SamplesNegative Samples
Patient AgePremierGDH/CultureSensitivity %95% CIPremierGDH/CultureSpecificity %95% CI
<2 years21/2391.3%73.2 - 97.6%54/5794.7%85.6 - 98.2%
≥2 years to 12years27/2896.4%82.3 - 99.4%118/12296.7%91.9 - 98.7%
>12 years to 21years10/1376.9%49.7 - 91.8%72/7596.0%88.9 - 98.6%
>21 years50/5394.3%84.6 - 98.1%345/36195.6%92.9 - 97.3%
Not Defined0/0N/AN/A1/1100.0%20.7 - 100.0%

Reproducibility

Reproducibility panels were performed by three clinical laboratories using blind coded panels. Samples were randomly sorted within each panel to mask identities. Each panel consisted of 3 contrived moderately positive specimens, 3 contrived low positive samples, 3 contrived high- negative specimens, and 1 natural negative specimen. Panels were tested at three independent laboratories by two operators at each laboratory, twice each day over 5 non-consecutive days. The expected results were obtained with all samples by all technologists at all test sites at each time interval. Reproducibility of the assay is 100% for moderate positive, high negative, and negative samples.

As seen in the following tables, the expected results were obtained 100% of the time.

{6}------------------------------------------------

510(k) A
Description
Identification
Date:

pplication Premier C. difficile GDH 510(k) Summary Premier C. difficile GDH ion: Attachment 002 ation: March 01, 2011

Table 7: Intra and inter assay variability data for all sites (single wavelength).

PanelMembersSampleNGrandMean ALBetween-DayBetween-RunBetween-SiteTotalResult
SD%CVSD%CVSD%CVSD%CV
PC301.9440.0874.5%0.1246.4%0.20710.6%0.27914.3%Positive
NC300.0530.00610.6%0.01019.6%0.01121.1%0.02036.6%Negative
MP 1301.9570.20110.3%0.25613.1%0.1939.9%0.35918.3%Positive
MP 2301.9360.20010.3%0.28214.9%0.28514.7%0.39820.6%Positive
MP 3301.7580.1005.7%0.20511.7%0.50528.7%0.52229.7%Positive
LP 1300.7050.0669.4%0.09613.6%0.09213.1%0.15321.7%Positive
LP 2300.6700.0669.9%0.09013.4%0.15022.4%0.18427.4%Positive
LP 3300.6940.0558.0%0.09513.7%0.12017.3%0.18927.2%Positive
HN 1300.0570.00610.7%0.00814.2%0.01627.7%0.01729.7%Negative
HN 2300.0580.00813.8%0.01118.5%0.01627.1%0.01729.5%Negative
HN 3300.0590.00813.1%0.01220.3%0.01524.6%0.02439.8%Negative
WN 1300.0500.0023.0%0.0024.4%0.01428.4%0.01325.4%Negative

Legend: PC=Positive control, NC=Negative control, MP=Moderate positive, HN=High negative, WN=Weak negative

Interpretation of Results at OD450:

Positive Control: ≥ 0.600 Negative Control: < 0.150 Positive: ≥ 0.200 Negative: <0.200

Table 8: Intra and inter assay variability data (single wavelength) - Site 1: CCHMC Lot 2 611096.004.

PanelSample NGrandMean ALBetween-DayBetween-TechTotalResult
MembersSD%CVSD%CVSD%CV
PC101.8450.1206.5%0.1095.9%0.1689.1%Positive
NC100.0640.00711.5%0.0058.3%0.00913.8%Negative
MP 1101.8710.27914.9%0.0522.8%0.30516.3%Positive
MP 2101.9470.1748.9%0.21711.2%0.28414.6%Positive
MP 3101.8960.1095.7%0.23712.5%0.29015.3%Positive
LP 1100.6730.0608.9%0.12618.7%0.13319.8%Positive
LP 2100.5760.0407.0%0.0518.9%0.10818.8%Positive
LP 3100.6490.17026.2%0.0487.4%0.17627.1%Positive
HN 1100.0740.0045.3%0.00810.7%0.01013.0%Negative
HN 2100.0740.0079.9%0.01114.3%0.01216.2%Negative
HN 3100.0750.02127.5%0.01824.5%0.03040.4%Negative
WN 1100.0660.0069.3%0.0012.1%0.00710.6%Negative

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo features a globe graphic on the left, followed by the word "Meridian" in a bold, sans-serif font. Below "Meridian" is the text "Bioscience, Inc." in a smaller, also bolded, sans-serif font. A horizontal line separates "Meridian" from "Bioscience, Inc."

510(k) Application Premier C. difficile GDH
Description:510(k) Summary Premier C. difficile GDH
Identification:Attachment 002
Date:March 01, 2011

Table 9: Intra and inter assay variability data (single wavelength) - Site 2: SHH Lot 1 611096.003.

PanelSample NGrandMean ALBetween-DayBetween-TechTotalResult
MembersSD%CVSD%CVSD%CV
PC101.8050.1448.0%0.0533.0%0.1679.3%Positive
NC100.0540.01834.3%0.02241.0%0.02954.5%Negative
MP 1101.8210.26214.4%0.18710.2%0.34418.9%Positive
MP 2101.6460.32519.7%0.0020.1%0.37622.8%Positive
MP 3101.1980.15613.0%0.0453.8%0.26021.7%Positive
LP 1100.6330.10716.9%0.0548.5%0.12720.0%Positive
LP 2100.5920.13122.1%0.0061.0%0.13923.5%Positive
LP 3100.6030.13322.1%0.0528.6%0.13422.2%Positive
HN 1100.0430.00920.2%0.00510.4%0.01330.1%Negative
HN 2100.0430.006-13.7%0.00613.5%0.00921.4%Negative
HN 3100.0450.01328.7%0.01022.4%0.01942.0%Negative
WN 1100.0390.0011.7%0.0012.5%0.0026.1%Negative

Legend: PC=Positive control, NC=Negative control, MP=Moderate positive, HN=High negative, WN=Weak negative

Interpretation of Results at OD450: Positive Control: ≥ 0.600

Negative Control: < 0.150 Positive: ≥ 0.200 Negative: <0.200

Table 10: Intra and inter assay variability data (single wavelength) - Site 3: NCH Lot 3 611096.005.

PanelGrandBetween-DayBetween-TechTotalResult
MembersSample NMean ALSD%CVSD%CVSD%CV
PC102.1810.22210.2%0.1275.8%0.31514.4%Positive
NC100.0420.0024.7%0.0024.4%0.0037.2%Negative
MP 1102.1780.22010.1%0.27812.8%0.34615.9%Positive
MP 2102.2150.23310.5%0.22610.2%0.33014.9%Positive
MP 3102.1790.2129.7%0.22410.3%0.40118.4%Positive
LP 1100.8090.10913.5%0.10713.3%0.15118.7%Positive
LP 2100.8440.09611.4%0.14316.9%0.16619.7%Positive
LP 3100.8300.08310.1%0.13916.8%0.18522.2%Positive
HN 1100.0530.00612.0%0.00714.1%0.01019.4%Negative
HN 2100.0560.01221.1%0.0058.5%0.01322.9%Negative
HN 3100.0580.0069.9%0.0023.9%0.00712.4%Negative
WN 1100.0450.0024.1%0.0013.2%0.0049.0%Negative

{8}------------------------------------------------

Meridian
Sioscience, Inc.

Application Premier C. difficile GDI1 40 1 . 0 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1
escription:C. difficile510(k) Summary Premie
identification ·AnaAttachment
Jate:COMAا الام ڪال اFebruary

Table 11: Reproducibility data (single wavelength) - Site 1: CCHMC, Lot 611096.00

Day 1Day 1Day 2Day 2Day 3Day 3Day 4Day 4Day 5Day 5
Sample IDSample Qual.ResultRun 1(DH)*Run 2(MJG)*Run 1(DH)*Run 2(MJG)*Run 1(DH)*Run 2(MJG)*Run 1(DH)*Run 2(MJG)*Run 1(DH)*Run 2(MJG)*Result
PCN/A1.7562.0381.8681.7251.4951.8201.8202.0811.8981.946Positive
NC0.0690.0830.0590.0530.0600.0720.0590.0670.0560.066Negative
MP 11.6941.5492.0992.0721.3071.7282.2411.9671.8292.223Positive
MP 21.9301.4911.9432.1472.2031.7312.3262.0521.8691.5452.161Positive
MP 31.5982.1112.0921.6981.4822.0931.9442.2091.5242.204Positive
LP 10.4840.5680.8870.7810.6840.4590.7310.5420.7230.5720.785Positive
LP 20.4520.6690.7260.5000.4840.5480.6010.6230.4340.719Positive
LP 30.4770.7490.5780.5000.5110.6060.5560.6180.9560.943Positive
HN 10.0630.0850.0700.0890.0660.0860.0750.0630.0690.076Negative
HN 20.0030.0740.0830.0650.1000.0610.0650.0690.0780.0640.082Negative
HN 30.0590.0610.0570.0690.0670.0640.0620.0860.0630.157Negative
WN 10.0020.0780.0750.0590.0660.0600.0640.0710.0620.0570.068Negative
Average high negative value0.0650.0760.0640.0860.0650.0720.0690.0760.0650.105
Average low positive value0.4990.7680.6950.5610.4850.6280.5660.6550.6540.816
Percent Correlation100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%
Correlation of cut off Specimens100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%

egend: PC=Positive control, NC=Negative control, MP=Moderate
Initials of person performing testing.

  • Materials or person performing testing.
  • Subject is resting.

interpretation of Results at ODaso:
Positive Control: 2 0.600
Negative Control: < 0.600
Positive Control: < 0.150
Negative: ≥ 0.200
Negative: <0.200

{9}------------------------------------------------

Image /page/9/Picture/9 description: The image shows the logo for Meridian Bioscience, Inc. The logo is vertically oriented, with the word "Meridian" stacked on top of "Bioscience, Inc." There is a graphic of a globe to the left of the text. The text is in a bold, sans-serif font.

510(k) Application Premier C. difficile GDH
Description:510(k) Summary Premier C. difficile GDH
Identification:Attachment 002
Date:February 08, 2011

Table 12: Reproducibility data (single wavelength) - Site 2: SHH, Lot 611096.00

Sample IDSample Qual.ResultDay 1Run 1(DM)*Day 1Run 2(JM)*Day 2Run 1(DM)*Day 2Run 2(JM)*Day 3Run 1(DM)*Day 3Run 2(JM)*Day 4Run 1(DM)*Day 4Run 2(JM)*Day 5Run 1(DM)*Day 5Run 2(JM)*Result
PCN/A1.8832.0572.0231.7951.5001.7361.7241.9151.7061.710Positive
NC0.0370.0380.0380.0390.0370.0770.0400.0660.0390.127Negative
MP 11.9251.5352.0082.1061.4571.9861.8522.3791.2051.761Positive
MP 21.9301.6791.4662.4001.7411.0151.5061.8561.8891.2751.636Positive
MP 31.1011.3631.8121.1041.0781.0970.9091.3161.2510.953Positive
LP 10.7240.5780.6570.8940.4320.5340.6210.7100.5430.640Positive
LP 20.4840.5710.3600.8190.7780.4650.5780.5810.6890.5030.576Positive
LP 30.4610.5120.7950.8640.5300.5530.5190.6850.5260.585Positive
HN 10.0430.0330.0390.0790.0440.0360.0390.0400.0360.045Negative
HN 20.0030.0430.0400.0380.0670.0370.0370.0390.0480.0370.043Negative
HN 30.0430.0390.0380.0990.0370.0390.0380.0450.0350.041Negative
WN 10.0020.0370.0420.0430.0370.0360.0410.0370.0400.0390.039Negative
Average high negative value0.0430.0370.0380.0820.0390.0370.0390.0440.0360.043
Average low positive value0.5850.4830.7570.8450.4760.5550.5740.6950.5240.600
Percent Correlation100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%
Correlation of cut off Specimens100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%
Legend: PC=Positive control NC=Negative control MP=Moderate positive LP=Low positive, HN=High negative, WN=Weak negative

egend: PC=Positive control, NC=Negative control, MP=Moderate positive, LP=Low positive, HN=High negative, WN=Weak negative

gend: PC=Positive control, NC=Nep

lnterpretation of Results at ODaso:
Positive Control: ≥ 0.600
Negative Control: < 0.600
Positive Control: < 0.150
Negative: ≥ 0.200
Negative: <0.200

{10}------------------------------------------------

Meridian oscience, Inc

10(k) Application------------------------------------------------------------------------------------------------------------------------------------------------------------------------------And and the first of the first of the first of the first of the first andAPremier C. difficile GD
Description:LE MINING AN A A510{k} Summary Premier C. difficile GDIAcres Active Active Act Act Act Act Act A
dentification:000Attachment
Date:February 08, 2011

Table 13: Reproducibility data (single wavelength) - Site 3: NCH, Lot 611096.00

Sample IDSample Qual.ResultDay 1Run 1(LS)*Day 1Run 2(KE)*Day 2Run 1(LS)*Day 2Run 2(KE)*Day 3Run 1(LS)*Day 3Run 2(KE)*Day 4Run 1(LS)*Day 4Run 2(KE)*Day 5Run 1(LS)*Day 5Run 2(KE)*Result
PCN/A1.9622.3652.1262.5512.6652.1102.1742.2131.5282.117Positive
NC0.0400.0420.0480.0400.0470.0410.0400.0400.0410.040Negative
MP 11.9952.3762.1332.8222.1992.1252.1572.2611.4252.290Positive
MP 21.9301.7552.5682.2692.5922.4662.2602.2022.361.5852.094Positive
MP 31.6672.8451.8922.3722.5892.0382.3732.3471.5822.083Positive
LP 10.5740.9020.7921.0470.9040.8660.8460.8470.5490.763Positive
LP 20.4840.5871.0370.7410.9760.9760.9090.8490.9690.5600.832Positive
LP 30.5470.9950.8061.0081.0380.7740.7380.9600.5310.907Positive
HN 10.0410.0580.0700.0590.0450.0560.0450.0570.0380.062Negative
HN 20.0030.0660.0610.0590.0840.0440.0460.0560.0600.0400.048Negative
HN 30.0560.0610.0740.0560.0510.0470.0610.0570.0560.061Negative
WN 10.0020.0420.0410.0530.0380.0450.0470.0440.0470.0450.046Negative
Average high negative value0.0540.0600.0680.0660.0470.0500.0540.0580.0450.057
Average low positive value0.5690.9780.7801.0100.9730.8500.8110.9250.5470.834
Percent Correlation100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%
Correlation of cut off Specimens100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%

egend: PC-Positive control, NC=Negative control, MP=Moderate positive, LN=High negative, WN=Weak negative
Initials of person performing testing.

  1. Number of persons performing testing at site:

lnterpretation of Results at ODaso:
Positive Control: ≥ 0.600
Negative Control: < 0.600
Positive Control: < 0.150
Negative: ≥ 0.200
Negative: <0.200

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo features a globe graphic on the left, followed by the word "Meridian" in a bold, italicized font. Below "Meridian" is the phrase "Bioscience, Inc." in a smaller, bold font. A horizontal line separates "Meridian" from "Bioscience, Inc."

510(k) Application Premier C. difficile GDH
Description:510(k) Summary Premier C. difficile GDH
Identification:Attachment 002
Date:February 08, 2011

Table 14: Intra and inter assay variability data for all sites (dual wavelength).

PanelSampleGrandBetween-DayBetween-RunBetween-SiteTotalResult
MembersNMean ALSD%CVSD%CVSD%CVSD%CV
PC301.8880.0884.7%0.1216.4%0.20811.0%0.27614.6%Positive
NC300.0120.00755.3%0.01086.7%0.00864.0%0.017147.6%Negative
MP 1301.8990.20010.5%0.25313.3%0.19610.3%0.35618.7%Positive
MP 2301.8810.19710.5%0.27915.2%0.28114.9%0.39320.9%Positive
MP 3301.6660.17910.7%0.23914.3%0.45127.1%0.54032.4%Positive
LP 1300.6590.06610.1%0.09714.8%0.09314.2%0.15523.6%Positive
LP 2300.6260.06610.5%0.09014.4%0.15023.9%0.18329.3%Positive
LP 3300.6470.0538.2%0.09214.3%0.11617.9%0.18628.8%Positive
HN 1300.0130.00535.9%0.00538.3%0.00648.2%0.01181.7%Negative
HN 2300.0160.00854.0%0.01169.8%0.00956.4%0.01487.3%Negative
HN 3300.0190.00948.2%0.01369.9%0.01262.8%0.025131.6%Negative
WN 1300.0080.0019.8%0.00221.0%0.00681.2%0.00676.3%Negative

Table 15: Intra and inter assay variability data (dual wavelength) - Site 1: CCHMC Lot 2 611096.004.

PanelMembersSample NGrandMean ALBetween-DaySDBetween-Day%CVBetween-TechSDBetween-Tech%CVTotalSDTotal%CVResult
PC101.7800.1166.5%0.0945.3%0.1639.1%Positive
NC100.0130.00319.5%0.0016.3%0.00320.9%Negative
MP 1101.8010.27315.1%0.0291.6%0.29416.3%Positive
MP 2101.8850.1729.1%0.20210.7%0.27814.7%Positive
MP 3101.8410.1085.9%0.24013.0%0.29816.2%Positive
LP 1100.6190.0569.1%0.12520.2%0.13221.3%Positive
LP 2100.5280.0438.2%0.0529.9%0.10920.6%Positive
LP 3100.5990.17028.4%0.0437.2%0.17729.5%Positive
HN 1100.0200.00629.1%0.0016.3%0.00841.6%Negative
HN 2100.0250.01041.2%0.00414.1%0.01350.6%Negative
HN 3100.0330.02781.0%0.01752.9%0.036109.3%Negative
WN 1100.0150.00213.7%0.00211.3%0.00429.3%Negative

Legend: PC=Positive control, NC=Negative control, MP=Moderate positive, HN=High negative, WN=Weak negative

Interpretation of Results at OD 450/630:

Positive Control: ≥ 0.600 Negative Control: < 0.100 Positive: Negative: < 0.150

{12}------------------------------------------------

Image: MeridianBioscience, Inc.510(k) Application Premier C. difficile GDH
Description: 510(k) Summary Premier C. difficile GDH
Identification: Attachment 002
Date: February 08, 2011

Table 16: Intra and inter assay variability data (dual wavelength) - Site 2: SHH Lot 1 611096.003.

PanelGrandBetween-DayBetween-TechTotalResult
MembersSample NMean ALSD%CVSD%CVSD%CV
PC101.7570.1418.0%0.0593.3%0.1659.4%Positive
NC100.0190.01897.6%0.022120.0%0.029157.3%Negative
MP 1101.7720.26014.7%0.18310.3%0.34119.2%Positive
MP 2101.5990.32220.1%0.0030.2%0.37023.1%Positive
MP 3101.1540.15413.4%0.0464.0%0.25622.2%Positive
LP 1100.5920.10617.9%0.0508.5%0.12521.1%Positive
LP 2100.5530.12923.4%0.0040.7%0.13724.9%Positive
LP 3100.5640.13323.5%0.0508.9%0.13323.5%Positive
HN 1100.0080.009106.4%0.00559.6%0.013152.1%Negative
HN 2100.0080.00682.6%0.00455.8%0.009120.3%Negative
HN 3100.0110.013125.0%0.00983.3%0.019179.1%Negative
WN 1100.0030.00124.2%0.00125.0%0.00142.1%Negative

Table 17: Intra and inter assay variability data (dual wavelength) - Site 3: NCH Lot 3 611096.005.

PanelGrandBetween-DayBetween-TechTotalResult
MembersSample NMean ALSD%CVSD%CVSD%CV
PC102.1280.21810.2%0.1326.2%0.31014.6%Positive
NC100.0040.00122.8%0.0007.9%0.00126.8%Negative
MP 1102.1260.21910.3%0.27713.0%0.34516.2%Positive
MP 2102.1600.23110.7%0.22710.5%0.32715.2%Positive
MP 3102.0030.47523.7%0.0492.4%0.57828.9%Positive
LP 1100.7660.11915.5%0.11915.5%0.15920.8%Positive
LP 2100.7980.09411.7%0.14718.4%0.16620.8%Positive
LP 3100.7790.09512.1%0.13317.1%0.18223.4%Positive
HN 1100.0110.00655.8%0.00436.7%0.00766.2%Negative
HN 2100.0140.01173.8%0.00964.3%0.01494.8%Negative
HN 3100.0140.00747.6%0.00214.3%0.00857.4%Negative
WN 1100.0050.00120.9%0.00120.2%0.00237.8%Negative

Legend: PC=Positive control, NC=Negative control, MP=Moderate positive, HN=High negative, WN=Weak negative

ﺎﻥ ·

Interpretation of Results at OD 450/630-

Positive Control: ≥ 0.600 Negative Control: < 0.100 Positive: Negative: < 0.150

{13}------------------------------------------------

Meridian
ioscience, Inc

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------510(k) Application------------------------------------------------------------------------------------------------------------------------------------------------------------------------------The first and the may be and the commend of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of thPremier C. difficile GDI
Description:510(k) Summary Premier C. difficile GD.------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
dentification:JULnanAttachment------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Dale:tFebruary 08, 2011

Table 18: Reproducibility data (dual wavelength) - Site 1: CCHMC, Lot 611096.00

Sample IDSample Qual.ResultDay 1Run 1(DH)*Day 1Run 2(MJG)*Day 2Run 1(DH)*Day 2Run 2(MJG)*Day 3Run 1(DH)*Day 3Run 2(MJG)*Day 4Run 1(DH)*Day 4Run 2(MJG)*Day 5Run 1(DH)*Day 5Run 2(MJG)*Result
PCN/A1.6841.9681.8271.6631.4321.7571.7701.9921.8571.854Positive
NC0.0110.0090.0120.0130.0120.0130.0160.0170.0130.018Negative
MP 11.6421.4692.0521.9851.2581.6602.1701.8961.7812.097Positive
MP 21.9301.4271.8862.0962.1201.6762.2612.0081.8011.5052.071Positive
MP 31.5242.0392.0451.6311.4192.0311.8882.1371.4812.213Positive
LP 10.5100.8400.7160.6220.4120.6810.4830.6700.5290.722Positive
LP 20.4840.3980.6260.6780.4450.4390.4930.5470.6020.3900.657Positive
LP 30.4230.7040.5300.4340.4690.5550.5090.5650.9100.886Positive
HN 10.0150.0080.0280.0120.0120.0320.0210.0190.0300.026Negative
HN 20.0030.0240.0070.0270.0530.0140.0160.0200.0350.0280.027Negative
HN 30.0190.0150.0180.0210.0150.0170.0260.0400.0250.133Negative
WN 10.0020.0200.0100.0120.0190.0060.0170.0160.0170.0150.018Negative
Average high negative value0.0190.0100.0240.0290.0140.0220.0220.0310.0280.062
Average low positive value0.4440.7230.6410.5000.4400.5760.5130.6120.6100.755
Percent Correlation100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%
Correlation of cut off Specimens100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%

egend: PC=Positive control, NC=Negative control, MP=Moderate positive,
Initials of person performing testing.

First person performing testing.

Interpretation of Results at OD 450/620:

iterpretation of Results at OD 450/630 :
Positive Control: ≥ 0.600
Negative Control: > 0.600
Positive Controli < 0.100
Negative: ≥ 0.150

Negative Control: < 0.100
Positive: >0.150

POSITIVE, 20,150
Negative, 20,150

egative; < 0.150

Page 14 of 17

{14}------------------------------------------------

Meridian
Bioscience, Inc 发

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------( C. difficile (510(k) Summary Premier------------------------------------------------------------------------------------------------------------------------------------------------------------------------------AAAttachmentFebruary 08, 2011
510(k) Application Premier C. difficile GDescription:Identification:Date:

Table 19: Reproducibility data (dual wavelength) - Site 2: SHH, Lot 611096.003

Day 1Day 2Day 3Day 4Day 5
Sample IDSample Qual.ResultRun 1(DM)*Run 2(JM)*Run 1(DM)*Run 2(JM)*Run 1(DM)*Run 2(JM)*Run 1(DM)*Run 2(JM)*Run 1(DM)*Run 2(JM)*Result
PCN/A1.8341.9981.9651.7731.4481.7091.6661.8641.6641.648Positive
NC0.0020.0040.0030.0040.0030.0430.0030.0290.0030.091Negative
MP 11.8821.4871.9552.0601.4121.9401.8042.3141.1571.705Positive
MP 21.9301.6301.4192.3331.6970.9781.4571.8151.8461.2291.585Positive
MP 31.0601.3141.7581.0631.0381.0490.8701.2731.2050.909Positive
LP 10.6850.5390.6170.8500.3950.4940.5820.6640.5040.591Positive
LP 20.4840.5330.3230.7790.7360.4290.5380.5440.6440.4650.535Positive
LP 30.4240.4740.7550.8250.4930.5130.4830.6380.4850.546Positive
HN 10.0080.0040.0040.0440.0050.0030.0040.0040.0030.004Negative
HN 20.0030.0090.0050.0040.0330.0030.0030.0050.0060.0020.006Negative
HN 30.0090.0050.0040.0650.0030.0050.0040.0050.0020.005Negative
WN 10.0020.0030.0020.0040.0040.0030.0030.0030.0030.0070.002Negative
Average high negative value0.0090.0050.0040.0470.0040.0040.0040.0050.0020.005
Average low positive value0.5470.4450.7170.8040.4390.5150.5360.6490.4850.557
Percent Correlation100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%
Correlation of cut off Specimens100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%

gend: PC=Positive control, NC=Nega
nitials of person performing testing.

umbers of person performing testing.

Interpretation of Results at OD 450/630:

Positive Control: >0.500

lnterpretation of Results at OD ๙๑๑๑๑๑๑๑๑
Positive Control: ≥ 0.600
Negative Control: ≥ 0.600
Negative Control: < 0.100
Negative:

{15}------------------------------------------------

Meridian
Bioscience, Inc. িনিট

510(k) Application Premier C. difficile GDI------------------------------------------------------------------------------------------------------------------------------------------------------------------------------A C A C L B . L . L
Description:510(k) Summary Premier C. difficile GDI
dentificationttachment
Date:t9044February 08, 201

Table 20: Reproducibility data (dual wavelength) - Site 3: NCH, Lot 611096.005

Sample IDSample Qual.ResultDay 1Run 1(LS)*Day 1Run 2(KE)*Day 2Run 1(LS)*Day 2Run 2(KE)*Day 3Run 1(LS)*Day 3Run 2(KE)*Day 4Run 1(LS)*Day 4Run 2(KE)*Day 5Run 1(LS)*Day 5Run 2(KE)*Result
PCN/A1.9112.3112.0752.4962.5872.0632.1232.1631.4762.071Positive
NC0.0030.0040.0040.0030.0060.0040.0030.0030.0030.003Negative
MP 11.9512.3222.0802.7622.1462.0762.0992.211.3712.238Positive
MP 21.9301.7112.5142.2052.5382.4112.2122.1362.3041.5352.038Positive
MP 31.6212.7841.8422.3212.5311.9902.3172.2921.5320.802Positive
LP 10.5330.8550.7471.0050.8580.8230.7940.8550.4760.710Positive
LP 20.4840.5390.9950.7020.9340.9320.8690.7800.9220.5210.790Positive
LP 30.4970.9500.7700.9630.9830.7340.6890.9170.4870.802Positive
HN 10.0080.0180.0220.0180.0040.0050.0040.0150.0020.012Negative
HN 20.0030.0100.0210.0140.0490.0050.0060.0070.0190.0030.009Negative
HN 30.0060.0220.0300.0190.0080.0060.0100.0160.0080.013Negative
WN 10.0020.0050.0030.0090.0030.0050.0070.0040.0040.0050.004Negative
Average high negative value0.0080.0200.0220.0290.0060.0060.0070.0170.0040.011
Average low positive value0.5230.9330.7400.9670.9240.8090.7540.8980.4950.767
Percent Correlation100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%
Correlation of cut off Specimens100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%100.0%

Neak nega Legend: PC=Positive control, NC=Negative control, MP=Moderate positive, LP=Low positive, HN=High negative,

  • Initials of person performing testing.

limits of person performing testing

lnterpretation of Results at OD «so/eso:
Positive Control: ≥ 0.600
Negative Control: ≥ 0.600
Negative :

Page 16 of 17

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo consists of a globe graphic on the left, followed by the word "Meridian" in a bold, sans-serif font. Below "Meridian" is the text "Bioscience, Inc." in a smaller, bold, sans-serif font. A horizontal line separates "Meridian" from "Bioscience, Inc."

510(k) Application Premier C. difficile GDH
Description:510(k) Summary Premier C. difficile GDH
Identification:Attachment 002
Date:February 08, 2011

Conclusions

The Premier C. difficile GDH assay can be used to screen human stool samples for C. difficile glutamate dehydrogenase and can be considered substantially equivalent to the reference method and predicate device.

{17}------------------------------------------------

Image /page/17/Picture/1 description: The image shows the logo for the Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES.USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird with its wings spread, facing to the right. The image is in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

Meridian Bioscience, Inc. c/o Ms. Susan Bogar Product Quality Assurance Manager 3471 River Hills Drive Cincinnati, OH 45244

MAY - 3 2011

Re: K110620

Trade/Device Name: Premier™ C. difficile GDH Assay Regulation Number: 21 CFR 866.2660 Regulation Name: Microorganism differentiation and identification device. Regulatory Class: Class I Product Code: MCB Dated: March 2, 2011 Received: March 3, 2011

Dear Ms. Bogar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter

{18}------------------------------------------------

Page 2 - Ms. Susan Bogar

will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Sallastors

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{19}------------------------------------------------

Indications for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Premier C. difficile GDH

Indications for Use:

Premier C. difficile GDH is a qualitative enzyme immunoassay screening test to detect Clostridium difficile antigen, glutamate dehydrogenase, in fecal specimens from symptomatic persons suspected of having C. difficile infection (CDI). This test does not distinguish between toxigenic and non-toxigenic strains of C. difficile. Samples from symptomatic patients that produce positive results with this test must be further tested with an assay designed to detect toxigenic C. difficile strains and assist with the diagnosis of CDI.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Kuddie Poole

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K118620

Page 1 of

§ 866.2660 Microorganism differentiation and identification device.

(a)
Identification. A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.