K Number
K110586

Validate with FDA (Live)

Date Cleared
2011-10-14

(227 days)

Product Code
Regulation Number
862.1825
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ADVIA Centaur Vitamin D Total (VitD) assay is for the in vitro diagnostic use in the quantitative determination of total 25 (OH) vitamin D in human serum and plasma (EDTA, lithium heparin, sodium heparin) using the ADVIA Centaur XP system. The ADVIA Centaur VitD assay is intended as an aid in the determination of vitamin D sufficiency.

The ADVIA Centaur Vitamin D Total (VitD) Calibrators is for in vitro diagnostic use in calibrating ADVIA Centaur® systems Vitamin D Total (VitD) assay.

The ADVIA Centaur Vitamin D Total (VitD) QC is for in vitro diagnostic use to monitor the precision and accuracy of the ADVIA Centaur® VitD assay on the ADVIA Centaur systems.

The ADVIA Centaur® Vitamin D Total (VitD) Master Curve Material is for in vitro diagnostic use in the verification of calibration and reportable range of the ADVIA Centaur VitD assay.

Device Description

The ADVIA Centaur Vitamin D Total assay is a one-pass 18 minute competitive immunoassay that uses an anti-fluorescein labeled (FITC) monoclonal antibody covalently bound to paramagnetic particles (PMP), one monoclonal antibody labeled with acridium ester (AE), and a Vitamin D analog labeled with fluorescein. An inverse relationship exists between the amount of Vitamin D present in the patient sample and the amount of relative light units (RLUs) detected by the system.

AI/ML Overview

Here's a summary of the acceptance criteria and study information for the ADVIA Centaur Vitamin D Total Assay, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied by the comparison to the predicate device and the successful results for various performance characteristics. The document states that "All of the evaluation studies gave acceptable results compared to the predicate device." Therefore, the reported performance is within the acceptable range as defined by the predicate device and established analytical standards.

Performance CharacteristicAcceptance Criteria (Implied by Predicate/Standards)Reported Device Performance (ADVIA Centaur Vitamin D Total Assay)
Analytical SensitivityAcceptable for Vitamin D test systems (vs. K021163)LoB = 1.60 ng/mL, LoD = 3.2 ng/mL, LoQ = 4.2 ng/mL
Imprecision (Total CV)Comparable to predicate for similar levelsLevel 11.7 ng/mL: 11.1%Level 18.0 ng/mL: 9.6%Level 32.4 ng/mL: 9.8%Level 49.9 ng/mL: 8.2%Level 55.8 ng/mL: 7.8%Level 132.1 ng/mL: 4.8%
Specificity (Cross-Reactivity)Low cross-reactivity for related compounds1,25 (OH)2 Vitamin D2: 4.0%1,25 (OH)2 Vitamin D3: 1.0%25 OH Vitamin D2: 104.5%25 OH Vitamin D3: 100.7%Paricalcitol: 0.1%3-epi-25-OH Vitamin D3: 1.1%Vitamin D2: 0.5%Vitamin D3: 0.3%
Interfering Substances≤ 10% change in results for specified concentrationsHemoglobin: 155 mg/dLTriglycerides: 540 mg/dLConjugated Bilirubin: 40 mg/dLUnconjugated Bilirubin: 40 mg/dLCholesterol: 350 mg/dLUric Acid: 20 mg/dLHuman Immunoglobulin: 12 g/dL
Method Comparison (Correlation with Predicate)Strong correlation (vs. K021163)n = 195, r = 0.960, Slope = 1.0, Y-intercept = +2.22
Serum/Plasma EquivalencyNo significant differences between tube typesR values: Serum vs. SST: 0.994, Serum vs. EDTA: 0.993, Serum vs. Lithium Heparin: 0.992, Serum vs. Sodium Heparin: 0.992

2. Sample Sizes Used for the Test Set and Data Provenance

  • Method Comparison: n = 195 samples. The provenance of this data (country of origin, retrospective/prospective) is not explicitly stated.
  • Imprecision: n = 160 (Total CV). It's typically from replicate measurements on control samples over multiple runs.
  • Specificity: Not a sample size in the traditional sense, but tested with specific concentrations of various compounds.
  • Interfering Substances: Tested as described in CLSI Document EP7-A2. Specific sample sizes for each interfering substance are not provided, but it would involve spiking known concentrations into samples.
  • Serum/Plasma Equivalency: n = 231 matched specimens drawn in different tube types. The provenance of this data (country of origin, retrospective/prospective) is not explicitly stated.

The document does not specify the country of origin of the data or whether the studies were retrospective or prospective for the various performance characteristics.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This type of information is generally not applicable to an in vitro diagnostic (IVD) assay like the ADVIA Centaur Vitamin D Total Assay. The "ground truth" for these assays is usually established through:

  • Reference Methods: For method comparison, the predicate device (IDS 25-OH Vitamin D EIA assay, K021163) or a recognized reference method like LC-MS/MS (mentioned for traceability of the ADVIA Centaur assay) serves as the comparator.
  • Known Concentrations: For analytical sensitivity, imprecision, specificity, and interfering substances, the "ground truth" is based on precisely prepared samples with known concentrations of the analyte or interfering substances.

Therefore, no panel of human experts was used to establish ground truth for this device.

4. Adjudication Method for the Test Set

Not applicable. As this is an IVD assay measuring biochemical analytes, there is no adjudication process involving human review for the "test set" in the way one might see in imaging studies. The results are quantitative and compared against established analytical methods or known values.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

No. An MRMC study is not relevant for this type of IVD device, which is an automated assay for blood analysis. MRMC studies are typically performed for diagnostic imaging devices where human interpretation is a critical component.

6. Standalone Performance Study

Yes, the studies described (Analytical Sensitivity, Imprecision, Specificity, Interfering Substances, Serum/Plasma Equivalency) represent the standalone performance of the ADVIA Centaur Vitamin D Total Assay. The "Method Comparison" against the predicate device also serves to demonstrate that the algorithm (assay) performs comparably to an already established method without human interpretation required for its output.

7. Type of Ground Truth Used

The ground truth used for this device includes:

  • Reference Assay/Predicate Device: For method comparison, the results generated by the IDS 25-OH Vitamin D EIA assay (K021163) served as the comparator.
  • Traceability to LC-MS/MS: The ADVIA Centaur assay's traceability to LC-MS/MS (Liquid Chromatography-Mass Spectrometry) indicates that for its fundamental accuracy and standardization, it references a highly precise and accurate analytical method often considered a gold standard for vitamin D measurement.
  • Known/Spiked Concentrations: For studies like analytical sensitivity, imprecision, specificity, and interfering substances, the ground truth is based on samples with precisely known concentrations of the analyte or potential interfering substances.

8. Sample Size for the Training Set

The document does not explicitly mention a "training set" in the context of machine learning or AI, as this is a traditional immunoassay. For such assays, method development and optimization phases involve extensive testing to establish reagent formulations, reaction conditions, and calibration algorithms. The data used for this developmental work is not typically reported as a "training set" in the same way it would be for an AI-based device.

9. How the Ground Truth for the Training Set Was Established

As this is a traditional immunoassay and not an AI/ML device, the concept of a "training set" with established ground truth as understood in AI development is not directly applicable. Instead, the assay's performance characteristics (calibration, linearity, imprecision, accuracy) would have been optimized and validated using a combination of:

  • Reference standards: Materials with assigned values, often traceable to higher-order reference methods like LC-MS/MS.
  • Clinical samples: Patient samples covering a range of analyte concentrations.
  • Spiked samples: Samples with known quantities of analyte added.
  • Internal quality control materials: Samples with established target values.

The "ground truth" during development would be established through careful analytical measurements using these standards and methods to ensure the assay accurately quantifies vitamin D.

{0}------------------------------------------------

K110586

OCT 1 4 2011

510(k) Summary

Submitter information
Contact person:Neil ParkerSenior Regulatory Affairs Specialist
Address:Siemens Healthcare Diagnostics511 Benedict AvenueTarrytown, NY 10591
Phone:914-524-2477914-524-2500 (fax)

Date summary prepared: February 28, 2011

Device Trade or Proprietary Names:

ADVIA Centaur Vitamin D Total Assay

ADVIA Centaur Vitamin D Total calibrators

ADVIA Centaur Vitamin D Total QC

ADVIA Centaur Vitamin D Total Master Curve Material

Device Common/Usual Name or Classification Name:

Vitamin D Test System Calibrator Quality Control Material (Assayed and Unassayed)

Classification Number / Class:

21 CFR 862.1825 - Vitamin D Test System - Class II 21 CFR 862.1150 - Calibrator - Class II 21 CFR 862.1660 - Quality Control Material (Assayed and Unassayed) - Class I

Siemens Healthcare Diagnostics ADVIA Centaur Vitamin D Total assay 510(k) Summary

{1}------------------------------------------------

Product code:

MRG - Vitamin D Test system JIT - Calibrator, Secondary JJX - single (specified) Analyte Controls (Assayed and Unassayed)

This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92.

The assigned 510(k) number is: __

Assay Predicate Device (including calibrators and controls):

Predicate Device
Device NameIDS 25-Hydroxy Vitamin D EIA
Common nameVitamin D Test System
510(k) NumberK021163
ManufacturerImmunodiagnostic Systems Ltd (IDS Ltd)

Master Curve Material Predicate Device

Predicate Device
Device NameThe VALIDATE Thyroid Calibration Verification TestSet
Common namesingle (specified) Analyte Controls (Assayedand Unassayed)
510(k) NumberK062501
ManufacturerMaine Standards Company

Device Description:

The ADVIA Centaur Vitamin D Total assay is a one-pass 18 minute competitive immunoassay that uses an anti-fluorescein labeled (FITC) monoclonal antibody covalently bound to paramagnetic particles (PMP), one monoclonal antibody labeled with acridium ester (AE), and a Vitamin D analog labeled with fluorescein. An inverse relationship exists between the amount of Vitamin D present in the patient sample and the amount of relative light units (RLUs) detected by the system.

{2}------------------------------------------------

Statements of Intended Use:

The ADVIA Centaur Vitamin D Total (VitD) assay is for the in vitro diagnostic use in the quantitative determination of total 25 (OH) vitamin D in human serum and plasma (EDTA, lithium heparin, sodium heparin) using the ADVIA Centaur XP system. The ADVIA Centaur VitD assay is intended as an aid in the determination of vitamin D sufficiency.

The ADVIA Centaur Vitamin D Total (VitD) Calibrators is for in vitro diagnostic use in calibrating ADVIA Centaur® systems Vitamin D Total (VitD) assay.

The ADVIA Centaur Vitamin D Total (VitD) QC is for in vitro diagnostic use to monitor the precision and accuracy of the ADVIA Centaur® VitD assay on the ADVIA Centaur systems.

The ADVIA Centaur® Vitamin D Total (VitD) Master Curve Material is for in vitro diagnostic use in the verification of calibration and reportable range of the ADVIA Centaur VitD assay.

{3}------------------------------------------------

Comparisons to the Predicate Devices: Assay Similarities

ItemsADVIA Centaur Vitamin D TotalassayIDS 25-OH Vitamin D EIAassay (Predicate Device)K021163
Similarity
Intended Use/Indication foruseFor the in vitro diagnostic use inthe quantitative determination oftotal 25 (OH) vitamin D in humanserum and plasma (EDTA,lithium-heparin, sodium-heparin)using the ADVIA Centaur XPsystem. The ADVIA Centaur VitDassay is intended as an aid in thedetermination of vitamin Dsufficiency.SimilarThe IDS 25-OH Vitamin D EIAkit is an enzymeimmunoassayintended for the quantitativedetermination of 25-Hydroxyvitamin D and otherhydroxylated metabolites inhuman serum and plasma.Results are to be used inconjunction with other clinicaland laboratory data to assistthe clinician in the assessmentof vitamin D sufficiency inadult populations.
Sample TypeSerum and Plasmasame
Reagent storage temperature2-8°Csame
Interpretation of resultsStandard curvesame
Expected ValuesDeficiency < 10 ng/mLInsufficiency 10-29 ng/mLSufficiency 30-100 ng/mLToxicity > 100ng/mLSame
Ranges established forapparently healthy adults not onsupplements:10.6 ng/mL to 43.4 ng/mLn = 542Ranges established adults19.1 ng/mL to 57.6 ng/mLn = 36

Assay Differences

ItemsADVIA Centaur Vitamin D TotalassayIDS 25-OH Vitamin D EIAassay (Predicate Device)K021163
Differences
PlatformADVIA Centaur XPELISA plate read on a platereader
Assay principleChemiluminescenceELISA
Assay procedureAutomatedManual
Approximate Assay time18 minute3.5 hours
Assay Range4.2 to 150 ng/mL5.9 to 120 ng/mL
TraceabilityTraceable to LC-MS/MSStandardized using UVquantification of 25-(OH)vitamin D

Siemens Healthcare Diagnostics
ADVIA Centaur Vitamin D Total assay
510(k) Summary

{4}------------------------------------------------

Calibrator Similarities

ItemsADVIA Centaur Vitamin D Totalassay calibratorsIDS 25-OH Vitamin D EIAassay calibrators (PredicateDevice) K021163
Similarity
Intended Use/Indication foruseFor in vitro diagnostic use incalibrating ADVIA Centaur®systems Vitamin D Total (VitD)assay.Similar
Antigen used in calibrators25-(OH) vitamin D3same
Storage temperature2-8°Csame.

Calibrator Differences

ItemsADVIA Centaur Vitamin D TotalassayIDS 25-OH Vitamin D EIAassay calibrators (PredicateDevice) K021163
Differences
Number of calibrators26
MatrixLyophilized human plasmaLyophilized human serum

{5}------------------------------------------------

QC Similarities

ItemsADVIA Centaur Vitamin D Totalassay controlsIDS 25-OH Vitamin D EIAassay controls (PredicateDevice) K021163
Similarity
Intended Use/Indication foruseFor in vitro diagnostic use tomonitor the precision andaccuracy of the ADVIA Centaur®VitD assay on the ADVIACentaur systems.Similar
Antigen used in controls25-(OH) vitamin D3same
Number of Levels2same
Storage temperature2-8°Csame

QC Differences

ItemsADVIA Centaur Vitamin D Totalassay controlsIDS 25-OH Vitamin D EIAassay controls (PredicateDevice) K021163
Differences
MatrixLyophilized human plasmaLyophilized human serum

· :

{6}------------------------------------------------

MCM Similarities

ItemsADVIA Centaur Vitamin D Totalassay MCMsMaine Standards CompanyVALIDATE Thyroid CalibrationVerification Test Set(K062501)
Similarity
Intended Use/Indication foruseThe ADVIA Centaur® Vitamin DTotal (VitD) Master CurveMaterial is for in vitro diagnosticuse in the verification ofcalibration and reportable rangeof the ADVIA Centaur VitDassay.The VALIDATE ThyroidCalibration Verification TestSet solutions are for in vitrodiagnostic use in theuantitative determination oflinearity, calibration verificationand verification of reportablerange in automated, semi-automated and manualchemistry systems.
Number of Levels5Same

MCM Differences

ItemsADVIA Centaur Vitamin D Total assay MCMsMaine Standards CompanyVALIDATE ThyroidCalibrationVerification Test Set(K062501)
Differences
MatrixHuman plasmaHuman Serum
FormLyophilizedliquid
Analyte25 (OH) vitamin DTriiodothyronine (T3),Thyroxine (T4), humanThyroid Stimulating Hormone(TSH), and Cortisol.

Performance:

Substantial equivalence for the ADVIA Centaur Vitamin D Total assay was demonstrated by testing several method performance characteristics including Analytical Sensitivity, linearity, imprecision, method comparison, interfering substances, and specificity. The following tables summarize the analytical sensitivity, linearity, precision (total), interfering substances, specificity, serum / plasma equivalency and method comparison results.

All of the evaluation studies gave acceptable results compared to the predicate device. These studies support that the ADVIA Centaur Vitamin D Total assay (including the

{7}------------------------------------------------

calibrators and controls) are substantially equivalent to the IDS 25-Hydroxy Vitamin D EIA that is currently marketed.

Analytical Sensitivity

LoB = 1.60 ng/mL LoD = 3.2 ng/mL LoQ = 4.2 ng/mL

Imprecision

ADVIA Centaur Vitamin DTotal AssayIDS 25-Hydroxy Vitamin D EIA
Level (ng/mL)Total CV (%)n = 160Level (ng/mL)Inter assay CV (%)n = 11
11.711.1
18.09.6
32.49.840.34.6
49.98.2
55.87.872.06.4
132.14.8132.08.7

Specificity

The ADVIA Centaur VitD Total assay shows high specificity for 25(OH) vitamin D2 and 25(OH) vitamin D3. The following compounds were tested with total 25(OH) vitamin D concentrations of 35 and 115 ng/mL. Percent change is calculated as: Percent cross-reactivity = (corrected assay value / amount of compound spiked) x 100

{8}------------------------------------------------

CompoundConcentration (ng/mL)Cross-Reactivity (%)
1, 25 (OH)2 Vitamin D21004.0
1, 25 (OH)2 Vitamin D31001.0
25 OH Vitamin D230104.5
25 OH Vitamin D330100.7
Paricalcitol240.1
3-epi-25-OH Vitamin D31001.1
Vitamin D210000.5
Vitamin D310000.3

Interfering Substances

Interfering substances were tested as described in CLSI Document EP7-A2 using the ADVIA Centaur VitD assay.

Specimens That AreDemonstrate ≤ 10% Change in Results Up To
hemolyzed155 mg/dL of hemoglobin
lipemic540 mg/dL of triglycerides
icteric40 mg/dL of conjugated bilirubin
icteric40 mg/dL of unconjugated bilirubin
Specimens That ContainDemonstrate ≤ 10% Change in Results Up To
cholesterol350 mg/dL
uric acid20 mg/dL
human immunoglobulin12 g/dL

Correlation

(y =ADVIA Centaur Vitamin D assay vs. x = IDS 25-OH Vitamin D EIA assay comparison method/system)

IDS vs Centaur
X AxisY AxisnrSlopeY-int
IDSCentaur VitD195.9601.0+2.22

{9}------------------------------------------------

Serum / Plasma (lithium heparin and EDTA)

The ADVIA Centaur VITD assay was evaluated using different sample matrices and tube collection types. A matrix study was performed using matched specimens drawn in different tube types,including serum red top, SST, EDTA, lithium heparin, and sodium heparin. Vitamin D values ranged from 11.9 to 136.9 ng/mL (29.8 to 342.3 nmol/L). Linear regression analysis was performed using the following:

  • · serum (x) vs. SST (y1)
  • · serum (x) vs. EDTA (y2)
  • · serum (x) vs. lithium heparin (y3)
  • · serum (x) vs. sodium heparin (y4)

No significant differences between tube types was observed. The following results were obtained:

Tube TypesNRange(ng/mL)SlopeInterceptR
Serum vs. SST23111.9 - 136.91.01-0.330.994
Serum vs. EDTA23111.9 - 136.91.09-0.170.993
Serum vs. Lithium Heparin23111.9 - 136.91.040.180.992
Serum vs. Sodium Heparin23111.9 - 136.91.040.900.992

Conclusions:

The Siemens Healthcare Diagnostics ADVIA Centaur Vitamin D Total assay (including calibrators and controls) is substantially equivalent to other products in commercial distribution intended for similar use. Most notably, it is substantially equivalent to the currently marketed IDS 25-Hydroxy Vitamin D EIA (K021163).

The Siemens Healthcare Diagnostics ADVIA Centaur Vitamin D Total assay MCMs are substantially equivalent to other products in commercial distribution intended for similar use. Most notably, it is substantially equivalent to the currently marketed VALIDATE Thyroid Calibration Verification Test Set (K062501).

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized image of an eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circle around the eagle. The eagle is black, and the text is also black.

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

Siemens Healthcare Diagnostics, Inc. c/o Mr. Neil Parker 511 Benedict Avenue Tarrytown, New York 10509

OCT 1 4 2011

Re: K110586

Trade Name: ADVIA Centaur Vitamin D Total (VitD) Assay ADVIA Centaur Vitamin D Total (VitD) Calibrators ADVIA Centaur Vitamin D Total (VitD) QC ADVIA Centaur Vitamin D Total (VitD) Master Curve Material Regulation Number: 21 CFR §862.1825 Regulation Name: Vitamin D Test System Regulatory Class: Class II Product Codes: MRG, JIS, JIX Dated: September 15, 2011 Received: September 16, 2011

Dear Dr. Neil Parker,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{11}------------------------------------------------

Page 2 -

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

signature

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{12}------------------------------------------------

Indication for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: ADVIA Centaur Vitamin D Total Assay Indication for Use:

The ADVIA Centaur Vitamin D Total (VitD) assay is for in vitro diagnostic use in the quantitative determination of total 25 (OH) vitamin D in human serum and plasma (EDTA, lithium-heparin, sodium-heparin) using the ADVIA Centaur XP system. The ADVIA Centaur VitD assay is intended as an aid in the determination of vitamin D sufficiency.

The ADVIA Centaur Vitamin D Total (VitD) Calibrators is for in vitro diagnostic use in calibrating ADVIA Centaur® systems Vitamin D Total (VitD) assay.

The ADVIA Centaur Vitamin D Total (VitD) QC is for in vitro diagnostic use to monitor the precision and accuracy of the ADVIA Centaur® VitD assay on the ADVIA Centaur systems.

The ADVIA Centaur® Vitamin D Total (VitD) Master Curve Material is for in vitro diagnostic use in the verification of calibration and reportable range of the ADVIA Centaur VitD assay.

Prescription Use X_ (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

CARS

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K110586.

Page 1 of

§ 862.1825 Vitamin D test system.

(a)
Identification. A vitamin D test system is a device intended for use in clinical laboratories for the quantitative determination of 25-hydroxyvitamin D (25-OH-D) and other hydroxylated metabolites of vitamin D in serum or plasma to be used in the assessment of vitamin D sufficiency.(b)
Classification. Class II (special controls). Vitamin D test systems must comply with the following special controls:(1) Labeling in conformance with 21 CFR 809.10 and
(2) Compliance with existing standards of the National Committee on Clinical Laboratory Standards.