K Number
K110536

Validate with FDA (Live)

Date Cleared
2012-03-19

(388 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner (21CFR 880.6250)

Device Description

A Powder-Free Vinyl Patient Examination Gloves is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner (21CFR 880.6250). Our Powder-Free Vinyl Patient Examination Gloves do not contain any UPS powder according to the results of testing conducted on the gloves. The residual powder testing is conducted per standards of ASTM IN-6124-06, which demonstrates that the powder free glove is less than 2mg/pc. Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powder-Free Vinyl Patient Examination Glove, 80LtZ, and meets all requirement of ASTM Standard D5250-06 (2011).

AI/ML Overview

This 510(k) submission is for Powder-Free Vinyl Patient Examination Gloves, which are Class I devices that typically do not require clinical studies. The submission focuses on demonstrating substantial equivalence to predicate devices and adherence to relevant ASTM standards. Therefore, many of the typical AI/software-based study questions are not applicable.

Here's an analysis of the provided text based on your request:

Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Conforms to ASTM-D-5250-06 (2011)Full conformity to ASTM-D-5250-06 (2011) standard
Meets FDA pinhole requirementsMeets pinhole FDA requirements
Meets biocompatibility requirementsMeets biocompatibility requirements (Primary Skin irritation and Skin Sensitization testing showed no adverse reactions)
Meets labeling claimsMeets labeling claims
Powder-free (residual powder < 2mg/pc)Residual powder testing per ASTM D6124-06 shows < 2mg/pc
Passes FDA 1000 ml Water Fill TestPass with samplings of AOL 2.5, Inspection Level I
No primary skin irritant or sensitization reactionsTesting conducted with results showing no primary skin irritant or sensitization reactions

Study Information

  1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size: For the FDA 1000 ml. Water Fill Test, the "samplings of AOL 2.5, Inspection Level I" were used. This indicates an acceptance quality limit (AQL) sampling plan, but the exact number of units sampled is not explicitly stated in the provided text. For Physical and Dimensions Testing, the "Inspection Level S-2, AOL 2.5" was used, again indicating a sampling plan rather than a fixed sample size. For Residual Powder testing, the standard ASTM D6124-06 was applied, but the specific sample size is not mentioned. For Biocompatibility testing, specific numbers of test subjects and samples are not provided but implied through "testing conducted".
    • Data Provenance: Not specified, but the manufacturing company is Zibo Tengyue Plastic Products Co., Ltd. from Zibo, Shandong, China. It is highly probable the testing was conducted in China or a contracted lab. The studies would be considered prospective for the specific batches tested.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. This device is a medical glove, and its performance is evaluated against engineering and biological standards, not by expert interpretation of images or clinical outcomes. Ground truth is established through standardized physical, chemical, and biological tests.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. Performance is determined by objective measurements and compliance with ASTM standards and FDA tests.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI/software device.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. This is not an AI/software device. The tests performed are standalone in the sense that they evaluate the device itself without human interaction as part of the performance metric.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Standards-based Ground Truth: The ground truth for this device is based on established engineering and biological standards. This includes:
      • ASTM D5250-06 (2011): Standard Specification for Poly(vinyl chloride) Gloves for Medical Application.
      • ASTM D6124-06: Standard Test Method for Residual Powder on Medical Gloves.
      • FDA 1000 ml. Water Fill Test: For pinhole detection.
      • Biocompatibility Standards: Implicitly for Primary Skin irritation and Skin Sensitization.
  7. The sample size for the training set

    • Not applicable. There is no "training set" in the context of this device. Device performance is based on manufacturing processes and quality control testing, not machine learning.
  8. How the ground truth for the training set was established

    • Not applicable. As there is no training set, this question is irrelevant.

{0}------------------------------------------------

KI10536

MARY 11 9, 2012

510 (K) SUMMARY

This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21CFR 807.92.

Submitter's Identification: 1.

Zibo Tengyue Plastic Products Co., Ltd. 77 Center of Zhangdian, Zhangdian District Zibo, Shandong, 255400

Contact Person: Jiang, Xiyan Product Manager Phone Number: 011-86-13355336122

Date summary prepared: August 18, 2011

2. Name of the Device:

Zibo Tengyue Plastic Products Co., Ltd. Powder-Free Vinyl Patient Examination Gloves, Non-Sterile

3. Predicate Device Informaton:

Shijiazhuang Star Plastic Co., Ltd. Vinyl Examination Gloves - Powder Free (K 100699)

Tangshan Zhonghong Pulin Food Products Co., Ltd. Vinyl Examination Gloves - Powder Free (K022091)

4. Device Description:

A Powder-Free Vinyl Patient Examination Gloves is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner (21CFR 880.6250). Our Powder-Free Vinyl Patient Examination Gloves do not contain any UPS powder according to the results of testing conducted on the gloves. The residual powder testing is conducted per standards of ASTM IN-6124-06, which demonstrates that the powder free glove is less than 2mg/pc. Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powder-Free Vinyl Patient Examination Glove, 80LtZ, and meets all requirement of ASTM Standard D5250-06 (2011).

{1}------------------------------------------------

Intended Use: 5.

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner (21CFR 880.6250)

Comparison to Predicate Devices: ર્.

Zibo Tengyue Plastic Products Co., Ltd. Powder-Free Vinyl Patient Examination Gloves are substantially equivalent in safety and effectiveness to Tangshan Zhonghong Pulin Food Products Co., Ltd. (K022091) and Shijiazhuang Star Plastic., Co., Ltd. (K100699) Powder-Free Viny) Patient Examination Gloves.

7. Discussion of Non-Clinical tests performed for Determination of Substantial Equivalence are as follows:

The standards used for Zibo Tengyue Plastic Products Co., Ltd. glove production are based on ASTM-D-5250-06 (2011). All testing meets requirements for Physical and Dimensions Testing conducted on gloves. Inspection Level S-2, AOL 2.5.

The FDA 1000 ml. Water Fill Test was also conducted with samplings of AOL 2.5, Inspection Level I, meeting these requirements, Primary Skin irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.

Our gloves meet standard of ASTM D6124-06 (Standard Test Method for Residual Powder on Medical Gloves).

Discussion of Clinical Tests Performed: 8.

Not Applicable - There is no hypoallergenic claim.

の。 Conclusions:

Zibo Tengyue Plastic Products Co., Ltd. Powder-Free Vinyl Patient Examination Gloves conforms fully to ASTM-D-5250-06 (2011) standard as well as applicable 21 CFR references, and, meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Sentic, 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Zibo Tengyue Plastic Products Company, Limited C/O Mr. Frank Liu President Basic Medical Industries, Incorporated 12390 East End Avenue Chino, California 91710

MAR 1 9 2012

Re: K110536

Trade/Device Name: Powder-Free Vinyl Patient Examination Gloves, Non-Sterile Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: February 10, 2012 Received: February 24, 2012

Dear Mr. Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Mr. Liu

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

G

for

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Anthony D. Watson, B.S., M.S., M.B.A.

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

Zibo Tengyue Plastic Products Co., Ltd. Applicant:

510(k) Number: KI 10236

Powder-Free Vinyl Patient Examination Gloves, Non-Sterile Device Name:

Indications of Use:

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner (21CFR 880.6250)

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over the Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

= trill = " = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: ___ KJ1 05 34

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.