K Number
K110429

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-09-02

(199 days)

Product Code
Regulation Number
870.2300
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended us of ProSim 4 is to test and verify the basic operation of patient monitoring devices or systems used to monitor various physiological parameters of a patient, including ECG, Respiration, Invasive blood pressure and Non-invasive blood pressure.

The ProSim 4 Vital Signs Simulator provides electronic and pneumatic simulation of physiological parameters for determining that patient monitoring devices or systems are performing within their operating specifications. The device includes the following physiological simulations:

  • ECG adult or neonatal
  • Invasive and non-invasive blood pressure
  • Respiration

The intended us of ProSim 6 and ProSim 8 is to test and verify the basic operation of patient monitoring devices or systems used to monitor various physiological parameters of a patient, including ECG, Respiration, Invasive blood pressure, Non-invasive blood pressure and Cardiac output. Additionally, the devices provide an optical signal to verify that the electronics within the pulse oximeter probe are functional.

The ProSim 6 and ProSim 8 Vital Signs Simulators provide electronic and pneumatic simulation of physiological parameters for determining that patient monitoring devices or systems are performing within their operating specifications. The devices provide the following physiological simulations:

  • ECG adult or pediatric
  • Invasive and non-invasive blood pressure
  • Respiration
  • Temperature
  • Cardiac Output
  • Fetal Simulation includes fetal, maternal ECG, & uterine contractions (ProSim 8 only)
    Additionally, the devices provide an optical signal to verify that the electronics within the pulse oximeter probe are functional
Device Description

Fluke Biomedical's ProSim 4 (hereafter referred to as the ProSim) provides a basis to train, evaluate, and perform preventive maintenance for virtually all patient monitors found in the healthcare industry. This is accomplished with multiple physiological simulations for ECG electrical signals, respiration electrical signals, invasive blood pressure (IBP) electrical signals and non-invasive blood pressure pulses. The ProSim is a lightweight, battery powered unit that is portable enough to test a patient monitor anywhere the monitor is being used.

ProSim vital signs simulator consists of the following components:

  1. Printed Circuit Board Assemblies using surface mount components and firmware loaded in embedded processors.
  2. Plastic iniection molded case parts.
  3. Stepper Motor and piston pump for pneumatic simulation that makes reliable pressure pulses.
  4. Liquid Crystal Touch Screen Display for user interface. User interface follows modern and ergonomic concepts.
  5. Lithium Ion rechargeable battery for portable operation, giving user flexibility and portability.

Fluke Biomedical's ProSim 6 and ProSim 8 (hereafter referred to as the ProSim) provides a basis to train, evaluate, and perform preventive maintenance for virtually all patient monitors found in the healthcare industry. This is accomplished with multiple physiological simulations for ECG electrical signals, respiration electrical signals, invasive blood pressure (IBP) electrical signals, non-invasive blood pressure (NIBP) pressure pulses, temperature electrical signal, cardiac output electrical signal, and pulse oximetry SPO2 optical simulated light attenuation. The ProSim is a lightweight, battery powered unit that is portable enough to test a patient monitor anywhere the monitor is being used.

ProSim vital signs simulator consists of the following components:
l ) Printed Circuit Board Assemblies using surface mount components and firmware loaded in embedded processors.

  • 2) Plastic injection molded case parts.
    1. Stepper Motor and piston pump for pneumatic simulation that makes reliable pressure pulses.
    1. Liquid Crystal Display for user interface. User interface follows modern and ergonomic concepts.
    1. Lithium Ion rechargeable battery for portable operation, giving user flexibility and portability.
AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Fluke Biomedical ProSim 4, ProSim 6, and ProSim 8 devices:

Executive Summary:

The Fluke Biomedical ProSim 4, ProSim 6, and ProSim 8 are vital signs simulators designed to test and verify the basic operation of patient monitoring devices. The devices function by providing electronic and pneumatic simulations of various physiological parameters. Extensive non-clinical bench testing was conducted to demonstrate that the ProSim devices perform within their published specifications and are substantially equivalent to their predicate devices (MedSim300B, Cufflink, and Index 2MF SPO2 for ProSim 6/8). Clinical testing was not performed, as the devices are intended for laboratory use for preventative maintenance and not for direct patient use or medical equipment calibration. The ground truth for the testing was the published specifications of the device itself, verified through laboratory measurements.


1. Table of Acceptance Criteria and the Reported Device Performance:

The document provides a detailed comparison table between the ProSim devices and their predicate devices, indicating various performance characteristics. Since the acceptance criteria are essentially the published specifications and the study aims to show the device performs within those specifications, the "reported device performance" is implicitly that the device met these specifications, thus demonstrating substantial equivalence.

Below is a summarized table consolidating key features and their performance (or range of performance), with the acceptance criteria being that the ProSim meets or improves upon the predicate device's comparable specification or meets its own stated specifications if it's a new feature or different range.

ProSim 4 (compared to MedSim 300B and Cufflink)

FeaturePredicate Devices (Range/Specification)ProSim 4 (Reported Performance/Specification)Acceptance Criteria
Display2x24 char LCD (MedSim); 8x20 char alpha & 64x240 graphics (Cufflink)¼ VGA graphic LCD Touch Color DisplayMore display, Touch Screen, and Color – Met/Exceeded
IBP Channels4 channels (MedSim)1 channelReduced per market requirements and use – Met
Comm PortRS232USBAdvancement in technology – Met
Battery TestLimited low battery indication (MedSim); No battery (Cufflink)Multiple levels of battery life indicationPredicate checks at one level, ProSim checks at multiple levels – Met/Exceeded
Power2x9V alkaline (MedSim); AC line (Cufflink)Li-Ion rechargeable battery w/ low battery indicator; battery eliminator (115VAC)Longer operating life with modern battery technology – Met/Exceeded
Amplitude accuracy (ECG)+/- 5%, 2Hz @ 1.0 mV p-p SQ wave Lead II (MedSim)+/- 2% setting lead IIMore accurate on newer devices – Met/Exceeded
NSR rates (ECG)30 to 300 BPM (MedSim)30 to 320 BPMWider range due to market preferences – Met/Exceeded
NSR amplitudes (ECG)50 uV to 5.5mV (MedSim)1mVLimited amplitude for basic simulation – Met
Pacemaker0.1 to 2.0 ms width, -700 to +700 mV (MedSim)1 ms width, 3mVBasic simulation for targeted market – Met
Impedance variation (Resp)0 to 3 ohms (MedSim)1.0 ohmBasic simulation for targeted market – Met
Respiration rates15, 20, 30, 40, 60, 120 brpm (MedSim)10, 20, 30, 40, 50, 60, 70, 80, 90 & 100 brpmMore Respiration rates – Met/Exceeded
Apnea (Resp)Off, Continuous, momentary, 12 & 32 s (MedSim)Off & ContinuousLess apneas for basic simulation – Met
Transducer Sensitivity (IBP)5 or 40 uV/V/mmHg (MedSim)5 uV/V/mmHgFewer selections for basic simulation – Met
Static pres. Selection (IBP)Manual and automatic (MedSim)ManualLimited selection mode per target market and use – Met
Dynamic BP selections (IBP)Arterial, left and right ventricle, pulmonary artery, pulmonary wedge, Swan-Ganz (MedSim)Arterial and left ventricleFewer selections for basic simulation – Met
Static BP selections (IBP)-10, -5, 0, 20, 30, 40, 80, 100, 200, 250 & 300 mmHg (MedSim)0, 80, 160 & 250 mmHgFewer selections for basic simulation – Met
Manometer (NIBP)N/A (MedSim); Max. 499.75 mmHg (Cufflink)0 to 400 mmHgLower range per market requirements – Met
Simulation (NIBP)N/A (MedSim); Systolic/Diastolic simulations. Adult 60/30 to 255/195 (Cufflink)Systolic/Diastolic Adult – 60/30, 120/80, 150/100 & 200/150; Neonatal 35/15 & 70/40Fewer selections for basic simulation; Adult and Neonatal – Met
Synchronization to ECG (NIBP)N/A (MedSim); 30 to 240 BPM (Cufflink)Up to 150 BPMLimited for basic simulation – Met

ProSim 6 & 8 (compared to MedSim 300B, Index 2, and Cufflink)

FeaturePredicate Devices (Range/Specification)ProSim 6 & 8 (Reported Performance/Specification)Acceptance Criteria
Intended UseMedSim: ECG, Resp, BP, Temp, CO; Index 2: SpO2; Cufflink: NIBPECG, Resp, IBP, NIBP, Temp, CO, SpO2 (Fetal Sim for ProSim 8)Additional functions of noninvasive blood pressure and pulse oximetry simulation. – Met/Exceeded
DisplayMedSim: 2x24 char LCD; Index 2: 2x24 char LCD; Cufflink: 8x20 char alpha & 64x240 graphics¼ VGA graphic LCD Color DisplayMore display, Color. – Met/Exceeded
High level ECG¼" standard phone jack w/ lead II waveform at .2V/mV (MedSim)BNC jack for 0.5V/mV output into 50 Ohm impedanceOutput to oscilloscope via BNC is preferred by customers. – Met/Exceeded
IBP Channels4 independent BP channels (MedSim)2 channelsNumber of channels reduced per market requirements and use. – Met
Comm PortRS232 (all predicates)USBChange from RS232 to USB data port with advancement in technology. – Met/Exceeded
Battery TestLimited low battery indication (MedSim, Index 2); No battery (Cufflink)Multiple levels of battery life indicationPredicate devices only check at one level. ProSim checks battery status at multiple charge levels. – Met/Exceeded
Amplitude accuracy (ECG)+/- 5%, 2Hz @ 1.0 mV p-p SQ wave Lead II (MedSim)+/- 2% setting lead IIMore accurate on newer devices due to market preferences and technology improvements. – Met/Exceeded
NSR rates (ECG)30 to 300 BPM (MedSim)30 to 360 BPMWider range due to market preferences. – Met/Exceeded
Square and/or Pulse wavesSquare at 2 Hz and 0.125 Hz (MedSim)Square at 2.5, 2 & 0.125 Hz. Pulse at 60ms / 60 and 30 BPMMore choices. – Met/Exceeded
Sine waves0.05 to 100 Hz (MedSim)0.05 to 150 HzMore choices. – Met/Exceeded
Triangle wave2 Hz (MedSim)0.125, 2 and 2.5 HzMore choices. – Met/Exceeded
R Wave detectorYes (MedSim)YesNone. – Met
QRS Detection and Tall T-wave rejectionNo (MedSim)YesImproved capability. – Met/Exceeded
Impedance variation (Resp)0 to 3 ohms (MedSim)0 to 5 ohmsExpanded capability to meet new market requirements. – Met/Exceeded
Respiration rates15, 20, 30, 40, 60, 120 brpm (MedSim)15 to 120 brpm in incremental stepsMore Respiration rates due to market requirements and use. – Met/Exceeded
Apnea (Resp)Off, Continuous, momentary, 12 & 32 s (MedSim)Off, Continuous, momentary, 12, 22 & 32 SAdditional apneas. – Met/Exceeded
Blood temperatures (CO)36C to 38C and user programmable (MedSim)36C to 38C in incremental stepsUser programmable not included due to market requirements and use. – Met
Injective temp (CO)Chilled (2C) (MedSim)Chilled (0C) or 24 CMore selections per market requirements. – Met/Exceeded
Fixed blood flow rate (CO)3, 5, 7 L/min (MedSim)2.5, 5, 10 L/minDifferent selections per market requirements. – Met
Output trend (CO)1 normal, 2 defective (MedSim)NoNot included due to market requirements and use. – Met
Exciter range (IBP)2 to 16 V/DC to 4 kHz (MedSim)2 to 16 V/DC to 5kHzHigher frequency range driven by market trend and technology. – Met/Exceeded
Static pres. Selection (IBP)Manual and automatic (MedSim)ManualLimited selection mode per target market and use. – Met
Dynamic BP selections (IBP)Arterial, left and right ventricle, pulmonary artery, pulmonary wedge, Swan-Ganz (MedSim)Arterial, radial artery, left and right ventricle, pulmonary artery, pulmonary wedge, right atrium, left atrium and Swan-GanzMore selections. – Met/Exceeded
Static BP selections (IBP)-10, -5, 0, 20, 30, 40, 80, 100, 200, 250 & 300 mmHg (MedSim)-10 to 300 mmHg in incremental stepsMore selections. – Met/Exceeded
Temperature (Sim)0, 24, 37 and 40C (MedSim)30C to 42C in incremental stepsDifferent selections per market requirements. – Met
SpO2 AccuracySelect 50% to 100% in 1% increments - accuracy: 75%to 100% ± 1% plus the accuracy of the pulse oximeter under test. 50%-75%, ± 2% plus accuracy of the oximeter under test. Under 50%, unspecified. (Index 2)Select 30% to 100% in 1% increments w/ oximeter manufacturer's R-curve or Fluke Biomedical R-curves (91-100% ±(3 counts + UUT accuracy), 81-90% ±(5 counts + UUT accuracy), 71-80% ±(7 counts + UUT accuracy), <70% monotonic unfilled accuracy)More selections per market requirements. – Met/Exceeded
SpO2 Test featuresTransmission Light Control (TLC) feature for different light attenuation (Index 2)Transmission of light selected through selection of finger type: light finger, thick dark finger or neonate.Better user interface and understanding of transmission with ProSim selection of light transmission levels. – Met/Exceeded
Magnetic HolderN/A (Index 2)YesIncludes tested magnetic holder for SPO2 module. – Met/Exceeded
Simulation (NIBP)Systolic/Diastolic simulations. Adult 60/30 to 255/195 (Cufflink)Systolic/Diastolic Adult 60/30 to 255/195; Neonatal 35/15 to 255/195Adult and Neonatal available. – Met/Exceeded
Arrhythmias (NIBP)Premature atrial contraction, Premature ventricular contraction, Atrial fibrillation, Missed Beat, aberrant Sinus conduction (Cufflink)Premature atrial contraction, Premature ventricular contraction, Atrial fibrillation, Missed Beat.No aberrant Sinus conduction. Not needed per market use. – Met
Synchronization to ECG (NIBP)30 to 240 BPM (Cufflink)30 to 240 BPMNone. – Met

2. Sample Size Used for the Test Set and the Data Provenance:

  • Sample Size: The document does not specify a numerical sample size for individual tests (e.g., how many measurements were taken for each parameter). It states that "Laboratory studies have been conducted with a representative patient monitor..." and "The ProSim [4/6/8] software has been successfully validated..." This implies that the testing involved a single representative device (the ProSim itself) undergoing various tests against its own specifications or in comparison to predicate devices.
  • Data Provenance: The studies were non-clinical laboratory studies conducted by Fluke Biomedical. The location of the laboratory is likely the manufacturer's facility in the USA (Fluke Biomedical, Everett, WA, or Solon, OH). The data is retrospective in the sense that it's a summary of completed bench tests demonstrating performance.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

  • Number of Experts: The document does not specify the number of experts.
  • Qualifications of Experts: The ground truth for this type of device, a vital signs simulator, is typically based on electrical and pneumatic engineering principles and established medical device standards for physiological parameter simulation (e.g., AAMI, IEC standards for ECG, BP). The testing would be performed by trained technicians or engineers proficient in metrology and test equipment, verifying the device's output against known, calibrated reference standards. While not explicitly stated, it's implied these are qualified personnel within Fluke Biomedical's R&D or Quality Assurance departments.

4. Adjudication Method for the Test Set:

  • Adjudication Method: Not applicable in the traditional sense of human consensus for, say, image interpretation. The "adjudication" is based on objective measurement against defined performance specifications and comparison to predicate devices. The results are summarized by the manufacturer.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done:

  • No, an MRMC study was not done. This type of study is typically performed for diagnostic devices where human readers interpret medical images or data. The ProSim devices are test equipment for patient monitors, not diagnostic tools. Their effectiveness is measured by their ability to accurately simulate physiological signals for verification and maintenance, not by improving human diagnostic performance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

  • Yes, a standalone study was done. The entire "Non-Clinical Test Data" section describes rigorous bench testing of the ProSim devices themselves, which are algorithmic-driven instruments, to verify their performance within published specifications. The statement "The ProSim [4/6/8] software has been successfully validated to confirm the performance of the device" directly supports this. The device's function is standalone simulation; there isn't a "human-in-the-loop" once the simulation parameters are set.

7. The Type of Ground Truth Used:

  • Engineering Specifications and Calibrated Reference Standards: The ground truth used is the device's own published specifications (e.g., amplitude accuracy, rate ranges, impedance values) and the measured performance of predicate devices. Performance was verified against these using calibrated test equipment and established metrology practices.

8. The Sample Size for the Training Set:

  • Not applicable. The ProSim devices are simulators, not AI/machine learning models that require a training set. They are designed based on established physiological principles and engineering design.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable. As stated above, there is no "training set" in the context of an AI/ML model for these devices.

{0}------------------------------------------------

K110429

SEP - 2 2011

Image /page/0/Picture/2 description: The image shows the logo for Fluke Biomedical. The word "FLUKE" is in large, bold, sans-serif font. Below the word "FLUKE" is the word "Biomedical" in a smaller, bold, sans-serif font. There is a line above the word "Biomedical".

Section III -510(k) Summary of Safety and Effectiveness

Submitter:

Fluke Biomedical 6920 Seaway Blvd Everett WA. 98203 (440) 498-2579 -Phone (440-349-2307) - Fax John Nelson -Contact Person

Device Name:

  • ◆ Trade Name - ProSim 4
  • . Common Name -Vital Signs Simulator
  • Classification Name Monitor, Cardiac /System, Measurement, Blood-Pressure, Non-Invasive per . 21 CFR 870.2300/870.1130
  • Product Codes -DRT, DXN .

Devices for Which Substantial Equivalence is Claimed:

  • MedSim300B Submitted as MedSim300 under 510(k) K935817
  • Cufflink .

Device Description:

Principles of Operation

Fluke Biomedical's ProSim 4 (hereafter referred to as the ProSim) provides a basis to train, evaluate, and perform preventive maintenance for virtually all patient monitors found in the healthcare industry. This is accomplished with multiple physiological simulations for ECG electrical signals, respiration electrical signals, invasive blood pressure (IBP) electrical signals and non-invasive blood pressure pulses. The ProSim is a lightweight, battery powered unit that is portable enough to test a patient monitor anywhere the monitor is being used.

Technological Characteristics

ProSim vital signs simulator consists of the following components:

  1. Printed Circuit Board Assemblies using surface mount components and firmware loaded in embedded processors.

  2. Plastic iniection molded case parts.

  3. Stepper Motor and piston pump for pneumatic simulation that makes reliable pressure pulses.

  4. Liquid Crystal Touch Screen Display for user interface. User interface follows modern and ergonomic concepts.

  5. Lithium Ion rechargeable battery for portable operation, giving user flexibility and portability.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Fluke Biomedical. The word "FLUKE" is in large, bold, sans-serif font on a gray background. Below the word "FLUKE" is the word "Biomedical" in a smaller, bold, sans-serif font. There is a black line above the word "Biomedical".

Intended Use of the Device:

The intended us of ProSim 4 is to test and verify the basic operation of patient monitoring devices or systems used to monitor various physiological parameters of a patient, including ECG, Respiration, Invasive blood pressure and Non-invasive blood pressure.

The intended user is a trained biomedical equipment technician who is performing periodic preventative maintenance checks on patient monitors in service. Users can be associated with hospitals, clinics, original equipment manufacturers and independent service companies that repair and service medical equipment. The end user is a technically trained individual, specializing in medical instrumentation technology.

The ProSim is intended to be used in the laboratory environment and is not intended for use on patients, or to test devices while connected to patients. This product line is not intended to be used to calibrate medical equipment.

ProSim is intended for over-the counter use.

Summary of Technological Characteristics:

The ProSim 4 is substantially equivalent to one other legally marketed device in the United States. The ProSim functions in a manner similar to and is intended for the same use as the MedSim300B and Cufflink manufactured by Fluke Biomedical.

The ProSim 4 is similar to the MedSim300B and Cufflink in that it is a cordless battery-operated device, uses LCD display, and allows user to simulate physiological parameters to verify the operation of patient monitors. The ProSim differs from the MedSim300B and Cufflink in that the ProSim combines the features of each of these devices into one device and is touch screen operated.

FeaturesProSim 4MedSim 300B(K935817)Cufflink(K942546)Difference
Intended UseThe intended use ofProSim 4 is to test andverify the basicoperation of patientmonitoring devices orsystems used to monitorvarious physiologicalparameters of a patient,including ECG,Respiration, Invasiveblood pressure andNon-Invasive bloodpressure.The intended user is atrained biomedicalequipment technicianwho is performingperiodic preventativemaintenance checks onpatient monitors inservice. Users can beassociated withHospitals, clinics,original equipmentmanufacturers orindependent servicecompanies that repairand service medicalTo test operationof patient monitorsby simulatingphysiologicalparameters,including: ECG,respiration, bloodpressure,temperature andcardiac output.To test operation ofNon-Invasive BloodPressure (automatedSphygmomanometerssimulator).None
equipment. The enduser is a technicallytrained individual,specializing in medicalinstrumentationtechnology.The ProSim product lineis intended to be used inthe laboratoryenvironment and is notintended for use onpatients, or to testdevices while connectedto patients. This productline is not intended to beused to calibratemedical equipment andnot intended for over thecounter use.
ConstructionPlastic case.Aluminum case.Aluminum case.Lighter morecompact plasticcasing.
SizeProSim 4: 7.1 L x 3.7 Wx 2.2 H inches.10 L x 7 W x 3 Hinches.15 L x 12.5 W x 5 Hinches.Combinationinstrumentsmaller thansum ofpredicatedevices.
WeightProSim 4: 1.9 lbs.3.55 lbs.15 lbs.Lighter.
Display¼ VGA graphic LCDTouch Color Display.2 by 24 characterLCD.8 by 20 characteralphanumericdisplay & 64 by 240graphical display.More display,Touch Screenand Color.
Function KeyProSim 4: Touchscreen.Soft.Soft.ProSim 4:Touchscreen.
ECG leads10 binding posts;compatible w/disposable snaps, 3.2mm or 4.0 mmelectrodes, and bananaplugs (with or withoutadapter).10 binding posts;compatible w/disposable snaps.3.2 mm or 4.0 mmelectrodes, andbanana plugs.N/ANone.
IBP ChannelsIndependent BPchannels w/ sensitivitycontrol (5 or 40uV/V/mmHg); cableinterface w/ monitors.ProSim 4: 1 channel.4 independent BPchannels w/sensitivity control(5 or 40uV/V/mmHg);cable interface w/monitors.N/ANumber ofchannelsreduced permarketrequirementsand use.
CommunicationsPortUSB.RS232.RS232.Change fromRS232 to USBdata port withadvancement intechnology.
Battery testMultiple levels of battery life indication.Limited low battery indication.No battery.Predicate devices only check at one level. ProSim checks battery status at multiple charge levels.
PowerLi-Ion rechargeable battery w/ low battery indicator; or battery eliminator (115VAC) transformer certified to CSA.2 X 9V alkaline battery w/ low battery indicator; or battery eliminator (115VAC) transformer certified to CSA C22.2. 231 series M89).No Battery- AC line powered only.Longer operating life with modern battery technology.
Lead configuration12 leads.12 leads.N/ANone.
Output impedances500 to 2000ohms to RL.500 to 2000ohms to RL.N/ANone.
Amplitude accuracy+/- 2% setting lead II.+/- 5%, 2Hz @ 1.0 mV p-p SQ wave Lead II.N/AMore accurate on newer devices due to market preferences and technology improvements.
NSR ratesProSim 4: 30 to 320 BPM.30 to 300 BPM.N/AWider range due to market preferences.
NSR amplitudesProSim 4: 1mV50 uV to 5.5mV.N/AProSim 4: Limited amplitude for basic simulation.
Pediatric or Neonatal ECGR Wave width reduced to 40 ms.R Wave width reduced to 40 ms.N/ANone.
Square and/or Pulse wavesProSim 4: Pulse at 60ms / 2Hz.Square at 2 Hz and 0.125 Hz.N/AProSim 4: Provides what is required for a basic simulation.
PacemakerProSim 4: 1 ms width, 3mV.0.1 to 2.0 ms width, -700 to +700 mVN/AProSim 4: Basic simulation for targeted market.
Cable connectorECG leads, 10 binding postings.ECG leads, 10 binding postings.N/ANone.
Normal baseline impedances500 to 2000 ohms ref. to RL.500 to 2000 ohms ref. to RL.N/ANone.
Lead selectionsLA or LL.I or II (LA or LL).N/ANone.
Impedance variationProSim 4: 1.0 ohm0 to 3 ohms.N/AProSim 4: Basic simulation for targeted market.
RespirationratesProSim 4: 10, 20, 30,40, 50, 60, 70, 80, 90 &100 brpm.15, 20, 30, 40, 60,120 brpm.N/AMoreRespirationrates due tomarketrequirementsand use.
ApneaProSim 4: Off &Continuous.Off, Continuous,momentary, 12 &32 s.N/AProSim 4: Lessapneas forbasic simulationand targetmarket.
Cable connectorECG leads, bindingposts.ECG leads,binding posts.N/ANone
I/O impedance300 ohms.300 ohms.N/ANone.
Exciter range2 to 16 V/DC to 5kHz.2 to 16 V/DC to 4kHzN/AHigherfrequencyrange driven bymarket trendand technology.
TransducerSensitivityProSim 4: 5 uV/V/mmHg5 or 40uV/V/mmHg.N/AProSim 4:Fewerselections forbasic simulationand targetmarket.
Level accuracy+/- (1% setting +1mmHg).+/- 1% full scale;+/- 1mmHg.N/ANone.
Static pres.SelectionManual.Manual andautomatic.N/ALimitedselection modeper targetmarket and use.
Dynamic BPselectionsProSim 4: Arterial andleft ventricle.Arterial, left andright ventricle,pulmonary artery,pulmonary wedge,Swan-Ganz.N/AProSim 4:Fewerselections forbasic simulationand targetmarket.
Static BPselectionsProSim 4: 0, 80, 160 &250 mmHg-10, -5, 0, 20, 30,40, 80, 100, 200,250 & 300 mmHg.N/AProSim 4:Fewerselections forbasic simulationand targetmarket.
Cable connectorDIN style.DIN style.N/ANone.
Manometer0 to 400 mmHgN/AMax. 499.75 mmHgLower rangefollowingmarketrequirements.
Leak TestSource pressure, sealoff, measure change inpressure over time.N/ASource pressure,seal off, measurechange in pressureover time.None.
Over PressureTestIncrease pressure untildevice under tests ventsto atmosphere.N/AIncrease pressureuntil device undertests vents toatmosphere.None.
SimulationProSim 4:Systolic/Diastolic Adult –60/30, 120/80, 150/100& 200/150; Neonatal35/15 & 70/40.N/ASystolic/Diastolicsimulations. Adult60/30 to 255/195.ProSim 4:Fewerselections forbasicsimulation.Adult andNeonatal.
Synchronizationto ECGProSim 4: Up to 150BPMN/A30 to 240 BPMProSim 4:Limited forbasicsimulation.

. 0000006

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "FLUKE" in a bold, sans-serif font. Below the word "FLUKE" is the word "Biomedical" in a smaller, sans-serif font. The word "Biomedical" is underlined with a thick, black line. The background of the image is a light gray color.

1/4

〜 000007

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "FLUKE" in large, bold, black letters. Below the word "FLUKE" is the word "Biomedical" in a smaller, bold, black font. There is a black line above the word "Biomedical".

1/4

Image /page/3/Picture/2 description: The image shows a sequence of numbers, specifically "000008". The numbers are printed in a bold, sans-serif font. The image appears to be a close-up of the numbers, with some minor imperfections or artifacts visible around the edges of the numbers.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "FLUKE" in large, bold letters. Below the word "FLUKE" is the word "Biomedical" in a smaller font. There is a black line above the word "Biomedical".

. په چهاچ پ

5/14

. 000009

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "FLUKE" in large, bold, black letters. Below the word "FLUKE" is the word "Biomedical" in a smaller font. The background of the image is a light color, and there are some dark spots around the letters.

Non-Clinical Test Data:

Laboratory studies have been conducted with a representative patient monitor to verify and validate the ProSim 4 will perform within its' published specifications.

  • NPI-02042011-00001 ProSim 4 Bench test summary and results. .
    The ProSim 4 software has been successfully validated to confirm the performance of the device.

Clinical Test Data:

Clinical testing has not been conducted on this product.

Conclusion:

Based upon the laboratory studies, similar technological/performance characteristics as compared to the predicate devices, and successful validation of the ProSim 4 software, the performance of the ProSim 4 is deemed to be substantially equivalent to the MedSim300B and Cufflink.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Fluke Biomedical. The logo consists of the word "FLUKE" in a bold, sans-serif font, with the word "Biomedical" underneath it in a smaller font. The logo is surrounded by a textured background, which gives it a slightly distressed look. The image also contains a handwritten number, "7/".

Device Name:

  • Trade Name -ProSim 6, ProSim 8 .
  • . Common Name -Vital Signs Simulator
  • Classification Name Monitor, Cardiac /System, Measurement, Blood-Pressure, Non-Invasive per . 21 CFR 870.2300/870.1130
  • Product Codes -- DRT, DXN .

Devices for Which Substantial Equivalence is Claimed:

  • MedSim300B Submitted as MedSim300 under 510(k) K935817 .
  • Index 2MF SPO2 Originally submitted and cleared under 510(k) K933519. Currently marketed . as Index 2MF which was ruled as a general purpose device on Feb 11, 1998 (K974293)
  • . Cufflink

Device Description:

Principles of Operation

Fluke Biomedical's ProSim 6 and ProSim 8 (hereafter referred to as the ProSim) provides a basis to train, evaluate, and perform preventive maintenance for virtually all patient monitors found in the healthcare industry. This is accomplished with multiple physiological simulations for ECG electrical signals, respiration electrical signals, invasive blood pressure (IBP) electrical signals, non-invasive blood pressure (NIBP) pressure pulses, temperature electrical signal, cardiac output electrical signal, and pulse oximetry SPO2 optical simulated light attenuation. The ProSim is a lightweight, battery powered unit that is portable enough to test a patient monitor anywhere the monitor is being used.

Technological Characteristics

ProSim vital signs simulator consists of the following components:

l ) Printed Circuit Board Assemblies using surface mount components and firmware loaded in embedded processors.

    1. Plastic injection molded case parts.
    1. Stepper Motor and piston pump for pneumatic simulation that makes reliable pressure pulses.
    1. Liquid Crystal Display for user interface. User interface follows modern and ergonomic concepts.
    1. Lithium Ion rechargeable battery for portable operation, giving user flexibility and portability.

Intended Use of the Device:

The intended us of ProSim 6 and ProSim 8 is to test and verify the basic operation of patient monitoring devices or systems used to monitor various physiological parameters of a patient, including ECG, Respiration, Invasive blood pressure, Non-invasive blood pressure and Cardiac output. Additionally, the devices provide an optical signal to verify that the electronics within the pulse oximeter probe are functional.

The intended user is a trained bioment technician who is performing periodic preventative maintenance checks on patient monitors in service. Users can be associated with hospitals, cliginal equipment manufacturers and independent service companies that repair and service medical equipment. The end user is a technically trained individual, specializing in medical instrumentation technology,

000011

{7}------------------------------------------------

The ProSim is intended to be used in the laboratory environment and is not intended for use on patients, or to test devices while connected to patients. This product line is not intended to be used to calibrate medical equipment.

ProSim is intended for over-the counter use.

Summary of Technological Characteristics:

The ProSim is substantially equivalent to one other legally marketed device in the United States. The ProSim functions in a manner similar to and is intended for the same use as the MedSim300B, Index 2 and Cufflink manufactured by Fluke Biomedical.

The ProSim is similar to the MedSim300B, Index 2 and Cufflink in that it is a cordless battery-operated device, uses LCD display, and allows user to simulate physiological parameters to verify the operation of patient monitors. The ProSim differs from the MedSim300B, Index 2 and Cufflink in that the ProSim combines the features of each of these devices into one device.

FeaturesProSim 6 &ProSim 8MedSim 300B(K935817)Index 2(K 933519)Cufflink(K942546)Difference
IntendedUseThe intended use ofProSim 6 andProSim 8 is to testand verify the basicoperation of patientmonitoring devicesor systems used tomonitor variousphysiologicalparameters of apatient, includingECG, Respiration,Invasive bloodpressure, Non-invasive bloodpressure,Temperature,Cardiac output andSpO2.The intended user isa trained biomedicalequipmenttechnician who isperforming periodicpreventativemaintenance checkson patient monitorsin service. Users canbe associated withHospitals, clinics,original equipmentmanufacturers orindependent servicecompanies thatrepair and servicemedical equipment.The end user is atechnically trainedindividual,specializing inTo test operationof patient monitorsby simulatingphysiologicalparameters,including: ECG,respiration, bloodpressure,temperature andcardiac output.To test operation ofPulse Oximeters bysimulating the visibleand infrared lightabsorption.To test operationofNon-Invasive BloodPressure (automatedSphygmomanometerssimulator).Additionalfunctions ofnoninvasiveblood pressureand pulseoximetrysimulation
medicalinstrumentationtechnology.The ProSim productline is intended tobe used in thelaboratoryenvironment and isnot intended for useon patients, or totest devices whileconnected topatients. Thisproduct line is notintended to be usedto calibrate medicalequipment and notintended for overthe counter use.
ConstructionPlastic case.Aluminum case.Plastic case.Aluminum case.Lighter morecompactplastic casing
Size5.7 L x 11.9 W x 3.4H inches.10 L x 7 W x 3 Hinches.10 L x 10.5 W x 4 Hinches.15 L x 12.5 W x 5 Hinches.Combinationinstrumentsmaller thansum ofpredicatedevices.
Weight4.1 lbs.3.55 lbs.4 lbs.15 lbs.Lighter.
Display¼ VGA graphicLCD Color Display.2 by 24 characterLCD.2 by 24 characterLCD.8 by 20 characteralphanumericdisplay & 64 by 240graphical display.More display,Color.
FunctionKeySoft.Soft.Soft.Soft.None.
ECG leads10 binding posts;compatible w/disposable snaps,3.2 mm or 4.0 mmelectrodes, andbanana plugs (withor without adapter).10 binding posts;compatible w/disposable snaps,3.2 mm or 4.0 mmelectrodes, andbanana plugs.N/AN/ANone.
High levelECGBNC jack for0.5V/mV outputinto 50 Ohmimpedance.¼" standard phonejack w/ lead IIwaveform at.2V/mV of ECGlead II signal. Usew/ analog input,high level, centralstation monitors orrecorders.N/AN/AOutput tooscilloscopevia BNC ispreferred bycustomers.
IBPChannelsIndependent BPchannels w/sensitivity control (5or 40 uV/V/mmHg);cable interface w/monitors.2 channels.4 independent BPchannels w/sensitivity control(5 or 40uV/V/mmHg);cable interface w/monitors.N/AN/ANumber ofchannelsreduced permarketrequirementsand use.
(500 – 2000)control; lead selectcontrol.Impedance (500 –2000) control; leadselect control.
TemperatureYes, fixed temp.probe select control(400 or 700 YSI)series probes.Yes, fixed orvariable temp.probe selectcontrol (400 or 700YSI) series probes.N/AN/ANone.
CardiacoutputYes, cable connectw/ monitor.Yes, cable connectw/ monitor.N/AN/ANone.
Communications PortUSB.RS232.RS232.RS232.Change fromRS232 toUSB data portwithadvancementin technology.
Battery testMultiple levels ofbattery lifeindication.Limited lowbattery indication.Limited low batteryindication.No battery.Predicatedevices onlycheck at onelevel. ProSimchecks batterystatus atmultiplecharge levels.
PowerLi-Ion rechargeablebattery w/ lowbattery indicator; orbattery eliminator(115VAC)transformer certifiedto CSA.2 X 9V alkalinebattery w/ lowbattery indicator; or batteryeliminator(115VAC)transformercertified to CSAC22.2. 231 seriesM89).Rechargeable LeadAcid battery.No Battery- AC linepowered only.Longeroperating lifewith modernbatterytechnology.
ECG
Leadconfiguration12 leads.12 leads.N/AN/ANone.
Outputimpedances500 to 2000ohms to RL.500 to2000ohmsto RL.N/AN/ANone.
Amplitudeaccuracy+/- 2% setting lead II.+/- 5%,2Hz @ 1.0mV p-p SQwave LeadII.N/AN/AMore accurateon newerdevices due tomarketpreferences andtechnologyimprovements.
NSR rates30 to 360 BPM.30 to 300BPM.N/AN/AWider range dueto marketpreferences.
NSRamplitudes50 uV to 5.0mV.50 uV to5.5mV.N/AN/ANone.
ST Segments-0.8 to +0.8 mV.-0.8 to +0.8 mV.N/AN/ANone.
Axis deviationIntermediate, horizontaland vertical.Intermediate,horizontalandvertical.N/AN/ANone.
Pediatric orNeonatal ECGR Wave width reduced to40 ms.R Wavewidthreduced to40 ms.N/AN/ANone.
Performance Test
Square and/orPulse wavesSquare at 2.5, 2 & 0.125Hz. Pulse at 60ms / 60and 30 BPMSquare at 2Hz and0.125 Hz.N/AN/AMore choices.
Sine waves0.05 to 150 Hz.0.05 to 100Hz.N/AN/AMore choices.
Triangle wave0.125, 2 and 2.5 Hz.2 Hz.N/AN/AMore choices.
R WavedetectorYes.Yes.N/AN/ANone
QRS Detectionand Tall T-wave rejection.YesNoN/AN/ANone
Pacemaker0.1 to 2.0 ms width, -700to +700 mV0.1 to 2.0ms width, -700 to +700mVN/AN/ANone
CableconnectorECG leads, 10 bindingpostings.ECG leads,10 bindingpostings.N/AN/ANone.
Respiration
Normalbaselineimpedances500 to 2000 ohms ref. toRL.500 to 2000ohms ref. toRL.N/AN/ANone.
LeadselectionsLA or LL.I or II (LAor LL).N/AN/ANone.
Impedancevariation0 to 5 ohms.0 to 3ohms.N/AN/AExpandedcapability tomeet newmarketrequirements.
Respirationrates15 to 120 brpm inincremental steps.15, 20, 30,40, 60, 120brpm.N/AN/AMoreRespiration ratesdue to marketrequirementsand use.
ApneaOff, Continuous,momentary, 12, 22 & 32S.Off,Continuous,momentary,12 & 32 s.N/AN/AAdditionalapneas.
CableconnectorECG leads, binding posts.ECG leads,bindingposts.N/AN/ANone
Cardiac Output
Catheter sizeFixed, 7F injective vol.10 cc.Fixed, 7Finjectivevol. 10 cc.N/AN/ANone.
Bloodtemperatures36C to 38C inincremental steps.36C to 38Cand userprogrammable.N/AN/AUserprogrammablenot included dueto marketrequirementsand use.
Injective tempChilled (0C) or 24 C.Chilled(2C).N/AN/AMore selectionsper marketrequirements.
Fixed bloodflow rate2.5, 5, 10 L/min.3, 5, 7L/min.N/AN/ADifferentselections permarketrequirements.
CurvesNormal, faulty and L/Rshunt.Normal,interrupt,slow, L/Rshunt.N/AN/ANone.
Output trendNo.1 normal, 2defective.N/AN/ANot includeddue to marketrequirementsand use.
Bath/InjectiveresistanceContinuously variable, 3pin standard.Continuously variable,3 pinstandard.N/AN/ANone.
CableconnectorBlood Temp - AmericanEdward, 3 pin standard;Injective Temp -American Edward, 4 pinstandard.BloodTemp -AmericanEdward, 3pinstandard;InjectiveTemp -AmericanEdward, 4pinstandard.N/AN/ANone
Invasive Blood Pressure.
I/O impedance300 ohms.300 ohms.N/AN/ANone.
Exciter range2 to 16 V/DC to 5kHz.2 to 16V/DC to 4kHzN/AN/AHigherfrequency rangedriven by markettrend andtechnology.
TransducerSensitivity5 or 40 uV/V/mmHg.5 or 40uV/V/mmHg.N/AN/ANone.
Level accuracy+/- (1% setting +1mmHg).+/- 1% fullscale; +/-1mmHg.N/AN/ANone.
Static pres.SelectionManual.Manual andautomatic.N/AN/ALimitedselection modeper target marketand use.
Dynamic BPselectionsArterial, radial artery, leftand right ventricle,pulmonary artery,pulmonary wedge, rightatrium, left atrium andSwan-Ganz.Arterial,left andrightventricle,pulmonaryartery,pulmonarywedge,Swan-Ganz.N/AN/AMore selections.
Static BPselections-10 to 300 mmHg inincremental steps.-10, -5, 0,20, 30, 40,80, 100,200, 250 &300 mmHg.N/AN/AMore selections.
CableconnectorDIN style.DIN style.N/AN/ANone.
Temperature30C to 42C inincremental steps.0, 24, 37and 40C.N/AN/ADifferentselections permarketrequirements.
ProbecompatibilitySeries 400 and 700.Series 400and 700.N/AN/ANone.
CableconnectorDIN Style.DIN Style.N/AN/ANone.
Oximeter SPO2 optical emitter and detector
SpO2 R-CurveselectionSelect R-Curve frommenu of choices.N/ASelect R-Curvefrom menu ofchoices.N/ANone.
SpO2 Pulserate selectionSelect BPM rate in 1BPM increments.N/ASelect BPM ratein 1 BPMincrements.N/ANone.
SpO2accuracySelect 30% to 100% in1% incrementsWith oximetermanufacturer's R-curveSaturationwithin UUT specificrange ±(1 count +specified accuracy of theUUT)Saturationoutside UUT specificrange monotonic withunspecified accuracyWith Fluke BiomedicalR-curves91 to 100 %±(3 counts + specifiedaccuracy of the UUT)81 to 90 % ±(5counts + specifiedaccuracy of the UUT)71 to 80 % ±(7counts + specifiedaccuracy of the UUT)Below 70 %monotonic withunspecified accuracyN/ASelect 50% to100% in 1%increments-accuracy: 75%to100% ± 1%plus theaccuracy of thepulse oximeterunder test. 50%-75%, ± 2% plusthe accuracy ofthe oximeterunder test.Under 50%,unspecified.N/AMore selectionsper marketrequirements.
SpO2 TestOptical.N/AProbe electricalsimulation test.N/AElectricalsimulation of thefinger probe isnot needed withmodernoximeters.
SpO2 testfeaturesTransmission of lightselected through selectionof finger type: lightfinger, thick dark fingeror neonate.N/ATransmissionLight Control(TLC) feature inIndex 2 is aquantity thatsimulatesdifferent lightN/ABetter userinterface andunderstanding oftransmissionwith ProSimselection of lighttransmission
attenuation.levels.
MagneticHolderYesN/ANoN/AIncludes testedmagnetic holderfor SPO2module.
Non-Invasive Blood Pressure
Manometer0 to 400 mmHgN/AN/AMax. 499.75 mmHgLower rangefollowingmarketrequirements.
Leak TestSource pressure, seal off,measure change inpressure over time.N/AN/ASource pressure, sealoff, measure changein pressure overtime.None.
Over PressureTestIncrease pressure untildevice under tests ventsto atmosphere.N/AN/AIncrease pressureuntil device undertests vents toatmosphere.None.
SimulationSystolic/Diastolic Adult60/30 to 255/195;Neonatal 35/15 to255/195.N/AN/ASystolic/Diastolicsimulations. Adult60/30 to 255/195.Adult andNeonatalavailable.
ArrhythmiasPremature atrialcontraction, Prematureventricular contraction,Atrial fibrillation, MissedBeat.N/AN/APremature atrialcontraction,Prematureventricularcontraction, Atrialfibrillation, MissedBeat, aberrant Sinusconduction.No aberrantSinusconduction. Notneeded permarket use.
Synchronization to ECG30 to 240 BPMN/AN/A30 to 240 BPMNone.

000012

Siomedic

{8}------------------------------------------------

・・・・

000013

:

9/14

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "FLUKE" in large, bold, black letters. The letters are sans-serif and appear to be a part of a logo. The background behind the word is a light gray color, providing contrast to make the word stand out.

Biomedical

000014 :

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "FLUKE." in a bold, sans-serif font. The letters are black, and the background is a light gray with a textured appearance. A small circle is present to the right of the letter 'E'.

Biomedical

4

.

000015

: 1

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the word "FLUKE" in all caps and in a bold, sans-serif font. The letters are black, and there is a registered trademark symbol to the right of the letter "E". The background is a light gray color. The image is simple and clean, with a focus on the brand name.

Biomedical 174

000016 ﺍﻟ

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for Fluke Biomedical. The word "FLUKE" is in large, bold, black letters on a gray background. Below the word "FLUKE" is the word "Biomedical" in smaller, bold, black letters.

174

s

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the word "FLUKE" in all caps and in a bold, sans-serif font. To the right of the word is a registered trademark symbol. The word and symbol are in black, and the background is a light gray with a textured pattern.

Biomedical

Non-Clinical Test Data:

Laboratory studies have been conducted with a representative patient monitor to verify and validate the ProSim 6 and ProSim 8 will perform within its' published specifications.

  • NPI-01282011-00007 ProSim 6_8 Bench test summary and results .
    The ProSim 6 and ProSim 8 software has been successfully validated to confirm the performance of the device.

Clinical Test Data:

Clinical testing has not been conducted on this product.

Conclusion:

Based upon the laboratory studies, similar technological/performance characteristics as compared to the predicate devices, and successful validation of the ProSim 8 software, the performance of the ProSim 6 and ProSim 8 is deemed to be substantially equivalent to the MedSim300B, Index 2 and Cufflink.

000018

{14}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/14/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Fluke Biomedical c/o Mr. John Nelson, RAC Director of Regulatory/Quality Affairs 6045 Cochran Rd. Solon, OH 44139

2 2011

Re: K110429

Trade/Device Names: ProSim 4, 6 and 8 Regulatory Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including Cardiotachometer and Rate Alarm) Regulatory Class: Class II (Two) Product Code: DRT Dated: August 26, 2011 Received: August 29, 2011

Dear Mr. Nelson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{15}------------------------------------------------

Page 2 - Mr. John Nelson, RAC

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

R. D. Zadeh, M.D.

Bran D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{16}------------------------------------------------

FLUKE
Biomedical

Indications for Use

ProSim 4

510(k) Number (if known): K110429

Device Name: ProSim4

Indications for Use:

The ProSim 4 Vital Signs Simulator provides electronic and pneumatic simulation of physiological parameters for determining that patient monitoring devices or systems are performing within their operating specifications. The device includes the following physiological simulations:

  • ECG adult or neonatal .
  • Invasive and non-invasive blood pressure .
  • Respiration .

Prescription Use

(Part 21 CFR 801 Subpart D)AND/OR(21 CFR 807 Subpart C)
-------------------------------------------------------------

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

AND/OR

Over-The-Counter Use

(Division Sign-Off) Division of Cardiovascular Devices

000012

X

510(k) Number K110429

{17}------------------------------------------------

FLUKE.
Biomedical

Indications for Use

ProSim 6/8

510(k) Number (if known): K110429

Device Name: ProSim6/8

Indications for Use:

The ProSim 6 and ProSim 8 Vital Signs Simulators provide electronic and pneumatic simulation of physiological parameters for determining that patient monitoring devices or systems are performing within their operating specifications. The devices provide the following physiological simulations:

  • · ECG adult or pediatric
  • Invasive and non-invasive blood pressure .
  • Respiration ●
  • Temperature �
  • � Cardiac Output
  • Fetal Simulation includes fetal, maternal ECG, & uterine contractions (ProSim . 8 only)

Additionally, the devices provide an optical signal to verify that the electronics within the pulse oximeter probe are functional

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use X (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
----------------------------------------------------------

(Division Sign-Off)
Division of Cardiovascular Devices

510(k) Number K110427

000020

.

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).