K Number
K110337

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-03-16

(406 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The I Macro Implant system is intended for implantation in the mandibular molar region where bone exists and the surgeon has determined that placement of a narrower diameter implant would increase the chance of failure due to poor primary stability, or increased surgical procedures leading to complications. This I Macro implant provides support for fixed or removable dental prostheses in a single tooth, partially edentulous prostheses, or complete arch prostheses. It further adds the option for immediate loading on single and splinted multiple unit restorations when good primary stability is achieved and with proper occlusal loading, to restore the chewing function.

Device Description

The I Macro implants are self-tapping; commercially pure; Titanium Alloy threaded screws, with light grit blasting or roughened surface treatment. These materials and procedures are exactly the same as cleared in the OCO Biomedical submission K090174. The I Macro Implant is available in a 6.0, 7.0, 8.0 & 9.0mm diameter and is available in 8, 10, and 12mm lengths.

AI/ML Overview

The provided text describes a medical device submission (K110337) for the "I Macro Implant System," a dental implant. The submission primarily focuses on demonstrating substantial equivalence to previously cleared predicate devices through comparisons of materials, design, processing, packaging, sterilization methods, and indications for use.

Based on the provided information, the device is a dental implant, and the "study" mentioned refers to various testing standards used to demonstrate the device's properties. These are not clinical studies in the typical sense of evaluating AI performance or diagnostic accuracy. Instead, they are engineering and biological evaluations.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Standard)Reported Device Performance (Compliance)
Fatigue Testing (ISO 14801:2007)Testing conducted according to ISO 14801. (No specific numerical results provided)
Biocompatibility Testing (ISO 10993-5, ISO 10993-1)Testing conducted according to ISO 10993-5 and ISO 10993-1.
Surface Analysis (SEM)SEM analysis of the implant surface after RBM blasting and acid etching performed. (No specific results provided)
Risk Management (ISO 14971:2007)Claims compliance with ISO 14971:2007.
Biological Evaluation (ISO 10993-1, -5, -10, -11, ISO 7405)Claims compliance with these ISO standards.
Packaging (ISO 11607-1, ISO 11607-2, EN 868-5, ASTM F 1980)Claims compliance with these standards.
Sterilization (ISO 11137-1, ISO 11137-2, ISO 11137-3, ISO 17665-1)Claims compliance with these standards.
Labeling (ISO 1041, ISO 980, ISO 15223)Claims compliance with these standards.
Clinical Investigation (ISO 14155-1:2003)Claims compliance with ISO 14155-1:2003.

Note: The document states that the fundamental scientific technology, including materials, processing, packaging, and sterilization methods, is "identical or very similar to referenced predicate devices." The acceptance criteria are largely met by demonstrating compliance with established international standards for medical device manufacturing and testing, rather than explicit numerical performance targets from a clinical trial.

2. Sample Size Used for the Test Set and the Data Provenance

This information is not applicable and not provided in the document. The document describes engineering and biological testing of the device itself (fatigue, biocompatibility, surface analysis), not a "test set" of patient data for evaluating an AI algorithm's performance.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This information is not applicable and not provided. This document does not describe the evaluation of an AI algorithm requiring a ground truth established by experts.

4. Adjudication Method for the Test Set

This information is not applicable and not provided. This document does not describe an AI evaluation requiring an adjudication method.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There is no mention of an MRMC comparative effectiveness study, or any AI component, in this document.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

There is no mention of a standalone algorithm performance study or any AI component in this document.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

This information is not applicable and not provided. The "ground truth" in this context would likely refer to the physical and chemical properties of the implant meeting the specifications outlined in the standards (e.g., fatigue strength exceeding a certain threshold, materials matching composition specifications).

8. The sample size for the training set

This information is not applicable and not provided. This document describes a medical device, not an AI algorithm that requires a training set.

9. How the ground truth for the training set was established

This information is not applicable and not provided. This document describes a medical device, not an AI algorithm.

{0}------------------------------------------------

K110337

MAR 1 6 2012

Contact Person:

David D. Dalise President/ Owner OCO Biomedical, Inc. 8500 Washington St. N.E., Suite A-1 Albuquerque, NM 87113 Phone: (505) 293-0025

Trade Name: I Macro Implant System Common Name: Dental Implant Classification Name: Dental Implant Endosseous, Root-Form

Substantial Equivalence to:

Immediate Stabilizing Implant (ISI) OCO 5.0mm Taper Implant OCO Biomedical TSI & ERI Mega Gen Implant Direct Southern Implants

K033392 (Cleared 12/11/03) K023336 (Cleared 10/9/02) K090174 (Cleared 9/14/09) K081302 (Cleared 8/15/08) K061319 (Cleared 9/29/06) K071161 (Cleared 11/16/07)

Description of Device:

The I Macro implants are self-tapping; commercially pure; Titanium Alloy threaded screws, with light grit blasting or roughened surface treatment. These materials and procedures are exactly the same as cleared in the OCO Biomedical submission K090174. The I Macro Implant is available in a 6.0, 7.0, 8.0 & 9.0mm diameter and is available in 8, 10, and 12mm lengths.

Titanium Chemical Compositions:

6AL/4V ELI Carbon - 0.08% Vanadium - 3.5%-4.5% Nitrogen - 0.05% Aluminum - 5.5%-6.5% Oxygen - 0.013 Hydrogen - 0.015% Iron - 0.25% Titanium -- Balance

{1}------------------------------------------------

The compatible abutments for the I Macro Implants where cleared under 510 K090174. Abutment Unit Numbers:

  • . 4 CB 5.5 TSI
  • 4 IOT 0 TSI .
  • 4100 TSI .
  • 5150 TSI �
  • 4200 TSI .
  • 5250 TSI .

(Please see Abutment Attachment)

Indications for Use:

The I Macro Implant system is intended for implantation in the mandibular molar region where bone exists and the surgeon has determined that placement of a narrower diameter implant would increase the chance of failure due to poor primary stability, or increased surgical procedures leading to complications. This I Macro implant provides support for fixed or removable dental prostheses in a single tooth, partially edentulous prostheses, or complete arch prostheses. It further adds the option for immediate loading on single and splinted multiple unit restorations when good primary stability is achieved and with proper occlusal loading, to restore the chewing function.

Substantial Equivalence:

OCO Biomedical, Inc. has submitted information to demonstrate that, for the purposes of FDA's regulation of medical devices, the I Macro Implants are substantially equivalent in indications and design principles to predicate devices previously cleared by the FDA: Immediate Stabilizing Implant (ISI) K033392 (Cleared 12/11/03), OCO Biomedical 5.0mm Taper Implant K023336 (Cleared 10/9/02). Mega Gen K081302 (Cleared 8/15/2008),OCO Biomedical TSI ERI K090174 (Cleared 9/14/09), Implant Direct K061319 (Cleared 9/29/06) and Southern Implants K071161 (Cleared 11/16/07).

The I Macro Implant has the following similarities to the predicate devices:

  • -has the same intended use
  • -incorporates the same materials and design
  • -is packaged and sterilized using the same materials and processes

Summary of Technological Comparison:

The fundamental scientific technology of the device is identical or very similar to referenced predicate devices. including materials, processing, packaging, and sterilization methods. See comparison table in the Executive Summary section of this application.

{2}------------------------------------------------

Fatigue testing was conducted according to ISO 14801, biocompatibility testing was conducted according to ISO 10993-5 and ISO 10993-1 and SEM analysis of the implant surface after RBM blasting and acid etching.

Conclusion:

OCO Biomedical, Inc. has submitted information to demonstrate that, for the purposes of FDA's regulation of medical devices, the I Macro Implants are substantially equivalent in indications and design principles to predicate devices previously cleared by the FDA.

Standards to which OCO Biomedical claims compliance:

  • ISO 14801:2007, Dentistry Implants Dynamic fatigue test for endosseous dental implants
  • ・ ISO 14971:2007, Medical devices - Application of risk management to medical devices
  • ISO 10993-1:2009, Biological evaluation of medical devices part 1: evaluation and testing
  • ISO 868-5:2009, Packaging materials and systems for medical devices which are to be sterilized; Part 5: Self-sealable pouches and reels of paper and plastic film construction; Requirements and test methods
  • -ISO 11607-1:2009. Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems
  • -ISO 11607-2:2006, Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes
  • -ASTM F 1980:2007, Standard guide for accelerated aging of sterile medical device packages
  • -ISO 11137-1:2006, Sterilization of health care products - Radiation Part 1: Requirements for development, validation, routine control of a sterilization process for medical devices.
  • -ISO 11137-2:2007, Sterilization of health care products - Radiation Part 2: Establishing the sterilization dose.
  • ISO 17665-1:2006 Sterilization of health care products Moist heat Part 1: -Requirements for the development, validation and routine control of a sterilization process for medical devices.
    • ISO 1041:2008, Information supplied by the manufacturer with medical devices
  • -ISO 980:2008, Graphical symbols for use in the labelling of medical devices
  • -ISO 15223:2007. Medical devices - Symbols to be used with medical device labels, labeling and information to be supplied
  • ISO 14155-1:2003. Clinical investigation of medical devices for human subjects --Part 1: General requirements
  • ISO 1041:2008 Information supplied by the manufacturer of medical devices -

{3}------------------------------------------------

  • ISO 11137-3 Sterilization of health care products Radiation -
  • ISO 10993-1:2009 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process
  • -ISO 10993-5:2009 Biological evaluation of medical devices
  • -ISO 10993-5:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization
  • -ISO 10993-11:2006 Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity
  • -ISO 7405:2008 Dentistry - Evaluation of biocompatibility of medical devices used in dentistry
  • ISO 11607 Packaging for Terminally Sterilized Medical Devices -
  • EN 868-5:2009 Packaging for terminally sterilized medical devices. Sealable pouches and reels of porous and plastic film construction. Requirements and test methods

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The eagle faces left and appears to be in flight. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.

Public Health Service

MAR 1 6 2012

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002

Mr. Jack Bloom Regulatory Manager OCO Biomedical 8500 Washington Street, NE Suite A-1 Albuquerque, New Mexico 87113

Re: K110337

Trade/Device Name: I Macro Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: March 10, 2012 Received: March 12, 2012

Dear Mr Bloom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Bloom

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm 1 1 5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Wh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number: K110337

Device Name: I Macro Implant System

Indications for use:

The I Macro Implant system is intended for implantation in the mandibular molar region where bone exists and the surgeon has determined that placement of a narrower diameter implant would increase the chance of failure due to poor primary stability, or increased surgical procedures leading to complications. This I Macro implant provides support for fixed or removable dental prostheses in a single tooth, partially edentulous prostheses, or complete arch prostheses. It further adds the option for immediate loading on single and splinted multiple unit restorations when good primary stability is achieved and with proper occlusal loading, to restore the chewing function.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

hn

(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anest Dested Devices Division Control, Dental Devices

510(k) Number. K110337

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.