K Number
K110289

Validate with FDA (Live)

Date Cleared
2011-08-30

(210 days)

Product Code
Regulation Number
882.1320
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Disposable Pre-gelled Ag/AgCl Surface Electrodes are intended for recording/stimulation and monitoring of the Electromygraphy (EMG), Electroencephalograph(EEG) and Evoked Potential(EP) signals.

Device Description

Friendship's Pre-gelled Ag/AgCl Surface Electrodes are disposable (for "single Use Only"). Used to detect electrophysiological signals or provide electrical stimulation subcutaneously. The electrodes are the interface medium between the diagnostic or monitoring equipment and the patient. The surface electrodes is comprised of top woven layer, carbon layer, Ag/AgCl layer and hydrogel layer on one end electrically connected to lead wire and a touch-proof connector on the other end.

AI/ML Overview

The provided document describes the "Friendship Pre-gelled Ag/AgCl Surface Electrodes," a disposable device for detecting electrophysiological signals or providing electrical stimulation subcutaneously.

Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

CriteriaAcceptance Criteria (Predicate Devices)Reported Device Performance (Friendship Pre-gelled Ag/AgCl Surface Electrodes)
BiocompatibilityKnown biocompatibility (Katecho hydrogel) for Cytotoxicity, Skin Irritation, and Skin Sensitization (ISO 10993-5, 10993-10)Katecho hydrogel is of known biocompatibility, and materials were tested for safety and biocompatibility by Katecho Inc. (Cytotoxicity, Skin Irritation, Skin Sensitization).
Electrical PerformanceMeet voluntary standard requirements under ANSI/AAMI EC12/2000 in compliance with IEC60601-1Tested and meets voluntary standard requirements under ANSI/AAMI EC12/2000 in compliance with IEC60601-1.
Device Description/FunctionalitySimilar materials, design, and function to predicate devices.Similar materials, design, and function to predicate devices.
Simulated UseFound acceptable for intended use by healthcare professionals.Evaluated by healthcare professionals during a simulated use test and found acceptable for its intended use.
Substantial EquivalenceComparable to predicate devices (Sunspot Pre-gelled Surface Electrodes K062198, Rhythmlink International Cutaneous Pad Electrodes K052188)Determined to be substantially equivalent to predicate devices.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated. The document mentions a "simulated use test" but does not quantify the number of electrodes or participants involved.
  • Data Provenance: The document does not specify the country of origin for the "simulated use test." It is implied to be related to the manufacturer's location in Xi'an, Shaanxi, China, but this is not explicitly stated as the origin of the test data. The biocompatibility data is from Katecho Inc. and refers to international ISO standards.
  • Retrospective or Prospective: The "simulated use test" sounds like a prospective test, but without more details, it's difficult to confirm.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

  • The document states the device "was evaluated by health care professionals during a simulated use test."
  • Number of Experts: Not specified.
  • Qualifications of Experts: It collectively refers to "health care professionals," but no specific qualifications (e.g., years of experience, specialty) are provided.

4. Adjudication Method for the Test Set

  • No specific adjudication method is mentioned for the "simulated use test." The evaluation by healthcare professionals seems to have led to a conclusion of "acceptable for its intended use" without detailing a multi-reader review process or conflict resolution.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

  • No, an MRMC comparative effectiveness study was not done. This device is a pre-gelled surface electrode, not an AI-powered diagnostic or assistive tool. Therefore, the concept of human readers improving with or without AI assistance is not applicable.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • No, a standalone (algorithm only) performance study was not done. This device is a physical electrode, not an algorithm.

7. The Type of Ground Truth Used

  • The "ground truth" for the device's performance relies primarily on adherence to established voluntary standards (ANSI/AAMI EC12/2000 and IEC60601-1 for electrical performance) and biocompatibility testing against ISO standards, along with subjective evaluation by healthcare professionals during a simulated use test for overall acceptability.
  • There is no mention of pathology, outcomes data, or expert consensus in the typical sense for a diagnostic device. The "ground truth" for substantial equivalence is based on direct comparison to already cleared predicate devices.

8. The Sample Size for the Training Set

  • Not applicable. This device is a physical medical device (electrode), not an algorithm or AI model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as there is no training set for this device.

{0}------------------------------------------------

K110289

AUG 300 2011

3. Attachment 1: 510(K) Summary

Summary Date:

This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of CFR 807.92.

May 6, 2011

Submitter's information:Zhai Ying ChuanGeneral ManagerXian Friendship Medical Electronics Co., Ltd.Gao Xin 1st Road,Hi-Tech Development Zone,Xi'an, Shaanxi,China, 710075Phone: (86) -29-88225200Fax: (86)- 29-88236285georgezhai2616@163.com
Trade Name:Friendship Pre-gelled Ag/AgCl Surface Electrodes
Common Name:Disposable Ag/AgCl Pre-gelled Surface Electrodes
Classification Name:21 CFR section 882.1320. Cutaneous Electrode.
Product Code:GXY
Predicate Devices:Sunspot Pre-gelled Surface Electrodes (K062198) manufacturedby Axon Systems, Inc.Rthymlink International Cutaneous Pad Electrodes (K052188)manufactured by Rhythmlink International, LLC

10.1 Device Description:

Friendship's Pre-gelled Ag/AgCl Surface Electrodes are disposable (for "single Use Only"). Used to detect electrophysiological signals or provide electrical stimulation subcutaneously. The electrodes are the interface medium between the diagnostic or monitoring equipment and the patient. The surface electrodes is comprised of top woven layer, carbon layer, Ag/AgCl layer and hydrogel layer on one end electrically connected to lead wire and a touch-proof connector on the other end. The surface electrodes are placed cutaneously by a licensed physician or technologist under the supervision of a

{1}------------------------------------------------

were already verified and validated.

10.6 Brief discussion of the clinical tests submitted:

Clinical studies were not deemed necessary regarding the surface electrodes due to their similarity in materials, design and function to those "predicate device". The device was evaluated by health care professionals during a simulated use test and was found to be acceptable for its intended use.

10.7 Biocompatibility testing

The contact material Katecho hydrogel is of known biocompatibility. Those materials were already tested for material safety and biocompatibility by the Katecho Inc.:

  • Cytotoxicity study ISO 10993-5 .
  • Skin irritation study ISO 10993-10 .
  • Skin sensitization study ISO 10993-10 .

10.8 Performance testing

The electrical performance of the Surface Electrode has been tested and meet the voluntary standard requirements under ANSI/AAMI EC12/2000 in compliance with IEC60601-1.

10.9 Conclusions:

Xian Friendship Medical Electronics Co., Ltd.'s Surface Electrodes are substantially equivalent to the predicate devices. Xian Friendship Medical Electronics Co., Ltd. manufactures the Surface Electrodes for Axon Systems, Inc. There are no new questions of safety or effectiveness raised or evident.

{2}------------------------------------------------

area @ Salaha The Partier States of the Parties of Children Come of the Children Comes of the ChildrenLA B SUNNER AT CREATINGﺎ ﻓﻲ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺎﺣﺔ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------CALL CLAIM CLAIM CLAIM CLAIM CLAIM CLASS COLUMN CALL CALIFORNIA100 mm 1 1 m1
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------and the first of the many of the many of the may be any and the may be and the may be any andTHE LEAST LEAST LEAST LEAST LEAST MAIL BE LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- LA-A-A-A-BERTIS -LA-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-Comments of the comments of theAnd Status of Acres of Cattle Company of
SpecificationEquivalent DevicePredicate Device 1 B
ManufacturerXian Friendship MedicalElectronics Co., Ltd.Axon Systems, Inc
Part number or 510(k)Part Number: SEAg-Cu-SFDA 510(K): K062198
Part numberSEAg-Cu-S-1500/2.0x2.7DSE2115
PartPre-gelled Surface ElectrodePre-gelled Surface Electrode
ClassificationIIII
DescriptionSingle and twisted pair, Pre-gelled Surface ElectrodeSingle and twisted pair, Pre-gelled Surface Electrode
Dimensions27 mm x 20 mm , varioussize27 mm x 20 mm , various size
Electrode shapesRound/oval design , variousDuck-foot design , various
Sensor MaterialAg/AgClAg/AgCl
Hydrogel typeSolid gelKatecho KM10E.Solid gelKatecho KM 10E
ConnectorDIN 42802DIN 42802
Lead LengthVarious: 1.5M, 2.0M, 2.5M,3.0M (other lengths may beadded)Various: 1.5M, 2.0M, 2.5M,3.0M (other lengths may beadded)
Lead wire colorMultiple colorsMultiple colors
Labeled asSingle use, disposableSingle use, disposable
Indications for useStimulation and recordingStimulation and recording
PouchAluminum polyesterAluminum polyester
Standard Electrode TypeAAMI EC12 2000AAMI EC12 2000
Stated potential adverse reactionsSkin irritationSkin irritation

.

.

{3}------------------------------------------------

PouchAluminum polyesterAluminum polyester
Standard Electrode TypeAAMI EC12 2000AAMI EC12 2000
Stated potential adverse reactionsSkin irritationSkin irritation
Packaging1 electrode/pouch1 electrode/pouch
Box24 electrodes/box24 electrodes/box
SpecificationEquivalent DevicePredicate Device 1 D
ManufacturerXian Friendship MedicalElectronics Co., Ltd.Axon Systems, Inc
Part number or 510(k)Part Number: SEAg-Cu-SFDA 510(K): K062198
Part numberSEAg-C 3.0DSE3003
PartPre-gelled Surface ElectrodePre-gelled Surface Electrode
ClassificationIIII
DescriptionSingle and twisted pair, Pre-gelled Surface ElectrodeSingle and twisted pair, Pre-gelled Surface Electrode
Dimensions30mm in diameter, roundsize50 mm x50 mm x 60 mm ,triangle
Electrode shapesRound/oval design , variousTriangle design , various
Sensor MaterialAg/AgClAg/AgCl
Hydrogel typeSolid gelSolid gel
ConnectorKatecho KM10E.Metal snapKatecho KM 10EMetal snap
Labeled asSingle use, disposableSingle use, disposable
Indications for useStimulation and recordingStimulation and recording
PouchAluminum polyesterAluminum polyester
Standard Electrode TypeAAMI EC12 2000AAMI EC12 2000

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an emblem featuring an abstract depiction of an eagle or bird-like figure with outstretched wings.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002

Xian Friendship Medical Electronics Co., Ltd. c/o Zhai Ying Chuan General Manager No. 9 Gao Xin 156 Road, Hi-Tech Development Zone Xi'an Shaanxi 710075 China

AUG 30 2011

Re: K110289

Trade/Device Name: Friendship Medical Disposable Pre-gelled Ag/AgCl Surface Electrode Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: Undated Received: July 21, 2011

Dear Mr. Zhai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. · Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Kesia Alexander

Image /page/5/Picture/7 description: The image shows a stylized cursive letter 'f'. The letter has a large, sweeping loop at the top, extending upwards and slightly to the right. A smaller loop is present at the bottom of the letter. The overall impression is of a handwritten signature or monogram.

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): K110289

Device Name: Disposable Pre-gelled Ag/AgCl Surface Electrode

Indications For Use:

Disposable Pre-gelled Ag/AgCl Surface Electrodes are intended for recording/stimulation

and monitoring of the Electromygraphy (EMG), Electroencephalograph(EEG)

and Evoked Potential(EP) signals.

Prescription Use X X

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D) C)

(21 CFR 801 Subpart

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Concurrence of CDRH, Office of Device Evaluation (ODE)

D. Dolan

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K110289

Page 1 of 1_________________________________________________________________________________________________________________________________________________________________

§ 882.1320 Cutaneous electrode.

(a)
Identification. A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.(b)
Classification. Class II (performance standards).