K Number
K110276

Validate with FDA (Live)

Device Name
POWERPRESS UNIT
Date Cleared
2011-11-09

(282 days)

Product Code
Regulation Number
870.5800
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Powerpress Unit Sequential Circulator is a manual, sequential, pneumatic compression device, intended for the primary or adjunctive treatment of primary or secondary lymphedema. The device is also intended for the additional or alternate treatment of venous insufficiency, and chronic venous stasis ulcers associated with venous insufficiency, as well as general treatment of swelling of the extremities. The device is intended for home or hospital use.

Device Description

Powerpress Unit is a sequential pneumatic compression device designed to apply compression to a limb. The device is composed of two components: Pneumatic Manual Pump and Limb Sleeve or garment composes of 4 chambers. Powerpress Unit enables different treatment pressure (30 ~ 100mmHg). Treatment pressure and time should be used according to physician prescription. When activated, air flows into chamber, the pump provides gradient pressurization to the chambers (sequential inflation of distal to proximal, with distal chambers inflated to a greater pressure than the proximal ones). After each chamber is inflated, the pressure is held constant until all chambers are inflated, in order to prevent reverse gradient flow. Once all chambers are inflated, they are then all released simultaneously, and the cycle repeats. Pressure within chambers are adjustable - pressure to chamber 1 is controlled by user-adjusted regulator on the pump. Pressure in chamber 2, 3 & 4 are individually lowered according to the default factory setting. Gradient: appx 7%, example: Foot 60mmHg - Ankle 56mmHg - Calf 52mmHg -Thigh 48mmHg Cycle Time: Inflation 36 sec / Deflation 24 sec Inflation time each chamber : 1st chamber(foot) - 36 sec, 2nd chamber(ankle) - 27 sec, 3rd chamber - 18 sec, 4th chamber(thigh) - 9 sec. A calibrated dial gauge displays pressure in the range of 0 ~ 160 mmHg.

AI/ML Overview

The provided text describes the 510(k) summary for the Powerpress Unit Sequential Circulator, which is a medical device intended for pneumatic compression therapy.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based only on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document states, "Bench testing and side by side comparisons were done with predicate devices to assure equivalence in performance." However, it does not explicitly list quantitative acceptance criteria for attributes like pressure accuracy, inflation/deflation timings, or gradient percentages, nor does it provide detailed performance results against such criteria. Instead, it relies on demonstrating substantial equivalence to a predicate device.

The reported performance characteristics of the Powerpress Unit are:

CharacteristicReported Performance (Powerpress Unit)
Treatment Pressure Range30 ~ 100 mmHg
Gradient (example)Foot 60mmHg - Ankle 56mmHg - Calf 52mmHg -Thigh 48mmHg (appx 7%)
Cycle TimeInflation 36 sec / Deflation 24 sec
Inflation Time per Chamber1st (foot) - 36 sec, 2nd (ankle) - 27 sec, 3rd - 18 sec, 4th (thigh) - 9 sec
Dial Gauge Display Range0 ~ 160 mmHg

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify any sample sizes for a test set. It mentions "bench testing and side by side comparisons," but details on the number of devices tested, the number of cycles performed, or any specific test parameters are absent. The concept of "data provenance" (country of origin, retrospective/prospective) is not applicable here as it's not a clinical study involving patient data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts:

This information is not provided. The assessment appears to be a technical comparison, not one requiring expert human interpretation of medical data.

4. Adjudication Method for the Test Set:

An "adjudication method" is not applicable here as there is no mention of a test set requiring expert review or consensus for ground truth.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, Effect Size of Human Improvement with AI vs. Without AI Assistance:

An MRMC study was not done. The device is a pneumatic compression unit, not an AI-assisted diagnostic or therapeutic tool with human-in-the-loop performance. Therefore, the concept of "effect size of how much human readers improve with AI vs. without AI assistance" is not relevant to this device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

The device is not an algorithm or AI system. It's a mechanical and pneumatic device. Therefore, the concept of "standalone performance" in this context is not applicable in the way it's typically understood for AI/algorithm-based devices. Its performance is assessed as a standalone physical product.

7. The Type of Ground Truth Used:

The "ground truth" for this device's performance assessment appears to be based on direct physical measurements and observation of its operational characteristics (e.g., pressure output, timing) compared to the known characteristics of the predicate device. It's essentially a comparison against established functional standards of similar devices, rather than expert consensus, pathology, or outcomes data in a clinical sense.

8. The Sample Size for the Training Set:

The concept of a "training set" is not applicable as this is not a machine learning or AI device.

9. How the Ground Truth for the Training Set Was Established:

This question is not applicable for the same reason as point 8.

{0}------------------------------------------------

Pg 1 of 2

510(K) SUMMARY

(as required by 21 CFR 807.92)

NOV - 9 2011

A. 510K Number : K110276

B. Applicant Company Name: Hanuri Distribution, Inc Adress: 9601 Owensmouth Ave. # 8 Chatsworth, CA 91311 USA Contact Person: Jung H Moon Phone: 818-998-1023, Fax: 818-998-0277

C. Proprietary and Established Names: Same as B

  • D. Regulatory Information
      1. Proprietary/Trade Name: Powerpress Unit
      1. Classification Name: Compressible Limb Sleeve
      1. Classification Panel : Cardiovascular
      1. Common / Usual Name: Sequential Compression Device
      1. Classification / Product Code: Class II / JOW 870.5800

E. Intended Use

The Powerpress Unit Sequential Circulator is a manual, sequential, pneumatic compression device, intended for the primary or adjunctive treatment of primary or secondary lymphedema. The device is also intended for the additional or alternate treatment of venous insufficiency, and chronic venous stasis ulcers associated with venous insufficiency, as well as general treatment of swelling of the extremities. The device is intended for home or hospital use.

F. Predicate Device

Powerpress Unit is substantially equivalent to the following

Predicate DeviceManufacturer510(k)#
SC-3008 Sequential CirculatorBio Compression System, IncK043423

G. Device Description

Powerpress Unit is a sequential pneumatic compression device designed to apply compression to a limb. The device is composed of two components.

  • . Pneumatic Manual Pump
  • . Limb Sleeve or garment composes of 4 chambers

Powerpress Unit enables different treatment pressure (30 ~ 100mmHg). Treatment pressure and time should be used according to physician prescription. When activated, air flows into chamber, the pump provides gradient pressurization to the chambers (sequential inflation of distal to proximal, with distal chambers inflated to a greater pressure than the proximal ones).

{1}------------------------------------------------

K110276
pg 2 of 2

After each chamber is inflated, the pressure is held constant until all chambers are inflated, in order to prevent reverse gradient flow. Once all chambers are inflated, they are then all released simultaneously, and the cycle repeats. Pressure within chambers are adjustable - pressure to chamber 1 is controlled by user-adjusted regulator on the pump. Pressure in chamber 2, 3 & 4 are individually lowered according to the default factory setting.

Gradient: appx 7%, example: Foot 60mmHg - Ankle 56mmHg - Calf 52mmHg -Thigh 48mmHg Cycle Time: Inflation 36 sec / Deflation 24 sec

Inflation time each chamber : 1* chamber(foot) - 36 sec, 2nd chamber(ankle) - 27 sec, 3rd chamber - 18 sec, 4th chamber(thigh) - 9 sec

A calibrated dial gauge displays pressure in the range of 0 ~ 160 mmHg

H. Technological Characteristics

The manufacturer believes that the technological characteristic of the Powerpress Unit is substantially similar to those of the predicate devices.

Powerpress Unit has very similar components to its predicate devices and very similar principles of operation.

I. Performance Standards

Bench testing and side by side comparisons were done with predicate devices to assure equivalence in performance. No new safety and/or effectiveness issues are raised.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with outstretched wings. The bird is enclosed within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper half of the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

NOV - 9 8011

Hanuri Distribution, Inc. % Mr. Jung Hyun Moon 9601 Owensmouth Avenue #8 Chatsworth, California 91311

Re: K110276

Trade/Device Name: Powerpress Unit Sequential Circulator Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: II Product Code: JOW, IRP Dated: October 20, 2011 Received: October 21, 2011

Dear Mr. Moon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Jung Hyun Moon

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Erin L. Keith

for Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

KIIOLIC

PJ 1 of 1

Indication for Use Statement

510(k) Number (if known) : K 110276

Device Name : Powerpress Unit

Indication For Use :

The Powerpress Unit Sequential Circulator is a manual, sequential, pneumatic compression device, intended for the primary or adjunctive treatment of primary or secondary lymphedema. The device is also intended for the additional or alternate treatment of venous insufficiency, and chronic venous stasis ulcers associated with venous insufficiency, as well as general treatment of swelling of the extremities. The device is intended for home or hospital use.

Prescription Use X

A

And/Or

Over-The Counter Use

(Part21 CFR801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Division Sign-Off

(Division Sigi Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K110276

§ 870.5800 Compressible limb sleeve.

(a)
Identification. A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.(b)
Classification. Class II (performance standards).