(178 days)
The ExploR® Radial Head Plating System is indicated for fractures, fracture dislocations, osteotomies and non-unions of the proximal radius.
The Explor® Radial Head Plating System is comprised of 3 styles of plates (neck plates, rim plates and rim long plates) in two sizes (small and standard) each and locking and non-locking screws in multiple lengths. Plate sizing and contouring was developed through the use of Biomet's IntelliFIT Technology which uses contour analysis to map patterns in complex bone on cadaveric specimens to determine plate sizing. (Note, the software was used to determine a set of pre-defined plate sizes and is not used to create individual, patient matched plates.)
Here's a summary of the device's acceptance criteria and the study details based on the provided text:
Device Name: ExploR® Radial Head Plating System
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristic (Acceptance Criteria) | Reported Device Performance |
|---|---|
| Plate Strength: Meet or exceed predicate | The new device met the predicate device's plate strength. |
Note: The document explicitly states "No mechanical or clinical testing was necessary for a determination of substantial equivalence." This implies that the primary acceptance criterion was demonstrating equivalence to predicate devices, particularly in mechanical performance.
Study Details:
The provided documents describe a premarket notification (510(k)) submission for the ExploR® Radial Head Plating System. The study conducted was primarily a non-clinical engineering analysis to demonstrate substantial equivalence to legally marketed predicate devices.
2. Sample size used for the test set and the data provenance:
- Test Set Sample Size: Not applicable. No "test set" in the sense of a dataset of patient cases was used. The evaluation centered on engineering analysis and comparison to predicate devices.
- Data Provenance: The engineering analysis was performed internally by Biomet Manufacturing Corp. As it's an engineering analysis, country of origin is not relevant in the sense of clinical data. It's retrospective in the sense that it relies on established engineering principles and comparisons to existing devices.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. There was no "test set" requiring ground truth established by experts in a clinical context. The "ground truth" for the engineering analysis would be the established performance characteristics of the predicate devices and accepted engineering principles, likely assessed by internal engineers and regulatory specialists.
4. Adjudication method for the test set:
- Not applicable. No clinical test set or adjudication process was described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC comparative effectiveness study was done. This device is a medical implant (bone fixation system), not an AI-powered diagnostic or assistive technology.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This device is a physical medical device, not an algorithm.
7. The type of ground truth used:
- Engineering Analysis: The "ground truth" for the performance evaluation was established through engineering analysis and comparison to the mechanical properties and established performance of existing legally marketed predicate devices. This is a comparative performance information summary rather than a clinical ground truth.
8. The sample size for the training set:
- Not applicable. This device is a physical medical implant, not a machine learning algorithm requiring a "training set." The design process for the plates involved "Biomet's IntelliFIT Technology which uses contour analysis to map patterns in complex bone on cadaveric specimens to determine plate sizing." However, this is for design, not for training an algorithm for a separate performance validation. The document states that the software is "not used to create individual, patient matched plates."
9. How the ground truth for the training set was established:
- Not applicable, as there was no training set in the context of an algorithm. However, for the design process involving IntelliFIT Technology, the "ground truth" would have been the anatomical patterns and complex bone contours observed in "cadaveric specimens." The establishment of this ground truth would involve anatomical dissection and imaging of these specimens.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Biomet Manufacturing Corp. The logo consists of the word "BIOMET" in a stylized font, with the words "MANUFACTURING CORP." printed below. The logo is black and white and appears to be a scan of a document.
JUL 2 1 2011
510(k) SUMMARY
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92
| Submitter Information | |
|---|---|
| Name | Biomet Manufacturing Corp. |
| Address | 56 East Bell DriveWarsaw, IN 46581-0857 |
| Phone number | (574) 267-6639 |
| Fax number | (574) 371-1027 |
| Establishment RegistrationNumber | 1825034 |
| Name of contact person | Patricia Sandborn BeresSenior Regulatory Specialist |
| Date prepared | January 3, 2011 |
| Name of device | |
| Trade or proprietaryname | Explor® Radial Head Plating System |
| Common or usualname | plate, fixation, bonescrew, fixation, bone |
| Classification name | Single/multiple component metallic bone fixation appliances and accessoriesSmooth or threaded metallic bone fixation fastener |
| Classification panel | Orthopedics |
| Regulation | 21 CFR 888.303021 CFR 888.3040 |
| Product Code(s) | HRSHWC |
| Legally marketed device(s) towhich equivalence is claimed | K062494 - EBI OptiLock® Upper Extremity Plating SystemK033456 - EVOLVE® Radial PlateK040777 - Synthes (USA) LCP Radial Head Plating System |
| Reason for 510(k) submission | New device |
| Device description | The Explor® Radial Head Plating System is comprised of 3 styles ofplates (neck plates, rim plates and rim long plates) in two sizes (smalland standard) each and locking and non-locking screws in multiplelengths. Plate sizing and contouring was developed through the useof Biomet's IntelliFIT Technology which uses contour analysis to mappatterns in complex bone on cadaveric specimens to determine platesizing. (Note, the software was used to determine a set of pre-definedplate sizes and is not used to create individual, patient matchedplates.) |
| Intended use of the device | Bone fixation |
Mailing Address:
Walling Address: 0587
Worldw.com (107.0687)
World Posts (100.000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000
Shipping Address:
56 East Bell Drive
Warsaw, IN 46582
F
i
p. 1 of 2
{1}------------------------------------------------
510(k) Summary ExploR® Radial Head Plating System Page 2 of 2
| Indications for use | The ExploR® Radial Head Plating System is indicated for fractures, fracture dislocations, osteotomies and non-unions of the proximal radius. | ||
|---|---|---|---|
| Summary of the technological characteristics of the device compared to the predicate | |||
| Characteristic | New Device | Predicate Device* | |
| Plate Design | Neck, Rim and Long Rim | K062494K033456K040777 | |
| Plate Material | Stainless Steel ASTM F138 or F139 | K062494 | |
| Plate Dimensions | Length: 18.2mm to 26.2mmWidth: 15.2 mm - 16.7mm | K062494K033456K040777 | |
| Screw Design | Locking and Non-Locking | K062494K033456K040777 | |
| Screw Material | Stainless Steel ASTM F138 or F139 | K062494 | |
| Screw Dimensions | Diameter: 2.0mmLength: 10-30mm | K033456K040777 | |
| PERFORMANCE DATA | |||
| SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIALEQUIVALENCE | |||
| Performance Test Summary-New Device | |||
| Characteristic | Standard/Test/FDA Guidance | Results Summary | |
| Plate Strength | Engineering Analysis | Meet or exceed predicate | |
| Comparative Performance Information Summary | |||
| Characteristic | Requirement | New Device | Predicate Device* |
| Plate Strength | Meet or exceed predicate | Meet | K062494 |
| SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIALEQUIVALENCE AND/OR OF CLINICAL INFORMATION | |||
| Clinical Performance Data/Information: None | |||
| MAGNETIC RESONANCE (MR) ENVIROMENT | |||
| Biomet® has performed non-clinical Magnetic Resonance Imaging (MRI) studies on Plating Systems manufactured of 316L Stainless Steel per ASTM F138. These Plating Systems are determined to be MR Conditional in accordance to ASTM F2503-08 Standard Practice for Marking Devices and Other Items for Safety in the Magnetic Resonance Environment. MR Conditional refers to an item that has been demonstrated to pose no known hazards in a specified MR environment with specified conditions of use. | |||
| CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA | |||
| No mechanical or clinical testing was necessary for a determination of substantial equivalence. The results of engineering analysis indicated the devices performed within the intended use, did not raise any new safety and efficacy issues and were found to be substantially equivalent to the predicate devices. | |||
| * Any statement made in conjunction with this submission regarding and/or a determination of substantial equivalence to any other |
product is intended only to relate to whether the product can be lawfully market approval or redassification and is not intended to be interpreted as an admission or any other type of evident infingenent litigation. [Establishment
Registration and Pernarket Notification Procedures, Final R
, 2f2
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" which is arranged in a circular fashion.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Biomet Manufacturing Corp. % Ms. Patricia Beres Senior Regulatory Specialist 56 East Bell Drive, P.O. Box 587 Warsaw, Indiana 46581-0587
JUL 2 1 2011
Re: K110201
Trade/Device Name: ExploR® Radial Head Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: June 22, 2011 Received: June 24, 2011
Dear Ms. Beres:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
Page 2 - Ms. Patricia Beres
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Eric L. Keith
For Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known):
Device Name: ExploR® Radial Head Plating System
Indications For Use:
:: .....
The ExploR® Radial Head Plating System is indicated for fractures, fracture dislocations, osteotomies and non-unions of the proximal radius.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ NO (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for M. Multerm
(Division Sign Oft) (Division Sigiron of Surgical, Orthopedic, Division of Suices
Page 1 of 1
510(k) Number K110201
N/A