(172 days)
When intended to promote fusion of the cervical spine, and the thoracic spine, (C3-T3), the Gibralt Spinal System is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion and/or tumors. The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C3-T3) spine.
The use of polyaxial screws is limited to placement in T1-T3 in treating thoracic conditions only. Polyaxial screws are not intended to be placed in the cervical spine.
This system can be used independently or in conjunction with Exactech 5.5mm or 6.0mm rod-based Thoraco-Lumbar Pedicle Screw Systems.
This submission proposes a new spinal fixation system. The proposed Gibralt Spinal System is a top-loading spinal fixation system that comprises various sizes of polyaxial screws, rods, hooks, and various connectors to provide immobilization and stabilization of spinal segments as an adjunct to fusion of the cervical, and/or upper thoracic spine. The Gibralt Spinal System components are manufactured from titanium alloy per ASTM F136. The system components are provided clean and non-sterile. The products must be steam sterilized by the hospital prior to use.
The Gibralt Spinal System includes a complete instrumentation system to assist the surgeon in the implantation of each component according to a traditional open surgical procedure.
The provided documentation describes the Exactech® Gibralt™ Spinal System, a medical device, and its 510(k) summary (K110197) for FDA clearance. The clearance is based on demonstrating substantial equivalence to legally marketed predicate devices.
This document does NOT describe an AI/ML powered device, nor does it detail a study involving human readers or AI algorithms as described in your request. The device is a spinal fixation system, and the "study" referred to is a series of mechanical tests and engineering analyses performed to demonstrate its physical performance and substantial equivalence to existing devices.
Therefore, many of the specific details requested regarding acceptance criteria, device performance in an AI context, sample sizes for test sets, data provenance, expert ground truth, adjudication methods, MRMC studies, or standalone algorithm performance are not applicable to this submission.
However, I can extract the information relevant to the device's performance demonstration as presented in the 510(k) summary.
Description of Acceptance Criteria and Performance (Non-AI Device)
The Exactech® Gibralt™ Spinal System is a spinal fixation system. Its "acceptance criteria" and "performance" are demonstrated through biomechanical testing and engineering analysis to show it is substantially equivalent to predicate devices. The core acceptance criterion is that the new system must withstand clinically relevant biomechanical loads, similar to the predicate devices.
Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria (Demonstration Goal) | Reported Device Performance (as demonstrated by) |
|---|---|
| Withstand clinically relevant biomechanical loads. | Mechanical testing and engineering analysis demonstrated that the proposed Gibralt Spinal System withstands clinically relevant biomechanical loads and that its mechanical performance is substantially equivalent to the cited predicate devices. |
| Equivalent intended use/indications for use. | The Gibralt Spinal System was shown to have equivalent indications for use and intended uses as the sited predicates for promoting fusion of the cervical and thoracic spine (C3-T3) for conditions like DDD, spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion, and/or tumors. |
| Composed of biocompatible materials conforming to industry standards. | The Gibralt Spinal System is composed of titanium alloy per ASTM F136, a biocompatible material, similar to the predicates. |
| Similar design features. | The Gibralt Spinal System was found to have similar design features to the predicates. |
| Dimensionally comparable to predicates. | The Gibralt Spinal System was found to be dimensionally comparable to the predicates. |
| Provided non-sterile for single use, and steam sterilizable by hospital. | The Gibralt Spinal System, like the predicates, is provided clean and non-sterile for single use only, and must be steam sterilized by the hospital prior to use using the same method as predicates. |
| No shelf-life expiration dating for the device itself. | Neither the Gibralt Spinal System nor cited predicates have shelf life expiration dating. |
| Satisfies specific ASTM standards for mechanical testing. | The device successfully underwent: - Rod-to-rod connector Flexural Grip and Torsion Grip testing in accordance with ASTM F1798. - Static Compression Bending, Static Torsion, and Dynamic Compression bending testing in accordance with ASTM F1717. - A biomechanical assessment comparing its mechanical performance to predicate devices. The results of these tests demonstrated substantial equivalence. |
Regarding the specific questions posed for AI/ML devices:
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- N/A (Not Applicable). This is a mechanical device. The "test set" consisted of physical prototypes or materials that underwent mechanical testing. The provenance relates to the manufacturing and testing procedures, not patient data.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- N/A (Not Applicable). Ground truth was established through adherence to recognized international biomechanical testing standards (ASTM F1798, ASTM F1717) and engineering analysis, not by human experts interpreting clinical data.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- N/A (Not Applicable). Adjudication methods are for human interpretation of data; this device underwent physical mechanical testing.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- N/A (Not Applicable). No AI or human reader study was conducted for this spinal fixation device.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- N/A (Not Applicable). No algorithm was involved.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- The "ground truth" for this mechanical device is defined by biomechanical engineering principles and established ASTM standards for spinal implant testing. The goal was to demonstrate that the device performs mechanically at least as well as, or comparably to, the predicate devices.
-
The sample size for the training set:
- N/A (Not Applicable). This device did not involve machine learning or a "training set."
-
How the ground truth for the training set was established:
- N/A (Not Applicable). No training set or ground truth in the context of ML was established.
{0}------------------------------------------------
510(k) Summary
JUL 15 2011
| Company: | Exactech ®, Inc2320 NW 66th CourtGainesville, FL 32653 |
|---|---|
| Date: | January 21, 2011 |
| Contact Person: | Vladislava Zaitseva |
Contact Person:
Regulatory Affairs Specialist Phone: (352) 377-1140 Fax: (352) 378-2617
Exactech® Gibralt™ Spinal System Proprietary Name:
Common Name: Spinal Fixation System
Classification Name:
- 21 CFR 888.3070 Pedicle screw spinal system, Class II, Product Code: MNI -. Orthosis, Spinal Pedicle Fixation
- 21 CFR 888.3050 Spinal interlaminal fixation orthosis, Class II, Product Code: . KWP - appliance, fixation, spinal interlaminal
Legally Marketed Devices to Which Substantial Equivalence Is Claimed
- S4 Spinal System (K050979) ●
- Vertex Reconstruction System (K003780) .
- CerviFix Axon System (K991089/K023675) .
Device Description
This submission proposes a new spinal fixation system. The proposed Gibralt Spinal System is a top-loading spinal fixation system that comprises various sizes of polyaxial screws, rods, hooks, and various connectors to provide immobilization and stabilization of spinal segments as an adjunct to fusion of the cervical, and/or upper thoracic spine. The Gibralt Spinal System components are manufactured from titanium alloy per ASTM F136. The system components are provided clean and non-sterile. The products must be steam sterilized by the hospital prior to use.
The Gibralt Spinal System includes a complete instrumentation system to assist the surgeon in the implantation of each component according to a traditional open surgical procedure.
Indications for Use
When intended to promote fusion of the cervical spine, and the thoracic spine, (C3-T3), the Gibralt Spinal System is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies),
{1}------------------------------------------------
K 110197
spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion and/or tumors. The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C3-T3) spine.
The use of polyaxial screws is limited to placement in TI-T3 in treating thoracic conditions only. Polyaxial screws are not intended to be placed in the cervical spine.
This system can be used independently or in conjunction with Exactech 5.5mm or 6.0mm rod-based Thoraco-Lumbar Pedicle Screw Systems.
Summary of Technological Characteristics
The rationale for substantial equivalence is based on consideration of the following characteristics:
. Intended Use/Indications for Use.
The Gibralt Spinal System has the equivalent indications for use and intended uses as the sited predicates.
Materials .
The Gibralt Spinal System and predicates are composed of titanium alloy, a biocompatible material conforming to a recognized industry standard for permanent implants.
. Design Features
The Gibralt Spinal System and predicates have similar design features.
. Dimensions
The Gibralt Spinal System and predicates are dimensionally comparable.
. Packaging and Sterilization
The Gibralt Spinal System and predicates are provided non-sterile for single use only, and will be steam sterilized by the hospital prior to use in the operating room using the same method.
Device Shelf Life .
Neither the Gibralt Spinal System nor cited predicates have shelf life expiration dating.
Performance specifications .
The Gibralt Spinal System and cited predicates all withstand clinically relevant biomechanical loads.
{2}------------------------------------------------
Substantial Equivalence Conclusion
The following mechanical testing and engineering analysis were conducted to demonstrate substantial equivalence of the proposed Gibralt Spinal System to the predicates:
- Rod-to-rod connector Flexural Grip and Torsion Grip testing in accordance with . ASTM F1798.
- . Static Compression Bending, Static Torsion, and Dynamic Compression bending testing in accordance with ASTM F1717.
- A biomechanical assessment comparing Gibralt Spinal System mechanical . performance to cited predicate devices.
The results of mechanical testing and analysis demonstrate that the proposed devices are substantially equivalent to cited predicates.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the symbol. The text is in all caps and is oriented to follow the curve of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Exactech, Inc. % Ms. Vladislava Zaitseva Regulatory Affairs Specialist 2320 NW 66th Court Gainesville, Florida 32653
JUL 15 2011
Re: K110197
Trade/Device Name: Exactech® Gibralt™ Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNI, KWP Dated: June 15, 2011 Received: June 16, 2011
Dear Ms. Zaitseva:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
Page 2 – Ms. Vladislava Zaitseva
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21. CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours
Sincerely yours,
Euid Keith
fo'
N
I
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use Statement
K110197 510(k) Number:
Device Name: Exactech® Gibralt™ Spinal System
INDICATIONS FOR USE:
When intended to promote fusion of the cervical spine, and the thoracic spine, (C3-T3), the Gibralt Spinal System is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion and/or tumors. The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C3-T3) spine.
The use of polyaxial screws is limited to placement in T1-T3 in treating thoracic conditions only. Polyaxial screws are not intended to be placed in the cervical spine.
This system can be used independently or in conjunction with Exactech 5.5mm or 6.0mm rod-based Thoraco-Lumbar Pedicle Screw Systems.
Prescription Use X and/or (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
Please do not write below this line - use another page if needed.
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
KI10197 510(k) Number.
N/A