K Number
K110177

Validate with FDA (Live)

Device Name
HELENA SYSTEM
Date Cleared
2011-04-21

(90 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Helena System is indicated for intervertebral body fusion of the lumbar spine, from L2 to Sl, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.

Device Description

The Helena System will be offered in various device configurations based on surgical approach and patient anatomy, and consist of: Lumbar interbody fusion device(s), which may be implanted bi-laterally via a posterior (PLIF) approach; as a single device via a transforaminal (TLIF) approach - as a single device via an anterior (ALIF) approach; or The Helena System implant components are made of polyether ether ketone (PEEK Optima LT1) that conforms to ASTM F2026. Additionally, the devices contain tantalum markers (ASTM F560) to assist the surgeon with proper placement of the device. The Helena System is implanted using a combination of device specific and universal class I instruments manufactured from stainless steel materials that conform to ASTM F899.

AI/ML Overview

The Helena System is an intervertebral body fusion device made of PEEK Optima LT1 with tantalum markers, designed for lumbar spine fusion from L2 to S1.

1. Acceptance Criteria and Device Performance

The acceptance criteria for the Helena System are based on non-clinical mechanical testing standards, demonstrating substantial equivalence to predicate devices. The device performance, as reported, met these criteria.

Acceptance CriteriaReported Device Performance
Static and Dynamic Compression Testing (per ASTM F2077-03)Substantially equivalent results compared to predicate devices, indicating acceptable mechanical integrity under physiological loads.
Subsidence Testing (per ASTM F2267-04)Substantially equivalent results compared to predicate devices, demonstrating acceptable resistance to subsidence into vertebral endplates.
Expulsion Testing (per ASTM Draft Standard F-04.25.02.02)Substantially equivalent results compared to predicate devices, indicating acceptable resistance to unintentional expulsion from the intervertebral space.

2. Sample Size and Data Provenance

  • Test Set Sample Size: Not explicitly stated in the provided documents. The testing was non-clinical, involving mechanical tests of the device itself rather than patient data.
  • Data Provenance: The data is based on laboratory non-clinical testing of the Helena System implants, following specified ASTM standards. No human or animal data provenance (e.g., country of origin, retrospective/prospective) is applicable to this type of testing.

3. Number and Qualifications of Experts for Ground Truth

Not applicable. The ground truth for the non-clinical mechanical tests is based on the performance standards defined by ASTM (American Society for Testing and Materials) specifications, not expert interpretation of medical images or patient data.

4. Adjudication Method for the Test Set

Not applicable. As the testing involved mechanical measurements against predefined standards, there was no need for expert adjudication. The results are quantitative and objective.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC study was performed or mentioned. This type of study is relevant for imaging-based diagnostic or prognostic AI tools that involve human interpretation, not for mechanical orthopedic implants.

6. Standalone Performance Study

A standalone performance study was done in the sense that the device's mechanical properties were tested independently against established standards (ASTM F2077-03, ASTM F2267-04, and ASTM Draft Standard F-04.25.02.02). The "algorithm only" concept is not applicable here as it's a physical medical device. The tests evaluated the device's ability to withstand various forces encountered in the body.

7. Type of Ground Truth Used

The ground truth used was objective engineering standards and performance criteria as defined by the ASTM standards (ASTM F2077-03 for static and dynamic compression, ASTM F2267-04 for subsidence, and ASTM Draft Standard F-04.25.02.02 for expulsion).

8. Sample Size for the Training Set

Not applicable. There is no "training set" in the context of mechanical testing for a physical medical device like this. Machine learning algorithms require training sets, but this regulatory submission concerns a physical implant.

9. How Ground Truth for the Training Set Was Established

Not applicable, as there is no training set.

{0}------------------------------------------------

K110177
Page 1 of 2

510(k) Summary Royal Oak Medical Devices Helena System

APR 2 1 2011

Premarket Notification

SUBMITTED BY

Royal Oak Medical Devices 39533 Woodward Avenue Bloomfield Hills, MI 48304

ESTABLISHMENT REGISTRATION NUMBER

Pending

OWNER/OPERATOR NUMBER

Pending

CONTACT PERSON

Matthew Kroll Vice President of Technical Services Royal Oak Medical Devices Phone: 248-628-2830 Fax: 248-969-8263

Intervertebral Body Fusion Device

888.3080 (Product Code: MAX)

SUBMISSION PREPARED BY

Lisa Peterson QA Consulting, Inc. Phone: 512-507-0746

March 17, 2011

Helena System

Class II

DATE PREPARED

CLASSIFICATION NAME

DEVICE CLASS

REGULATION NUMBER

COMMON NAME

PROPRIETARY NAME

IDENTIFICATION OF PREDICATE DEVICE(S)

Predicate devices include various cleared interbody fusion systems:

Intervertebral Fusion Device with Bone Graft, Lumbar

  • RAY® Threaded Fusion Cage (P950019) -
  • -Lumbar I/F Cage (P960025)
  • -Eminent Spine System (K090064)
  • SpineSmith Cynch System (K090376) -
  • BAK Vista (P950002) -

{1}------------------------------------------------

KII017

page 2 of 2

DEVICE DESCRIPTION

The Helena System will be offered in various device configurations based on surgical approach and patient anatomy, and consist of:

Lumbar interbody fusion device(s), which may be implanted

  • bi-laterally via a posterior (PLIF) approach;
  • as a single device via a transforaminal (TLIF) approach -
  • as a single device via an anterior (ALIF) approach; or

The Helena System implant components are made of polyether ether ketone (PEEK Optima LT1) that conforms to ASTM F2026. Additionally, the devices contain tantalum markers (ASTM F560) to assist the surgeon with proper placement of the device.

The Helena System is implanted using a combination of device specific and universal class I instruments manufactured from stainless steel materials that conform to ASTM F899.

INDICATIONS

The Helena System is indicated for intervertebral body fusion of the lumbar spine, from L2 to Sl, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.

SUMMARY OF TECHNOLOGICAL CHARACTERISTICS

The purpose of this premarket notification is to obtain clearance to market the Helena System. The Helena System is comprised of various device configurations designed to accommodate patient anatomy and provide the surgeon with different surgical approach options.

The Helena System implant components are made of polyether ether ketone (PEEK Optima LT1) that conforms to ASTM F2026. Additionally, the devices contain tantalum markers (ASTM F560) to assist the surgeon with proper placement of the device.

The subject system has similar technological characteristics as the predicate devices identified above. Specifically, the following characteristics support this conclusion:

  • Intended for use at either one level or two contiquous levels from L2-S1 for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis at the involved level(s).
  • Substantially equivalent results of non-clinical testing relative to static and dynamic testing -(per ASTM F2077-03), subsidence (per ASTM F2267-04), and expulsion (per ASTM Draft Standard F-04.25.02.02)

DISCUSSION OF NON-CLINICAL TESTING

The following non-clinical tests were conducted:

  • Static and dynamic compression testing, conducted in accordance with ASTM F2077-03 ।
  • Subsidence testing, conducted in accordance with ASTM F2267-04 ।
  • Expulsion testing, conducted in accordance with ASTM Draft Standard F-04.25.02.02 -

CONCLUSIONS

The subject and predicate device(s) share the same intended use, primary implant design and equivalent material of manufacture. The non-clinical mechanical test results demonstrate that any minor differences do not impact device performance as compared to the predicates and demonstrate that the Helena System is substantially equivalent to the predicate device.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three parallel lines forming the staff and a single snake winding around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 .

Royal Oak Medical Devices % Mr. Matthew Kroll Vice President of Technical Services 39533 Woodward Avenue Bloomfield Hills, Michigan 48304

APR 2 1 2011

Re: K110177

Trade/Device Name: Helena System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: March 21, 2011 Received: March 23, 2011

Dear Mr. Kroll:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{3}------------------------------------------------

Page 2 -- Mr. Matthew Kroll

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/McdicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Hing B. Rha

Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

K110177 510(k) Number (if known):

Device Name: Helena System

Indications for Use:

The Helena System is indicated for intervertebral body fusion of the lumbar spine, from L2 to SI, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiquous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-Off

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

KI10177 510(k) Number_________________________________________________________________________________________________________________________________________________________________

39533 Woodward Avenue Bloomfield Hills, Michigan 48304 Royal Oak Medical Devices

Phone: 248-628-2830 Fax: 248-969-8263

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.