K Number
K110156

Validate with FDA (Live)

Date Cleared
2011-07-27

(189 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ABPM50 Automatic Blood Pressure Monitor is intended to monitor the systolic, diastolic and mean blood pressure as well as pulse rate via non-invasive oscillometric technique in which an inflatable cuff is wrapped around the upper arm. It can be used on adult, pediatric and neonatal individuals.

Device Description

The proposed device, ABPM50 Automatic Blood Pressure Monitor, is battery driven automatic non-invasive Blood Pressure Monitor. It can automatically complete the inflation and BP measurement, which can measure systolic, diastolic and mean blood pressure as well as the pulse rate at upper arm within its claimed range and accuracy via the oscillometric technique. User can select the unit of the measurement: mmHg or kPa. ABPM50 can be used on adult, pediatric, and neonatal individuals.

The device has the data storage function for data review including measurement time, systolic blood pressure, diastolic blood pressure, mean blood pressure and pulse rate. These records can be uploaded to PC via USB and processed with the PC software.

ABPM50 has physiological alarm function which can be turned on or off by users. When the measurement results exceed the alarm limit, the physiological alarm function will be triggered. The alarm limit can be set by users, and the low limit must be lower than the corresponding high limit. In addition, it has technical alarm function which will be triggered when the battery voltage is lower than 2.3V , and this alarm can not be cancelled unless being closed or the power replaced.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the ABPM50 Automatic Blood Pressure Monitor, based on the provided text:

Acceptance Criteria and Device Performance

The provided document states that the clinical tests were conducted following the ANSI/AAMI SP10:2002+A1:2003+A2:2006 standard for manual, electronic, or automated sphygmomanometers. While the specific numerical acceptance criteria are not explicitly detailed in the provided text, this standard generally outlines requirements for accuracy of blood pressure measurements. The document does state that the device "complies with the standard requirements and the accuracy the manufacture declared." Without the explicit declared accuracy from the manufacturer, we can infer that the device meets the general accuracy requirements of the ANSI/AAMI SP10 standard.

General Acceptance Criteria (Inferred from ANSI/AAMI SP10):
The ANSI/AAMI SP10 standard typically requires that:

  • The mean difference between the device measurement and reference measurement (mercury sphygmomanometer) should be ± 5 mmHg or less.
  • The standard deviation of the differences should be ± 8 mmHg or less.
  • A minimum number of subjects (e.g., 85 adult subjects) should be tested.
Acceptance Criteria Category (Inferred from ANSI/AAMI SP10)Reported Device Performance
Mean difference (device vs. reference) for BPComplies with standard
Standard deviation of differences for BPComplies with standard
Accuracy of measurement (Systolic, Diastolic, Mean BP, Pulse Rate)Complies with standard and manufacturer's declared accuracy

Study Details

  1. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated. The document refers to "clinical tests following ANSI/AAMI SP10:2002+A1:2003+A2:2006". This standard specifies a minimum number of subjects (e.g., 85 for adults, with specific requirements for pediatric and neonatal populations, including a minimum of 15 subjects for each age group within those populations). Given it covers "adult, pediatric, and neonatal individuals," it would involve multiple age cohorts.
    • Data Provenance: The study was conducted in Qinhuangdao Maternal and Child Health Hospital. This indicates a prospective study conducted in China.
  2. Number of experts used to establish the ground truth for the test set and their qualifications:

    • Not specified in the provided text. The ANSI/AAMI SP10 standard typically requires simultaneous auscultation by two trained observers for ground truth, often using a Y-tube and a third supervisor, but the exact number and qualifications are not detailed here.
  3. Adjudication method for the test set:

    • Not specified. Assuming adherence to ANSI/AAMI SP10, the ground truth would typically be established by trained observers taking manual readings, often with a method to resolve discrepancies between observers, such as averaging or a third adjudicator.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    • No, a MRMC comparative effectiveness study was not conducted. This type of study is more common for diagnostic imaging AI systems where human reader performance is a key metric. This study focused on the standalone accuracy of the blood pressure monitor against a reference method.

    • Effect size of human readers improvement with AI vs. without AI assistance: Not applicable, as no MRMC study was performed.

  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, the study described is effectively a standalone performance study of the ABPM50 device. It tested the device's accuracy in measuring blood pressure and pulse rate (algorithm only) compared to a reference standard (human-measured manual blood pressure, implied by ANSI/AAMI SP10).
  6. The type of ground truth used:

    • Expert concensus / Manual auscultatory readings from trained observers. The ANSI/AAMI SP10 standard dictates that the reference blood pressure measurements are taken by trained observers using a mercury sphygmomanometer (or an equivalent validated reference device) simultaneously with the device under test.
  7. The sample size for the training set:

    • Not applicable. The provided document describes a clinical validation study for a device, not a machine learning model that requires a distinct training set. The device operates based on oscillometric principles, not a deep learning algorithm trained on a large dataset in the typical sense.
  8. How the ground truth for the training set was established:

    • Not applicable, as a training set for a machine learning model is not explicitly mentioned or implied for this device's function. The "training" for such a device would instead relate to the engineering and calibration processes during development, which aren't typically described with ground truth in the same way as AI models.

{0}------------------------------------------------

JUL 2 7 2011

Attachment II 510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) Number: __________________________________________________________________________________________________________________________________________________

  1. Date of Submission: December 27, 2010
    1. Sponsor
      CONTEC MEDICAL SYSTEMS CO., LTD NO.24 Huanghe West Road, Economic & Technical Development Zone, Qinhuangdao, Hebei Province, 066004,China Establishment Registration Number: 3006979678 Contact Person: Xueyong Li Position: Quality Manager Tel: +86-335-8015490 Fax: +86-335-8015489 Email: lxy1011@163.com
    1. Submission Correspondent Ms. Diana Hong & Mr. Lee Fu Mid-Link Consulting Co., Ltd P.O. Box 237-023, Shanghai, 200237, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net
      Proposed Device Identification 4.

Proposed Device Name: Automatic Blood Pressure Monitor Proposed Device Model: ABPM50 Classification: Class II Product Code: DXN Regulation Number: 21 CFR 870.1130 Review Panel: Cardiovascular Intended Use Statement:

ABPM50 Automatic Blood Pressure Monitor is intended to monitor the systolic, diastolic and mean blood pressure as well as pulse rate via non-invasive oscillometric technique in which an inflatable

{1}------------------------------------------------

cuff is wrapped around the upper arm. It can be used on adult, pediatric and neonatal individuals.

  • Predicate Device Identification 5.
    510(k) Number: K091068 Product Name: Accutorr V Monitor Manufacturer: Datascope Patient Monitoring, Mindray DS USA, Inc.

  • Device Description 6.
    The proposed device, ABPM50 Automatic Blood Pressure Monitor, is battery driven automatic non-invasive Blood Pressure Monitor. It can automatically complete the inflation and BP measurement, which can measure systolic, diastolic and mean blood pressure as well as the pulse rate at upper arm within its claimed range and accuracy via the oscillometric technique. User can select the unit of the measurement: mmHg or kPa. ABPM50 can be used on adult, pediatric, and neonatal individuals.

The device has the data storage function for data review including measurement time, systolic blood pressure, diastolic blood pressure, mean blood pressure and pulse rate. These records can be uploaded to PC via USB and processed with the PC software.

ABPM50 has physiological alarm function which can be turned on or off by users. When the measurement results exceed the alarm limit, the physiological alarm function will be triggered. The alarm limit can be set by users, and the low limit must be lower than the corresponding high limit. In addition, it has technical alarm function which will be triggered when the battery voltage is lower than 2.3V , and this alarm can not be cancelled unless being closed or the power replaced.

Test Conclusion 7.

Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

IEC 60601-1: 1988 +A1: 1991 +A2: 1995, Medical Electrical Equipment - Part 1: General requirements for safety.

IEC 60601-1-2: 2007, Medical Electrical Equipment - Part 1: General requirements for safety-2, Collateral Standard: Electromagnetic compatibility - Requirements and tests.

ANSI/AAMI SP10: 2002 & A1: 2003 & A2: 2006 , Manual, electronic, or automated sphygmomanometers

The clinical tests following ANSI/AAMI SP10:2002+A1:2003+A2:2006 are conducted on proposed II-2

{2}------------------------------------------------

device in Qinhuangdao Maternal and Child Health Hospital. The test results demonstrated that the proposed device comply with the standard requirements and the accuracy the manufacture declared.

    1. Substantially Equivalent Conclusion
      The proposed device, ABPM50 Automatic Blood Pressure Monitor, is determined to be Substantially Equivalent (SE) to the predicate device, Accutorr V Monitor (K091068), in respect of safety and effectiveness.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird with three wing-like shapes. The logo is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a sans-serif font.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Contec Medical Systems Co., Ltd. c/o Diana Hong Midlink Consulting Co., Ltd. PO Box 237-023 Shanghai, China 200237

JUL 2 7 2011

Re: K110156

Trade/Device Name: ABPM50 Automatic Blood Pressure Monitors Regulation Number: 21 CFR 870.1130 Regulation Name: Non-invasive blood pressure measurement systems Regulatory Class: Class II (two) Product Code: DXN Dated: July 18, 2011 Received: July 21, 2011

Dear Ms. Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Diana Hong

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Attachment III Indications for Use

510(k) Number: K110156

Device Name: Automatic Blood Pressure Monitor, ABPM50

Indications for Use:

ABPM50 Automatic Blood Pressure Monitor is intended to monitor the systolic, diastolic and mean blood pressure as well as pulse rate via non-invasive oscillometric technique in which an inflatable cuff is wrapped around the upper arm. It can be used on adult, pediatric and neonatal individuals

& PRESCRIPTION USE (Part 21 CFR 801 Subpart D) [ ]OVER-THE-COUNTER USE (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

(Division Sign-Off)
Division of Cardiovascular Devices

510(k) Number K//0/56

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).