K Number
K110124

Validate with FDA (Live)

Date Cleared
2011-02-04

(17 days)

Product Code
Regulation Number
884.2660
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Fetal Doppler JPD-100S is a hand-held, battery powered audio Doppler device used for detecting fetal heart beats.

Device Description

JPD-100S Fetal Doppler is a prescription from licensed physician in hospitals, clinics and private offices. It is a hand-held, battery powered audio Doppler device integrated with 3MHz probe, used for detecting fetal heart beats. And the device is normally applied to pregnancy for more than twelve weeks.

AI/ML Overview

The provided document describes the acceptance criteria and a study for the Fetal Doppler JPD-100S. However, the study conducted is primarily for substantive equivalence to a predicate device, focusing on electrical safety, EMC, and biological compatibility, rather than a clinical performance study with defined performance metrics like sensitivity or specificity.

Here's an analysis based on the information provided, highlighting what is present and what is not:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not present a table of specific clinical acceptance criteria (e.g., minimum sensitivity, specificity, accuracy for detecting fetal heartbeats). Instead, the "acceptance criteria" are implied by the standards the device was tested against and found compliant.

Acceptance Criterion (Standard Compliance)Reported Device Performance (Compliance Status)
IEC60601-1 (1988; Amendment 1, 1991-11, Amendment 2, 1995): Medical electrical equipment - Part 1: General requirements for basic safety and essential performanceCompliant
IEC60601-1-2 (2001): Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and testsCompliant
IEC 61266 (1994): Ultrasonic - Hand-held probe Doppler fetal heartbeat detectors - Performance requirements and methods of measurement and reportingCompliant
ISO 10993-5 (2009): Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicityCompliant
ISO 10993-10 (2002; Amd. 1:2006): Biological evaluation of medical devices - Part 10: Tests for irritation and delayed-type hypersensitivityCompliant
Transducer Acoustic Output (Isata and frequency)Compliant with standard criteria

2. Sample Size Used for the Test Set and Data Provenance

This information is not provided in the document. The "Test Conclusion" section mentions "Laboratory testing" but does not detail a clinical test set, sample size of patients, or data provenance. The assessment focuses on hardware compliance with standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This information is not provided. As no clinical test set for detecting fetal heartbeats is described, there's no mention of experts or their qualifications for establishing ground truth.

4. Adjudication Method for the Test Set

This information is not provided. No adjudication method is mentioned as a clinical test set is not described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The study described is a submission for substantial equivalence based on technical specifications and compliance with standards, not a clinical trial comparing human readers with and without AI assistance, or comparing this device to others clinically.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The device is a "hand-held, battery powered audio Doppler device used for detecting fetal heart beats." This implies human-in-the-loop operation, and the documentation does not describe an "algorithm only" standalone performance evaluation. The "detection" is performed by the device, and the output (audio) is interpreted by a human user. The study focuses on the device's technical specifications and safety, not its standalone diagnostic accuracy as an autonomous algorithm.

7. The Type of Ground Truth Used

The concept of "ground truth" as typically understood in AI/clinical performance studies (e.g., pathology, outcomes data) is not applicable to the data presented. The "ground truth" for the tests performed was compliance with engineering and safety standards (e.g., electrical safety tests, EMC tests, acoustic output measurements, biocompatibility tests). For the function of detecting fetal heartbeats, the "performance requirements" of IEC 61266 (1994) would define what constitutes "detection," but the document doesn't detail how this was clinically verified beyond stating compliance with the standard.

8. The Sample Size for the Training Set

This information is not provided. There is no mention of a "training set," as the device described is a hardware medical device (fetal Doppler) and not an AI/machine learning algorithm requiring a training phase for its core function.

9. How the Ground Truth for the Training Set Was Established

This information is not provided and is not applicable as there is no mention of a training set for an AI/ML algorithm.

{0}------------------------------------------------

. . .. ..

.

K110124

Attachment II 510(k) Summary

As Required by CFR 870.92

Preparing DateNovember 23, 2010FEB - 4 2011
SponsorShenzhen Jumper Medical Equipment Co., Ltd
5th Floor, Building No.34, Baoyuan Industrial Zone, Xixiang Street,
Baoan District, Shenzhen, Guangdong 518067, P.R. China
Boley He Quality Manager
Tel: +86-755-2669 6279
Fax: +86-755-2685 2678
Email: rd@jumper-medical.com
SubmissionCorrespondentMs. Diana Hong / Mr. Lee Fu
Shanghai Mid-Link Business Consulting Co., Ltd
P.O. BOX 237-023
Shanghai, 200237, China
Tel: +86-21-64264467
Fax: 240-238-7587
Email: diana.hong@mid-link.net
Proposed Product
Trade NameFetal Doppler
ModelJPD-100S
Product Code:KNG
Regulation Number:21 CFR 884.2660
Device Class:Class II
Submission Purpose:New Device
Submission TypeTraditional 510(k)
Predicate Device:
Device NameContec Pocket Fetal Doppler Baby Sound A
K NumberK082480
ManufacturerContec Medical Systems Co., LTD
Intended UseThe Fetal Doppler JPD-100S is a hand-held, battery powered audio
Doppler device used for detecting fetal heart beats.
Device DescriptionJPD-100S Fetal Doppler is a prescription from licensed physician inhospitals, clinics and private offices. It is a hand-held, battery poweredaudio Doppler device integrated with 3MHz probe, used for detectingfetal heart beats. And the device is normally applied to pregnancy formore than twelve weeks.
Test ConclusionLaboratory testing was conducted to validate and verify that theproposed devices met all design specifications, including electricalsafety, EMC, specification per following standards.
IEC60601-1(1988; Amendment 1, 1991-11, Amendment 2, 1995):Medical electrical equipment -Part 1: General requirements for basicsafety and essential performanceIEC60601-1-2 (2001): Medical electrical equipment - Part 1-2:General requirements for basic safety and essential performance -Collateral standard: Electromagnetic compatibility -Requirements andtestsIEC 61266 (1994 ): Ultrasonic - Hand-held probe Doppler fetalheartbeat detectors -Performance requirements and methods ofmeasurement and reportingISO 10993-5 (2009): Biological evaluation of medical devices -- Part5: Tests for In Vitro cytotoxicityISO 10993-10 (2002; Amd. 1:2006): Biological evaluation of medicaldevices - Part 10: Tests for irritation and delayed-type hypersensitivityResults of these tests demonstrate compliance to the requirements ofall consensus standards.
Comparison SummaryThe proposed device shares same classification, intended use, similartechnical specifications, same safety specifications and skin-contactmaterial with the predicate device. The main difference includesbattery, operation and storage conditions, transducer frequency andIsata, but the difference is considered to have no effect oneffectiveness and safe. The SE analysis of the difference is provided asfollowing.The batteries of proposed and predicate device are different. But bothof them comply with the requirements of electrical safety and EMC ofIEC 60601-1 and IEC 60601-1-2. Therefore, this difference is

.

{1}------------------------------------------------

イ・・・・

11-2

{2}------------------------------------------------

considered has no effect on effectiveness and safety.

The operation and storage conditions of the proposed device have slight differences with those of the predicate device. But these conditions are considered to be able to cover the normal working and storage environment. In addition, these conditions are clearly on the proposed labeling, to remind the user handle and operate with these conditions. Therefore, this difference is considered to have no effect on effectiveness and safe.

The transducer frequency of proposed device and predicate device is different. But the proposed transducer was tested for its acoustic output, and the result demonstrates that it complies with the standard criteria. In addition, 3MHz transducer of fetal doppler is very common in clinical use, such as, 3MHz normal transducer of Contec Pocket Fetal Doppler SONOLINE A and SONOLINE B (K082480). Therefore, this difference is considered to have no effect on effectiveness and safe.

The Isata of proposed device and predicate device is different. But the proposed transducer was tested for its acoustic output, and the result demonstrates that it complies with the standard criteria. In addition, transducer of fetal doppler with Isata <5 mW/cm² is very common in clinical use, such as, CMS-150-T2 and CMS-150-T5 transducer of Contec Pocket Fetal Doppler SONOLINE A and SONOLINE B (K082480). Therefore, this difference is considered to have no effect on effectiveness and safe.

SE Determination

The proposed device, Fetal Doppler JPD-100S, is substantially equivalent (SE) to the predicate device Contec Pocket Fetal Doppler Baby Sound A (K082480).

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - U.S.A." is arranged in a circular pattern around the bird-like figure. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

Shenzhen Jumper Medical Equipment Co., Ltd % Mr. Marc M. Mouser Engineering Leader/FDA Office Coordinator Underwriters Laboratories, Inc. 2600 N.W. Lake Road CAMAS WA 98607-8542

FEB - 4 2011

Re: K110124

Trade/Device Name: Fetal Doppler JPD-100s Regulation Number: 21 CFR 884.2660 Regulation Name: Fetal ultrasound monitor and accessories Regulatory Class: II Product Code: KNG Dated: December 17, 2010 Received: January 18, 2011

Dear Mr. Mouser:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the Fetal Doppler JPD-100s, as described in your premarket notification:

Transducer Model Number

3.0MHz CW

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

and the country of the county of

{4}------------------------------------------------

Page 2 - Mr. Mouser

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

If you have any questions regarding the content of this letter, please contact Lauren Hefner at (301) 796-6881.

Sincerely Yours,

Amharl. D. O'Rhum for

Mary Pastel, ScD. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure(s)

{5}------------------------------------------------

Section II Indications for Use Statement

510(k) Number: Device Name:

Fetal Doppier JPD-100S

Indications for Use:

The Fetal Doppler JPD-100S is a hand-held, battery powered audio Doppler device used for detecting fetal heart beats.

Prescription Use > X AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THE LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)


(Division Sign Off)

(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) Number K110124

Page 1 of 2

{6}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

Device Name: Fetal Doppler JPD-100S

Transducer: 3.0MHz CW inherent of the main unit

Intended Use: Detect fetal heart beats as follows:

Clinical ApplicationMode of Operation
General(Track 1 only)Specific (Track 1&3)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalN
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Specify)

N= new indication: P= previously cleared by FDA; E= added under Appendix E Comments:

(PLEASE DO NOT WRITE BELOW THE LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

_Page 2 of 2 II-2

Division of Radiological Devices Office of In Vitro Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety

510K K110621

§ 884.2660 Fetal ultrasonic monitor and accessories.

(a)
Identification. A fetal ultrasonic monitor is a device designed to transmit and receive ultrasonic energy into and from the pregnant woman, usually by means of continuous wave (doppler) echoscopy. The device is used to represent some physiological condition or characteristic in a measured value over a period of time (e.g., perinatal monitoring during labor) or in an immediately perceptible form (e.g., use of the ultrasonic stethoscope). This generic type of device may include the following accessories: signal analysis and display equipment, electronic interfaces for other equipment, patient and equipment supports, and component parts. This generic type of device does not include devices used to image some relatively unchanging physiological structure or interpret a physiological condition, but does include devices which may be set to alarm automatically at a predetermined threshold value.(b)
Classification. Class II (performance standards).