(501 days)
HL568 Self-Monitoring Blood Glucose (SMBG) System is self-test medical device and intended for single patient home-use to monitor the blood glucose (B-D-glucose) levels in quantitative measurement from fresh capillary whole blood obtained from the finger tip. It is intended for use outside the body (in vitro diagnostic use) by diabetics at home to measure the glucose concentration for aiding diabetes management. It is not intended for the diagnosis of or screening for diabetes mellitus.
HL568 Self-Monitoring Blood Glucose (SMBG) System is intended to be used by a single person and should not be shared and it is also not intended for use on neonates and should be used with HL568 Blood glucose test strip and HL568 Control Solution.
HL568 Blood glucose test strips are for use with the HL568 Blood glucose meter to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertip.
HL568 Control Solutions are for use with the HL568 Blood glucose meter and HL568 Blood glucose test strips to check that the meter and test strips are working together properly and that the test is performing correctly.
The device is capable of transferring the storage data to the connected personal computer (PC) via USB cable. In addition, HL568 Self-Monitoring Blood Glucose System is featured with audio function; it could help the user to know the measured result by hearing but is not intended for visually impaired users.
The HLS68 Self-Monitoring Blood Glucose System is comprised of the HL568 Blood Glucose Meter, HL568 Blood Glucose Test Strip, HL568 Control Solution (2 levels), a lancing device and lancets. All the measured values can be read out in one LCD panel, battery operated and the blood glucose meter is portable. The Subject Device is self-test IVD medical device which is intended for quantitative measurement of blood glucose (B-D-glucose) levels from fresh capillary whole blood obtained from the fingertip and used outside the body only ( in vitro diagnostic use). It is intended for over-the-counter, home use by single patient to measure the glucose concentration for aiding diabetes management.
The test principle based on the electrochemical biosensor technology using glucose oxidase method. Each test strip reacts with glucose in the blood sample to produce a current proportional to the blood glucose level. This reaction is measured by the meter and displayed as the blood glucose result.
The Subject Device is designed to have USB computer interface to allow users make use of their storage memory data more personalized. Users can transfer the readings of their blood glucose meters to the connected personal computer (PC) via USB cable to save and manage their personal data.
Subject Device, HL568 Blood Glucose Meter is featured with an audio function for use to easily and efficiently perform the glucose testing. It provides an auditory aid for the users to know the measured result by hearing and to use the device step by step.
Here's a summary of the acceptance criteria and study information for the HL568 Self-Monitoring Blood Glucose System, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
The document states that the HL568 Self-Monitoring Blood Glucose System is compliant with ISO 15197-2003: In vitro diagnostic test systems - Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. While the document doesn't explicitly list the detailed acceptance criteria and specific performance metrics achieved (e.g., mean absolute relative difference percentage), it broadly indicates that "All the relevant activities were performed by designate individual(s) and the results demonstrated that the predetermined acceptance criteria were fully met."
Therefore, for the purpose of this request, the acceptance criteria are considered to be those outlined in ISO 15197-2003. Without the full standard or detailed report, the specific numerical acceptance criteria and the device's precise performance against them cannot be provided in a table. However, the study concludes that these criteria were met.
General Performance Characteristics (Extracted from the "Comparison of Subject Devices and predicate device" table, implying parity with predicate):
| Characteristic | Acceptance Criteria Implied by Predicate (and Met by Subject Device) | Reported Device Performance (Implied as met) |
|---|---|---|
| Detection method | Amperometry: current generated by oxidation of reduced mediator | Same |
| Enzyme | Glucose Oxidase | Same |
| Mediator | Ferricyanide | Same |
| Open Stability | 3 months | Same |
| Sample Type | Capillary whole blood | Same |
| Calibration | Plasma equivalence | Same |
| Test range | 20 - 600 mg/dL | Same |
| Coding Method | Coding by button | Same |
| Altitude limit | >10,000 feet (predicate is 10,000 feet) | 10,335 feet |
| Sample volume | ≥ 1µL | > 1µL |
| Hematocrit Range | 30 - 55% | Same |
| Test time | 5 seconds | 5 seconds (audio function off) |
| Operating Humidity | R.H. ≤ 80% (predicate is 10% to 90%) | R.H. ≤ 80% |
| Day average (reporting) | Including 14, 30-day averages | 7, 14, 28, 60, 90-day average |
2. Sample size used for the test set and the data provenance:
The document mentions compliance with ISO 15197-2003 and other CLSI standards (EP05-A2, EP06-A, EP07-A2, EP09-A2). These standards outline specific requirements for sample sizes in various tests (e.g., precision, linearity, interference, method comparison). However, the specific number of samples used for the test set is not explicitly stated in the provided text.
The data provenance (e.g., country of origin, retrospective or prospective) is also not explicitly stated.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not provided in the document. For blood glucose monitoring systems, ground truth is typically established using a laboratory reference method, not expert consensus in the way image analysis or clinical diagnosis might.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This information is not applicable and therefore not provided in the document, as the ground truth for blood glucose measurements is generally established by a laboratory reference method, not by expert adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
A multi-reader multi-case (MRMC) comparative effectiveness study is not mentioned and is not typically relevant for a self-monitoring blood glucose system. This device is an automated measurement system, not an AI-assisted diagnostic tool that humans read.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
The device described is a standalone algorithm/device in the sense that it provides a direct blood glucose measurement without requiring human interpretation of visual data. The performance described (compliance with ISO 15197-2003) refers to the performance of the device itself.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
While not explicitly stated in detail, for blood glucose monitoring systems, the ground truth is established by a laboratory reference method, typically a central laboratory analyzer that provides highly accurate and precise glucose measurements. This is standard practice for studies complying with ISO 15197. The document mentions compliance with "CLSI/NCCLS EP09-A2: Method Comparison and Bias Estimation Using Patient Samples," which further supports the use of a reference method for comparison.
8. The sample size for the training set:
Blood glucose monitoring systems based on electrochemical biosensor technology, like the HL568, are typically developed and validated using well-established biochemical principles and sensor design, rather than large-scale machine learning training sets in the same way an AI image analysis algorithm would be. Therefore, the concept of a "training set" in the context of machine learning, with a distinct sample size, is not applicable or mentioned for this device. Calibration and initial validation would involve a range of glucose concentrations and interfering substances, but not in the "training set" paradigm.
9. How the ground truth for the training set was established:
As explained above, the concept of a "training set" as it applies to AI/ML is generally not relevant for this type of device. The accuracy of the device is assessed against a laboratory reference method during its development and validation phases.
{0}------------------------------------------------
MAY 2 5 2012
REMARKET NOTIFICATION
510(k) SUMMARY
(As Required By 21 CFR 807.92)
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: K110074
Date: 2012.02.29 revised
1: Submitter:
Health & Life Co., Ltd. 9F, No.186, Jian Yi Road, Chung-Ho District, New Taipei City, Taiwan, R.O.C. TEL: +886-2-8227-1300 FAX: +886-2-8227-1301
Contact person: Sarah Su/ Regulatory Affairs Dept. E-mail: sarah.su@hlmt.com.tw Tel: 886-2-8227-1300 ext.1201 Fax: 886-2-8227-1301
2. Name of the Device:
Proprietary Name: HL568 Self-Monitoring Blood Glucose System Common Name: Blood Glucose Monitoring System Classification Name: Glucose Test System Classification: 21 CFR 862.1345 Glucose Test System, Class II
21 CFR 862.1660 Quality Control Material, Class I
Product Code: NBW, CGA, JJX
Panel: Clinical Chemistry (75)
{1}------------------------------------------------
3. Information for the 510(k) Cleared Device (Predicate Device):
ONE TOUCH ULTRA blood Glucose Monitoring System
- a. Device name: ONE TOUCH ULTRA blood Glucose Monitoring System
- b. Classification Name: Glucose Test System
- c. Common Name: Glucose Test System
- d. Device Classification : Class II ; Class I
- Regulatory Information: 21 CFR 862.1345 ; 21 CFR 862.1660 e.
- Panel: Clinical Chemical (75) ﻟﻠﻨﻔﺴﺎ
- Product code: NBW g.
- h. 510(K) Number: K062195
4. Device Description:
4.1 Blood Glucose Concentration Measurement
The HLS68 Self-Monitoring Blood Glucose System is comprised of the HL568 Blood Glucose Meter, HL568 Blood Glucose Test Strip, HL568 Control Solution (2 levels), a lancing device and lancets. All the measured values can be read out in one LCD panel, battery operated and the blood glucose meter is portable. The Subject Device is self-test IVD medical device which is intended for quantitative measurement of blood glucose (B-D-glucose) levels from fresh capillary whole blood obtained from the fingertip and used outside the body only ( in vitro diagnostic use). It is intended for over-the-counter, home use by single patient to measure the glucose concentration for aiding diabetes management.
4.2 Test Principle
The test principle based on the electrochemical biosensor technology using glucose oxidase method. Each test strip reacts with glucose in the blood sample to produce a current proportional to the blood glucose level. This reaction is measured by the meter and displayed as the blood glucose result.
4.3 USB data transmission Function
The Subject Device is designed to have USB computer interface to allow users make use of their storage memory data more personalized. Users can transfer the readings of their blood glucose meters to the connected personal computer (PC) via USB cable to save and manage their personal data. Once the software program included in the accompany software CD is installed, users can manage their measurement data personally, such as transferring data, deleting data, viewing the readings on a chart and restoring data. (Please refer to Attachment A-4-1, A-4-7(SDS, SRS), A-4-11(Functionality validation report) and A-5-5(D-FMEA) for more detail information)
{2}------------------------------------------------
4.4 Audio function
Subject Device, HL568 Blood Glucose Meter is featured with an audio function for use to easily and efficiently perform the glucose testing. It provides an auditory aid for the users to know the measured result by hearing and to use the device step by step. When your or symbol appears on the screen, it indicates the audio function is on with volume from high to low. User can use this optional feature at their discretion. (Please refer to Attachment A-4-1, A-4-7 (SDS, SRS), A-4-11 (Functionality validation report) and A-5-5(D-FMEA) for more detail information)
4.5 Special condition for use statement(s)
HL568 Blood Glucose Test Strip and HL568 Control Solution are designed to be used only with the HL568 Blood Glucose Meter. Also, the proposed device is not intended for use in the alternative site testing other than the fingertip.
5. Intended Use
HL568 Self-Monitoring Blood Glucose (SMBG) System is self-test medical device and intended for single patient home-use to monitor the blood glucose (β-D-glucose) levels in quantitative measurement from fresh capillary whole blood obtained from the finger tip. It is intended for use outside the body (in vitro diagnostic use) by diabetics at home to measure the glucose concentration for aiding diabetes management. It is not intended for the diagnosis of or screening for diabetes mellitus.
HL568 Self-Monitoring Blood Glucose (SMBG) System is intended to be used by a single person and should not be shared and it is also not intended for use on neonates and should be used with HL568 Blood glucose test strip and HL568 Control Solution.
HL568 Blood glucose test strips are for use with the HL568 Blood glucose meter to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertip.
HL568 Control Solutions are for use with the HL568 Blood glucose meter and HL568 Blood glucose test strips to check that the meter and test strips are working together properly and that the test is performing correctly.
The device is capable of transferring the storage data to the connected personal computer (PC) via USB cable. In addition, HL568 Self-Monitoring Blood Glucose System is featured with audio function; it could help the user to know the measured result by hearing but is not
{3}------------------------------------------------
intended for visually impaired users.
-
- Comparison of Subject Devices and predicate device:
- Technological Characteristics Comparison Table of HL568 Self-Monitoring Blood Glucose System and ONE TOUCH ULTRA blood Glucose Monitoring System (K062195)
Similarities ( 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - Similarities , '
| Item | Subject Device | Predicate Device |
|---|---|---|
| HL568 SMBG System | OneTouch Ultra Blood GlucoseMonitoring System | |
| Detectionmethod | Amperometry: current isgenerated by oxidation ofreduced mediator | same |
| Enzyme | Glucose Oxidase | same |
| Mediator | ferricyanide | same |
| Open Stability | 3 months | same |
| Sample Type | Capillary whole blood | same |
| Calibration | Plasma equivalence | same |
| Test range | 20 - 600 mg/dL | same |
| Coding Method | Coding by button | same |
| Altitude limit | 10,335 feet | 10,000 feet |
| Sample volume | > 1µL | Minimum 1µL |
| Hematocrit Range | 30 - 55% | same |
| Test time | 5 seconds | same |
| (audio function off) | ||
| Operating Humidity | R.H. ≤ 80% | R.H.:10% to 90% |
| Day average | 7, 14, 28, 60, 90-day average | 14, 30-day average |
| Meter DimensionL x W x H | 3.90" x 1.96 x 0.87" | 3.12" x 2.25' x 0.85" |
| Meter Weight | 1.94 ounces(without batteries) | 1.5 ounces with battery |
5-4
{4}------------------------------------------------
7. Discussion of Clinical Tests Performed:
HL568 Self-Monitoring Blood Glucose System (Subject Device) is compliant to the standard of ISO 15197-2003: In vitro diagnostic test systems - Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. All the relevant activities were performed by designate individual(s) and the results demonstrated that the predetermined acceptance criteria were fully met.
8. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
The subject device was tested to evaluate its safety and effectiveness, including the followings:
- 15197:2003 In vitro diagnostic test systems Requirements for blood-glucose a. monitoring systems for self-testing in managing diabetes mellitus
- IEC 60601-1:1988 + A1:1991 + A2:1995 Medical electrical equipment Part 1: b. General requirements for safety
- IEC 60601-1-2:2007 Medical electrical equipment Part 1-2: General C. requirements for basic safety and essential performence - Collateral standard: Electromagnetic compatibility - Requirements and tests
- CLSI/NCCLS EP05-A2: Evaluation of Precision Performance of Quantitative d. Measurement Methods
- CLSI EP06-A: Evaluation of the Linearity of Quantitative Measurement. e.
- CLSI EP07-A2: Interference Testing in Clinical Chemistry f.
- CLSI/NCCLS EP09-A2: Method Comparison and Bias Estimation Using Patient ഈ Samples
- FDA Guidance: Guidance for the Content of Premarket Submissions for Software h. Contained in Medical Devices , May 11, 2005
- FDA Guidance: Total Product Cycle Life for Portable Invasive Blood Glucose i. Monitoring System, October 24, 2006
{5}------------------------------------------------
9. Conclusions:
. . . . . . . . .
. .
:
.
.
.
:
.
:
The subject device was tested and fulfilled the requirements from those standards mentioned above, and it's concluded that the subject device is substantially equivalent to the predicate device.
.
:
.
:
:
:
.
、
. .
.
and the country of the country of the country
:
.
{6}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
10903 New Hampshire Avenue Silver Spring, MD 20993
Health & Life Co., Ltd. c/o Ms. Sarah Su 9F, No. 186, Jian Yi Road Chung-Ho City, Taipei China (Taiwan) 235
Re: K110074
Trade Name: HL568 Self-Monitoring Blood Glucose System Regulation Number: 21 CFR §862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Codes: NBW, CGA, JJX Dated: May 22, 2012 Received: May 25, 2012
MAY 2 5 2012
Dear Ms. Su:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{7}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you t the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical
Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ...
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm
Sincerely yours,
Signature
Counney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Indication for Use
KII 0074 510(k) Number (if known):
Device Name:
HL568 Self-Monitoring Blood Glucose System
Indications for Use:
HL568 Self-Monitoring Blood Glucose (SMBG) System is self-test medical device and intended for single patient home-use to monitor the blood glucose (B-D-glucose) levels in quantitative measurement from fresh capillary whole blood obtained from the finger tip. It is intended for use outside the body (in vitro diagnostic use) by diabetics at home to measure the glucose concentration for aiding diabetes management. It is not intended for the diagnosis of or screening for diabetes mellitus.
HL568 Self-Monitoring Blood Glucose (SMBG) System is intended to be used by a single person and should not be shared and it is also not intended for use on neonates and should be used with HL568 Blood glucose test strip and HL568 Control Solution.
HL568 Blood glucose test strips are for use with the HL568 Blood glucose meter to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertip.
HL568 Control Solutions are for use with the HL568 Blood glucose meter and HL568 Blood glucose test strips to check that the meter and test strips are working together properly and that the test is performing correctly.
The device is capable of transferring the storage data to the connected personal computer (PC) via USB cable. In addition, HL568 Self-Monitoring Blood Glucose System is featured with audio function; it could help the user to know the measured result by hearing but is not intended for visually impaired users.
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Rute Chale
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K110074
Page 1 of
V
§ 862.1345 Glucose test system.
(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.