K Number
K110057

Validate with FDA (Live)

Device Name
MINI MASSEUSE
Date Cleared
2011-10-28

(291 days)

Product Code
Regulation Number
890.5850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Mini, Model EC12-01 is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. Mini, Model EC12-01 is not intended to be used as a therapy for any medical condition. None of the Mini programs are designed for injured or ailing muscles and their use on such muscles is contraindicated. Mini, Model EC12-01 is intended for use by healthy persons to apply transcutaneous electrical muscle stimulation(EMS) through skin contact electrodes. Indications for use are as follows: Mini is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. Mini is not intended to as a therapy for any medical condition. Please read the indications, precautions and warnings carefully to determine whether any of them apply to you. Mini stimulates action potentials in the motor nerves supplying muscle. Different patterns of muscle activity can be imposed on the target muscle, depending on the timing and intensity of the stimulation signal. Mini contains programs for exercise warm-up, muscle performance improvement and exercise recovery.

Device Description

The Mini is a powered muscle stimulator for muscle conditioning. It is a device used for other than medical purposes to apply an electrical current to electrodes on a person's skin to temporarily affect the stimulated muscle's contractile properties, force output, and/or fatique resistance. Unlike the classified powered muscle stimulator devices intended for use in physical medicine and rehabilitation, this device is not intended for use in patients with medical conditions and is intended only for muscle conditioning purposes. The Mini is a 9 volt battery powered hand-held device, with detachable electrodes. The electrodes are placed over the muscles which are to be stimulated. The muscles are excited by electric pulses which are created by an electronic circuit. This circuit is software driven, to the user can adjust the stimulation. Mini is pre-programmed with 6 stimulation programs. Programs duration varies depending on the selected mode, which can be reduced by user inputs.

AI/ML Overview

The provided document is a 510(k) Summary for the "Mini Model EC12-01" Powered Muscle Stimulator. It claims substantial equivalence to a predicate device, the Slendertone 561 (K071666). However, the document does not describe an acceptance criteria or explicitly detail a performance study proving the device meets specific performance criteria through empirical testing with human subjects or a defined dataset.

Instead, the submission focuses on demonstrating substantial equivalence to the predicate device by comparing technical specifications and intended use, and by showing compliance with relevant safety standards.

Here's a breakdown based on the information provided, highlighting what is present and what is missing:

1. A table of acceptance criteria and the reported device performance:

This information is not explicitly provided in terms of discrete acceptance criteria and corresponding reported performance metrics (e.g., accuracy, sensitivity, specificity) for device functionality related to muscle stimulation efficacy.

The document does include a table comparing the EC12-01 to the predicate device (MODEL561) across various parameters (Mode/Function, Duration, Intensity Range, Waveform, etc.). This table serves to demonstrate that the devices are technically similar, implying similar performance. The "Discussion" column in this table addresses potential differences and argues that they do not constitute safety or effectiveness differences, thereby serving a similar purpose to a performance claim.

Table of Technical Comparison (serving as indirect performance comparison):

ItemsEC12-01MODEL561 (K071666)Discussion (Implying Equivalence/Performance)
Mode/Function66Same. Basing on the above comparison and detailed description, the subject device's 6 mode is totally same as the predict device's mode Program1-6.
Duration12-30Mins12-30MinsSame.
Intensity Range1-10 (No sub-level)1-99 (subdivided into 10 for each 110, so total 199)Similar but all matching standards. According to the IEC60601-2-10, each stepped voltage and current does not exceed 10%, the highest and lowest values should less than 2%. The EC12-01 classified the intensity 110 according to the standards, the predict device is same as the EC12-01, has 110 but subdivided each into 10, so total 1~99. According to the standards, the intensity's safety and effectiveness for both devices is same.
Trigger FunctionNoYesThe predicate device has trigger function, when key it without release, into contraction cycle; but release it, into relaxation cycle. The EC12-01 has no this function.
Output Channels12Although the number is different, but the two channel of predict device are same, all are synchronous, so the output has the same effect.
Intend for UseEMSEMSSame. EMS (electrical muscle stimulation). Intend to stimulate healthy muscles in order to improve or facilitate muscle performance.
WaveformBiphasicBiphasicSame.
Net Charge per pulse0mC0mCSame.
Positive Pulse Width200-300us200-300usSame.
Negative Pulse Width200-300us200-300usSame.
Interphase interval100us100usSame.
Regulated CurrentYesYesSame.
Microprocessor ControlYesYesSame.
Automatic Shut OffYesYesSame.
Patient Override ControlNoNoSame.
Compliance with 21 CFR 898YesYesSame.
BatteryIDX9V 6F22IDX9V 6F22Same.
Electrodes /Pads60x50mm85x50mmDifferent, but do not exceed safety limit.
Weight (gram)87gram90gramWeight different
Dimensions (in.) [W x H x D]2.4' x 4.7' x1.3'2.4'x3.9' x1.2'Dimensions different.
Housing Materials and ConstructionABS plasticABS plasticSame.
Low power indication when lower than 3.3VYesYesSame.
Lock functionYesYesSame.
Time countdownYesYesSame.
Indication of current modeYesYesSame.
Indication of VoltageNoYesEC12-01 does not have indication of voltage.
Intensity barYesYesSame.
Pad contact indicatorYesYesSame.
PauseYesYesSame.

The document also lists "Performance and Safety Testing" by referencing compliance with several IEC and ISO standards and an FDA guidance document. This implies that the device's performance aligns with the requirements of these standards.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

No information provided. The document does not describe a test set, human subjects study, or clinical trial to assess the device's efficacy in improving muscle performance. The submission relies on claiming substantial equivalence to a predicate device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

This is not applicable/information not provided as no test set requiring expert ground truth was described for efficacy.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

This is not applicable/information not provided as no test set requiring adjudication was described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable. The device is an electrical muscle stimulator, not an AI-powered diagnostic or interpretive tool that would involve human readers or AI assistance in that context.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This is not applicable. The device is a physical muscle stimulator, not an algorithm. Its operation inherently involves human interaction (placing electrodes, selecting programs, adjusting intensity). The "software" mentioned drives the electrical pulses, it's not a standalone diagnostic or interpretive algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

This is not applicable/information not provided as no study requiring such ground truth was described for the device's performance in improving muscle performance. The "ground truth" for the submission's argument is the predicate device's established safety and effectiveness.

8. The sample size for the training set:

Not applicable/information not provided. There is no mention of a "training set" in the context of device performance assessment. The "programming" of the CMOS chip is mentioned, but this refers to the internal software development rather than machine learning training.

9. How the ground truth for the training set was established:

Not applicable/information not provided. As there is no training set described for performance assessment, there is no mention of how ground truth would have been established for it.

{0}------------------------------------------------

510(k) Summary

Submitter's References
MJ Specialty Retail
5198 NORTH US HWY 65
WALNUT SHADE, MO
Post code: 65771
Tel: (417)561-1216
Fax: (417)561-1218
Official Correspondent
Mr. Leiping Wang
No. 56, Luoming South Road, Guanhaiwei
Town, Cixi
Zhehang, Ningbo, China 65771
Tel: (417)561-1216
Fax: 86-574-63671001
Summary Contents
• Submitted Device
• Device Description
• Statement of Intended Use
• Technological Equivalence
• Safety
• Effectiveness
• Conclusions

{1}------------------------------------------------

1. Submitted Device

Trade NameMini
Common NameSport Muscle Stimulator
Classification NamePower Muscle Stimulator

1.1. Predicate/Legally Marketed Device

ModelManufacturerK NumberSubmitted Device
SLENDERTONEENERVIVE,MODEL 561BIO-MEDICALRESEARCHCORP.K071666Mini , Model EC12_01
TOP-RANKADHESIVEELECTRODETOP-RANKHEALTH CAREEQUIPMENT CO.,LTD.K070612Accessory Electrodes toMini

1.2. Proposed Indications

The proposed indications for the predicate Slendertone 561 and the submitted Mini EC12-01 are the same. The design, stimulation energy, stimulation times, stimulation frequency and intended use are exactly the same.

Mini is intended for used by healthy persons to apply transcutaneous electrical muscle stimulation(EMS) through skin contact electrodes. Indications for use are as follows:

Mini is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. Mini is not intended to as a therapy for any medical condition. Please read the indications, precautions and warnings carefully to determine whether any of them apply to you.

Mini stimulates action potentials in the motor nerves supplying muscle. Different patterns of muscle activity can be imposed on the target muscle, depending on the timing and intensity of the stimulation signal. Mini contains programs for exercise warm-up, muscle performance improvement and exercise recovery.

{2}------------------------------------------------

K110057

2. Device Description

Image /page/2/Picture/2 description: The image shows an advertisement for the "Mini Model EC12-01", an electronic stimulation unit. The advertisement claims the unit uses copyright-protected ultra-smooth, deep tissue feeling sensations similar to technology used exclusively by chiropractors, physical therapists, and sports trainers around the world. The advertisement also states that the unit is available in three colors. The image shows three of the units.

{3}------------------------------------------------

2.1 Description :

The Mini is a powered muscle stimulator for muscle conditioning. It is a device used for other than medical purposes to apply an electrical current to electrodes on a person's skin to temporarily affect the stimulated muscle's contractile properties, force output, and/or fatique resistance. Unlike the classified powered muscle stimulator devices intended for use in physical medicine and rehabilitation, this device is not intended for use in patients with medical conditions and is intended only for muscle conditioning purposes.

The Mini is a 9 volt battery powered hand-held device, with detachable electrodes. The electrodes are placed over the muscles which are to be stimulated. The muscles are excited by electric pulses which are created by an electronic circuit. This circuit is software driven, to the user can adjust the stimulation. Mini is pre-programmed with 6 stimulation programs. Programs duration varies depending on the selected mode, which can be reduced by user inputs.

2.2 Structure

The case of the device is made of ABS plastic. Unit dimensions are . 120mm x 60mm x 32mm (L x W x H).

  • There is an LCD display on the face for user information. ●
  • . Five keys on the face of the device permit user inputs.
  • Two electrode inputs are located on the device. Electrodes and . leads are supplied as accessories to the device.

2.3 Electronics and Software

Electronics

The functions of the Mini are driven by a CMOS chip, which has a Flash ROM to accommodate the programming. The chip is programmed to deliver pulses to the electronic circuit, corresponding to the 6 stimulation programs. User interface inputs are input into the chip via the interface buttons on the face of the device. The user can manipulate the program, intensity and time duration of the stimulation. Digital signals are sent to the chip and the program responds to the variable inputs. The output of the stimulator is limited by programming and the electronic circuit design. The device is not capable of delivering harmful levels of energy. Software

Programming for the device is done in Assembly language on the CMOS chip. The program was tested and validated for proper operation.

Page 4 of 11

{4}------------------------------------------------

3. Statement of Intended Use

Mini . Model EC12-01 is intended to stimulate healthy muscles in order to improve or facilitate muscle performance.

Mini Model EC12-01 is not intended to be used as a therapy for any medical condition. None of the Mini Masseuse programs are designed for injured or ailing muscles and their use on such muscles is contraindicated.

Mini, Model EC12-01 is intended for use by healthy persons to apply transcutaneous electrical muscle stimulation(EMS) through skin contact electrodes. Indications for use are as follows:

Mini is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. Mini is not intended to as a therapy for any medical condition. Please read the indications, precautions and warnings carefully to determine whether any of them apply to you.

Mini stimulates action potentials in the motor nerves supplying muscle. Different patterns of muscle activity can be imposed on the target muscle, depending on the timing and intensity of the stimulation signal. Mini contains programs for exercise warm-up, muscle performance improvement and exercise recovery.

3.1 Training Programs

PROGRAM AND OBJECTIVE

EC12-01 has two programs for muscle performance improvement, as well as programs for exercise preparation and post-exercise active recovery. Depending on your objective your objective, you may select the appropriate program as follows:

ObjectiveProgram numberIntensity RangeDuration
Exercise PreparationMode 11-1012
Active Recovery 1Mode 21-1030
Active Recovery 2Mode 31-1020
Active Recovery 3Mode 41-1020
Build EnduranceMode 51-1020
Build StrengthMode 61-1020

Page 5 of 11

{5}------------------------------------------------

K110057

Mode 1: Build Endurance (20 Min.)

This program gently warms up the muscles prior to exercise. It uses a technique called ""twitch potentiation" to prepare muscle fibers for exercise. It feels like a rhythmic massage. Increase the intensity unit you get a strong, but comfortable muscle movement.

Mode 5: Build Endurance (20 Min.)

This program uses a low frequency pulse train which favors twitch fibers. It develops their aerobic capacity and capillary supply and thereby exercises. The exercise comprises an alternating sequence of work and rest and deep muscle contraction. Do not exceed your comfort level.

Mode 6: Muscle Strengthening(20 Min.)

This program uses a pulse frequency appropriate to fast twitch muscle fibers It develops their aerobic capacity and is used for improving maximum muscle strength. The exercise comprises a sequence of work phases separated by longer relaxation phases. Increase the stimulation intensity until you get a strong and deep contraction. Do not exceed your comfort level.

Mode 2: Active Recovery 1 (30 Min.)

This program produces muscle twitches at very low frequency and it feels like a tapping massage. This stimulates blood flow and helps speed up the removal of lactic acid as it accelerates the exchange between the blood supply and the muscle fibers. As a result, the muscles recover more quickly from fatigue, becoming more relaxed with reduced stiffness. Use it after intense exercise to promote recovery and relaxation.

Mode 3: Active Recovery 2 (20Min.)

This 20-minute program is similar to Active Recovery 1 except that the muscle twitch rate slows down during the session. It feels like a tapping massage, but softer than Active Recovery 1.

Mode 4: Active Recovery 3 (20 Min.)

This from of stimulation activates the muscle in a rhythmic motion in a short contraction/relaxation cycle. It feels like a kneading massage and is smoother than the other active recovery programs.

Page 6 of 11

{6}------------------------------------------------

4. Substantial Equivalence

The submitted Mini EC12-01 is constructed to perform like the predicate Slendertone 561. Both devices have six modes of operation and deliver the same electrical stimulation. The electrical stimulation delivered is alike in pulse frequency, voltage amplitude, pulse width, ramp-up/ramp-down, contraction time, relaxation time, times of stimulation and both have the same claimed effects from the stimulation.

ItemsEC12-01MODEL561(K071666)Discussion
Mode/Function66Same.Basing on the above comparison anddetailed description, the subjectdevice's 6 mode is totally same as thepredict device's mode Program1-6;
Duration12-30Mins12-30MinsSame.
Intensity Range1-10(No sub-level)1-99(subdividedinto 10 foreach 110,so total 199)Similar but all matching standards.According to the IEC60601-2-10,each stepped voltage and currentdoes not exceed 10%, the highestand lowest values should less than2%. The EC12-01 classified theintensity 110 according to thestandards, the predict device is sameas the EC12-01, has 110 butsubdivided each into 10, so total1~99. According to the standards, theintensity's safety and effectiveness forboth devices is same.
Trigger FunctionNoYesThe predicate device has triggerfunction, when key it without release,into contraction cycle; but release it,into relaxation cycle. The EC12-01has no this function.
Output Channels12Although the number is different, butthe two channel of predict device aresame, all are synchronous, so theoutput has the same effect.
Intend for UseEMSEMSSame.EMS (electrical muscle stimulation).Intend to stimulate healthy muscles inorder to improve or facilitate muscleperformance.
WaveformBiphasicBiphasicSame.

Page 7 of

{7}------------------------------------------------

K110057

ItemsEC12-01MODEL561(K071666)Discussion
ShapeRectangularRectangularSame.
Net Charge per pulse0mC0mCSame.
Positive Pulse Width200-300us200-300usSame.
Negative Pulse Width200-300us200-300usSame.
Interphase interval100us100usSame.
Regulated CurrentYesYesSame.
MicroprocessorControlYesYesSame.
Automatic Shut OffYesYesSame.
Patient OverrideControlNoNoSame.
Compliance with 21CFR 898YesYesSame.
Battery1X9V 6F221X9V 6F22Same.
Electrodes /Pads60x50mm85x50mmDifferent, but do not exceed safetylimit.
Weight (gram)87gram90gramWeight different
Dimensions (in.)[W x H x D]2.4' x 4.7' x1.3'2.4'x3.9' x1.2'Dimensions different.
Housing Materials andConstructionABS plasticABS plasticSame.
Low power indicationwhen lower than 3.3VYesYesSame.
Lock functionYesYesSame.

Page & of II

{8}------------------------------------------------

Time countdown-------------------------------------------------------------------------------------YesYes.Same.
Value Career------------------------------------------------------------------------------------------------------------------------------------------------------------------------------MALE A BREAK A FREE A
ItemsEC12-01MODEL561(K071666)Discussion
Indication of currentmodeYesYesSame.
Indication of VoltageNoYesEC12-01 does not have indication ofvoltage.
intensity barYesYesSame.
pad contact indicatorYesYesSame.
pauseYesYesSame.

Differences

As noted in the table above, the EC12-01 has ten (10) levels of intensity, and the predicate has 99 levels. Both devices deliver from 0- 40V and 0-80mA. The predicate has control that is more precise in the range of control, but both deliver the same energy for stimulation. This does not constitute a safety or effectiveness difference.

The Slendertone 561 has 2 output channels, while the submitted Mini EC12-01 has a single channel of output. This difference does not constitute a difference in safety or effectiveness. This simply means the predicate device can deliver stimulation to two areas at a time, while the submitted device can deliver stimulation to one at a time.

{9}------------------------------------------------

5. Performance and Safety Testing

Mechanical, environmental safety and performance testing have been accomplished according to the table below to demonstrate this product is substantially equivalent to the predicate device. The submitted device passed all of these tests.

no.Serial Number/versionStandard and description
1IEC 60601-1Medical electrical equipment - Part 1:General requirements for safety
2IEC 60601-1-2Medical electrical equipment - Part 1-2:General requirements for safety - Collateralstandard: Electromagnetic compatibility -Requirements and tests
3IEC 60601-2-10Medical electrical equipment - Part 2-10:Particular requirements for the safety ofnerve and muscle stimulators
4ISO10993-5Biological Evaluation Of Medical Devices -Part 5: Tests For Cytotoxicity: In VitroMethods
5ISO10993-10Biological Evaluation Of Medical Devices -Part 10: Tests For Irritation And Delayed-Type Hypersensitivity
6Guidance Document forPowered Muscle Stimulator510(k)sFDA Guidance document for PoweredMuscle Stimulators

Page 10 of 11

{10}------------------------------------------------

K110057

6 Conclusions:

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification The MJ Specialty Retail. Mini ,Model EC12_01is substantially equivalent to the predicate device as described herein.

Page 11 of 11

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird, with its wings spread and head facing to the right. The bird is composed of three curved lines that suggest feathers or wings.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

OCT 2 8 2011

MJ Specialty Retail % IRC Mr. Charles Mack 77325 Joyce Way Echo, Oregon 97826

Re: K110057

Trade/Device Name: Mini Model EC12 01 Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: II Product Code: NGX Dated: October 21, 2011 Received: October 25, 2011

Dear Mr. Mack:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not micleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your advine can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{12}------------------------------------------------

Page 2 - Mr. Charles Mack

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its to!!-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Eunice Keith

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{13}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: Mini, Model EC12-01

Indications For Use:

Mini, Model EC12-01 is intended to stimulate healthy muscles in order to improve or facilitate muscle performance.

Mini, Model EC12-01 is not intended to be used as a therapy for any medical condition. None of the Mini programs are designed for injured or ailing muscles and their use on such muscles is contraindicated.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-Off

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

KI10057 510(k) Number .

Page 1 of

N/A