K Number
K110016

Validate with FDA (Live)

Date Cleared
2011-08-23

(232 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder Free Vinyl Patient Examination Gloves, Colored (Yellow) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Powder free Vinyl Patient Examination Gloves, Colored (Yellow) are disposable devices which made of poly(vinyl chloride) polymer compound, intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between pattent and examiner and they meets all of the requirements of ASTM standard D 5250-06 el .

AI/ML Overview

1. Acceptance Criteria and Reported Device Performance

CharacteristicsAcceptance Criteria (Standard)Device Performance
DimensionASTM standard D 5250-06e1Meets
Physical PropertiesASTM standard D 5250-06e1Meets
Freedom from pinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 5250-06e1 and D6124-01Meets (<2mg/glove)
Primary Skin IrritationISO 10993-10 (Passes in rabbits)Not a Primary Skin Irritation
Dermal sensitizationISO 10993-10 (Passes in guinea pig)Not a Dermal sensitization

2. Sample Size and Data Provenance

The provided text does not specify the exact sample sizes used for each test. However, it indicates that the tests were performed according to established standards like ASTM and ISO, which would inherently involve specific sample sizes.

The data provenance is not explicitly stated in terms of country of origin, but the submission is from RONGYAO PLASTIC PRODUCE CO., LTD in China. The tests are non-clinical, implying they are laboratory-based and likely conducted prospectively to evaluate the device's conformance to standards.

3. Number of Experts and Their Qualifications

This information is not applicable. The device is a "Powder free Vinyl Patient Examination Gloves," and its performance is evaluated against technical standards and biocompatibility tests, not clinical assessment by experts.

4. Adjudication Method

This information is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies where expert consensus is needed to establish ground truth from ambiguous or complex data, such as medical images. For the technical performance of examination gloves, the outcome is determined by objective measurements against predefined standards.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. An MRMC comparative effectiveness study is designed to evaluate the improvement in human reader performance (e.g., radiologists interpreting images) with and without AI assistance. This type of study is not relevant for a patient examination glove, which is a physical device and not an AI diagnostic tool.

6. Standalone (Algorithm Only) Performance Study

No. This study type is for AI algorithms. The device in question is a physical glove, and its "standalone" performance is assessed by meeting the listed technical and biological compatibility criteria.

7. Type of Ground Truth Used

The ground truth used is based on established industry standards and regulatory requirements. Specifically, these include:

  • ASTM standards (D 5250-06e1, D6124-01) for physical characteristics.
  • 21 CFR 800.20 for freedom from pinholes.
  • ISO 10993-10 for biocompatibility (primary skin irritation and dermal sensitization).

8. Sample Size for the Training Set

This information is not applicable. Training sets are used for machine learning algorithms. This device is a manufactured product and does not involve AI algorithms that require training data.

9. How the Ground Truth for the Training Set was Established

This information is not applicable, as there is no training set for this device.

{0}------------------------------------------------

510(K) Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is: (110016 ."

Premarket Notification [510(k)] Summary

Submitter's name :RONGYAO PLASTIC PRODUCE CO., LTD
Submitter's address :#58 BINHU ROAD, HENGSHUI CITY, 053000,CHINA
Phone number :(86) 318-2126583
Fax number :(86) 318-2126588
Name of contact person:Mr. Zhao Gang
Date the summary was prepared:Jul. 25, 11
Device Name:Powder free Vinyl Patient Examination Gloves, Colored(Yellow)
Proprietary/Trade name:Powder Free Vinyl Patient Examination Gloves
Common Name:Patient examination glove
Classification Name:Patient examination glove
Device Classification:I
Regulation Number:21 CFR 880.6250
Panel:General Hospital (80)
Product Code:LYZ

Class I* powder free vinyl patient examination gloves , colored (yellow) that meets all of the requirements of ASTM standard D 5250-06e1.

Predicate device : Powder free Vinyl Patient Examination Gloves(White, Yellow), SHIJIAZHUANG MANFUL LIGHT INDUSTRIAL PRODUCTS CO., LTD. K051156 .

Device Description:Powder free Vinyl Patient Examination Gloves, Colored (Yellow) are disposable devices which made of poly(vinyl chloride) polymer compound, intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between pattent and examiner and they meets all of the requirements of ASTM standard D 5250-06 el .

Device Intended Use(Indication for use ): Powder free Vinyl Patient Examination Gloves, Colored (Yellow) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

A summary of the technological characteristics of new device compared to the predicate device.

The powder free vinyl patient examination gloves, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

{1}------------------------------------------------

CharacteristicsStandardDevice performance
DimensionASTM standard D 5250-06e1.Meets
Physical PropertiesASTM standard D 5250-06e1.Meets
Freedom from pinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 5250-06e1and D6124-01Meets<2mg/glove
Primary Skin Irritation inrabbitsPasses
BiocompatabilityISO 10993-10Not a Primary Skin Irritation
Dermal sensitization in theguinea pigPasses
ISO 10993-10Not a Dermal sensitization

A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Powder free Vinyl Patient Examination Gloves, Colored (Yellow) meet requirements per ASTM D5250-06e1, per ASTM D6124-01, per 21 CFR 800.20 and ISO10993-10:2002/Amd1:2006.

The performance test data of the non-clinical tests that support a determination of substantial equivalent is the same as mentioned immediately above.

A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

The conclusions

It can be concluded that the Powder Free Vinyl Patient Examination Gloves meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims .

It can be concluded that the Powder free Vinyl Patient Examination Gloves, Colored (Yellow) is as safety, as effective and performs as well as the predicate devices Powder free Vinyl Patient Examination Gloves(White, Yellow), SHIJIAZHUANG MANFUL LIGHT INDUSTRIAL PRODUCTS CO., LTD. K051156 .

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized human figure with three arms, representing service, knowledge, and health. The figure is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MID 20993-0002

Rongyao Plastic Produce Co. Ltd. % Mr. Chu Xiaoan Room 1606 Bldg 1 Yuan #209 Bei Si Huan Zhong Road Beijing, PR China 100083

AUG 2 3 2011

Re: K110016

Trade/Device Name: Powder-Free Vinyl Patient Examination Gloves (Colored: vellow) Regulation Number: 21 CFR §880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: July 25, 2011 Received: July 27, 2011

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2-Mr. Chu Xiaoan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony O. Wainer

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K 110016

INDICATIONS FOR USE

Applicant: RONGYAO PLASTIC PRODUCE CO., LTD

K110016 510(k) Number (if known):_*

Device Name: Powder Free Vinyl Patient Examination Gloves, Colored (Yellow)

Indications For Use:

Powder Free Vinyl Patient Examination Gloves, Colored (Yellow)_is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use × (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

SURER JUANSERCUITENCE of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: Sulloly

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.