(107 days)
DBX Putty is intended for use as a Demineralized Bone Matrix for voids or gaps that are not intrinsic to the stability of the bony structure. DBX Putty is indicated for treatment of surgically created osseous defects or osseous defects created from traumatic injury.
DBX Putty can be used alone in the posterolateral spine. DBX Putty can also be used as an extender in the spine with autograft or allograft. DBX Putty can be used with bone marrow. DBX Putty is for single patient use only.
DBX Inject is intended for use as a Demineralized Bone Matrix for voids or gaps that are not intrinsic to the stability of the bony structure. DBX Inject is indicated for treatment of surgically created osseous defects or osseous defects created from traumatic injury.
DBX Inject can be used alone in the posterolateral spine. DBX Inject can also be used as . an extender in the spine with autograft or allograft. DBX Inject can be used with bone marrow. DBX Inject is for single patient use only.
DBX Putty is composed of Demineralized Bone Matrix and sodium hyaluronate. The demineralized bone allograft in this product is prepared from tissue procured from a deceased donor using aseptic surgical techniques. The bone used in DBX Putty is cortical bone. Sodium hyaluronate is a naturally derived material that is biocompatible and biodegradable. The sodium hyaluronate is mixed in a phosphate buffered saline and is added to the demineralized bone to aid in maintaining physiological pH as well to improve the handling characteristics of demineralized bone.
DBX Inject is composed of Demincralized Bone Matrix and sodium hyaluronate. The demineralized bone allograft in this product is prepared from tissue procured from a deceased donor using aseptic surgical techniques. The bone used in DBX Inject is cortical bone. Sodium hyaluronate is a naturally derived material that is biocompatible and biodegradable. The sodium hyaluronate is mixed in a phosphate buffered saline and is added to the demineralized bone to aid in maintaining physiological pH as well to improve the handling characteristics of demineralized bone.
The provided text is a 510(k) summary for two medical devices: DBX® Demineralized Bone Matrix Putty and DBX® Demineralized Bone Matrix Inject. It describes their intended use, device description, and safety/effectiveness information.
However, the document does not contain the acceptance criteria or a study proving that the device meets specific performance criteria in the way that would typically be presented for an AI/ML device (e.g., sensitivity, specificity, F1 score, etc.). Instead, this document focuses on the bio-compatibility, osteoinductive potential, and viral clearance of the bone matrix products, essentially demonstrating that they are safe and effective for their intended biological function as bone void fillers.
Here's an analysis of the provided information in relation to your request, acknowledging that much of the requested detail for an AI/ML device is not applicable or present in this type of submission.
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Biocompatibility: Safe for use in the human body. | Established through long history of safe and effective clinical use of DBX® materials and supported by laboratory testing per ISO 10993. |
| Sterility: Free from viable microorganisms. | Single-donor processed using aseptic techniques and tested for sterility per current USP <71>. |
| Osteoinductive Potential (Lot Release): Ability to induce bone formation. | Every lot of final DBX Putty/Inject product must test positive in an athymic mouse model or an alkaline phosphatase assay (which has a positive correlation with the athymic mouse model). Note: "It is unknown how the osteoinductive potential, measured in the athymic mouse model or the alkaline phosphatase assay, will correlate with clinical performance in human subjects." This is a key limitation acknowledged by the manufacturer. |
| Viral Clearance and Inactivation: Reduce viral load to safe levels. | Processing methods for the DBM contained in DBX® were evaluated, demonstrating significant viral inactivation potential for a wide range of potential model human viruses. |
| Indications for Use (Expansion): Safe and effective for use as an extender in the posterolateral spine with allograft. | Based on "a recent in vivo study" (details not provided in this summary). The fundamental scientific technology is described as the same as the predicate (K040262). |
Important Note: The document focuses on the biological and safety aspects of a medical device (demineralized bone matrix) rather than analytical performance associated with an AI/ML diagnostic or predictive tool. There are no performance metrics like sensitivity, specificity, or F1 score.
Given that this is a 510(k) summary for a bone void filler and not an AI/ML device, most of the remaining requested information is not applicable. I will address what can be inferred or state its non-applicability.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Test Set Sample Size: Not explicitly stated for specific studies. The lot release testing for osteoinductive potential is conducted on "every lot of final DBX Putty product," implying continuous testing rather than a single pre-market test set. The "recent in vivo study" mentioned for the expanded indications does not specify a sample size.
- Data Provenance: The document does not specify the country of origin for the "recent in vivo study" or the "long history of safe and effective clinical use." The athymic mouse model is a laboratory model.
- Retrospective or Prospective: Not specified. The "long history of safe and effective clinical use" would imply retrospective data accumulation, but the "recent in vivo study" could be prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- This is not applicable as the studies described are biological/in vivo and laboratory tests, not assessments requiring expert interpretation for ground truth establishment in the context of an AI/ML inference.
- For the osteoinductive potential in the athymic mouse model or alkaline phosphatase assay, the ground truth is established by the assay's biochemical or biological outcome.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. This concept relates to resolving discrepancies in expert interpretations, which is not relevant to the types of studies described (in vivo biological models, laboratory assays).
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This device is a bone void filler, not an AI/ML diagnostic or assistive tool for human readers. Therefore, no MRMC study or AI assistance effect size is mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. There is no algorithm; this is a biological product.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Biocompatibility: Established through adherence to ISO 10993 standards and a history of clinical use. This effectively uses a combination of established standards and historical outcomes data.
- Sterility: USP <71> microbiological testing standards.
- Osteoinductive Potential: The positive outcome in an athymic mouse model or an alkaline phosphatase assay serves as the "ground truth" for lot release.
- Viral Clearance: Laboratory assessment of viral inactivation potential using model viruses.
- Expanded Indications: Based on "a recent in vivo study," implying direct biological outcomes in a living system (likely animal model as it's not a human clinical trial summary).
8. The sample size for the training set
- Not applicable. This medical device is not an AI/ML algorithm that requires a training set.
9. How the ground truth for the training set was established
- Not applicable, as there is no training set for an AI/ML algorithm.
{0}------------------------------------------------
103795 DBX® 510(k)
XII. 510(k) SUMMARY OF SAFETY & EFFECTIVENESS
APR 1 3 2011
| PROPRIETARY NAME: | DBX® Demineralized Bone Matrix Putty |
|---|---|
| COMMON NAME: | Bone Void Filler Containing Human DemineralizedBone Matrix (DBM) |
| PROPOSED REGULATORYCLASS: | Class II |
| CLASSIFICATIONIDENTIFICATION: | 21 C.F.R. §888.3045 Resorbable calcium salt bonevoid filler device |
| PRODUCT CODE: | MBP, MQV |
| SPONSOR: | Musculoskeletal Transplant Foundation125 May StreetEdison, NJ 08837732-661-0202 |
INDICATIONS FOR USE:
DBX Putty is intended for use as a Demineralized Bone Matrix for voids or gaps that are not intrinsic to the stability of the bony structure. DBX Putty is indicated for treatment of surgically created osseous defects or osseous defects created from traumatic injury.
DBX Putty can be used alone in the posterolateral spine. DBX Putty can also be used as an extender in the spine with autograft or allograft. DBX Putty can be used with bone marrow. DBX Putty is for single patient use only.
DEVICE DESCRIPTION:
DBX Putty is composed of Demineralized Bone Matrix and sodium hyaluronate. The demineralized bone allograft in this product is prepared from tissue procured from a deceased donor using aseptic surgical techniques. The bone used in DBX Putty is cortical bone. Sodium hyaluronate is a naturally derived material that is biocompatible and biodegradable. The sodium hyaluronate is mixed in a phosphate buffered saline and is added to the demineralized bone to aid in maintaining physiological pH as well to improve the handling characteristics of demineralized bone.
{1}------------------------------------------------
SAFTEY AND EFFECTIVENESS INFORMATION:
This 510(k) was submitted for an addition to the indications for use of DBX® Putty in the posterolateral spine. Based on a recent in vivo study, DBX® Putty can be used as an extender in the spine with allograft. The fundamental scientific technology of the DBX® Putty with expanded indications for use in the posterolateral spinc is the same as the technology for the predicate DBX® Putty (FDA cleared, K040262).
Biocompatibility of DBX® materials has been established through their long history of safe and effective clinical use, further supported by laboratory testing conducted per ISO 10993. DBX® is single-donor processed using aseptic techniques and is tested for sterility per current USP <71>.
OSTEOINDUCTIVE POTENTIAL
DBX Demineralized Bone Matrix is osteoconductive, and has been shown to have osteoinductive potential in an athymic mouse model. Every lot of final DBX Putty product is tested in an athymic mouse model or in an alkaline phosphatase assay, which has been shown to have a positive correlation with the athymic mouse model, to ensure the osteoinductive potential of the final product. Standard testing performed in an athymic mouse or alkaline phosphatase assay must prove positive for lot release. It is unknown how the osteoinductive potential, measured in the athymic mouse model or the alkaline phosphatase assay, will correlate with clinical performance in human subjects.
VIRAL CLEARANCE AND INACTIVATION:
The method for processing the DBM contained in DBX® was evaluated for its viral inactivation potential. A panel of model potential human viruses representing various virus types, sizes, shapes and genomes were evaluated. The DBM processing methods were determined to provide significant viral inactivation potential for a wide range of potential viruses.
{2}------------------------------------------------
XII. 510(k) SUMMARY OF SAFETY & EFFECTIVENESS
| PROPREIETARY NAME: | DBX® Demineralized Bone Matrix Inject |
|---|---|
| COMMON NAME: | Bone Void Filler Containing Human DemineralizedBone Matrix (DBM) |
| PROPOSED REGULATORYCLASS: | Class II |
| CLASSIFICATIONIDENTIFICATION: | 21 C.F.R. §888.3045 Resorable calcium salt bonevoid filler device |
| PRODUCT CODE: | MBP, MQV |
| SPONSOR: | Musculoskeletal Transplant Foundation125 May StreetEdison, NJ 08837732-661-0202 |
INDICATIONS FOR USE:
DBX Inject is intended for use as a Demineralized Bone Matrix for voids or gaps that are not intrinsic to the stability of the bony structure. DBX Inject is indicated for treatment of surgically created osseous defects or osseous defects created from traumatic injury.
DBX Inject can be used alone in the posterolateral spine. DBX Inject can also be used as . an extender in the spine with autograft or allograft. DBX Inject can be used with bone marrow. DBX Inject is for single patient use only.
DEVICE DESCRIPTION:
DBX Inject is composed of Demincralized Bone Matrix and sodium hyaluronate. The demineralized bone allograft in this product is prepared from tissue procured from a deceased donor using aseptic surgical techniques. The bone used in DBX Inject is cortical bone. Sodium hyaluronate is a naturally derived material that is biocompatible and biodegradable. The sodium hyaluronate is mixed in a phosphate buffered saline and is added to the demineralized bone to aid in maintaining physiological pH as well to improve the handling characteristics of demineralized bone.
{3}------------------------------------------------
SAFTEY AND EFFECTIVENESS INFORMATION:
This 510(k) was submitted for an addition to the indications for use of DBX® Inject in the posterolateral spine. Based on a recent in vivo study, DBX® Inject can be used as an extender in the spine with allograft. The fundamental scientific technology of the DBX® Inject with expanded indications for use in the posterolateral spine is the same as the technology for the predicate DBX® Putty (FDA cleared, K040262).
Biocompatibility of DBX® Inject matcrials has been established through their long history of safe and effective clinical use, further supported by laboratory testing conducted per ISO 10993. DBX® Inject is single-donor processed using aseptic techniques and is tested for sterility per current USP <71>.
OSTEOINDUCTIVE POTENTIAL
DBX Inject is osteoconductive, and has been shown to have osteoinductive potential in an athymic mouse model. Every lot of final DBX Inject product is tested in an athymic mouse model or in an alkaline phosphatase assay, which has been shown to have a positive correlation with the athymic mouse model, to ensure the osteoinductive potential of the final product. Standard testing performed in an athymic mouse or alkaline phosphatase assay must prove positive for lot release. It is unknown how the osteoinductive potential, measured in the athymic mouse model or the alkaline phosphatase assay, will correlate with clinical performance in human subjects.
VIRAL CLEARANCE AND INACTIVATION:
The method for processing the DBM contained in DBX® Inject was evaluated for its viral inactivation potential. A panel of model potential human viruses representing various virus types, sizes, shapes and genomes were evaluated. The DBM processing methods were determined to provide significant viral inactivation potential for a wide range of potential viruses.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is positioned to the right of a circular seal. The seal contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Musculoskeletal Transplant Foundation % Ms. Nancy Joy 125 May Street Edison, New Jersey 08837
APR, 1 3 2311
Re: K103795
Trade/Device Name: DBX® Demineralized Bone Matrix Putty and Inject Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV. MBP Dated: March 07, 2011 Received: March 10, 2011
Dear Ms. Joy:
We have reviewed-your-Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{5}------------------------------------------------
Page 2 - Ms. Nancy Joy
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Aiz B. M. he
fer
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
IV. INDICATIONS FOR USE
510(k) Number (if known): K103795
Device Name: DBX® Demineralized Bone Matrix Putty
Indications for Use:
DBX Putty is intended for use as a Demineralized Bone Matrix for voids or gaps that are not intrinsic to the stability of the bony structure. DBX Putty is indicated for treatment of surgically created osseous defects or osseous defects created from traumatic injury.
DBX Putty can be used alone in the posterolateral spine. DBX Putty can also be used as an extender in the spine with autograft or allograft. DBX Putty can be used with bone marrow. DBX Putty is for single patient use only.
' OR Prescription Use _ X (Per 21 CFR 801 Subpart D)
Over-The-Counter Use (Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign Off
(Division Sign-Off (Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K103795
{7}------------------------------------------------
IV. INDICATIONS FOR USE
510(k) Number (if known): K103795
Device Name: DBX® Demineralized Bone Matrix Inject
Indications for Use:
DBX Inject is intended for use as a Demineralized Bone Matrix for voids or gaps that are not intrinsic to the stability of the bony structure. DBX Inject is indicated for treatment of surgically created osseous defects or osseous defects created from traumatic injury.
DBX Inject can be used alone in the posterolateral spine. DBX Inject can also be used as an extender in the spine with autograft or allograft. DBX Inject can be used with bone marrow. DBX Inject is for single patient use only.
OR Prescription Use _ X (Per 21 CFR 801 Subpart D)
Over-The-Counter Use (Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative of Surgical, Original, Oricand Restorative Devices
510(k) Number K103795
N/A