K Number
K103688

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-01-14

(28 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The USGI g-Prox EZ Endoscopic Grasper is intended for use in minimally Invasive procedures to facilitate tissue grasping and manipulation.

Device Description

The g-Prox EZ Endoscopic Grasper is a sterile, single patient use device used for tissue grasping and mobilization. It is available in two jaw lengths. It includes a lumen that can accept the g-Cath Tissue Anchor Delivery Catheter and other small diameter instruments. It is comprised of a polycarbonate proximal handle, flexible shaft made of medical grade polymers and distal stainless steel jaws. It has a nominal working length and outer diameter of 111 cm and 5.5 mm, respectively.

AI/ML Overview

This 510(k) summary (K103688) describes the g-Prox EZ Endoscopic Grasper, a sterile, single-patient-use device intended for tissue grasping and mobilization in minimally invasive procedures.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Device IntegrityNon-clinical performance testing, including bond joint strength testing, was conducted. Test data confirmed that the modified g-Prox EZ device demonstrated equivalent safety and performance to the predicate device.
FunctionalityNon-clinical performance testing was conducted. Test data confirmed that the modified g-Prox EZ device demonstrated equivalent safety and performance to the predicate device. This likely includes the ability to grasp and mobilize tissue, and for the g-Prox EZ, specifically the ability to cut G-Cath anchor suture tails.
In Vivo Simulated UseNon-clinical performance testing, including in vivo simulated use testing, was conducted. Test data confirmed that the modified g-Prox EZ device demonstrated equivalent safety and performance to the predicate device.
Equivalence to Predicate Device (g-Prox Endoscopic Grasper - K093018)The g-Prox EZ Endoscopic Grasper demonstrated equivalent safety and performance to the predicate device in terms of intended use, technology, materials, and overall function. Specific comparisons include: - Identical shaft and distal jaw materials and sizes. - Identical intended use and principles of operation. - Identical lumen for accepting other instruments. - New suture cutting component in g-Prox EZ jaw.

2. Sample Size Used for the Test Set and Data Provenance

The submission states that "Non-clinical performance testing was conducted." While it details the types of tests performed (device integrity, functionality, in vivo simulated use), it does not explicitly state the sample size for these tests or the data provenance (e.g., country of origin, retrospective or prospective). This information is typically found in the detailed test reports which are part of the full 510(k) submission but not included in this summary.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This document does not describe the use of human experts to establish ground truth for the non-clinical performance testing. The evaluation focused on engineering and performance characteristics of the device itself rather than human interpretation of medical images or conditions.

4. Adjudication Method for the Test Set

Not applicable. This device's performance was evaluated through non-clinical engineering and functional testing, not through a process requiring expert adjudication of results.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not performed. This device is a surgical instrument (grasper), not an AI-powered diagnostic or assistive tool for human readers. Therefore, the concept of improving human readers with AI assistance does not apply.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a physical surgical instrument and does not involve an algorithm or AI component in its operation. Its performance is entirely mechanical/physical, used by a human operator.

7. The Type of Ground Truth Used

The "ground truth" for this device's testing was based on engineering specifications, material properties, and functional requirements. For example:

  • Device Integrity: "Ground truth" would be the predetermined acceptable limits for bond joint strength or other integrity measures.
  • Functionality: "Ground truth" would be the successful demonstration of tissue grasping, mobilization, and suture cutting (for the EZ model) according to design specifications.
  • In Vivo Simulated Use: "Ground truth" would involve successful and safe operation within a simulated biological environment, meeting predetermined performance criteria.

8. The Sample Size for the Training Set

Not applicable. As described in point 6, this device is a physical surgical instrument and does not involve any AI algorithms or machine learning that would require a "training set."

9. How the Ground Truth for the Training Set was Established

Not applicable. As described in point 8, there is no training set for this device.

{0}------------------------------------------------

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

K103688
Page 1/2

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and CFR 807.92

1.SUBMITTER INFORMATIONJAN 14 2011
a. Company Name:USGI Medical
b. Company Address:1140 Calle CordilleraSan Clemente, CA 92673
c. Telephone:Fax:(949) 369-3890(949) 369-3891
d. Contact Person:Mary Lou MooneyVice President of Clinical,Regulatory & Quality
e. Date Summary Prepared:January 14, 2011
2.DEVICE IDENTIFICATION
a. Trade/Proprietary Name:g-Prox EZ Endoscopic Grasper
b. Common Name:Grasper
c. Classification Name:Endoscope and accessories, 21CFR 876.1500

3. IDENTIFICATION OF PREDICATE DEVICES

g-Prox Endoscopic Grasper

USGI Medical (K093018)

DESCRIPTION OF THE DEVICE 4.

The g-Prox EZ Endoscopic Grasper is a sterile, single patient use device used for tissue grasping and mobilization. It is available in two jaw lengths. It includes a lumen that can accept the g-Cath Tissue Anchor Delivery Catheter and other small diameter instruments. It is comprised of a polycarbonate proximal handle, flexible shaft made of medical grade polymers and distal

{1}------------------------------------------------

K103688
Page 2/2

stainless steel jaws. It has a nominal working length and outer diameter of 111 cm and 5.5 mm, respectively.

న్. STATEMENT OF INTENDED USE

The g-Prox EZ Endoscopic Grasper is intended for use in minimally invasive procedures to facilitate tissue grasping and mobilization.

6. COMPARISON WITH PREDICATE DEVICES

The g-Prox EZ Endoscopic Grasper is comparable to the predicate device in terms of intended use, technology, and materials. The shaft and distal jaw materials are identical to those used in the predicate device and are provided in the same sizes. The g-Prox EZ device handle is made of polycarbonate instead of Delrin. Both the g-Prox EZ and the predicate device have the identical intended use and principals of operation. Both the g-Prox EZ and the predicate device include a lumen that can accept the g-Cath Tissue Anchor Delivery Catheter and other small diameter instruments. The g-Prox EZ includes a suture cutting component in the jaw to allow the user to cut the g-Cath anchor suture tail with the g-Prox EZ.

7. SUMMARY OF PERFORMANCE DATA

Design control activities for the described modifications were completed in accordance with 21CFR 820.30 and USGI Medical documented design control procedures. Human factors considerations were also incorporated into the device modifications. Risk analysis was performed in accordance with USGI Medical procedures to identify any risks associated with the modifications.

Non-clinical performance testing was conducted to ensure that the device performs as intended when used in accordance with its instructions for use and to demonstrate equivalence to the predicate device. Testing focused on those performance features related to the modifications and included device integrity testing (i.e., bond joint strength), functionality and in vivo simulated use testing. Test data confirmed that the modified g-Prox EZ device demonstrated equivalent safety and performance to the predicate device.

In conclusion, non-clinical performance data demonstrate that the g-Prox EZ Endoscopic Grasper is as safe, effective and performs as well as the g-Prox predicate device.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

USGI Medical % Ms. Mary Lou Mooney Vice President of Clinical, Regulatory and Quality 1140 Calle Cordillera San Clemente, California 92673

JAN 1 4 201

Re: K103688

Trade/Device Name: g-Prox EZ Endoscopic Grasper Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: OCZ, HET Dated: December 16, 2010 Received: December 17, 2010

Dear Ms. Mooney:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Mary Lou Mooney

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electionic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.html far the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Sincerely, yours,

AS B. nh

Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

K103688 510(k) Number (if known):

Device Name: g-Prox Endoscopic Grasper

Indications For Use 510(k) Number (if known):

The USGI g-Prox EZ Endoscopic Grasper is intended for use in minimally Invasive procedures to facilitate tissue grasping and manipulation.

Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel Knaus

Page 1 of (

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K103688

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.