K Number
K103410

Validate with FDA (Live)

Device Name
NEOPAP SYSTEM
Date Cleared
2011-03-30

(128 days)

Product Code
Regulation Number
868.5905
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NeoPAP System is intended to provide continuous positive airway pressure (CPAP) for use in hospitals to treat newborns and infants, weighing less than 5 kg, with RDS or who are recovering from RDS. (Respiratory Distress Syndrome).

Device Description

The Children's Medical Ventures LLC NeoPAP System provides nasal CPAP (continuous positive airway pressure) as a non-invasive method of breathing support for infants and newborns with compromised respiratory systems who weigh less than 5 kg.

The NeoPAP System provides CPAP therapy delivered to patient through a nasal cannula or mask. The system has a feature called Baby-Trak which actively regulates pressure at all times by monitoring pressure at patient's nose and adjusting flow from oxygen and air gas supplies using proportional solenoids. The Baby-Trak feature eliminates the need for a closely-fitted patient interface that can damage fragile tissues.

In addition to CPAP mode, the NeoPAP System offers a constant flow mode where the device delivers a fixed flow of humidified air/oxygen mixture through the nasal prongs or mask. There is also a resuscitation mode where the device delivers a fixed flow of unhumidified air/oxygen through the resuscitation port for use with a resuscitation or hyperinflation bag. The percentage of oxygen can be adjusted while in any mode by using a rotary knob on the user interface.

The NeoPAP System alarms when pressure levels or oxygen concentration are above or below alarm limits. The system also detects occlusion, disconnect, and other alarm conditions. The system displays a real-time bar graph that indicates delivered pressure.

AI/ML Overview

The provided document describes the NeoPAP System, a medical device for providing continuous positive airway pressure (CPAP) to newborns and infants. It focuses on demonstrating substantial equivalence to a predicate device (Guardian Neonate CPAP / Humidification System, K040862) after modifications.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The submission states that "All tests that were created for the modified NeoPAP System had passing results with acceptance criteria successfully met, which demonstrates the safety & effectiveness of the system." However, the specific quantitative acceptance criteria and detailed performance results for each test are not explicitly listed in the provided text. The document primarily focuses on what was tested and the conclusion that criteria were met, rather than presenting the criteria and performance data in a tabular format.

A general summary of some "performance aspects" that are implied to meet criteria through comparison to the predicate device would be:

Feature/AspectAcceptance Criteria (Implied)Reported Device Performance
Intended UseUnchanged from predicate (K040862): provide CPAP for newborns/infants < 5 kg with/recovering from RDS in hospitals.Unchanged from predicate, with re-wording to clarify.
Operating PrincipleSame as predicate.Same as predicate.
TechnologySame as predicate.Same as predicate.
Manufacturing ProcessSame as predicate.Same as predicate.
Software FunctionalityModifications (e.g., occlusion alarms, user confirmation, configuration menu changes) do not impact safety or effectiveness.Software code reviews, unit testing, integration testing conducted with passing results. Occlusion alarms correctly modified to halt gas delivery only for high manifold/patient pressure and auto-reset. User confirmation for rotary knobs implemented.
BiocompatibilityMeets required standards.Testing conducted with passing results.
Environmental PerformanceMeets required standards.Testing conducted with passing results.
User Manual/LabelingCompliant and accurate.Inspection conducted with passing results.
Drawing InspectionsDesign specifications met.Inspections conducted with passing results.
Clinical Simulation (Usability)Performs equivalently to predicate.Confirmed that NeoPAP System performs equivalently to predicate device (K040862).
Flow Mode - Delivery LocationDelivers fixed flow of humidified air/oxygen mixture through nasal cannula/mask.Changed from oxygen hood (predicate) to nasal cannula/mask. Verified functionality.
Resuscitation Mode - %O2 RangeDelivers 21-100% oxygen.Changed from 100% (predicate) to 21-100%. Verified functionality.
Humidification MethodCompatible with an external humidifier (Fisher & Paykel MR850 Humidifier, K073706).Demonstrated compatibility.
Stand-alone Battery Life2 hours.Extended from 1 hour (predicate) to 2 hours. Verified functionality.
Standby ModeFunctionality implemented.Added to accommodate setup/care procedures. Verified functionality.
Patient Interface - Pressure Measurement LocationProvides most accurate readings.Moved to most proximal point of patient interface. Verified improved accuracy.
Patient Interface - Relief Valve Opening Pressure15 cmH2O.Reduced from 18 cmH2O (predicate) to 15 cmH2O. Verified functionality.

2. Sample Size Used for the Test Set and the Data Provenance

The document mentions "Design verification tests," "Formal test protocols," "bench verification testing," "clinical simulation (usability testing)." However, it does not provide specific sample sizes for any of these test sets. It also does not specify the data provenance (e.g., country of origin, retrospective/prospective) for any of the testing conducted. The "clinical simulation" implies prospective testing, but details are absent.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

The document mentions "clinical simulation (usability testing confirmed that the NeoPAP System performs equivalently to the predicate device (Guardian Neonate CPAP / Humidification System, K040862))." However, it does not specify the number of experts involved in this usability testing or their qualifications. It also doesn't explicitly state that they established the "ground truth" in a formal sense, but rather that they confirmed equivalence.

4. Adjudication Method for the Test Set (e.g., 2+1, 3+1, none)

The document does not specify any adjudication method used for the test set.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is not applicable as the NeoPAP System is a medical device (CPAP system) and not an AI-powered diagnostic or assistive technology for human readers. Therefore, an MRMC study comparing human reader performance with and without AI assistance was not performed.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

This section is not applicable as the NeoPAP System is a physical medical device (CPAP system) with embedded software, not a standalone algorithm. The "algorithm" (software) functions as part of the overall device. The software was tested through "software code reviews, software unit testing, software integration testing," which represent standalone testing of the software components.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc)

The concept of "ground truth" as typically defined for AI/diagnostic studies (e.g., pathology for disease diagnosis) is not directly applicable here. The "truth" for this device's performance is established by:

  • Engineering specifications and standards conformance: For safety, electrical, mechanical, environmental, and biocompatibility aspects.
  • Functional performance testing: Verifying that the device operates as intended (e.g., delivers correct pressure, flow, O2 concentration, alarms correctly).
  • Equivalence to predicate: The "ground truth" for the modified device's effectiveness appears to be its performance being equivalent to the legally marketed predicate device (K040862) through "clinical simulation (usability testing)."

8. The Sample Size for the Training Set

The document does not mention a "training set" in the context of machine learning or AI. This is a medical device submission focused on verification and validation of a hardware and software system, not a data-driven AI model that requires training data.

9. How the Ground Truth for the Training Set Was Established

As there is no mention of a "training set" for an AI model, this question is not applicable.

{0}------------------------------------------------

TAB 5

K103410
NeoPAP System

MAR 3 0 2011

510(K) SUMMARY

Official ContactScott WrightSenior Regulatory Affairs EngineerChildren's Medical Ventures LLC191 Wyngate DriveMonroeville, PA 15146Scott.Wright1@philips.com
412-380-8812 t724-681-2407 c412-380-8850 f
Date of SubmissionNovember 22, 2010
Classification Reference21 CFR 868.5905
Product CodeBZD - ventilator, non-continuous (respirator)
Common/Usual NameCPAP System
Proprietary NameNeoPAP System
Predicate Device(s)Guardian Neonate CPAP / Humidification System (K040862)
Reason for submissionModified Device

Substantial Equivalence

The modified device has the following similarities to the previously cleared predicate device:

  • Similar intended use. 0
  • Same operating principle. ם
  • Same technology. 0
  • ם Same manufacturing process.

Design verification tests were performed on the Children's Medical Ventures LLC NeoPAP System as a result of the risk analysis and product requirements. Children's Medical Ventures LLC has

{1}------------------------------------------------

determined that the modifications have no impact on the safety and effectiveness of the device. In summary, the device described in this submission is substantially equivalent to the predicate device.

The modified device complies with the applicable standards referenced in the Guidance for FDA Reviewers and Industry "Guidance for the Content of Pre-market Submissions for Software Contained in Medical Devices," May 2006.

Formal test protocols were written and executed to verify and validate the modified NeoPAP System. All tests that were created for the modified NeoPAP System had passing results with acceptance criteria successfully met, which demonstrates the safety & effectiveness of the system.

Testing included software code reviews, software unit testing, software integration testing, bench verification testing, biocompatibility testing, environmental testing, user manual/labeling inspection, drawing inspections, and a clinical simulation (usability testing confirmed that the NeoPAP System performs equivalently to the predicate device (Guardian Neonate CPAP / Humidification System, K040862). All tests had acceptable results or discrepancies that do not impact safety or effectiveness.

Intended Use

The NeoPAP System is intended to provide continuous positive airway pressure (CPAP) for use in hospitals to treat newborns and infants, weighing less than 5 kg, with RDS or who are recovering from RDS. (Respiratory Distress Syndrome). The intended use of the NeoPAP System is unchanged from the predicate device (Guardian Neonate CPAP / Humidification System, commonly referred to as NeoPAP System, K040862).

Device Description

The Children's Medical Ventures LLC NeoPAP System provides nasal CPAP (continuous positive airway pressure) as a non-invasive method of breathing support for infants and newborns with compromised respiratory systems who weigh less than 5 kg.

The NeoPAP System provides CPAP therapy delivered to patient through a nasal cannula or mask. The system has a feature called Baby-Trak which actively regulates pressure at all times by monitoring pressure at patient's nose and adjusting flow from oxygen and air gas supplies using proportional solenoids. The Baby-Trak feature eliminates the need for a closely-fitted patient interface that can damage fragile tissues.

In addition to CPAP mode, the NeoPAP System offers a constant flow mode where the device delivers a fixed flow of humidified air/oxygen mixture through the nasal prongs or mask. There is also

{2}------------------------------------------------

a resuscitation mode where the device delivers a fixed flow of unhumidified air/oxygen through the resuscitation port for use with a resuscitation or hyperinflation bag. The percentage of oxygen can be adjusted while in any mode by using a rotary knob on the user interface.

The NeoPAP System alarms when pressure levels or oxygen concentration are above or below alarm limits. The system also detects occlusion, disconnect, and other alarm conditions. The system displays a real-time bar graph that indicates delivered pressure.

Predicate Comparison

The fundamental scientific technology of the NeoPAP System is unchanged from the predicate device (K040862). Children's Medical Ventures LLC has made the following changes to the previously cleared NeoPAP System to be considered for this submission:

  • Flow Mode -- Flow Mode for the previously cleared system only delivered flow into an oxygen . hood, whereas the modified system enables the device to deliver a constant flow of oxygenated gas through a nasal cannula or mask. In the modified system, the process for adjusting flow settings was modified to match the process of adjusting pressure settings in CPAP Mode. Both processes now utilize a rotary knob.
  • Resuscitation Mode The previously cleared system could only deliver 100% oxygen to the . patient while in Resuscitation Mode. Since research has shown that high levels of oxygen for an extended period of time can lead to eye damage, the modified system is capable of delivering a range of oxygen concentration between 21 – 100% while in Resuscitation Mode.
  • Aesthetics The external design of the device has been modified to better align with the . Philips brand image.
  • User Interface The user interface (UI) of the modified device was changed to emphasize . ease of use and to be more intuitive than the original design. Key changes are the addition of a flashing alarm bar, removal of the flow segment display, replacement of the majority of membrane keys with capacitive touch keys, and the addition of pressure, FiQ2, and battery status LEDs changing from white to red when a specific parameter causes an alarm. These changes allow for a less cluttered appearance and more visible feedback for the user to determine what parameter is in an alarm state.
  • Occlusion Alarms Software was modified so that occlusion alarms only cause gas delivery . to stop in the case of high manifold or patient pressure. It is no longer halted in the case of flow readings being lower than expected. The software was also modified so that all occlusion alarms auto-reset. Once the trigger for the alarm is no longer present, the occlusion flag is clear and gas delivery will automatically restart if it had been halted.

{3}------------------------------------------------

  • User Confirmation After Turning Rotary Knobs Software was modified so that the user must . confirm that settings should be changed after turning the O2 or pressure/flow setting control knobs. This will prevent inadvertent changes to settings.
  • Humidification The original NeoPAP System utilized a Vapotherm humidification cartridge . which was internal to the device and a delivery circuit that utilized a heated water jacket to minimize rainout within the delivery line. However, due to Vapotherm's recall of the cartridge for potential of bacterial growth, the NeoPAP System was voluntarily recalled after less than a year on the market. For this reason, the modified NeoPAP System does not contain an internal humidifier, but is instead compatible with an external humidifier that is currently being demonstrated to be compatible with the NeoPAP System is the Fisher & Paykel MR850 Humidifier (K073706).
  • Battery Life— The stand-alone battery life of the modified system is 2-hours instead of 1-hour t as in the original NeoPAP System. The longer battery life will be useful in the event that power is lost and a back-up generator is not available or that the device is used during intrahospital transport with no AC outlet available.
  • Standby Mode The original NeoPAP System did not have a Standby Mode. It was added . to the modified NeoPAP System to accommodate system set-up prior to patient receiving therapy and care procedures where the patient is taken off of the device for a period of time.
  • Configuration Menu For both the original and modified NeoPAP System it is necessary to . power off the device in order to enter the configuration menu. In order to allow the clinician to more easily adjust alarm parameters, several items were moved from the configuration menu to the delivery mode menu. The items that were moved are leak delay, high pressure delay, low pressure delay, pressure alarm band, and FiO2 alarm band.
  • Device Packaging -- The original device was packaged using foam inserts. The modified . device utilizes suspension packaging which suspends the device in the airspace of the shipping container between two layers of highly resilient, low-slip film.
  • Patient Interface Circuit The original NeoPAP System patient interface consisted of a . "generator" circuit into which the clinician would plug nasal prongs or nasal masks of an appropriate size for their patient. The tubing came over the top of the patient's head and between the eyes. The modified patient interface (commonly referred to as the PI2 System) consists of a one-piece cannula and one-piece mask. The tubing comes from both sides of the patient's head similar to an adult patient on oxygen. Additionally, a pressure tap was added to the most proximal part of the patient interface (at the patient) to allow the NeoPAP

{4}------------------------------------------------

System to better respond to changes in patient pressure to maintain the pressure level set by the clinician.

  • Patient Interface Sizes/Shape Masks are still available in 2 sizes and are unchanged from . the original NeoPAP System in terms of overall footprint/outline. Cannulas are still available in 5 sizes and are unchanged from the original NeoPAP System in terms of prong spacing and prong diameter. The shape of the nasal prongs have been modified from a flare shape to a tapered shape for ease of manufacturing.
  • Patient Interface Tubing and Connections –The pressure and delivery tubing in the original . NeoPAP patient interface was surrounded by a water jacket. It was connected via a slip fit to the patient tubing. The PI2 System utilizes corrugated delivery tubing packaged standard for Fisher & Paykel RT324 humidification circuits with a heated wire to minimize condensation so that a water jacket is not necessary. The pressure line is permanently bonded to the exhalation body to prevent misconnection. The patient tube is permanently bonded to a cork that plugs into the exhalation body.
  • Patient Interface Exhalation The original NeoPAP patient interface had a series of holes in . the tubing located about 7 inches from the patient. Instead of holes in the tubing, the P12 has an exhalation body located about 15 inches from the patient with three large openings on different faces of the body so that the exhalation is not easily blocked regardless of orientation.
  • Patient Interface Pop Off Valve The original NeoPAP patient interface had an umbrella .
  • valve located directly at the patient in the generator which held the prongs or mask. The Pl2 . System utilizes the exact same valve made by the same supplier, but it is located in the exhalation body delivery air path.
  • Patient Interface Bonnet The original NeoPAP patient interface included a cotton bonnet . available in 12 sizes to secure the "generator" circuit. The Pl2 System includes a bonnet available in 5 sizes with a hook and loop material to adjust the fit. It has a unique foam material on the inside lining to minimize shifting of the bonnet and adjustable bonnet clips to hold the PI2 tubing in place.
  • Patient Interface Usage Type: The device manual of the origina! NeoPAP System specified . that the delivery tube, patient tube, and bonnet were single patient use while the nasal prongs and mask were single use. The device manual for the modified NeoPAP System indicates that entire patient interface is single patient use and lists acceptable agents/procedure for cleaning the exterior of the patient interface.

{5}------------------------------------------------

and the country of the country of the county of the county of the county of the county of

Traditional 510(k) Tab 5 – 510(k) Summary

Table 1 (follows) compares the NeoPAP System with the predicate device.

Table 1 – Device Comparison
Guardian Neonate CPAP /Humidification System (K040862)NeoPAP System
Intended UseIntended to provide CPAP for usein hospitals to treat newborns andinfants with RDS or are recoveringfrom RDS. (Respiratory DistressSyndrome) May or may not includehumidification capabilitiesUnchanged from K040862, other thanrewording.The NeoPAP System is intended toprovide continuous positive airwaypressure (CPAP) for use in hospitalsto treat newborns and infants,weighing less than 5 kg, with RDS orwho are recovering from RDS(Respiratory Distress Syndrome).
Intended Environment of UseHospitalUnchanged from K040862.
Patient PopulationInfants with or recovering from RDSweighing less than 5 kg.Unchanged from K040862.
Product CodeBZDUnchanged from K040862.

Operating Mode Descriptions and Ranges

Gas DeliveredAir / Oxygen MixtureUnchanged from K040862.
CPAP Mode - DescriptionDevice actively regulates pressureat all times by monitoring pressureand adjusting flow from air and O2supplies using proportionalsolenoids.Unchanged from K040862.
CPAP Mode - DeliveryLocationNasal prongs / mask.Unchanged from K040862.
CPAP Mode - Range ofPressure2 to 10 cmH2OUnchanged from K040862.
CPAP Mode - %O2 Range21 – 100%Unchanged from K040862.
Flow Mode - DescriptionDevice delivers fixed flow ofhumidified air/oxygen mixture.Unchanged from K040862.

스 """

. ---

:

{6}------------------------------------------------

の 2017年 10:00 PM 2017 10:

Table 1 - Device Comparison
Guardian Neonate CPAP /Humidification System (K040862)NeoPAP System
Flow Mode - DeliveryLocationOxygen hood.Nasal cannula / mask.
Flow Mode – Range of Flow5 to 15 L/min5 to 10 L/min
Flow Mode - %O2 Range21 - 100%Unchanged from K040862.
Resuscitation Mode -DescriptionResuscitation Mode provides un-humidified gas through theresuscitation port for use with aresuscitation or hyperinflation bag.Unchanged from K040862.
Resuscitation Mode - FixedFlow10 L/minUnchanged from K040862.
Resuscitation Mode - %O2Range100%21 - 100%
Standby ModeNoYes
Humidification MethodVapotherm Microporous MembraneThe Fisher & Paykel MR850 Humidifier(K073706) has been demonstrated tobe compatible with the NeoPAPSystem in this submission.
Range of Temperature ofGas Delivered33 to 41 °CThe F&P MR850 humidifier does nothave communications with the NeoPAPdevice so the temperature is set on thehumidifier instead of the NeoPAPdevice.

User Interface / Display of Measured Data

Circuit Pressure (bar graphdisplay) Range$0-12 cmH_2O$Unchanged from K040862.
%O₂ (window display) Range21 – 100%Unchanged from K040862.
Gas Temperature Range10 to 50 °CThe F&P MR850 humidifier does nothave communication with the NeoPAPdevice so the measured gas

. . . . . . . . .

{7}------------------------------------------------

. . . . . .

Tab 5 - 510(k) Summary
Table 1 – Device Comparison
Guardian Neonate CPAP /Humidification System (K040862)NeoPAP System
temperature will only be displayed onthe humidifier and not on the NeoPAPdevice.
Flow (bar graph)Indicator only – indicates thatpatient is inhaling / exhaling butdoes not give actual flow reading.Flow bar graph removed to preventconfusion and provide a less clutteredappearance.
AlarmsHigh / Low Air Supply PressureHigh / Low O₂ Supply PressureHigh / Low FiO2High / Low Flow (Flow mode only)High / Low Pressure (CPAP modeonly)High / Low TemperatureHigh / Low Water PressureNo Water SupplyNo AC PowerNo O₂ SensorBattery Not PresentLow BatteryReplace BatteryPressure Unstable (CPAP modeonly)Partial OcclusionOcclusionLeakUnexpected RestartMiscellaneous System Error AlarmsUnchanged from K040862 with theexception that the NeoPAP device doesnot have communication with the F&FMR850 humidifier so the NeoPAP willnot alarm for high / low gastemperature, high / low water pressureor no water supply. However, thehumidifier has its own alarming systemfor high / low gas temperature.
Alarm Visual IndicatorBacklit outline of a red bell.Flashing alarm bar. Additionally,pressure and FiO2 readings that aretypically displayed using white LEDschange to red so that the user canmore easily determine which of the
Table 1 – Device Comparison
Guardian Neonate CPAP /Humidification System (K040862)NeoPAP System
System Halt ConditionsMains power disconnectedAir or O2 supply pressure over 93psiInternal battery voltage depletedHigh pressure detected duringstartupDelivery tube disconnected duringoperationAmbient temperature out of rangeTube occluded during system startupOther miscellaneous halt conditionsUnchanged from K040862.
Input TypesRotary knobs and membrane keys.Rotary knobs, membrane keys, andcapacitive keys.
Configuration Menu AccessTurn off device and power on whileholding the Alarm Reset key.Unchanged from K040862.
Configuration Menu ItemsAlarm Sound GapAuto-Reset SettingPatmLanguageLeak DelayHigh Pressure DelayLow Pressure DelayPressure Alarm BandTemperature Alarm BandFiO₂ Alarm BandAlarm Sound Gap, Auto-Reset Setting,Patm, and Language remain inConfiguration Menu, and Leak Delay,High Pressure Delay, Low PressureDelay, Pressure Alarm Band, and FiO₂Alarm Band where moved to theDelivery Mode Menu so they could beaccessed without powering off thedevice. Temperature Alarm Band itemdoes not exist in current software sincethe F&P MR850 humidifier does nothave communication with the NeoPAPdevice so it will not alarm for high / lowgas temperature.
Table 1 – Device Comparison
Guardian Neonate CPAP /Humidification System (K040862)NeoPAP System
PowerAC and BatteryUnchanged from K040862.
Stand-alone Battery Life1 hour2 hours
Oxygen SensorVentrex Oxygen Sensor (K963415)Maxtec Oxygen Sensor (K972992).
Oxygen Sensor LocationMixing block upstream of vaportransfer cartridgeMoved to pressure manifold block toallow FiO2 to also be measured inResuscitation Mode.
Patient Interface
Usage TypeDelivery tube, patient tube, bonnet,are single patient use. Nasalprongs / mask are single use.All components are single patient use.The entire patient interface can becleaned and reused on the samepatient.
Sterility ConditionNon-sterileUnchanged from K040862.
Interface DesignNasal prongs/mask is snapped into"generator" which is connected todual lumen tubing that is routedabove the nose (secured by a clipintegral to the hat) to over the top ofthe head.Nasal cannula/mask is one-piecemolding that is connected to dual lumepressure/delivery tube and singlelumen exhalation tube that are routedto the sides of the head (secured bybonnet clips) and then over the head.
Exhalation ManagementSeries of holes in tubingapproximately 7 inches frompatient. Tubing must be orientedcorrectly so that holes are notblocked.Exhalation body approximately 15inches from patient with 3 largeopenings on different facets of theexhalation body so that exhalation isnot easily blocked regardless oforientation.
Patient PressureMeasurement LocationAt connection betweendelivery/pressure tubing and patienttube (approximately 10 inches frompatient).At most proximal point of patientinterface (closest to patient) to providemost accurate readings.
Pressure Relief ValveLocationLocated directly at the patient in the"generator" which holds nasalLocated in the exhalation body deliveryair path approximately 15 inches
Table 1 - Device Comparison
Guardian Neonate CPAP /Humidification System (K040862)NeoPAP System
prongs or maskupstream from patient.
Relief Valve OpeningPressure (Measured atPatient)18 cmH2O15 cmH2O
Nasal Cannula SizesExtra large, large, medium, small,and extra small.Unchanged from K040862.
Nasal Mask SizesLarge and small.Unchanged from K040862.
Bonnet DesignCotton bonnet without mechanismto adjust fit of bonnet. Lacesthread through bonnet eyelets tosecure patient interface.Foam bonnet with hook and loopmechanism to adjust fit of bonnet.Adjustable bonnet clips to securepatient interface.
Bonnet SizesSizes 0 to 11.Extra large, large, medium, small, andextra small.

and the comments of the production of the comments of the comments of the

.

:- -

{8}------------------------------------------------

. .

and the country of the country of the county


  • --- · · · · ·

:

{9}------------------------------------------------

. . . . . . . .


'

·

{10}------------------------------------------------

.

:

. .

:

.

. . . . . .


.

:

:

·

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three tail feathers and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circle around the eagle. The eagle is black, and the text is also black.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002

Children's Medical Ventures, LLC C/O Mr. Scott Wright Senior Regulatory Affairs Engineer 191 Wyngate Drive Monroeville, Pennsylvania 15146

MAR 3 0 2011

Re: K103410

Trade/Device Name: NeoPAP System Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: March 18, 2011 Received: March 23, 2011

Dear Mr. Wright:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{12}------------------------------------------------

Page 2 - Mr. Wright

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutlFDA/CentersOffices/CDRH/CDRHOffices/ucm] 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Nh for

Anthony Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{13}------------------------------------------------

Indications for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: _ _ _ NeoPAP System_

Indications for Use:

・・

The NeoPAP System is intended to provide continuous positive airway pressure (CPAP) for use in hospitals to treat newborns and infants, weighing less than 5 kg, with RDS or who are recovering from RDS. (Respiratory Distress Syndrome).

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

L. Schulte

on of Anesthesioner. General Infection Control and De

§ 868.5905 Noncontinuous ventilator (IPPB).

(a)
Identification. A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.(b)
Classification. Class II (performance standards).